• When Not Using IRXross for a Long Period of Time
« IRXross with Standard Specification and without Optional Dehumidifier »
1. When Sample compartment window (STD) is used.
Sample compartment window (STD) is made of KBr.
Remove the sample compartment window from the interferometer and keep it in the dedicated case or a desiccator.
Sample compartment window (STD) is made of KBr.
Remove the sample compartment window from the interferometer and keep it in the dedicated case or a desiccator.
When storing in the dedicated case, check the silica gel inside the dedicated case and replace it if necessary.
Attach the Lid of sample compartment window over the hole made by removing the sample compartment window.
Replace to new Silica gel. Click here for Replacement of Silica Gels.
Inspect according to "Periodic Inspections" once a week, and then replace to new Silica gel if necessary.
Attach the Lid of sample compartment window over the hole made by removing the sample compartment window.
Replace to new Silica gel. Click here for Replacement of Silica Gels.
Inspect according to "Periodic Inspections" once a week, and then replace to new Silica gel if necessary.
Click here for Periodic Inspection.
2. When Sample compartment window (AH) is used.
Sample compartment window (AH) is made of KRS-5.
Replace to new Silica gel. Click here for Replacement of Silica Gels.
Inspect according to "Periodic Inspections" once a week, and then replace to new Silica gel if necessary.
Sample compartment window (AH) is made of KRS-5.
Replace to new Silica gel. Click here for Replacement of Silica Gels.
Inspect according to "Periodic Inspections" once a week, and then replace to new Silica gel if necessary.
Click here for Periodic Inspection.
« IRXross with Optional Dehumidifier »
1. When Sample compartment window (STD) is used.
Sample compartment window (STD) is made of KBr.
Even if the IRSpirit is not used for long term, AC power must still be supplied. The dehumidifier constantly runs, desiccating the inside of the interferometer and protecting the optical elements against humidity.
However, remove the sample compartment window from the interferometer and keep it in the dedicated case or a desiccator.
Sample compartment window (STD) is made of KBr.
Even if the IRSpirit is not used for long term, AC power must still be supplied. The dehumidifier constantly runs, desiccating the inside of the interferometer and protecting the optical elements against humidity.
However, remove the sample compartment window from the interferometer and keep it in the dedicated case or a desiccator.
When storing in the dedicated case, check the silica gel inside the dedicated case and replace it if necessary.
Attach the Lid of sample compartment window over the hole made by removing the sample compartment window.
Inspection according to "Periodic Inspections" must be done once a week.
Attach the Lid of sample compartment window over the hole made by removing the sample compartment window.
Inspection according to "Periodic Inspections" must be done once a week.
Click here for Periodic Inspection.
2. When Sample compartment window (AH) is used.
Sample compartment window (AH) is made of KRS-5.
Even if the IRSpirit is not used for long term, AC power must still be supplied. The dehumidifier constantly runs, desiccating the inside of the interferometer and protecting the optical elements against humidity.
Inspection according to "Periodic Inspections" must be done once a week.
Sample compartment window (AH) is made of KRS-5.
Even if the IRSpirit is not used for long term, AC power must still be supplied. The dehumidifier constantly runs, desiccating the inside of the interferometer and protecting the optical elements against humidity.
Inspection according to "Periodic Inspections" must be done once a week.
Click here for Periodic Inspection.
• If a Power Outage Occurs
« IRXross with Standard Specification and without Optional Dehumidifier »
1. When Sample compartment window (STD) is used.
Sample compartment window (STD) is made of KBr.
When AC power cannot be supplied because of periodic inspection of utilities etc., remove the sample compartment window from the interferometer and keep it in the dedicated case or a desiccator.
When AC power cannot be supplied because of periodic inspection of utilities etc., remove the sample compartment window from the interferometer and keep it in the dedicated case or a desiccator.
When storing in the dedicated case, check the silica gel inside the dedicated case and replace it if necessary.
Attach the Lid of sample compartment window over the hole made by removing the sample compartment window.
Replace to new Silica gel. Click here for Replacement of Silica Gels.
Inspect according to "Periodic Inspections" once a week, and then replace to new Silica gel if necessary.
Attach the Lid of sample compartment window over the hole made by removing the sample compartment window.
Replace to new Silica gel. Click here for Replacement of Silica Gels.
Inspect according to "Periodic Inspections" once a week, and then replace to new Silica gel if necessary.
Click here for Periodic Inspection.
2. When Sample compartment window (AH) is used.
Sample compartment window (AH) is made of KRS-5.
When AC power cannot be supplied because of periodic inspection of utilities etc., replace to new Silica gel.
Sample compartment window (AH) is made of KRS-5.
When AC power cannot be supplied because of periodic inspection of utilities etc., replace to new Silica gel.
Click here for Replacement of Silica Gels.
Inspect according to "Periodic Inspections" once a week, and then replace to new Silica gel if necessary.
Inspect according to "Periodic Inspections" once a week, and then replace to new Silica gel if necessary.
Click here for Periodic Inspection.
« IRXross with Optional Dehumidifier »
1. When Sample compartment window (STD) is used.
Sample compartment window (STD) is made of KBr.
When AC power cannot be supplied because of periodic inspection of utilities etc., remove the sample compartment window from the interferometer and keep it in the dedicated case or a desiccator.
When AC power cannot be supplied because of periodic inspection of utilities etc., remove the sample compartment window from the interferometer and keep it in the dedicated case or a desiccator.
When storing in the dedicated case, check the silica gel inside the dedicated case and replace it if necessary.
Attach the Lid of sample compartment window over the hole made by removing the sample compartment window.
Exchange to the original Silica gel case and then, replace to new Silica gel. Click here for Replacement of Silica Gels.
Inspect according to "Periodic Inspections" once a week, and then replace to new Silica gel if necessary.
Attach the Lid of sample compartment window over the hole made by removing the sample compartment window.
Exchange to the original Silica gel case and then, replace to new Silica gel. Click here for Replacement of Silica Gels.
Inspect according to "Periodic Inspections" once a week, and then replace to new Silica gel if necessary.
Click here for Periodic Inspection.
2. When Sample compartment window (AH) is used.
Sample compartment window (AH) is made of KRS-5.
When AC power cannot be supplied because of periodic inspection of utilities etc., exchange to the original Silica gel case.
Replace to new Silica gel. Click here for Replacement of Silica Gels.
Inspect according to "Periodic Inspections" once a week, and then replace to new Silica gel if necessary.
When AC power cannot be supplied because of periodic inspection of utilities etc., exchange to the original Silica gel case.
Replace to new Silica gel. Click here for Replacement of Silica Gels.
Inspect according to "Periodic Inspections" once a week, and then replace to new Silica gel if necessary.
Click here for Periodic Inspection.