• When Not Using IRTracer-100 for a Long Period of Time (Over One Month)
Even if the IRTracer-100 is not used for long term, AC power must still be supplied. The dehumidifier constantly runs, desiccating the inside of the interferometer and protecting the optical elements against humidity.
However, remove the sample compartment window from the interferometer and keep the beam splitter in the beam splitter case or a desiccator. Attach the included sample compartment window cover over the hole made by removing the sample compartment window.
Inspection according to "Periodic Inspections" must be done once a week.
However, remove the sample compartment window from the interferometer and keep the beam splitter in the beam splitter case or a desiccator. Attach the included sample compartment window cover over the hole made by removing the sample compartment window.
Inspection according to "Periodic Inspections" must be done once a week.
Click here for Periodic Inspection.
• If a Power Outage Occurs
The dehumidifier cannot desiccate when AC power cannot be supplied because of periodic inspection of utilities etc. Remove the beam splitter and sample compartment window from the interferometer and keep the beam splitter in the beam splitter case or a desiccator. Attach the included sample compartment window cover over the hole made by removing the sample compartment window.
The dehumidifier cannot desiccate when AC power cannot be supplied because of periodic inspection of utilities etc. Remove the beam splitter and sample compartment window from the interferometer and keep the beam splitter in the beam splitter case or a desiccator. Attach the included sample compartment window cover over the hole made by removing the sample compartment window.