• When Not Using IRAffinity-1S for a Long Term (Over a Month)
Even if the IRAffinity-1S is not used for long term, AC power must still be supplied. The Dehumidifier works, desiccating the inside of the interferometer and protecting the optical elements against humidity.
Inspection according to the "Periodic Inspection" must be done once a week.
Inspection according to the "Periodic Inspection" must be done once a week.
Click here for Periodic Inspection.
• For Brownout (If a Power Outage Occurs)
When AC power cannot be supplied because of periodic inspection of utilities etc., the Dehumidifier does not desiccate. Replace Silica gel packages in the FTIR once a week during brownout.
When AC power cannot be supplied because of periodic inspection of utilities etc., the Dehumidifier does not desiccate. Replace Silica gel packages in the FTIR once a week during brownout.
Click here for Replacement of Silica Gels.