If your chromatography has recently started to demonstrate tailing peaks; this could be related to common causes described below.
Causes | Solutions |
Secondary interactions | - For bases increase pH (as permitted); for acids decrease pH; increase ionic strength of buffer (as permitted); change column type. |
Dead volume | - Reconnect the column with the fitting to reduce dead volume. |
Column degradation | - Replace the column. |
Column void | - Fill void (previous performance unlikely to be fully recovered). |
Interfering peak | - Use a longer column; further method development. |
Wrong mobile phase pH | - Adjust pH (2 clear pH units from pKa recommended). |
Sample chelating to active sites | - Limit interaction via ion pair reagent, modifier or sequester agent, change column or post injector wettable flow-path. |
Inadequate buffering | - Use 50-100 mM buffer concentration (UV methods). |
Sample loading | - Reduce sample concentration. |