If your LC system is demonstrating lower than expected system pressure; the following causes may be responsible.

| Causes | Solutions |
| Partial leak in system | - Check all connections and retighten any which have leaks. |
| Flow rate | - Check the method has the correct flow rate. - Test the flow rate accuracy using a calibrated flow rate meter or collect a specific volume and monitor the time required. |
| Method | - Check if method is using correct temperature and correct solvents. |
| Incorrect column | - Use correct column with correct dimensions and particle geometry. |
| Column temperature too high | - Set adequate column temperature and check no column damage if exceeded column temperature limit. |
| Sensor malfunction | - Repair or replace pressure sensor. |