If you notice your LC system is experiencing fluctuating pressure; the following causes may be responsible.
Causes | Solutions |
Air bubbles | - Purge the solvent lines to remove the air bubbles. |
Worn pump seals | - Replace seals. |
Check valves | - Sonicate the check valves in isopropanol. - Change the check valves if problem persists. |
Leaks | - Degradation of pump seals could cause small leaks. Replace the seals. Check connections. |
Inadequate degassing | - Degas solvent; replace mobile phase frits; repair degasser (if applicable). |
Using a gradient elution | - Pressure cycling caused by viscosity changes is normal but, use adequately sized mixer volume. |