Although applications can vary. The recommended seal rinse solutions for reversed phase and normal phase are listed below.
Mobile Phase |
LC-40D X3 / XS / XR |
LC-40D |
||
Necessary |
Recommended rinsing |
Necessary |
Recommended rinsing |
|
Buffer solution |
Yes |
10 % |
Yes |
Water |
Normal phase solvent (Hexane) |
No |
- |
No |
- |
Other solutions |
Yes |
10 % |
Not specified |
Water (10 % |
Please note: Replace seal rinse solutions regularly. In the absence of organic solvent most aqueous media can suffer from microbial growth which in turn can contaminate your system.