• Rinsing the Online Flow Line
NOTE: Rinsing is normally not necessary because the instrument is designed so that water passes through even when measurement is not being performed.
1. Press
- [Maintenance].
2. Press [Machine Sequence].
The [Machine Sequence] screen appears.
Make sure that the sample liquid is flowing to the online side.
The [Machine Sequence] screen appears.
Make sure that the sample liquid is flowing to the online side.
3. Press [Flow Sequence-Main Unit].

NOTE: Note that rinsing cannot be stopped once the button is pressed. It takes approximately 5 minutes to complete.
Flow line rinsing starts automatically.
4. The [Machine Sequence] screen is displayed when rinsing is finished.
The rinsing process is complete.
NOTE: Perform rinsing multiple times if necessary, such as when a high-concentration sample was measured recently.
The rinsing process is complete.
NOTE: Perform rinsing multiple times if necessary, such as when a high-concentration sample was measured recently.
• Rinsing the Vial Sampler Flow Line
1. Press
- [Maintenance].
2. Press [Machine Sequence].
The [Machine Sequence] screen appears.
The [Machine Sequence] screen appears.
3. Prepare four vials and load them into the vial sampler.
NOTE:
• See "Using the Vial Sampler" before loading the vials into the vial sampler.
• Use pure water with a TOC of less than 100 μg/L. "Preparing the Blank Water".
• Use pure water with a TOC of less than 100 μg/L. "Preparing the Blank Water".
Click here for "Calibration (TOC-1000e)" ("Using the Vial Sampler" and "Preparing the Blank Water")
4. Press [Flow Sequence-Vial Unit].

NOTE: Note that rinsing cannot be stopped once the button is pressed. It takes approximately 20 minutes to complete.
Flow line rinsing starts automatically.
5. The [Machine Sequence] screen is displayed when rinsing is finished.
The rinsing process is complete.
NOTE: Perform rinsing multiple times if necessary, such as when a high-concentration sample was measured recently.
The rinsing process is complete.
NOTE: Perform rinsing multiple times if necessary, such as when a high-concentration sample was measured recently.