In the event of a power outage, turn off the main switch.
Turn the main switch back on after power has recovered.
NOTE: If power went out while measurement was running, the results of measurement steps completed up to that point are retained.
Turn the main switch back on after power has recovered.
NOTE: If power went out while measurement was running, the results of measurement steps completed up to that point are retained.
1. Loosen the retaining screw and remove the waterproof cover.

2. Turn off the main switch on the left side and disconnect the power cord from the outlet.
3. When the power recovers, plug the power cord into the outlet and turn on the main switch.
Restart the instrument.
A system check runs when the instrument is restarted.
If the power outage occurred while a measurement is ongoing, the measurement will automatically resume. The following information is retained:
• Analog output and contact output settings (Analog synchronized stream identification)
• History of instrument alarms and measurement value alarms
• Contact output settings for instrument alarms and measurement value alarms
3. When the power recovers, plug the power cord into the outlet and turn on the main switch.
Restart the instrument.
A system check runs when the instrument is restarted.
If the power outage occurred while a measurement is ongoing, the measurement will automatically resume. The following information is retained:
• Analog output and contact output settings (Analog synchronized stream identification)
• History of instrument alarms and measurement value alarms
• Contact output settings for instrument alarms and measurement value alarms
4. Attach the waterproof cover.