[CALIBRATION]
• Conductivity calibration
147 mS/cm KCL Conductivity Std. Set (Part No.: 638-60254-91)(Required Blank Water: 3 vials)
• TOC calibration
• TOC calibration
0-250-500 μg/L Sucrose TOC Std. Set (Part No.: 638-60255-91)(Required Blank Water: 3 vials) or
0-500-1000-2000 μg/L Sucrose TOC Std. Set (Part No.: 638-60257-91)Required Blank Water: 4 vials)
0-500-1000-2000 μg/L Sucrose TOC Std. Set (Part No.: 638-60257-91)Required Blank Water: 4 vials)
NOTE: Always use the vials in a standard solution set collectively. Exchanging blank water between different standard solution sets may prevent acquisition of correct measurements.