CARx prim is out of range [Error Code:9063 9064 9065 9066]
Carrier gas is not supplied correctly. Check the amount of remaining gas in the cylinder and tubing.
-
Detector is not controlled [Error Code:9021]
Ignition or illumination was started when the detector control was set to [Off]. <GC instrument monitor setting> Press [HOME] > [Detector] on GC instrument monitor. Turn [On] the detector c...
-
Can not switch gas because of gas selector# sub error occurring [Error Code:2372 2373]
You tried to set the supply gas to sub while there is an error with the sub of the gas selector. Check whether the sub gas is supplied and whether gas is not leaking from connections. After r...
-
Can not switch gas because of gas selector# main error occurring [Error Code:2370 2371]
You tried to set the supply gas to main while there is an error with the main of the gas selector. Check whether the main gas is supplied and whether gas is not leaking from connections.After re...
-
SCD induc. tube use time warn. [Error Code:9076]
The number of analysis exceeded the set threshold. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacem...
-
SCD detector abs. press is low [Error Code:9080]
The SCD detector absolute pressure sensor has failed. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.
-
SCD flow controler#2 error [Error Code:6178]
An unexpected error has occurred on the flow controller for SCD gas control. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.
-
SCD flow controler#1 error [Error Code:6177]
An unexpected error has occurred on the flow controller for SCD gas control. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representa...
-
DET#1 2 3 4 failed to start plasma[Error Code:4208 4209 4210 4211]
BID Plasma cannot be turned on. • Check that the detector gas flow is appropriate. • Ignition may be difficult when gas replacement is insufficient immediately after tubing connection. Supply th...
-
System is not ready [Error Code:9010]
The analysis was started while the system was not ready. Start the analysis after the system is ready. If this message appears when all the parameters including temperature and flow are ready...