  • (GCMS) Filament ON Errors

    • "Filament consumed" message appears.   Filament installation is not correctly.   Filament malfunction.   Click here for Filament Replacement Procedure   • "Trap/Total current insufficient" messag...

  • (GCMS) Vacuum System Auto Startup Errors

    Cause 1: Does Rotary Pump1 (RP1) operate? Action 1: Connect the RP1 power cable to the back panel of the MS, and turn on the RP1 switch. If not resolved, contact a Shimadzu sales or service represe...

  • (GCMS) GCMSsolution Software Startup Errors

    1. PC and MS or MS and GC connected properly? All connections must be made properly 2. Is GC power on? Turn on GC power supply and restart. 3. Is MS power on? Turn on MS power supply and restart. 4...

  • (GCMS) Error status indicated by LED on MS

    Normal LED Indications:   Switch MS Unit → All 3 LED light → Green and Red LED goes out → Yellow LED remains ON   LED Error Indications   Yellow LED dose not light: Appropriate power is not supplie...

  • (GCMS) Checking for Leaks

    This section describes how to check the instrument for leaks. If an air leak is clearly present, locate the leak as described in "2. MS Vacuum Leak Check", and take corrective measures.   Operating...

  • (GCMS) Measurres durriing Powerr Outtage

    For power outages of 10 minutes or less, normally no particular procedure is required as long as the [Vacuum Restart Mode] checkbox is selected in the window opened by clicking [Option] in the [Vac...

  • (GCMS) Chemical Ionization(CI, NCI) Reagent Gas Setup Procedure

    Parts Number: 225-07776-91 PLUG KIT (Used as a stop cock when no reagent gas is used.)   1. Automatically stop the GC/MS in GCMSsolution.  Wait until the "system shutdown" message appears      2. C...

  • (GCMS) Chemical Ionization(CI, NCI) When signal intensity is extremely weak

    Cause 1: Either the cylinder valve of the gas cylinder, the control knob of the pressure regulator, or the outlet valve of the pressure regulator is closed. The solenoid valve of the reagent gas co...

  • (GCMS) Chemical Ionization(CI, NCI) When good spectrum cannot be obtained

    There may be leak of the reagent gas if the conditions are not satisfied in spite that the reagent gas is being supplied. The possible causes of leak are described below: Cause 1: Either the cylind...

  • (GCMS) DI-2010 Operation Methods

    Parts Number: 225-04209  CAP (Cap for the DI-2010) 225-09344  SAMPLE VIAL 100 PC (DI-2010 standard accessories (10 pcs):Sample pot made of Pyrex) 225-09470  QUARTZ VIAL (10pcs.)   The following des...