Gas Selector# purge control irregularity[Error Code:2360 2361]
The gas selector does not operate correctly. Check that the supply pressure to the gas selector is within the range below. Check that the gas supply pressure is stable. If the error appears ...
-
Gas Selector# main control irregularity[Error Code:2358 2359]
Did not operate correctly. Check that the set pressure for the gas selector is lower than the supply pressure. Check that the supply pressure to the gas selector is within the range below....
-
PURn was overpressure[Error Code:2218 2219 2220 2221]
Excessively high pressure is applied to the purge line. Check that the purge line is not blinded. Do not open or close the main valve of the supply gas during AFC operation. Restart the GC wi...
-
CARn was overpressure[Error Code:2214 2215 2216 2217]
Excessively high pressure is applied to the injection port. These errors may occur when supply gas pressure is changed during AFC operation. Do not open or close the main valve of the supply gas d...
-
CARm AFC set point not reached[Error Code:2130 2131 2204 2205]
The pressure or flow rate may have not reached the set value due to the following reasons. The gas supply pressure is insufficient. There is a leak at tubing connections. The resistance of supply ...
-
Analysis waiting[Error Code:5112]
The analysis was started while the PREP RUN function is active and the instrument is in the waiting status. Set [Prep Run Wait Time] longer.The PREP RUN is activated when using the following f...
-
CARm calc. ratio out of range[Error Code:5075 5076 5108 5109]
The carrier gas total flow calculated from split ratio program, etc. is outside the set range. Change the split ratio so that the following equation is satisfied. (Total flow) = (Column flow) +...
-
PURn calc. prss out of range[Error Code:5042 5043 5044 5067 5068 5069 5103 5104]
The inputted purge flow is outside the set range. Set the purge flow within the AFC control range.
-
DET#n FTD overcurrent detected[Error Code:4105 4106 4107 4108]
Detector current is abnormal. If the resistance of the filament becomes abnormally high when the FTD is in use, the protection circuit is actuated to prevent damage to the filament. Turn off t...
-
DET#n TCD overcurrent detected[Error Code:4101 4102 4103 4104]
Detector current is abnormal. If the resistance of the filament becomes abnormally high when the TCD is in use, the protection circuit is actuated to prevent damage to the filament. Turn off t...