DetAPCn ECDG was overpressure[Error Code:2268 2269 2270 2271]
ECD detector gas flow exceeded the specified value. These errors may occur when supply gas pressure is changed during APC operation. Do not open or close the main valve of the supply gas during ...
-
DetAPCn DCG was overpressure[Error Code:2264 2265 2266 2267]
BID detector gas flow exceeded the specified value. These errors may occur when supply gas pressure is changed during APC operation. Do not open or close the main valve of the supply gas during ...
-
DetAPCn air was overpressure[Error Code:2224 2227 2230 2233]
Air flow exceeded the specified value. These errors may occur when supply gas pressure is changed during APC operation. Do not open or close the main valve of the supply gas during APC operation...
-
DetAPCn hydrogen was overpressure[Error Code:2223 2226 2229 2232]
Hydrogen flow exceeded the specified value. These errors may occur when supply gas pressure is changed during APC operation. Do not open or close the main valve of the supply gas during APC operat...
-
DetAPCn makeup was overpressure[Error Code:2222 2225 2228 2231]
Makeup flow exceeded the specified value. These errors may occur when supply gas pressure is changed during APC operation. Do not open or close the main valve of the supply gas during APC operat...
-
DAMCn reference set point not reached[Error Code:2276 2277 2278 2279]
The pressure or flow rate may have not reached the set value due to the following reasons. The gas supply pressure is insufficient. There is a leak at tubing connections. The resistance of suppl...
-
DAMCn makeup set point not reached[Error Code:2272 2273 2274 2275]
The pressure or flow rate may have not reached the set value due to the following reasons. The gas supply pressure is insufficient. There is a leak at tubing connections. The resistance of supp...
-
DetAPCn makeup valve not completely closed[Error Code:2057 2060 2063 2066]
There is a leak inside the makeup gas control valve of the APC for the detector. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.
-
CARx purge valve not completely closed[Error Code:2048 2049 2050 2202]
There is a leak inside the purge flow control valve of the AFC. Check that the purge line outlet is not blinded. If the error appears repeatedly even after checking the above, turn off the main...
-
Gas Selector# sub control irregularity[Error Code:2362 2363]
Did not operate correctly. Check that the set pressure for the gas selector is lower than the supply pressure. Check that the supply pressure to the gas selector is within the range below. Ch...