Rob Stimpson
- Total activity 889
- Last activity
- Member since
- Following 0 users
- Followed by 0 users
- Votes 0
- Subscriptions 444
Articles
Recent activity by Rob Stimpson-
Pre-Run program time error [Error Code:4304]
Program time is longer than the upper limit "9999.99 min". Do not set the total time longer than 9999.99 min. The analysis can be performed even after this message is displayed, however, the...
-
Clean up program time error [Error Code:4303]
Program time is longer than the upper limit "9999.99 min". Do not set the total time longer than 9999.99 min. The analysis can be performed even after this message is displayed, however, the progr...
-
Clean up program start error [Error Code:4305]
Clean up program cannot be started because oven temperature is not ready. Set the initial temperature of the column oven at higher values than room temperature by more than 2℃. (Example, Set the ...
-
COL CRG use time warning [Error Code:9000]
Exceeded the set threshold of CRG coolant consumption (Operating Time). Replace the coolant and then reset the counter. <GC instrument monitor setting> Press HOME button > [Function] > [Diagno...
-
Fan use time warning [Error Code:9002]
Usage time exceeds the set value. Risk of current leakage, heat, smoke, and ignition increases. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/serv...
-
CARx septum counter warning [Error Code:9004 9005 9006 9052]
Exceeded the set threshold of inlet septum counter. Replace the septum and then reset the counter. Counter reset: <GC instrument monitor setting> Press HOME button > [Function] > [Diag...
-
CARx insert counter warning [Error Code:9007 9008 9009 9053]
Exceeded the set threshold. Replace the insert and then reset the counter. Counter reset: <GC instrument monitor setting> Press HOME button > [Function] > [Diagnosis] > [Analysis counte...
-
DET#n ignition failed. [Error Code:4203 4204 4205 4206]
FID or FPD ignition failed. Check the following items. Detector gas flow rate: Press (HOME) - [Detector] button on GC instrument monitor (HOME) and display the folloeing screen. Make sure that...
-
Ignition finished(retried) [Error Code:9011]
Detector ignition sequence was re-attempted, because the ignition failed initially. Check the following items. ・Check for gas leaks. ・Verify the gas flows. ・Diagnose ignition pulse and...
-
xxx sensor use time warning [Error Code:9012 9013 9014 9015 9016 9017 9018 9019 9024]
Risk of current leakage, heat, smoke, and ignition increases because usage time exceeds the set value. Turn off the main power of the instrument, stop gas supply, and contact your Shimadz...