Rob Stimpson
- Total activity 889
- Last activity
- Member since
- Following 0 users
- Followed by 0 users
- Votes 0
- Subscriptions 444
Articles
Recent activity by Rob Stimpson-
Gas Selector# PCB error [Error Code:91 92]
PCB may have failed. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement.
-
TC was broken [Error Code:79]
Thermocouple for overheat protection may be broken. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacemen...
-
Hydrogen sensor init error [Error Code:77]
Hydrogen sensor may have failed. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement.
-
Atm. press is out of range [Error Code:6]
If this error is displayed when the actual room temperature and air pressure is within the operation range, sensors installed on the instrument may have failed. Turn off the main power of the i...
-
Room temperature is out of range [Error Code:5]
If this error is displayed when the actual room temperature and air pressure is within the operation range, sensors installed on the instrument may have failed. Turn off the main power of ...
-
DC24V is out of range [Error Code:2]
The power source may have failed. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement.
-
DC5V is out of range [Error Code:1]
The power source may have failed. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement.
-
Clock is initialized [Error Code:29]
Built-in clock is reset. Battery for the clock is depleted or the hardware fails. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service repres...
-
COL A/D converter error [Error Code:30]
A/D converter may have failed. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement.
-
Room temp, Atm. press A/D error [Error Code:31]
A/D converter may have failed. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement.