Program time error [Error Code:4302]
Program time is longer than the upper limit "9999.99 min". Do not set the total time longer than 9999.99 min.
-
Settings were changed [Error Code:4301]
Setting is changed while program is running. If you changed the part which is not executed, the instrument continues the analysis according to the changed program. Check the setting.
-
DET1 2 3 4 BID short-circuit in discharge cell [Error Code:4129 4130 4131 4132]
Output from plasma generation circuit of BID shorted out. Turn off the main power and then turn on the instrument again. If the error occurs repeatedly, cables may be damaged. Turn off the mai...
-
DAMCn makeup control irregularity [Error Code:2087 2088 2089 2090 2094 2095 2096 2097]
Component does not operate correctly. Check that the supply pressure to each unit is within the range below. Check that the gas supply pressure is stable. If the error appears repeatedly ...
-
xxx temperature controller error [Error Code:1037 1038 1039 1040 1041 1042 1043 1044 1057]
Temperature control circuit has a problem. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.
-
Overheat is sensed [Error Code:1036]
Temperature control circuit has a problem. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.
-
SCD ozonier voltage fell below lower limit [Error Code:6123]
The voltage of the ozone generator is lower than the lower limit. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for p...
-
SCD ozonier voltage was exceeded upper limit [Error Code:6122]
The voltage of the ozone generator exceeds the upper limit. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replac...
-
SCD fan error [Error Code:6121]
The cooling fan does not operate correctly. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement.
-
TCD signal was out of range [Error Code:4201]
Zero adjustment cannot be performed because the TCD baseline is unstable. The detector controller unit may have failed. Perform zero adjustment again after the TCD baseline becomes stable. If it do...