CARn hydrogen flowing over the specified value [Error Code:2296 2297 2298 2299]
There is a program that requires hydrogen gas flow more than the specified value, or the component does not operate correctly. Ensure that the set values in programs such as Column Oven Tempera...
-
CARm AFC control irregularity [Error Code:2134 2135 2136 2137 2208 2209 2210 2211]
Each unit does not operate correctly. Check that the supply pressure to the AFC is within the range below. Check that the supply pressure is sufficient for the AFC gas to flow. Che...
-
Hydrogen sensor is not installed [Error Code:78]
Hydrogen sensor is set to be used although hydrogen sensor is not connected - When not using hydrogen sensor, set hydrogen sensor at [N/A]. The hydrogen sensor may be not recognized - When using ...
-
SCD AD converter initialization was failed [Error Code:6145]
A/D converter may have failed. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement.
-
SCD detector leak error [Error Code:6144]
The pressure of the detector part is high. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. ...
-
SCD furnace leak error [Error Code:6143]
The pressure of the furnace is high. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. ...
-
SCD H2 solenoid valve has been fixed to ON [Error Code:6142]
The solenoid valve cannot be closed. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement. ...
-
SCD thermistor error [Error Code:6141]
The thermistor may have failed. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement.
-
SCD thermistor temperature exceeds the limit [Error Code:6135]
The thermistor detected abnormal high temperature. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replac...
-
SCD PSU-24V error [Error Code:6124]
The power source may have failed. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement.