Flow# command stopped due to error[Error Code:4021 4022 4023 4024 4025 4026]
A command is sent to the flow controller while the flow controller has an error. When any of these errors occurs, another error should have occurred regarding the flow controller. Check th...
-
AOC link error[Error Code:4013]
Communication with AOC failed or an error occurred during communication. These errors are issued not only when a communication failure or device malfunction occurs, but when the cable is unplugge...
-
AOC2 transmission error[Error Code:4012]
Communication with AOC failed or an error occurred during communication. These errors are issued not only when a communication failure or device malfunction occurs, but when the cable is unplugge...
-
AOC time out[Error Code:4011]
Communication with AOC failed or an error occurred during communication. These errors are issued not only when a communication failure or device malfunction occurs, but when the cable is unplugge...
-
AOC set value out of range[Error Code:4010]
Communication with AOC failed or an error occurred during communication. These errors are issued not only when a communication failure or device malfunction occurs, but when the cable is unplugged ...
-
AOC command is invalid[Error Code:4009]
Communication with AOC failed or an error occurred during communication. These errors are issued not only when a communication failure or device malfunction occurs, but when the cable is unplugge...
-
AOCm syringe counter warning [Error Code:9059 9060]
Exceeded the set threshold. Replace the syringe and then reset the counter. <GC instrument monitor setting> Press HOME button > [Function] > [Diagnosis] > [Analysis counter], and then di...
-
Temp. is not controlled [Error Code:9020]
Ignition or illumination was started when the GC did not start temperature control. Set [Start Temp/Det] at [Enable] on [GC Start Sequence] screen, start the GC, and then start ignition or illu...
-
PURn is not controlled [Error Code:9055 9056 9057 9058]
The analysis was started when the purge control was set to [Off]. Set [Septum Purge] to [On] on [Purge Flow Program] screen, and then start the analysis. <GC instrument monitor setting> ...
-
SCD furnace temp sensor warn. [Error Code:9075]
The count exceeded the set threshold value. The risk of current leakage, heat, smoke, and ignition increases. Turn off the main power of the instrument, stop gas supply, and contact your Shimadz...