DET#1 2 3 4 FTD adjustment failed [Error Code:4149 4150 4151 4152]
It cannot be adjusted because FTD power (%) is [0] or [100]. Check that connection of electrode connector is correct and collector fixing knurled screws are tightened. When connections are correc...
-
DET#1 2 3 4 DTCD overvoltage detected in cell power [Error Code:4133 4134 4135 4136]
The voltage of the PTCD cell power is abnormal. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement.
-
DET#1 2 3 4 FPD overcurrent detected [Error Code:4125 4126 4127 4128]
FPD photomultiplier current is abnormal. Ensure that strong light does not leak in photomultiplier current from the connection.If the error cannot be resolved after you cut off the light with black...
-
APCa-b PCB error [Error Code:23 24 25 26 27 28]
PCB may have failed. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement.
-
DetAPCn PCB error [Error Code:13 14 15 16]
PCB may have failed. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative to ask for part replacement.
-
APCn hydrogen flowing over the specified value[Error Code:2304 2305 2306 2307 2308 2309 2310 2311 2312 2313 2314 2315 2316 2317 2318 2319 2320 2321]
Hydrogen flow exceeded the specified value. These errors may occur when supply gas pressure is changed during APC operation. Do not open or close the main valve of the supply gas during APC operat...
-
DetAPCn DCG control irregularity[Error Code:2152 2153 2154 2155]
The flow of the flow controller to control the BID detector gas is unstable. Check that the BID detector gas supply pressure is within the range below. Check that the supply pressure is suf...
-
DetAPCn DCG valve not completely closed[Error Code:2148 2149 2150 2151]
The sweep gas flow supplied from the bypass line is unstable, or there is a leak inside the BID gas control valve of the APC for the detector. Close the knob of the bypass unit of BID totally ...
-
CARx prim is out of range[Error Code:2138 2139 2140 2212]
The carrier gas supply pressure is incorrect. Check that the carrier gas supply pressure is within the primary pressure range in the configuration described below. An error will occur if the ...
-
APCx was overpressure[Error Code:2234 2235 2236 2237 2238 2239 2240 2241 2242 2243 2244 2245 2246 2247 2248 2249 2250 2251]
Excessively high pressure is applied to the AUX-APC. These errors may occur when supply gas pressure is changed during AUX-APC operation. Do not open or close the main valve of the supply gas...