(GC) RS232 Connection Troubleshooting with GC-2010
You should be able to determine the COM port it's connected to in Device Manager within Windows. From there, in the LabSolutions RealTime software go to the assistant toolbar, Main -> System Config...
-
(GCMS) Automatic Library Search in GCMSSolution PostRun - Registry File Toggle
The function to automatically library search peaks is not switched on by default. With PostRun closed, in the C: Drive > GCMSSolutions > Regist please double click the file "AutoSpcTableLibSearch_O...
-
(GCMS) Creating/Adding to a Private Library in GCMSSolutions
Adding to the Library Data Analysis Load spectra Tab or side bar > Library Editor Right click on spectra > Target Mode – Loads new spectra File > Create private library File > Open libra...
-
(GCMS) Prep Run Wait Time inGCMSSolution
When the system becomes 'stuck' in Prep Run mode (e.g. manual sampling and the time between samples is longer than the system expects), we can increase the time to stop the system from defaulting t...
-
(GCMS) NIST Library: Reset Library Files within GCMSSolution
When the library files are defaulting to e.g. NIST08 and we want to force the system to use the new libraries, please follow these steps: Ensure that all .lib files are saved in the C:\GCMSsolutio...
-
(GC/GCMS) AOC-6000 Plus in Post Run Mode
AOC-6000 Plus stuck in Post Run Mode When the AOC-6000 Plus remains in ‘Post-run’ and doesn’t enter ‘Ready’ mode, please perform a full power cycle in the following order: shut down the instrument,...