1. Remove the inlet tube fixing screw.
2. Loosen the outlet tube fixing screw.
NOTE: To Loosen a little is enough. Do not remove it.
NOTE: To Loosen a little is enough. Do not remove it.
3. Remove the reactor fixing screw.
4. Lift up the guide (1) and remove the upper chamber (2).
NOTE: The lower chamber (3) does not have to be removed.
5. When placing the upper chamber on a desk or other surface, place a spacer (4) on the desk first and then place the upper chamber on top.
NOTE: The lower chamber (3) does not have to be removed.
5. When placing the upper chamber on a desk or other surface, place a spacer (4) on the desk first and then place the upper chamber on top.