In the event of an error, LabSolutions may report an error code. Please quote this error code when reporting the issue to Shimadzu. A description of each error code is listed below:
Additional LabSolutions Error Codes (0100 - 2958)
Code Message Reason Action
3006 |
Difference between Start and End Scan Wavelength is less than Scan Step. |
----- |
----- |
3007 |
Start Scan Wavelength is over final value. |
----- |
----- |
300C |
Please set the 'STOP' command. |
----- |
----- |
3062 |
The LC Stop Time should be larger than the maximum time of the FRC Time Program.\nPlease enter an enough time in the LC |
----- |
----- |
3020 |
Memory Error! Cannot read the 'Sample Pre-treatment' program. |
A significant error occurred. There may be insufficient memory. |
Close all unnecessary applications. |
3021 |
Cannot use this function. |
The specified function cannot be used for current instrument configuration. |
Specify the appropriate function. |
3022 |
Upper/lower limit value error. Please set a value within the range %g to %g. |
----- |
----- |
3023 |
Cannot use this command in the system._\nThis command is skipped automatically. |
The specified function cannot be used for current instrument configuration. |
Specify the appropriate function. |
3024 |
Please set the 'STOP' command after xxx min. |
----- |
----- |
3025 |
Please set the 'LC Program' command between 0.01 and 9999.9 min. |
----- |
----- |
3026 |
Cannot use this unit. Select another unit. |
Specify the instrument which does not exist in hardware configuration. |
Cannot specify the instrument which does not exist in hardware configuration. If there exists the instrument in the hardware configuration, confirm the connection between instruments or instrument and PC. |
3027 |
Invalid value! Please set the value only %g. |
----- |
----- |
3030 |
Syntax error! Please check the command parameters. |
----- |
----- |
3031 |
Value is out of range. |
----- |
----- |
3032 |
Cannot use this function. |
The specified function cannot be used for current instrument configuration. |
Specify the appropriate function. |
3033 |
Memory error! Cannot read the 'Sample Pre-treatment' program. |
A significant error occurred. There may be insufficient memory. |
Close all unnecessary applications. |
3034 |
The 'FOR' command loop is too deep to compile. |
----- |
----- |
3035 |
Cannot find end of the 'FOR' command. |
----- |
----- |
3036 |
This error code is not defined. |
A significant error occurred. |
Restart the PC and try again. If the attempt failed, reinstall the application. |
3037 |
Invalid value! Please set a value within the range xxx to yyy. |
----- |
----- |
3038 |
The program is not set to the specified start page. Set the program. |
----- |
----- |
3040 |
Memory error! Cannot read the 'Sample Pre-treatment' program. |
A significant error occurred. There may be insufficient memory. |
Close all unnecessary applications. |
3041 |
Cannot use this function. |
The specified function cannot be used for current instrument configuration. |
Specify the appropriate function. |
3042 |
Upper/lower limit value error. Please set a value within the range %g to %g. |
----- |
----- |
3043 |
Please set the command within the range %g min to %g min. |
----- |
----- |
3044 |
Please set command within LC Stop Time. |
----- |
----- |
3050 |
[Switch Interval Time] must be greater than [End Time(xxx) + Injection Delay Time(yyy)]. |
In the setting of CDD detector, the summation of End Time and Injection Delay Time is larger than Switch Interval |
In data Acquisition tab and Detector tab in the Instrument Parameter View, these valuesmust be specified as described in message. |
3051 |
Cell temperature upper/lower limit value error. Set the setting value to withinmin. to max. degree. When the cell temperature is automatically set, set the oven temperature to -3degree. of this range. |
----- |
----- |
3052 |
Incorrect combination of the values of total capacity, concentration, flow, switch interval time and regeneration time. |
Incorrect combination of the values of total capacity, concentration, flow, switch interval time and regeneration time. |
Total capacity(C_VOL), concentration of mobile phase(MPCNC), flow(FLOW), switch interval time(SINTVL) and regeneration time(REGENTM) should fill the next formula \n(formula) A*C_VOL*MPCNC*FLOW*SINTVL/REGENTM<=180 mA (regeneration current)\nA: When cartridge type is 206 6, |
3053 |
The suppressor switching cycle of the CDD detector or the flow rate of the solvent delivery pump for suppressor cartridge regeneration is incorrectly set. |
----- |
----- |
3055 |
The suppressor switching cycle of the CDD detector or the flow rate of the solvent delivery pump for suppressor cartridge regeneration is incorrectly set. |
The setting of The suppressor switching cycle of the CDD detector or the flow rate of the solvent delivery pump for suppressor cartridge regeneration is not suitable. |
Total capacity(C_VOL), concentration of mobile phase(MPCNC), flow(FLOW), switch interval time(SINTVL) and regeneration time(REGENTM) should fill the next formula \n(formula) A*C_VOL*MPCNC*FLOW*SINTVL/REGENTM<=180 mA (regeneration current)\nA: When cartridge type is 206 6, |
3056 |
CDD detector regeneration time upper/lower limit value error. Change [Regenerate Time] or [Switching Interval Time] on the detector tab page, or change the flow rate of the solvent delivery pump for suppressor cartridge regeneration. |
CDD detector regeneration time upper/lower limit value error. |
Total capacity(C_VOL), concentration of mobile phase(MPCNC), flow(FLOW), switch interval time(SINTVL) and regeneration time(REGENTM) should fill the next formula \n(formula) A*C_VOL*MPCNC*FLOW*SINTVL/REGENTM<=180 mA (regeneration current)\nA: When cartridge type is 206 6, 312 3 |
3057 |
When the concentration setting of the CDD detector suppressor cartridge is set to "0", the flow rate of the solvent delivery pump for suppressor cartridge regeneration is set. Correctly set the CDD detector suppressor cartridge concentration, or change the flow rate to "0". |
----- |
----- |
3058 |
Incorrect combination of the values of flow switch interval time and regeneration time. |
Incorrect combination of the values of flow switch interval time and regeneration time. |
Flow(FLOW),_switch interval time (SINTVL) and regeneration time (REGENTM) should fill the next formula.\n(formula)_REGENTM<=SINTVL-1/FLOW |
3060 |
Program is not correct. |
There is a rate or value error in the Time Program or Flow Program. |
Check the values and correct the error. For example, if a positive rate value is set, but the oven temperature is set to a lower value that the current value, the temperature rate cannot be positive. |
3205 |
The End wavelength must be greater than Start wavelength. |
Invalid values are set to the acquisition wavelength range. |
Set Start wavelength and End wavelength correctly. |
3206 |
The End time must be greater than Start time. |
Invalid values are set to the acquisition time range. |
Set Start time and End time correctly. |
3209 |
Cannot perform background compensation because wavelength table does not match. |
Wavelength table does not match between 2 PDA data files. |
Use data acquired by the same PDA detector to perform background subtraction. |
320B |
Multi-chromatogram table of this file includes wavelength out of the stored wavelength range. |
Multi-chromatogram table of this file includes wavelength out of the stored wavelength range. |
Set the Multi-chromatogram table correctly. |
3219 |
The wavelength calibration results are not correct. A probable cause is that the optical system has not been correctly |
The result of having performed wavelength calibration was out of the range of the element. |
Re-adjust the optical system. |
4000 |
Selected library is a public library. |
No change including writing can be made on public libraries. |
Specify a private library. |
4001 |
Now searching. The operation is not available for now. |
Libraries are searched. |
Wait for a while. After search finishes, try the same operation again. |
4002 |
No structural formula is registered. |
The structural formula has not been registered to the library. |
The structural formula of the compound cannot be displayed or printed. |
4004 |
Search parameters are not set. |
Search parameters for index search have not been set. |
Set search parameters on the Index Search Parameters Table in the [MS Library Editor] sub-window. |
4005 |
There is no compound hit. |
There was no compound that met the search conditions. |
The message is for confirmation. |
4006 |
No search was performed. |
Since library search was not performed on spectra registered on the Spectrum Process Table or Identified Results Table, no results could be displayed. |
Perform library search. |
4007 |
Ratio between values before and after hydrogen compensation should be between 0.5 and 2.0. |
Hydrogen compensation can be set up to x2. |
Adjust values before and after the compensation. The values normally should be within this range. |
4009 |
Either m/z value or intensity is set to 0. The spectrum peak will be deleted. Do you want to |
----- |
----- |
4011 |
The start m/z value is set larger than the end m/z value. |
The start m/z value is set larger than the end m/z value. |
Set the start m/z value smaller than the end m/z value. |
4012 |
Further search for subtraction result spectra is impossible. |
The number of repeated subtraction for library search has exceeded the specified limit. |
----- |
4013 |
Spectra in the library cannot be copied to other libraries. |
Spectra included in the library are protected so that they cannot be copied to other libraries. |
----- |
4014 |
The selected compound will be deleted from this library. Deleted a compound cannot be recovered. \nAre you sure you want to continue? |
----- |
----- |
4015 |
Select library file. |
The library function is not selected. |
Check the condition of the library search. |
4080 |
The tuning file has not been tuned according to conditions specified in the analysis method. Therefore, perform auto tuning. |
The tuning file may not have been tuned according to the specified interface, polarity, or data acquisition mode. |
Create a new turning file in the [Tuning] sub-window and then perform auto tuning. |
4081 |
Mass compensation of Q-array, DL, or deflector voltage has not been performed. |
----- |
----- |
4082 |
Tuning by Quick scan is not performed. |
Tuning by [Quick scan] is not performed even though [Quick scan] is set for [MainRod Bias] of MS instrument parameters. |
Set [Scan Mode] for [Main-Rod Bias] on the [H/W Parameters] tab page in the [Manual Tuning] sub-window. If necessary, click the [Detail] button, make settings, and perform tuning. Then, save the tuning file. |
4083 |
Flow rate in the tuning file is different between positive and negative polarities. Perform autotuning or manual tuning. |
Nebulizing gas flow rate is different between tuning with the positive polarity and tuning with the negative polarity. |
Perform tuning with the same flow rate for positive and negative polarities. |
4084 |
Tuning file version is different. \n Perform auto-tuning or manual tuning. |
An attempt was made to use a tuning file created with a different MS type. |
Perform auto-tuning or manual tuning. |
4085 |
Temperature and flow rate in the tuning file are different for positive and negative polarities. \n Perform auto-tuning or manual tuning. |
The polarity cannot be switched during data acquisition because the temperature and nebulizing gas flow rate are different for positive and negative polarities. This check is performed when the nebulizing gas flow rate during data acquisition is set to be determined by the tuning result. (Setting is made on the [Interface] tab page in [Instrument Parameter View] in the |
Enter correct values in the [Manual Tuning] sub-window. |
4086 |
A type of instrument on which tuning was performed is different than the current instrument. Perform tuning again. |
You are attempting to use a tuning file tuned for a different type of instrument. |
Perform tuning again for the current instrument. |
4100 |
Perform optimization when the acquisition mode is set to SIM. |
Acquisition mode is not set to SIM. Method optimization can be performed only when the acquisition mode is set to SIM. |
When performing method optimization, set the acquisition mode to SIM. |
4101 |
Set up to two events for one segment. |
Three or more events are set for one segment. Method optimization can be performed when the number of events is two at max. |
Set up to two events for one segment. |
4102 |
Setting of the MS instrument parameters does not match the subject method. Do you want to apply the parameters? \n When they are applied, the existing MS instrument parameters for the subject method will be deleted. |
For the optimized method and subject method (method open in the [Data Acquisition] sub-window), the number of segments, the number of events, or the combination of channel number and m/z value does not match. |
Select [Yes] to replace the MS instrument parameters for the subject method with the MS instrument parameters for the optimized method. Select [No] otherwise. |
4103 |
MS Instrument is not registered in the Configuration File. |
----- |
----- |
4104 |
Event time is not enough.Please change SIM channel number or long event time. |
The event time value is below the minimum limit. |
Decrease the number of SIM channel or increase the event time. |
4105 |
Please change event time less or equal to 655sec. |
The event time specified exceeds the allowable range. |
Use a lower time setting within 655 sec. |
4106 |
MS acquisition time cannot be shortened. |
The acquisiton time cannot be shortened. |
Please do not input short time at former acquisition time. |
4107 |
Method file is not optimized. |
An attempt was made to display the optimization results of a method file that has not been optimized. |
Optimize the file before displaying the results. |
4108 |
Method file including the Survey/Dependent event can not be optimized. |
The method file could not be optimized since it contains a Survey Event or Dependent Event. |
To optimize the method file, delete the Survey Event or Dependent Event from the method. |
4109 |
Method file including the AD output can not be optimized. |
The method file could not be optimized since the AD output is set to the method |
To optimize the method file, delete the AD output from the method. |
410A |
Precursor m/z dose not match the Condition of auto selection for product m/z. |
The automatic selection conditions for the product m/z contradict the precursor m/z settings. |
Set an appropriate precursor m/z or automatic selection conditions for the product m/z.\nWhen "Optimize MRM event from precursor ion search" mode is selected, precursor m/z is defined by molecular weight and adduct |
410B |
The maximum number of events is exceeded. Please decrease the number of method files to be |
All events cannot be set since there are too many events set to the method files to be added. |
Reduce the number of events by reducing the number of method files to be added. |
410C |
The maximum number of compounds is exceeded. Please decrease the number of method files to be added. |
All compounds cannot be set since there are too many compounds set to the method files to be added. |
Reduce the number of compounds by reducing the number of method files to be added. |
410D |
The MRM event cannot be added by the method with the analog output. Please delete the analog |
The MRM event cannot be added since analog output is set. |
Delete analog output before adding the MRM event. |
410E |
Time cannot be adjusted by the method with the analog output. Please delete the analog output. |
The time cannot be adjusted since analog output is set. |
Delete the analog output. |
410F |
The maximum number of compounds is exceeded. Please review the compound table. |
All compounds cannot be set since there are too many. |
Check the Compound Table. |
4110 |
Failed to compress data. Data has been saved by nocompression. |
Because it was not possible to preserve it by compressing the data file, it preserved it without compressing it. |
----- |
4200 |
This file has the parameters tuned with the current tuning conditions. Do you invalidate parameters and change the |
This message is displayed when the parameter of tuned file is changed. |
Do not change the parameters, if you don't want to invalidate the tuning result. |
4201 |
Use the file as the default tuning file? |
----- |
If this file is specified as the default tuning file, it is used as the tuning file whenever a tuning file is not specified for single run or realtime batch. Click [Yes] to use it as the default tuning file. Otherwise, click [No]. |
4202 |
Use the file as the default tuning file? |
The message is displayed when deleting the currently selected monitor group. |
Click [OK] to delete and [Cancel] not to delete. |
4203 |
The group name is already registered. |
Registration of a new monitor group was attempted under a name already |
Change the monitor group name. |
4204 |
Groups are already registered to the maximum (16). |
Up to 16 monitor groups including 2 standard groups can be registered. |
Delete unnecessary monitor group and register new one. |
4205 |
The default setting group cannot be deleted. |
Deletion of a standard monitor group was attempted. |
The default monitor group cannot be deleted. |
4206 |
The Calibration Table will be supplemented. \nDo you |
The Calibration Table must be supplemented since it has been changed. |
To supplement, click [Yes], and to cancel supplementation, click [No]. |
4207 |
The Calibration Table will be initialized. \nAre you sure you want to perform? |
This message is displayed for confirmation before the Calibration Table is initialized. |
To initialize, click [Yes], and to cancel initialization, click [No]. |
4208 |
Do you want to save peak profiles and spectra of the standard sample in this setting? |
This message is for confirming whether or not to save the profile and spectra of the standard sample under the preset |
To save, click [Yes], and to cancel the save, click [No]. |
4209 |
To perform Auto-Tuning, use ESI or DUIS interface. |
Auto tuning cannot be performed when an APCI (APPI) interface is mounted. |
Change the interface to the one that auto-tuning can be performed with. |
420A |
The Setting of EEPROM was transmitted. |
----- |
----- |
420B |
[untranslated]Because this tuning file have not been tuned on positive polarity, you can't make the conditions to tune only negative polarity. \nPlease check positive polarity also. |
Auto-tuning cannot be performed using negative settings only when tuning with positive ion polarity has not been completed. |
Perform tuning by positive polarity in the manual tuning mode, or specify tuning by auto-tuning. |
420C |
To perform Edit Mass for Negative Calibration, Use ESI or DUIS interface. |
Edit Mass for Negative Calibration cannot be performed, because the interface is set as APCI by configuration. |
Change the interface to ESI or DUIS. |
420D |
Auto-Tuning Started.(Discriminative Level) |
----- |
----- |
420E |
Auto-Tuning has been completed.(Discriminative Level) |
----- |
----- |
420F |
Auto-Tuning was terminated.(Discriminative Level) |
----- |
----- |
4210 |
Auto-Tuning could not complete.(Discriminative Level) |
----- |
----- |
4211 |
The extended parameter was transmitted. |
----- |
----- |
4212 |
The low m/z value is set larger than the high m/z value. |
The low m/z value is set larger than the high m/z value. |
Set the low m/z value smaller than the high m/z value. |
4213 |
The product m/z value is set larger than the precursor m/z |
The product m/z value is set larger than the precursor m/z value. |
Set the product m/z value smaller than the precursor m/z value. |
4214 |
The losses of value is set larger than the precursor m/z value. |
The losses of value is set larger than the precursor m/z value. |
Set the losses of value smaller than the precursor m/z value. |
4215 |
The influence goes out to Scan Speed in the [MS] tab page when Settling Time is changed. Do you wish to continue?\nThe [MS] tab page, click [Normal] or [Advanced] in the [Instrument Parameters View]. |
Settling Time is changed. |
To download, click [Yes], and to cancel download, click [No]. |
4216 |
The influence goes out to Loop Time in the [MS] tab page when Polarity switch waiting time is changed. Do you wish to continue?\nThe [MS] tab page, click [Normal] or [Advanced] in the [Instrument Parameters |
Polarity switch waiting time is changed. |
To download, click [Yes], and to cancel download, click [No]. |
4217 |
Please activate the voltage value only 2 points. |
----- |
----- |
4218 |
Please set the m/z calibration value so as to increase |
----- |
----- |
4219 |
To perform Auto-Tuning, Use ESI probe. |
----- |
Change the interface to the one that auto-tuning can be performed with. |
421A |
To perform Edit Mass for Negative Calibration, Use ESI probe. |
Edit Mass for Negative Calibration cannot be performed, because the interface is set as APCI by configuration. |
Change the interface to ESI. |
421B |
Fail to initialize a tuning file. |
Factory Setting is not set. |
Create a new tuning file. If the phenomenon reoccurs, set Factory Setting with Instrument Maintenance Tools again. If the phenomenon still reoccurs, contact your Shimadzu representative. |
4A15 |
Number of Reference Ions exceeds five |
Up to 5 reference ions can be entered. |
Decrease the number of reference ions in the table. |
4A16 |
Please select target ion. |
No target ion has been set. |
The target needs to be set in all occasions. Change [Type] of the target ion. |
4A2C |
Compounds are already registered to the maximum. |
Up to 1,000 compounds can be registered. |
Delete unnecessary compounds and then register new ones. |
4A2E |
Peak integration for quantitation cannot be performed. |
----- |
---- |
4A31 |
Incorreect retention time is set. |
The retention time is invalid. |
Enter the correct RT. |
4A34 |
Need to set m/z. |
There is no compound between start and end time. |
Change end time to click on chromatogram, |
4A35 |
The number of channel in a line of SIM table will be over 64 by changing the end time. \nYou should set the end time so that the number of channel will be less than or equal to 64. |
The changed end time would result in the number of channels included on a single row of the SIM table to exceed 64. |
If the SIM table is automatically created, all the compounds cannot be registered on the channels. Change the end time so that the number of channels is 64 maximum. |
4A36 |
The number of channel in a line of SIM table will be over 64. You had better set the SIM table again. |
When the MS parameters were created automatically from the Compound Table, the number of ion channels exceeded the maximum. |
Either decrease the number of compounds and reference ions, or specify processing start and end times in the Compound Table so that the processing times of individual compounds do not overlap. |
4A37 |
The compound that retention time exceeds the acquisition time of the method file is not reflected in SIM table. |
The retention tme of target compound is out of qcquisiton time range. |
Revise the settings of compound table or change the acqusition time. |
4A38 |
The compound whose event number did not exist on the method file has been deleted. |
An event# that is bigger than oner that exists in origical MS instrument parameter is registered in the compound table. |
Confirm all of event# in the compound table. |
5100 |
Communication error. |
An error occurred at the transfer protocol level. |
Make sure that the analytical instrument is not turned OFF, and that connection cables is not pulled out. \nContact your Shimadzu representative when this problem can not be |
5101 |
GC instrument error occurred. |
An error was notified from the GC instrument. |
This error is accompanied by a message sent from the GC instrument. Check the content of this error message and remedy the error according to the Instruction Manual provided with the GC instrument. |
5102 |
The GC program has been started with data acquisition cannot performed. |
This error occurs when operations to start up a GC program (e.g. by pressing the start button on the instrument) are performed when the instrument is not ready for data acquisition. |
Operate after preparing for data acquisition. |
5103 |
GC is not connected. |
Failed to connect to the GC. |
First, check that the GC power supply is allocated, that the power is turned ON, that cables between the GC and PC are properly connected, and that the operating environment (OS, drive settings for the communication devices, etc.) is normal. Also, check the communications port specification is not in error at "Communication Settings" (displayed by double-clicking the [Instrument] icon on the "Modules Used for Analysis", or by highlighting and clicking [Properties]) that is called from "System Configuration". If this message is displayed even though the above is normal, contact your |
5104 |
The AOC is not connected. Check the connections between the CBM and AOC, and restart the CBM and GC. |
The AOC is not actually connected even though the AOC is set to be used in the system configuration. |
A probable cause is that CBM and AOC are not connected correctly, or that AOC is set to be used even though there is no AOC. Check the connection between CBM and AOC, or make sure that the [AOC Connection] checkbox that is displayed by clicking [CBM-102 Settings] under <System Configuration> - <Properties> is set correctly. |
5105 |
The GC type does not match. |
The actually connected GC type does not match the GC type set in the instrument properties in the system configuration. |
Retry setting the system configuration according to the type of currently connected GC. |
5106 |
The column could be damaged if you start the GC system while the carrier gas flow controller is Off (its On/Off button is on the instrument monitor bar). \nAre you sure you want to start the GC |
GC system may not be started while the flow controller is Off. \nSometimes the column can be damaged in this condition. |
Please check the condition and setting if you want to continue this operation or select [No] to terminate this operation. |
5107 |
GC instrument warning occurred. |
A warning was notified from the GC instrument. |
This error is accompanied by a message sent from the GC instrument. Check the content of this error message and remedy the error according to the Instruction Manual provided with the GC instrument. |
5110 |
Autosampler settings are not correct. |
The autosampler related settings in the system configuration are in error. |
Retry setting the system configuration of theAutosampler. |
5111 |
Please set the column information. |
The column length, inner diameter, and upper limit temperature are not set. |
Click the [Select] button and select the column currently used in the [Column List] dialog box. If the column in use is not registered, click the [Add] button and enter the inner diameter, length and other details. |
5112 |
Please select the detector. |
The detector is not selected. |
Be sure to select a detector. |
5113 |
The same"Analog Input"is set to two or more detectors. |
The same "Analog Input (AD1/AD2)" is set to 2 or more detectors selected for the analysis line. |
Click "Yes" in the message display sub-window. Set "Analog Input" for the detector not registered in the analysis line to "None". |
5114 |
The same ""Analog Input"" is set to two or more detectors. \nDo you change the analog input setting of the detector not used for data acquisition to ""None""? |
The same"Analog Input (AD1/AD2)"is set to two or more detectors. |
Click "Yes" in the message display sub-window. Set "Analog Input" for the detector not selected in the analysis line to "None". |
5115 |
Change the"Filter Time Const"setting of the detector tonmsec to match the type of injection unit? |
Processing for changing the default detector filtering time constant according to the type of injection unit on the same analysis line will be performed. |
Select [Yes] to change the filtering time constant to the displayed value. |
5116 |
HSS is not connected. |
The HSS is not connected to the PC. |
Make sure that the HSS is turned ON and that it is operating normally. Also, check the connection cables. |
5117 |
An attempt was made to process unusable parameters. |
An attempt was made to send parameters that cannot be used in the current system configuration. Or, an attempt was made to send parameters during operation. |
Make sure that the system configuration and method are set within the correct ranges. |
5118 |
Failed to load the HSS system configuration. |
The HS system configuration could not be loaded. |
Make sure that the HSS is turned ON and that it is operating normally. Also, check the connection cables. |
5119 |
Failed to load the HSS status. |
The HS status could not be loaded. |
Make sure that the HSS is turned ON and that it is operating normally. Also, check the connection cables. |
511A |
Failed to set the HSS method. |
The HS method is not set. |
Set the HSS method. |
511B |
Failed to execute the HSS. |
The HS system could not accept the execution command. |
Make sure that the HSS is turned ON and that it is operating normally. Also, check the connection cables. |
511C |
Failed to stop the HSS. |
The HS system could not accept the stop command. |
Make sure that the HSS is turned ON and that it is operating normally. Also, check the connection cables. |
511D |
Failed to reset the HSS. |
The HS system could not accept the reset command. |
Make sure that the HSS is turned ON and that it is operating normally. Also, check the connection cables. |
511E |
Failed to set the HSS ready. |
The HS system could not accept the command during stop processing. |
Make sure that the HSS is turned ON and that it is operating normally. Also, check the connection cables. |
5126 |
The HSS is currently processing. |
The HS system could not accept the command since it was processing. |
Wait a while before trying again. |
5127 |
HSS system memory error |
An error was found in the HS system memory. |
When this error persists after processing, contact your Shimadzu representative. |
5128 |
Needle motor time-out error |
The needle motor did not reach the specified position during the specified |
When this error persists after processing, contact your Shimadzu representative. |
5129 |
Vial lift time-out error |
The vial lift did not reach the specified position during the specified time. |
When this error persists after processing, contact your Shimadzu representative. |
512A |
Magazine motor time-out error |
The magazine motor did not reach the specified position during the specified |
When this error persists after processing, contact your Shimadzu representative. |
512B |
Oven motor time-out error |
The oven motor did not reach the specified position during the specified time. |
When this error persists after processing, contact your Shimadzu representative. |
512C |
The magazine is full. |
There is no empty vial position for unloading the sample vial from the oven. |
Remove the vial from the magazine. When this error persists after processing, contact your Shimadzu |
512D |
Unexpected error occurred. |
Error of unknown cause occurred. |
Turn the system OFF then ON again. When this error persists after processing, contact your Shimadzu |
512E |
Oven motor startup error |
The oven motor stopped at the start stage. |
When this error persists after processing, contact your Shimadzu representative. |
512F |
Magazine motor startup error |
The magazine motor stopped at the start stage. |
When this error persists after processing, contact your Shimadzu representative. |
5130 |
Oven heater controller error |
The oven heater electrical controller is in error. |
When this error persists after processing, contact your Shimadzu representative. |
5131 |
Needle heater controller error |
The needle heater electrical controller is in error. |
When this error persists after processing, contact your Shimadzu representative. |
5132 |
Transfer heater controller error |
The transfer heater electrical controller is in error. |
When this error persists after processing, contact your Shimadzu representative. |
5134 |
Oven heater sensor is shortcircuited |
A short-circuit occurred on the oven heater temperature sensor. |
When this error persists after processing, contact your Shimadzu representative. |
5135 |
Needle heater sensor is shortcircuited |
A short-circuit occurred on the needle heater temperature sensor. |
When this error persists after processing, contact your Shimadzu representative. |
5136 |
Transfer heater sensor is shortcircuited |
A short-circuit occurred on the transfer heater temperature sensor. |
When this error persists after processing, contact your Shimadzu representative. |
5138 |
HSS oven heater sensor lead is broken |
The oven heater temperature sensor did not receive the signal. |
When this error persists after processing, contact your Shimadzu representative. |
5139 |
Needle heater sensor discontinuity |
The needle heater temperature sensor did not receive the signal. |
When this error persists after processing, contact your Shimadzu representative. |
513A |
Transfer heater sensor discontinuity |
The transfer tube heater temperature sensor did not receive the signal. |
When this error persists after processing, contact your Shimadzu representative. |
513C |
HSS oven heater fuse has melted down. |
The oven heater fuse melted down. The temperature limit was exceeded. |
When this error persists after processing, contact your Shimadzu representative. |
513D |
Needle monitor is short-circuited |
A short-circuit occurred on the needle unit. |
When this error persists after processing, contact your Shimadzu representative. |
513E |
Transfer monitor is short-circuited |
A short-circuit occurred on the transfer tube. |
When this error persists after processing, contact your Shimadzu representative. |
513F |
Needle monitor sensor discontinuity |
The needle heater monitor did not receive the signal. |
When this error persists after processing, contact your Shimadzu representative. |
5140 |
Transfer monitor sensor discontinuity |
The transfer tube monitor did not receive the signal. |
When this error persists after processing, contact your Shimadzu representative. |
5141 |
Oven unload error |
The vial could not be found when the sample vial was unloaded from the oven. |
When this error persists after processing, contact your Shimadzu representative. |
5142 |
Operator loading error |
Failed to load the sample vial. |
When this error persists after processing, contact your Shimadzu representative. |
5143 |
Command was ignored. |
The command was ignored since the command was sent during processing that does not accept commands. |
Wait a while before trying again. |
5144 |
The HSS system is not connected to the controller. |
A probable cause is that discontinuity occurred midway. |
Restart the software. |
5145 |
No vial. |
The vial No. differs from the batch setting. |
Make sure that there are no omissions when setting the vials in place. |
5146 |
Waiting for HSS external ready signal. |
The externally connected instrument is not ready though the injection time has been reached. |
Operation is started when the externally connected instrument is ready. A probable cause is that the heat retention time does not reach the preset value. For details, refer to the report for the HS system. |
5147 |
An error has occurred on the HS (Please check the HS state). |
An error occurred on the HS system, and the HS system stopped in this state. |
If absence of a vial is the cause, cancel data acquisition, set a vial in place, and perform data acquisition again. If something else is the cause, refer to [Troubleshooting] in the Instruction Manual of the HS. |
5148 |
The HSS heat retention time has been extended. |
The retention time was extended since the externally connected instrument is not ready though the injection time has been |
There is a possibility that the heat retention time does not reach the set value. For details, refer to the report for the HS system. |
2149 |
Please unify the HS parameters in the method file. |
The HS parameters are not unified in the method file used in batch processing. |
Set all method files to the same value. |
514A |
The HS parameter [GC Cycle Time] is set to the"GC Analysis time + 0.5 minutes"or less. |
The method HS parameter [GC Cycle Time] is set to the "GC Analysis time + 0.5 minutes" or less. |
Set so that [GC Cycle Time] is longer than the"GC Analysis time + 0.5 minutes". |
514B |
The HS parameter [GC Cycle Time] is set to the"GC Analysis time + cooling time"or less. Are you sure? |
The method HS parameter [GC Cycle Time] is set to the "GC Analysis time + cooling time" or less. (The cooling time is calculated as 10 minutes when the initial temperature of the column oven is 50 degree or more, and as 15 minutes when the column oven is less than 50 degree.) |
Set so that [GC Cycle Time] is longer than the "GC Analysis time + cooling time". |
514C |
The HS parameter [Thermostatic Time] is set to the"GC Analysis time + 0.5 minutes"or less. |
With data acquisition in the MHE mode, the method HS parameter [Thermostatic Time] is set to the "GC Analysis time + 0.5 minutes" or less. |
Set so that [Thermostatic Time] is longer than the"GC Analysis time + 0.5 minutes". |
514D |
The HS parameter [Thermostatic Time] is set to the"GC Analysis time + cooling time"or less. Are you sure? |
With data acquisition in the MHE mode, the method HS parameter [Thermostatic Time] is set to the "GC Analysis time + cooling time" or less. (The cooling time is calculated as 10 minutes when the initial temperature of the column oven is 50 degree or more, and as 15 minutes when the column oven is less than 50 degree.) |
Set so that [Thermostatic Time] is longer than the "GC Analysis time + cooling time". |
514E |
The values of the HS parameter [Temperature Zone (Oven, Needle, Transfer)] are not in ascending order. Are you sure? |
The values of the method HS parameter [Temperature Zone (Oven, Needle, Transfer)] and the GC parameter injection unit temperature values are not in ascending order. |
Set the HS zone temperature setting and GC injection unit temperature value as follows. (oven temperature) < (needle temperature)=<(transfer temperature)=<(GC injection unit temperature) |
514F |
Please set the vial Nos. starting with 1 up to the maximum number of vials. |
The vial No. of the batch is set to a value greater than the maximum number of vials of the HS. |
Set the vial No. to the maximum number of vials or less. |
5150 |
Please set the vial Nos. in ascending order. |
The vial No. of the batch must be set in ascending order. |
Set the vial Nos. in ascending order. |
5151 |
Please unify the number of repeats for the vial Nos. |
With data acquisition in the MHE mode, the number of repeats of the vial No. for the batch is not unified. |
Unify the number of repeats for the vial Nos. |
5152 |
Please set the number of repeats for the vial Nos. up to 9. |
With data acquisition in the MHE mode, the number of repeats of the vial No. for the batch is set higher than 9. |
Set the number of repeats for the vial Nos. to 9 or less. |
5153 |
Cannot start up simultaneously. |
An attempt was made to use AOC and HS simultaneously. |
Perform data acquisition on either of the lines since AOC and HS cannot be used simultaneously. |
5154 |
The same vial number cannot be set to lines 1 and 2. |
The same vial No. is set to lines 1 and 2. |
The sample in the same vial cannot be poured simultaneously on lines 1 and 2. Set different vial Nos. |
5155 |
GC for communication is not connected. |
Cannot communicate with GC, and connection failed. |
Check the cable and power supply. |
5160 |
Turn the HSS heater on? |
This message is displayed for confirmation when the turning heater of HS-20 on. |
Click "yes" if turning the heater on and click "No" if keeping the heater off. |
5161 |
Are you sure you want to turn the HSS heater off? |
This message is displayed for confirmation when the turning heater of HS-20 off. |
Click "yes" if turning the heater off and click "No" if keeping the heater on. |
5162 |
Additional Flow configuration is not correct. |
The same Additional flow that is used for HS-20 is configured. |
Configure the different additional flow for each port of HS20. |
5163 |
HS-20 error occurred. |
An error was notified from the HS-20 instrument. |
This error is accompanied by a message sent from the HS20 instrument. Check the content of this error message and remedy the error according to the Instruction Manual provided with the HS-20 instrument. |
5164 |
HS-20 warning occurred. |
A warning was notified from the HS-20 instrument. |
This warning is accompanied by a message sent from the HS-20 instrument. Check the content of this warning message and remedy the warning according to the Instruction Manual provided with the HS-20 instrument. |
5165 |
HS-20 is not connected. |
Failed to connect to the HS-20. |
First, check that the HS-20 power supply is allocated, that the power is turned ON, that USB cables between the HS20 and PC are properly connected, and that the operating environment (OS, drive settings for the communication devices, etc.) is normal. Also, check whether the serial number of HS-20 is displayed or not at "Communication Settings" (displayed by double-clicking the [Instrument] con on the "Modules Used for Analysis", or by highlighting and clicking [Properties]) that is called from "System |
5166 |
The HS-20 parameter [GC Cycle Time] is set to the "GC Analysis time + 0.5 minutes" or less. |
The method HS parameter [GC Cycle Time] is set to the "GC Analysis time + 0.5 minutes" or less. |
Set so that [GC Cycle Time] is longer than the"GC Analysis time + 0.5 minutes". |
5167 |
The HS-20 parameter [GC Cycle Time] is set to the "GC Analysis time + cooling time" or less. Are you sure? |
The method HS parameter [GC Cycle Time] is set to the "GC Analysis time + cooling time" or less. (The cooling time is calculated as 10 minutes when the initial temperature of the column oven is 50 degree or more, and as 15 minutes when the column oven is less than 50 degree) |
Set so that [GC Cycle Time] is longer than the "GC Analysis time + cooling time". |
5168 |
Please set the number of repeats for the vial Nos. up to 9. |
When the same vial No for the batch is repeated, it is recogniozed as MHE mode. The number of repeats is set higher than |
Set the number of repeats for the vial Nos. to 9 or less. |
5169 |
Please set the batch Line up to 1000. |
The number of batch lines must be set less than 1000. |
The number of batch lines must be set less than 1000 for HS-20.If you want to set more than 1000 lines, use "batch |
516A |
Cannot start up simultaneously. |
An attempt was made to use AOC and HS-20 simultaneously. |
Perform data acquisition on either of the lines since AOC and HS-20 cannot be used simultaneously. |
516B |
Please unify the HS parameters in the method file. |
The HS-20 parameters are not unified in the method file used in batch processing. |
Set all method files to the same value. |
516C |
HS-20 Communication error. |
Communication error between HS-20 and PC occurred. |
USB cables between the HS-20 and PC are properly connected, and that the operating environment (OS, drive settings for the communication devices, etc.) is normal. |
516D |
Check the condition of the power plug. |
It Continues using a lot of electricity. |
Confirm the plug for power supply because it may cause ignition unless it is not plugged in properly. |
5173 |
HS-10 error occurred. |
An error was notified from the HS-10 instrument. |
This error is accompanied by a message sent from the HS10 instrument. Check the content of this error message and remedy the error according to the Instruction Manual provided with the HS-10 instrument. |
5174 |
HS-10 warning occurred. |
A warning was notified from the HS-10 instrument. |
This warning is accompanied by a message sent from the HS-10 instrument. Check the content of this warning message and remedy the warning according to the Instruction Manual provided with the HS-10 instrument. |
5175 |
HS-10 is not connected. |
Failed to connect to the HS-10. |
First, check that the HS-10 power supply is allocated, that the power is turned ON, that USB cables between the HS10 and PC are properly connected, and that the operating environment (OS, drive settings for the communication devices, etc.) is normal. Also, check whether the serial number of HS-10 is displayed or not at "Communication Settings" (displayed by double-clicking the [Instrument] con on the "Modules Used for Analysis", or by highlighting and clicking [Properties]) that is called from "System Configuration". |
5176 |
The HS-10 parameter [GC Cycle Time] is set to the "GC Analysis time + 0.5 minutes" or less. |
The method HS parameter [GC Cycle Time] is set to the "GC Analysis time + 0.5 minutes" or less. |
Set so that [GC Cycle Time] is longer than the "GC Analysis time + 0.5 minutes". |
5177 |
The HS-10 parameter [GC Cycle Time] is set to the "GC Analysis time + cooling time" or less. Are you sure? |
The method HS parameter [GC Cycle Time] is set to the "GC Analysis time + cooling time" or less. (The cooling time is calculated as 10 minutes when the initial temperature of the column oven is 50 degree or more, and as 15 minutes when the column oven is less than 50 degree) |
Set so that [GC Cycle Time] is longer than the "GC Analysis time + cooling time". |
5178 |
Please set the batch Line up to 240. |
The number of batch lines must be set less than 240. |
The number of batch lines must be set less than 240 for HS-10.If you want to set more than 240 lines, use "batch queue". |
5179 |
Cannot start up simultaneously. |
An attempt was made to use AOC and HS-10 simultaneously. |
Perform data acquisition on either of the lines since AOC and HS-10 cannot be used simultaneously. |
517B |
HS-10 Communication error. |
Communication error between HS-10 and PC occurred. |
USB cables between the HS-10 and PC are properly connected, and that the operating environment (OS, drive settings for the communication devices, etc.) is normal. |
517C |
Half turn the Vial Tray? |
This message is displayed for confirmation when the halof turn the tray of HS-10. |
Click "yes" if half turn the tray, and click "No" if doesn't half turn the tray. |
517D |
HS-10 is not connected because of maintenance. |
When HS-10 is on maintenance working, the PC cannot connect with HS. |
Connect the HS-10 after the maintenace end. |
517E |
Please set the number of repeats for the vial Nos. up to 10. |
When the same vial number for the batch is repeated with MHE mode, the number of repeats is set higher than 10. |
Set the number of repeats for the vial Nos. to 10 or less. |
5300 |
The system controller has not been set. Are you sure you want to continue processing? |
----- |
----- |
5301 |
Not ready for data acquisition. |
The instrument is readying for data acquisition. |
Wait for data acquisition to become ready before starting data acquisition. |
5302 |
Failed to move the vial. |
The connection between the PC and LC instrument or each instrument and controller may be wrong. |
Try again after reconstructing the connection accurately. |
5304 |
Please upgrade ROM version of CBM-20A/lite. |
Rom version of CBM-20A/lite don't suppurt the unit which tries to use. |
Make ROM version of CBM-20A/lite a latest version. |
5306 |
The integrated PDA detector can not be selected in UV detector model. |
Using UV detector model, the integrated PDA detector is selected on PDA table item in the [Instrument] sub-window. |
Please double-click [Instrument (Communication Setting)] in the [System Configuration] sub-window, and eliminate integrated PDA detector from PDA table entry in the [Instrument] sub-window. |
5307 |
The integrated UV detector or another optional PDA detector can not be selected in PDA detector model. |
Using PDA detector model, integrated UV detector is selected in [System Configration] sub-window, or PDA detector other than [LC-2030 PDA Detector] is selected in the [Instrument] sub-window. |
Please eliminate integrated UV detector in [System Configration] sub-window, or select [LC-2030 PDA Detector] on PDA table item in the [instrument] subwindow. |
5308 |
The integrated PDA detector in PDA detector model is not selected. |
Using PDA detector model, other than the [LC-2030 PDA Detector] is selected on PDA table item in the [Instrument] subwindow. |
Please double-click [Instrument (Communication Setting)] in the [System Configuration] sub-window, and eliminate [LC-2030 PDA Detector] from PDA table entry in the [Instrument] sub-window. |
5309 |
Failed the communication between the SFC and pump because the connection cable between the SFC and pump is not connected properly or the |
The connection cable between the SFC and pump is not connected properly or the pump is turned off. |
After checking the connections, perform the [Auto Configuration] again. |
5312 |
The sample injection is paused. Analysis is waiting until it becomes ready. |
The sample injection is paused. |
When the Autosampler is set to be used, make sure that the Autosampler is operating normally. |
5313 |
The sample injection was failed and data acquisition was |
Data acquisition was not started in the specific period of time. |
When the Autosampler is set to be used, make sure that the Autosampler is operating normally. |
5315 |
Please set the Wash command one minute or more after the End command time. |
----- |
----- |
53F1 |
SP CTRL:Invalid operation. Fullregenerating. [1001] |
While full-regenerating, Activate on or switch cartridge command was operated. |
Do not operate those commands while full_regenerating. |
53F2 |
SP CTRL:Invalid operation. Now ACT ON. [1002] |
Full-regeneration was operated while HPLC system was activated. |
Do not operate full-regeneration while HPLC system is activated. |
53F3 |
SP CTRL: There is a contradiction in the suppressor control parameter settings. [1003] |
The suppressor-related parameters are not correct. |
Check the values and set so that values are correct. |
53F4 |
SP CTRL:Suppressor Ch1-L cannot regenerate. [1004] |
Suppressor cartridge current did not become constant. |
The suppressor cartridge may have become degraded. Replace the cartridge. |
53F5 |
SP CTRL:Suppressor Ch1-R cannot regenerate. [1005] |
Suppressor cartridge current did not become constant. |
The suppressor cartridge may have become degraded. Replace the cartridge. |
53F6 |
SP CTRL:Suppressor Ch2-L cannot regenerate. [1006] |
Suppressor cartridge current did not become constant. |
The suppressor cartridge may have become degraded. Replace the cartridge. |
53F7 |
SP CTRL:Suppressor Ch2-R cannot regenerate. [1007] |
Suppressor cartridge current did not become constant. |
The suppressor cartridge may have become degraded. Replace the cartridge. |
53F8 |
CTO: Door is open. Please close the door. [1008] |
Column oven door is open. |
Close the door. |
53F9 |
CTO: Door is open too long. [1009] |
Column oven door was open for too long time while using suppressor. |
Close the door while using suppressor. |
53FA |
SP CTRL: Cartridge regeneration was canceled. [100A] |
----- |
----- |
5403 |
Controller: Data not protected. Method and sequence files have been initialized. [0003] |
Method and sequence files have been initialized. |
After making the appropriate settings for the connected components in the "Configuration" tab page, reset the methods and sequences in all the relevant windows. |
5404 |
Controller: Data not protected. Method and sequence files have been restored. [0004] |
The method and sequence files have been returned to the state they were in the previous time the system controller was started up. |
Check the methods and sequences. |
5405 |
Controller: Illegal backup parameter. Method and sequence files have been |
Some parameters have been initialized. |
Check the methods and sequences. |
5406 |
Controller: Illegal backup parameter. System information and user information have been changed. [0006] |
Some information has been initialized. |
Check the system and user information. |
5407 |
Controller: ROM4 error occurred. [0007] |
----- |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
5408 |
Controller: Signal received from ALARM INPUT terminal. [0008] |
An error signal was sent from the external device connected to the external input/output terminal [IN]. |
Inspect the external device, and eliminate the cause of the error. |
540A |
Contoller:Link time out error. [000A] |
The fiber-optic communication link has been disconnected. (This error is suppressed when link check is turned OFF in the Configuration screen.) |
Ensure the connection of the fiber-optic cable between the fraction collector and the system controller. |
540B |
Controller: Component fixed on Configuration screen has not been linked. [000B] |
Component fixed on configuration screen has not been linked and analysis started. |
Check the connections. Perform system configuration again. |
540C |
Controller: Incorrect ID from sub- CPU. [000C] |
----- |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
540E |
Controller: Incorrect data from sub-CPU. [000E] |
----- |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
540F |
FRC or SIL: Invalid vial number. [000F] |
Invalid sample(vial) number. |
Set values within the valid range for the sample rack used. |
5410 |
SIL:Needle X axis home position error. [0010] |
Move to X (right and left) direction was faulty. |
Check if nothing is stuck in the movable part.\nIf nothing is stuck, contact your Shimadzu representative. |
5411 |
SIL:Needle Y axis home position error. [0011] |
Move to Y (front and back) direction was faulty. |
Check if nothing is stuck in the movable part.\nIf nothing is stuck, contact your Shimadzu representative. |
5412 |
SIL:Needle Z axis home position error. [0012] |
Move to Z (up and down) direction was faulty. |
Check if nothing is stuck in the movable part.\nIf nothing is stuck, contact your Shimadzu representative. |
5413 |
SIL:6-port valve home position error. [0013] |
Rotation of 6-port valve motor was faulty. |
Remedy this problem referring to the following procedure. 1. Check if the rotor seal has reached its service life. Replace the rotor seal if it has reached its service life. 2. Disassemble and clean. The torque required for turning the valve sometimes increases due to foreign objects in the mobile phase and impurities in the sample. Remove these foreign matters by disassembling and cleaning the valve. (At this time, replace scratched seals with new seals.) If the above procedure does not remedy the problem, contact |
5414 |
SIL:3-way valve home position error. [0014] |
Rotation of 3-way valve motor was faulty. |
Please perform actions using the following procedures.\n1. Check that the rotor seal has not exceeded its service life.\n - If it has, replace it with a new rotor seal.\n2. Disassemble the valve and clean it.\n - Due to mobile phase contaminants or sample impurities, the torque necessary for valve rotation may increase excessively. Remove these contaminants by disassembling the valve and cleaning it. (If any scratches are found on the rotor seal surface, replace it with a new rotor seal.)\nIf the error is not corrected after performing the actions above, contact |
5415 |
SIL:Syringe home position error. [0015] |
Motion of syringe plunger driving motor was faulty. |
The friction between the syringe tip and barrel may have excessively increased.\nRemove the syringe and aspirate/discharge water or alcohol with the syringe repeatedly, allowing the syringe tip seal surface to become acclimated to the barrel. \nIf the error is not corrected after performing the action above, contact your Shimadzu representative. |
5416 |
SIL: Pretreatment program error! An error occurred while pretreatment program was |
There is a line which cannot be executed in the pretreatment program. |
Check set parameter for the line. |
5417 |
SIL: Pretreatment program does not exist. Please edit a pretreatment program. [0017] |
Pretreatment program to be executed does not exist. |
Check if file number to be executed is specified correctly. |
5418 |
SIL:Pretreatment file exceeds the maximum size. [0018] |
Total number of bytes in file exceeds memory capacity. |
Up to 250 byte is available in a file. Divide to several files using CALL and GOTO command if program is too large. |
5419 |
SIL: Undefined rack is used. |
Rack is not set, or rack detection failed. |
Set values within the valid range for the sample rack used. |
541A |
SIL:6-position valve home position error. [001A] |
6-port valve does not rotate to the correct positions. |
Please perform actions using the following procedures.\n1. Check that the rotor seal has not exceeded its service life.\n - If it has, replace it with a new rotor seal.\n2. Disassemble the valve and clean it.\n - Due to mobile phase contaminants or sample impurities, the torque necessary for valve rotation may_increase excessively. Remove these contaminants by disassembling the valve and cleaning it. (If any scratches are found on the rotor seal surface, replace it with a new rotor seal.)\nIf the error is not corrected after performing the actions above, contact |
541B |
SIL:Metering pump home position error. [001B] |
Metering pump home position error has occurred. |
Remedy this problem referring to the following procedure. Check if the plunger seal has reached its service life. Replace the plunger seal if it has reached its service life. The torque required for turning the valve sometimes increases due to foreign objects in the rinse solution and impurities in the sample. Remove these foreign matters by disassembling and cleaning the valve. At this time, replace scratched seals with new seals. If the above procedure does not remedy the problem, contact your Shimadzu |
541C |
SIL:Invalid aspiration/discharge volume setting in pretreatment program. [001C] |
Suction of more volume than syringe capacity was attempted in pretreatment program. |
If suction volume is larger than syringe capacity, reset it to a smaller volume than syringe capacity.\nIf discharge volume is larger than suction volume, reset discharge volume to a smaller volume than sucked volume, or reset |
541D |
SIL:Invalid vial number. [001D] |
Invalid sample(vial) number. |
Confirm whether a standard rack is set. |
541E |
SIL:Invalid aspiration/discharge volume setting of metering pump in pretreatment program. [001E] |
Invalid aspiration/discharge volume setting of metering pump in pretreatment program. |
If suction volume is larger than syringe capacity, reset it to a smaller volume than syringe capacity.\nIf discharge volume is larger than suction volume, reset discharge volume to a smaller volume than sucked volume, or reset |
5420 |
FRC:Arm X axis home position error. [0020] |
Motor in X direction (right and left) slipped. |
Turn the power OFF and check if the arm is in contact with something. |
5421 |
FRC:Arm Y axis home position error. [0021] |
Motor in Y direction (front and back) slipped. |
Turn the power OFF and check if the arm is in contact with something. |
5422 |
FRC:Arm Z axis home position error. [0022] |
Motor in Z direction (up and down) slipped. |
Turn the power OFF and check if the arm is in contact with something. |
5424 |
FRC:Undefined rack. User rack is not defined. Teach it and activate. [0024] |
An attempt was made to use an unregistered rack. |
Set rack number properly. |
5425 |
FRC:Too many peaks. 3rd peak detected during a delay time. |
An attempt was made to collect the third peak during delay time for the first peak. |
Change the time program so that collection of the third peak is not made in the delay time for the first peak. |
5426 |
FRC:Invalid vial number. Set value within the valid range for the sample rack. [0026] |
Fraction collector attempted to move to a vial number which does not exist in the rack. |
Set vial number in the time program correctly. |
5430 |
Pump: Pressure has exceeded P.Max. Check flow lines for clogging. If P.Max is too low, reset it. [0030] |
Pump discharge pressure has exceeded the set max. pressure limit, [P.Max]. Pumping will automatically stop, unless [SPROT] function has been activated, in which case pumping will continue at half |
Inspect the clogging in the flow line. If no clogging is found, set the appropriate [P.Max] value. |
5431 |
Pump:Pressure does not rise to P.Min. Check flow lines for leaks. If P.Min is too high, reset it. |
Pump discharge pressure has fallen below the set min. pressure limit, [P.Min]. Pumping will automatically stop. |
Check the flow line for leaks. If no leak is found, the [P.Min] setting is inappropriate, and must be changed. |
5432 |
Pump:Home position can not be detected. [0032] |
The motor home position cannot be detected, the motor does not run, or the motor is slipping. |
Turn the power OFF, then ON again, and restart pumping.\nIf the message appears again, turn the power OFF and contact your Shimadzu representative. |
5433 |
Pump:Operation mode is local. Set the operation mode to remote. [0033] |
Operation mode is local. |
Set the operation mode to remote. |
5434 |
Pump:Leak detected. Eliminate the leak in the pump's interior or exterior. [0034] |
Leak sensor has detected leakage.\nSensor is too sensitive. |
Inspect the leakage in the flow line. Wipe away any liquid around the leak sensor.\nAdjust the leak sensor actuation level. |
5440 |
Detector: Grating home position error. [0040] |
Motor home position sensor does not operate properly. |
Contact your Shimadzu representative. |
5441 |
Detector:Filter home position error. [0041] |
Filter home position sensor does not operate properly. |
Contact your Shimadzu representative. |
5442 |
Detector:Lamp home position error. [0042] |
Lamp home position sensor does not operate properly. |
Contact your Shimadzu representative. |
5443 |
Detector:D2 lamp does not light. [0043] |
Fault in deuterium lamp or in its circuits. |
Check the total lamp illumination time, and replace the lamp.\nIf the error occurs after lamp replacement, contact your Shimadzu representative. |
5444 |
Detector: D2 lamp is over current. [0044] |
Abnormally high current flowing through lamp. |
Check the total lamp illumination time, and replace the lamp.\nIf the error occurs after lamp replacement, contact your Shimadzu representative. |
5445 |
Detector: Leak is detected. [0045] |
Eluent leak can be detected in the detector. |
Stop the pump flow and then execute any action to stop the eluent leak. See detail Instrument manual. |
5446 |
Detector:Operation mode is local. Set operation mode to remote. [0046] |
Operation mode is local. |
Set the operation mode to remote. |
5447 |
Detector:Ex grating motor home position error. [0047] |
Ex grating motor has slipped. |
Contact your Shimadzu representative. |
5448 |
Detector:Em grating motor home position error. [0048] |
Em grating motor has slipped. |
Contact your Shimadzu representative. |
5449 |
Detector:Auto zero level is out of range. Confirm the installation of the cell, no bubble in the cell, and the settings of SENS and GAIN parameters. [0049] |
Signal level is too high. |
Confirm the installation of the cell, no bubble in the cell.\nSet SENS lower. |
544A |
Detector:Wavelength is inaccurate. Run WAVE CALIB in SPD to calibrate wavelength. [004A] |
Wavelength discrepancy exceeds 1nm. |
1. Cell is not installed properly\n- Install cell properly\n2. Error in wavelength calibration\n- Run [WAVE CALIB.] (VP function) to calibrate wavelength, then run [WAVE CHECK] to check wavelength accuracy.\n3. Sample in cell is highly absorptive of UV and visible light, or large air bubbles in sample greatly restrict amount of light transmitted through cell at emission line wavelengths (656 & 254nm) of deuterium lamp or mercury lamp.\n- Flush the inside of the cell with a mobile phase solvent that does not absorb light in the vicinity of above wavelength.\n- Purge cell interior with air or nitrogen.\nIf [CHECK NO GOOD] appears again when [WAVE CHECK] is run after the above action has been taken, turn the power OFF and contact your |
544B |
Detector:Detector temperature has exceeded 63 C. [004B] |
The temperature of the measuring point rises to 63C or more. |
Check the operating environment. If you cannot find any problems, contact your Shimadzu representative. |
544C |
Detector: Xe lamp does not light. [004C] |
Fault in Xe lamp or in its circuits. |
Check the Xe lamp operating time and replace it.\nIf the error persists, turn the power OFF and contact your Shimadzu representative. |
544D |
Detector: REF DATA OVERFLOWED. [004D] |
----- |
----- |
5450 |
Detector: NO SCAN DATA. [0050] |
Spectrum data was requested for output from a file which dose not contain scanned |
Check the file number set when scanning was executed, and use that file. |
5451 |
Detector: OVEN OVERHEATED. [0051] |
----- |
Turn the power OFF and contact your Shimadzu representative. |
5452 |
Detector: THERMO MODULE OVER-CURRENT. [0052] |
----- |
Turn the power OFF and contact your Shimadzu representative. |
5453 |
Detector: TEMP SENSOR. [0053] |
----- |
Turn the power OFF and contact your Shimadzu representative. |
5460 |
Detector: PMT OVER- CURRENT. [0060] |
----- |
Turn the power OFF and contact your Shimadzu representative. |
5461 |
Detector: NEGATIVE VOLTAGE SOURCE. [0061] |
----- |
Turn the power OFF and contact your Shimadzu representative. |
5462 |
Detector: TEMP FUSE BLOWN OFF. [0062] |
----- |
Turn the power OFF and contact your Shimadzu representative. |
5463 |
Detector: HARDWARE ERROR. [0063] |
----- |
Turn the power OFF and contact your Shimadzu representative. |
5470 |
CTO: Oven temperature has exceeded T.Max. [0070] |
The oven temperature has exceeded the value set for [T.MAX] . |
1. Lower the oven set temperature.\n2. Set [T.MAX] to a value 5degreehigher than the set temperature. An overshoot may occur before the temperature reaches the |
5471 |
CTO: Pt sensor reading abnormally high. [0071] |
Thermosensor readings are abnormally high. There may be a short circuit in the |
Turn the power OFF and contact your Shimadzu representative. |
5472 |
CTO: Pt sensor reading abnormally low. Pt sensor is not connected or is short-circuited. |
Thermosensor readings are abnormally low. The sensor may be disconnected. |
Turn the power OFF and contact your Shimadzu representative. |
5473 |
CTO: Leak detected. Eliminate the leak in the oven's interior or exterior. [0073] |
Density of solvent vapor inside column oven has exceeded leak sensor actuation level. |
1. Mobile phase has leaked inside the oven.\n- Take action to stop the leak.\n2. There is solvent vapor in the air around the unit.\n- Provide adequate\n3. Sensor sensitivity is incorrectly adjusted.\n- Readjust the sensitivity.\nTake the appropriate action described in the table above. Then run the fan for 5 minutes with the right door open. The leak sensor is unstable for 60 seconds after starting up the column oven and so errors are not detected during this |
5474 |
CTO: Fan speed does not increase. [0074] |
Fan speed does not increase to the set value. |
If the cause is obstruction of the fan by tubing or other object, turn the power OFF and remove the obstruction. If the object cannot be removed, or if the error is due to some other cause, contact your Shimadzu representative. |
5475 |
CTO: Operation mode is local. Set operation mode to remote. |
Operation mode is local. |
Set the operation mode to remote. |
5476 |
CTO: Leak sensor wire is broken. [0076] |
Leak sensor malfunctions (probably due to broken/disconnected wire). |
Turn the power OFF and contact your Shimadzu representative. |
5477 |
CTO: Low set temperature. [0077] |
Set temperature value is inappropriate, in one of following ways:\n(1) When use of cooler is enabled, set temperature is room temperature minus 10C or lower.\n(2) In other situations, set temperature is lower than the room temperature. |
Set temperature higher than room temperature minus 10C if using the cooler, or higher than room temperature if not using the cooler.\n(Temperature regulation is not stopped.) |
5478 |
CTO: High set temperature. [0078] |
Set temperature value is room temperature plus 60C or higher. |
Set a temperature lower than room temperature plus 60C.\n(Temperature regulation is not stopped.) |
5480 |
RV.A: Rotary valve A home position error. [0080] |
No Motor Home Position Error of RV.A has occurred. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
5481 |
RV.B:Rotary valve B home position error. [0081] |
No Motor Home Position Error of RV.B has occurred. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
5482 |
RV.C:Rotary valve C home position error. [0082] |
No Motor Home Position Error of RV.C has occurred. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
5483 |
RV.D:Rotary valve D home position error. [0083] |
No Motor Home Position Error of RV.D has occurred. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
5484 |
Syringe Unit: Valve home position error. [0084] |
Valve home position error has occurred. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
5485 |
Syringe Unit:Syringe home position error. [0085] |
No Motor Home Position Error of Syringe unit has occurred. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
5486 |
Syringe Unit:The setting for suction/discharge is out of range. [0086] |
The setting for suction/discharge of Syringe Unit is out of the range. |
Check the setting for suction/discharge of Syringe Unit. |
548C |
SUBC: Leak is detected. [008C] |
SUBC:Leak sensor was detected leakage. |
Inspect the leakage from the flow line. Wipe out spilled liquid after fixing the leakage. |
5490 |
LINK TIME OUT (CBM). [0090] |
----- |
Check the power supply of the instrument and cable connections. Check the network environment. Restart CBM. If the error persists, contact your Shimadzu |
5491 |
SIL:No vial is set on the rack. [0091] |
No sample vial was placed in the rack position. |
Remedy this problem referring to the following procedure. 1. The sample vial installation location is incorrect. Place the vial at the correct location. 2. A probable cause is that a vial of height lower than a standard Shimadzu sample vial is in use. In this case, set the sample vial automatic detection function OFF referring to the Instruction Manual. If the above procedure does not remedy the problem, |
5492 |
SIL: Leak detected. Please eliminate the leak. [0092] |
Concentration of organic phase vapor inside the unit has exceeded the leak sensor actuation level. |
Check for liquid leakage from the following locations. 1. Injection port. If liquid is leaking from the injection port, replace the needle seal. 2. Valve piping joint. If liquid is leaking from the valve piping joint, re-tighten the piping, or, replace the piping and the joint parts with new parts. 3. Sample loop. If liquid is leaking from the sample loop, replace with new parts. 4. Leak sensor sensitivity adjustment. If the leak sensor is over-responding, calibrate the sensor, and re-adjust the threshold value. If the above procedure does not remedy the problem, contact your |
5493 |
SIL: PUMP DROP DETECTED. [0093] |
----- |
Turn the power OFF and contact your Shimadzu representative. |
54A0 |
Detector: Zero glass home position error (RID). [00A0] |
The home position of the zero glass drive unit cannot be properly detected. |
Contact your Shimadzu representative. |
54A1 |
Detector: Leak is detected. Please eliminate the leak. [00A1] |
Leak detected. |
Inspect and repair plumbing. Wipe away leakage. |
54A2 |
Detector: Insufficient lamp light intensity. [00A2] |
----- |
Check the total operating time of the lamp, and replace the lamp. If this does not remedy the problem, contact your Shimadzu representative. |
54A3 |
Detector: Over-heated (RID). |
Electric circuit was damaged. |
Contact your Shimadzu representative. |
54A4 |
Detector:Over-heated.(RID) [00A4] |
The temperature of the measuring point rises to 63C or more. |
Contact your Shimadzu representative. |
54B0 |
Controller: Leak is detected. Please eliminate the leak. [00B0] |
Leak was detected on an instrument. |
Remove the cause of leakage, and wipe off the liquid on the leak sensor. |
54C0 |
CDD:Valve1 home position error. [00C0] |
Valve is not operating properly due to one of the following reasons.\n(1) The valve is stuck due to dirt or precipitates.\n(2) The valve cable is not connected properly.\n(3) Problem in the valve motor or drive circuit. |
(1) Disassemble the valve and clean or replace the rotor and stator.\n(2) Check the connection of the cables.\n\nIf those are not the cause, contact your Shimadzu representative. |
54C1 |
CDD:Valve2 home position error. [00C1] |
Valve is not operating properly due to one of the following reasons.\n(1) The valve is stuck due to dirt or precipitates.\n(2) The valve cable is not connected properly.\n(3) Problem in the valve motor or drive circuit. |
(1) Disassemble the valve and clean or replace the rotor and stator.\n(2) Check the connection of the cables.\nIf those are not the cause, contact your Shimadzu representative. |
54C2 |
CDD:Suppressor Ch1 Voltage too high. [00C2] |
The actual current passing through the suppressor cartridge did not reach the preset value one minute after the start of regeneration. |
The suppressor cartridge may have become degraded. Replace the cartridge. |
54C3 |
CDD:Suppressor Ch2 Voltage too high. [00C3] |
The actual current passing through the suppressor cartridge did not reach the preset value one minute after the start of regeneration. |
The suppressor cartridge may have become degraded. Replace the cartridge. |
54C4 |
CDD:Ch1 Background is too high. [00C4] |
GAIN1 was set to 1 and the background level exceeded 100uS/cm when autozero was performed. |
If background level is more than 100 uS/cm, set GAIN1 to 2 or larger. |
54C5 |
CDD:Ch2 Background is too high. [00C5] |
GAIN2 was set to 1 and the background level exceeded 100uS/cm when autozero was performed. |
If background level is more than 100 uS/cm, set GAIN2 to 2 or larger. |
54D0 |
LC is not connected. |
LC is not connected. |
Check the instrument status and the cable connections. |
54D1 |
The CBM is not connected to the network. |
----- |
Check the power supply of the instrument and cable connections. Check the network environment. Restart CBM. If the error persists, contact your Shimadzu |
54D2 |
The CBM is not linked to the SPD-M20A. |
----- |
Check the connections between the PDA detector and CBM. Check the network environment. Restart the PDA detector, CBM, and PC. |
54D3 |
Configuration settings are not correct. |
Instrument configuration settings do not match with the actual installation. |
Check the instrument configuration and optional settings. |
54D9 |
Batch table setting error. Injection volume or vial number is not correct. |
Batch table settings are not correct. |
Check injection volume or vial number in the batch table. |
54DA |
LC system is not ready to start batch processing. |
LC system is not ready to start batch processing. |
Check the instrument status and restart the batch when it becomes ready. |
54E0 |
RS-232C communication error occurred. |
RS-232C communication error has occurred. |
Check the instrument status and the cable connections. |
54E1 |
SCSI communication error |
SCSI communication error has occurred. |
Check the instrument status and the cable connections. |
54E2 |
LC instrument error occurred. |
LC instrument error has occurred. |
Check the error code and take appropriate actions. |
54E3 |
Buffering error |
There is a buffer overflow. |
Use faster PC, or decrease the number of systems or detectors to decrease transmission traffic. |
54E4 |
Memory loss error |
Memory loss error has occurred. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
54E5 |
Maintenance logs error |
Failed to write to maintenance log. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
54E6 |
Error logs error |
Failed to write to error log. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
54E7 |
Operation error |
Failed to write to operation log. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
54E8 |
Failed to initialize the connection to the LC system. |
Failed to initialize a connection to the LC system. |
(1) Check the instrument status and the cable connections.\n(2) Check COM port number and controller type are set correctly in the configuration settings. |
54F1 |
There are no vials usable for fraction. |
----- |
----- |
54F2 |
Detector: The temperature sensor has failed.[8728] |
The temperature sensor in the detector has failed. |
Turn the power OFF and contact your Shimadzu representative. |
54F3 |
Detector: The PCB or other circuit has failed.[8729] |
The PCB or other circuit of the detector has failed. |
Turn the power OFF and contact your Shimadzu representative. |
5500 |
PDA is not connected. |
PDA is not connected. |
Check the instrument status and the cable connections. |
550C |
PDA for communication is not connected. |
PDA for communication is not connected. |
Check the instrument status and the cable connections. |
550D |
Failed to initialize the connection to the PDA. |
Failed to initialize a connection to the PDA. |
(1) Check the instrument status and the cable connections.\n(2) Check SCSI board number and SCSI ID are set correctly in the configuration settings. |
550E |
Failed to disconnect from the |
Failed to disconnect from the PDA. |
Check the instrument status and the cable connections. |
5511 |
This PDA detector cannot be used simultaneously from |
An attempt was made to use the PDA detector at the same time as CLASS-VP. |
Exit CLASS-VP when the PDA detector is used. |
5512 |
This operation is currently not available. |
You try to execute the operation which prohibit during data acquisition. |
Try again after terminating data acquisition. |
5513 |
This PDA detector is already used. |
This PDA detector is used elsewhere. |
Re-connect after disconnecting the destination where the PDA detector is used. |
5514 |
This PDA detector is already connected. |
This PDA detector is already connected. |
This PDA detector can be used. |
5515 |
Not connected to the PDA |
PDA board is not connected. |
Check the instrument status and the cable connections. |
5516 |
The PDA detector command was executed twice or more. |
The PDA detector is executing another command. |
Try again after the currently executing command ends. |
5517 |
Incorrect device code in PDA detector command response packet. |
An instrument different from the registered PDA detector is connected. |
Disconnect and re-connect. |
5518 |
The PDA detector command execution result status is in error. |
An attempt was made to execute command that cannot be processed by the PDA detector. |
Disconnect and re-connect. |
5519 |
Instrument type does not match. Re-configure the instrument. |
The actually connected PDA type does not match the PDA type set in the instrument properties in the system configuration. |
Retry setting the system configuration according to the type of currently connected PDA. |
551A |
The LC-2030 LC system is not configured. The LC-2030 LC system is necessary to use the LC-2030 PDA Detector. |
LC-2030 PDA Detector is selected without configuring LC-2030 LC system. |
Please double-click [Instrument (Communication Setting)] in the [System Configuration] sub-window, and select LC2030 LC system in [Instrument] sub-window. |
5520 |
SCSI communication error occurred. |
An error occurred in SCSI communication with the PDA detector. |
Check the power supply of the instrument, cable connections and terminator settings. |
5521 |
TCP/IP communication error occurred. |
An error occurred in TCP/IP communication with the PDA detector. |
Check the power supply of the instrument and cable connections. |
5550 |
PDA detector cannot start after the analysis runs. |
SPD-M10A analysis did not start. |
Make sure that the system controller relays are set appropriately in the system configuration, and that the cables are connected correctly. |
5551 |
PDA detector cannot start after the analysis runs. |
SPD-M20A analysis did not start. |
Before starting the PDA detector by an event cable connection, check the following details. \n1. Make sure that the system controller relays are set appropriately in the system configuration. \n2. Make sure that the controller and the PDA detector are correctly connected by the event cable. \n\n Before starting the PDA detector by an Ethernet connection, check the following details. \n1. Make sure that the controller and the PDA detector are in the same subnet. \n2. Make sure that the LAN cable is connected correctly to the controller and the PDA detector. \n3. Make sure that the same group name is set to the controller and the PDA detector. \n4. Make sure that the controller's system name is set to the master server name setting field |
5556 |
PDA signal saturated. The PDA exposure time may not be adjusted properly. |
The signal is saturated because the exposure time for the PDA elements is inappropriate. |
Perform exposure time optimization using SPD-M30A Utility. |
5557 |
Specified wavelength range is not available. The PDA may not be adjusted or calibrated properly. |
Wavelength calibration was not performed properly. |
Perform wavelength calibration using SPD-M30A utilities. |
5601 |
PDA Detector: System error. [0001] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5602 |
PDA Detector: System error. [0002] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5603 |
PDA Detector: System error. [0003] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5604 |
PDA Detector: System error. [0004] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5605 |
PDA Detector: System error. [0005] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5606 |
PDA Detector: EEPROM read error has occurred. [0006, 8902] |
EEPROM read error has occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5607 |
PDA Detector: EEPROM write error has occurred. [0007, 8903] |
EEPROM write error has occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5608 |
PDA Detector: System error. [0008] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5609 |
PDA Detector: System error. [0009] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
560A |
PDA Detector: System error. [000A] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
560B |
PDA Detector: System error. [000B] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
560C |
PDA Detector: System error. [000C] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
560D |
PDA Detector: System error. [0010] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
560E |
PDA Detector: Wavelength calibration was not performed. [0011, 8901] |
Wavelength calibration was not executed. |
Perform wavelength calibration. |
5611 |
PDA Detector: System error. [1001] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5612 |
PDA Detector: System error. [1002] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5613 |
PDA Detector: System error. [1010] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5614 |
PDA Detector: PDA board is not connected. [1020, 8911] |
The PDA circuit board is not connected. |
Turn the power OFF and contact your Shimadzu representative. |
5615 |
PDA Detector: Detector is not ready. [1030, 8912] |
Measurement was started while another error was unresolved. Due to network malfunction, etc., previous measurement was not completed normally. |
Resolve the previous error. Then restart the instrument. |
5616 |
PDA Detector: System error. [1040] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5617 |
PDA Detector: System error. [1050] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5621 |
PDA Detector: Leak is detected. [2001, 8921] |
Mobile phase has leaked inside the instrument. |
Check the plumbing. |
5622 |
PDA Detector: System error. [2010] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5623 |
PDA Detector: System error. [2011] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5631 |
PDA Detector: Communications to PC data buffer overflow. [4001, 8941] |
There is a PC data buffer overflow. |
Check the PC status and the cable connections. If the error persists, turn the power OFF and contact your Shimadzu representative. |
5632 |
PDA Detector: System error. [4002] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5633 |
PDA Detector: System error. [4010] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5634 |
PDA Detector: System error. [4011] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5635 |
PDA Detector: System error. [4012] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5636 |
PDA Detector: System error. [4020] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5637 |
PDA Detector: System error. [4021] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5641 |
PDA Detector: SCSI communication error. [7004] |
SCSI communication error has occurred. |
Check the instrument status and the cable connections. |
5642 |
PDA Detector: SCSI communication error. [7005] |
SCSI communication error has occurred. |
Check the instrument status and the cable connections. |
5643 |
PDA Detector: SCSI communication error. [7006] |
SCSI communication error has occurred. |
Check the instrument status and the cable connections. |
5644 |
PDA Detector: SCSI communication error. [7007] |
SCSI communication error has occurred. |
Check the instrument status and the cable connections. |
5645 |
PDA Detector: SCSI communication error. [7008] |
SCSI communication error has occurred. |
Check the instrument status and the cable connections. |
5646 |
PDA Detector: SCSI communication error. [7009] |
SCSI communication error has occurred. |
Check the instrument status and the cable connections. |
5647 |
PDA Detector: SCSI communication error. [700C] |
SCSI communication error has occurred. |
Check the instrument status and the cable connections. |
5648 |
PDA Detector: SCSI communication error. [700D] |
SCSI communication error has occurred. |
Check the instrument status and the cable connections. |
5649 |
PDA Detector: SCSI communication error. [700E] |
SCSI communication error has occurred. |
Check the instrument status and the cable connections. |
564A |
PDA Detector: SCSI communication error. [700F] |
SCSI communication error has occurred. |
Check the instrument status and the cable connections. |
564B |
PDA Detector: SCSI communication error. [7010] |
SCSI communication error has occurred. |
Check the instrument status and the cable connections. |
564C |
PDA Detector: SCSI communication error. [7011] |
SCSI communication error has occurred. |
Check the instrument status and the cable connections. |
564D |
PDA Detector: SCSI communication error. [7020] |
SCSI communication error has occurred. |
Check the instrument status and the cable connections. |
564E |
PDA Detector: SCSI communication error. [7021] |
SCSI communication error has occurred. |
Check the instrument status and the cable connections. |
564F |
PDA Detector: SCSI communication error. [7022] |
SCSI communication error has occurred. |
Check the instrument status and the cable connections. |
5650 |
PDA Detector: SCSI communication error. [7030] |
SCSI communication error has occurred. |
Check the instrument status and the cable connections. |
5651 |
PDA Detector: SCSI communication error. [7031] |
SCSI communication error has occurred. |
Check the instrument status and the cable connections. |
5652 |
PDA Detector: SCSI communication error. [7040] |
SCSI communication error has occurred. |
Check the instrument status and the cable connections. |
5653 |
PDA Detector: SCSI communication error. [7050] |
SCSI communication error has occurred. |
Check the instrument status and the cable connections. |
5654 |
PDA Detector: SCSI communication error. [7051] |
SCSI communication error has occurred. |
Check the instrument status and the cable connections. |
5655 |
PDA Detector: SCSI communication error. [7060] |
SCSI communication error has occurred. |
Check the instrument status and the cable connections. |
5656 |
PDA Detector: SCSI communication error. [7070] |
SCSI communication error has occurred. |
Check the instrument status and the cable connections. |
5661 |
PDA Detector: RS232C communication error. [8001] |
RS-232C communication error has occurred. |
Check the instrument status and the cable connections. |
5662 |
PDA Detector: RS232C communication error. [8002] |
RS-232C communication error has occurred. |
Check the instrument status and the cable connections. |
5663 |
PDA Detector: RS232C communication error. [8003] |
RS-232C communication error has occurred. |
Check the instrument status and the cable connections. |
5664 |
PDA Detector: System error. [9001] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5671 |
PDA Detector: Changing lamp condition. [C001, 89C1] |
Changing lamp condition. |
Then restart PDA Detector and PC. |
5672 |
PDA Detector: D2 lamp is over current. [C002, 89C2] |
D2 lamp malfunctioned. Lighting circuit malfunctioned. |
Replace the D2 lamp. If this does not remedy the problem, contact your Shimadzu representative. |
5673 |
PDA Detector: There is a problem with the D2 lamp. Check the lamp connections and the lamp for defects. [C003, 89C3] |
Probable causes are the D2 lamp connector is not connected, the D2 lamp malfunctioned, or the lighting circuit malfunctioned. |
Check the D2 lamp connections. Check the total operating time of the lamp, and replace the lamp. If this does not remedy the problem, contact your Shimadzu representative. |
5674 |
PDA Detector: System error. [C004] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5675 |
PDA Detector: System error. [C005] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5676 |
PDA Detector: There is a problem with the W lamp. Check the lamp connections and the lamp for defects. [C006, 89C4] |
Probable causes are the W lamp connector is not connected, or the W lamp malfunctioned. |
Check the W lamp connections. Check the total operating time of the lamp, and replace the lamp. If this does not remedy the problem, contact your Shimadzu representative. |
5677 |
PDA Detector: Starting up D2 lamp. [C007, 89C5] |
Starting up D2 lamp. |
Then restart PDA Detector and PC. |
5678 |
PDA Detector: System error. [C008] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5679 |
PDA Detector: System error. [C009] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
567A |
PDA Detector: System error. [C00B] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
567B |
PDA Detector: The light source compartment cover is open. Close the cover. [C00C, 89C6] |
The lamp cover was not in place at instrument startup. |
Mount the lamp cover, and turn the lamp on again. |
567C |
PDA Detector: D2 lamp heater is over current. [C00D, 89C7] |
A probable cause is that the D2 lamp heater circuit malfunctioned. |
Check the total operating time of the lamp, and replace the lamp. If this does not remedy the problem, contact your Shimadzu representative. |
567D |
PDA Detector: Lamp housing is overheated. [C00E, 89C8] |
A probable cause is that the cooling fan blast flow is poor, or the cooling fan or heater circuit malfunctioned. |
Ensure at least 15 mm of space on the sides of the PDA detector. If this does not remedy the problem, contact your Shimadzu representative. |
567E |
PDA Detector: W lamp is overcurrent. [C00F, 89C9] |
A probable cause is that the W lamp malfunctioned. |
Check the total operating time of the lamp, and replace the lamp. If this does not remedy the problem, contact your Shimadzu representative. |
5681 |
PDA Detector: System error. [D001] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5682 |
PDA Detector: Failed to detect light shutter/filter origin. [D002, |
Failed to detect light shutter/filter origin. |
Turn the power OFF and contact your Shimadzu representative. |
5683 |
PDA Detector: System error. [D003] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5691 |
PDA Detector: System error. [E001] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5692 |
PDA Detector: Parameter set error; co-efficient of elementwave. [E002, 89E1] |
Parameter set error; co-efficient of element-wave. |
Then restart PDA Detector and PC. |
5693 |
PDA Detector: System error. [E003] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5694 |
PDA Detector: System error. [E010] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5695 |
PDA Detector: System error. [E011] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5696 |
PDA Detector: System error. [E020] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
56A1 |
PDA Detector: System error. [FF01] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
56A2 |
PDA Detector: System error. [FF02] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
56A3 |
PDA Detector: System error. [FF03] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
56A4 |
PDA Detector: System error. [FF04] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
56A5 |
PDA Detector: System error. [FF05] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
56A6 |
PDA Detector: System error. [FF06] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
56A7 |
PDA Detector: System error. [FF07] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
56A8 |
PDA Detector: System error. [FF08] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
56A9 |
PDA Detector: System error. [FF20] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
56AA |
PDA Detector: System error. [FF21] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
56AB |
PDA Detector: System error. [FF80] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
56AC |
PDA Detector: System error. [FF81] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
56AD |
PDA Detector: System error. [FF82] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
56B1 |
PDA Detector: Invalid EEPROM parameter. [0013, 8904] |
Invalid EEPROM parameter. |
Turn the power OFF and contact your Shimadzu representative. |
56B5 |
PDA Detector: System check error has occurred. [2100, 8922] |
Error occurred during system check. |
Turn the power OFF and contact your Shimadzu representative. |
56B7 |
PDA Detector: Hard disk cannot be initialized. [4031, 8942] |
Hard disk initialization failed. |
Turn the power OFF and contact your Shimadzu representative. |
56B8 |
PDA Detector: Insufficient hard disk space. [7032, 8943] |
Hard disk is full. |
Turn the power OFF and contact your Shimadzu representative. |
56B9 |
PDA Detector: Buffer memory overflow in detector during analysis. It is not possible to send the data again. [4003, 8944] |
Detector buffer memory overflowed. |
Turn the power OFF and contact your Shimadzu representative. |
56BA |
PDA Detector: SNTP synchronization error. [5001, 8951] |
The time error with the SNTP server exceeds 1 second. The time is not correctly adjusted. |
Check the power supply of the instrument and cable connections. Check the network environment. Restart the PDA detector. If the error persists, contact your Shimadzu representative. |
56BB |
PDA Detector: TCP/IP communications timed out. [5002, 8952] |
Ethernet communication with the PC timed out. |
Check the power supply of the instrument and cable connections. Restart the PDA detector. If the error persists, contact your Shimadzu representative. |
56BC |
PDA Detector: The cell is overheated. [6001, 8961] |
Temperature-controlled flow cell malfunctioned, control circuit |
Turn the power OFF and contact your Shimadzu representative. |
56BD |
PDA Detector: The cell is over current. [6002, 8962] |
Temperature-controlled flow cell malfunctioned, control circuit |
Turn the power OFF and contact your Shimadzu representative. |
56BE |
PDA Detector: Invalid slit position. [7101, 8971] |
Invalid slit position. |
Then restart PDA Detector and PC. |
56BF |
PDA Detector: Invalid response. [7102, 8972] |
Invalid response. |
Then restart PDA Detector and PC. |
56C0 |
PDA Detector: Invalid sampling rate. [7103, 8973] |
Invalid sampling rate. |
Then restart PDA Detector and PC. |
56C1 |
PDA Detector: Invalid cell temperature. [7104, 8974] |
Invalid cell temperature. |
Then restart PDA Detector and PC. |
56C2 |
PDA Detector: Invalid analog range. [7106, 8976] |
Invalid analog range. |
Then restart PDA Detector and PC. |
56C3 |
PDA Detector: Invalid polarity. [7106, 8976] |
Invalid polarity. |
Then restart PDA Detector and PC. |
56C4 |
PDA Detector: Invalid wavelength. [7107, 8977] |
Invalid wavelength. |
Then restart PDA Detector and PC. |
56C5 |
PDA Detector: Leak sensor is short-circuited. [7881, 8978] |
Leak sensor is short. |
Turn the power OFF and contact your Shimadzu representative. |
56C6 |
PDA Detector: Leak sensor is open. [7882, 8979] |
Leak sensor is open. |
Turn the power OFF and contact your Shimadzu representative. |
56C7 |
PDA Detector: Lamp housing temperature sensor is open. [7883, 897A] |
Lamp housing temperature sensor is open. |
Turn the power OFF and contact your Shimadzu representative. |
56D2 |
PDA Detector: D2 lamp heater current is low. [C010, 89CA] |
A probable cause is that the D2 lamp heater circuit malfunctioned. |
Turn the power OFF and contact your Shimadzu representative. |
56D3 |
PDA Detector: Failed to detect D2 lamp. [C012, 89CB] |
The D2 lamp connector is not connected. Or, a wire is cut, or the circuit board malfunctioned, etc. |
Connect the D2 lamp connector. |
56D4 |
PDA Detector: Failed to detect W lamp. [C013, 89CC] |
The W lamp connector is not connected. Or, a wire is cut, or the circuit board malfunctioned, etc. |
Connect the W lamp connector. |
56D6 |
PDA Detector: System error. [89FE] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
56D7 |
PDA Detector: TCP/IP cancelled. [5003, 8953] |
An error has occurred during TCP/IP communication with PDA detector. |
Verify that the instrument is turned ON and the cables properly connected.\nIf the error is not corrected, contact your Shimadzu representative. |
56D8 |
PDA Detector: CBM LAN timeout. [5004, 8954] |
Ethernet communication with the CBM timed out. |
Check the network environment.\nTurn the power of PDA detector and CBM OFF once and turn it ON again.\nIf the error persists, turn the power OFF and contact your Shimadzu representative. |
56D9 |
PDA Detector: SNTP timeout error has occurred. [5005, 8955] |
SNTP timeout error has occurred. |
Check the network instruments. Then restart PDA Detector and PC. |
56DE |
PDA Detector: Failed to capture the PDA signal.[89E2] |
Malfunction inside the instrument |
Turn the power OFF and contact your Shimadzu representative. |
56E0 |
PDA Detector: Low set temperature.[8963] |
The set temperature is not appropriate. |
Set the temperature more than 5C higher than the ambient temperature. |
56E1 |
PDA Detector: Installed cell is not registered. Register the cell ID in PDA detector, then adjust the exposure time and calibrate the wavelength.[8965] |
1)An unregistered cell was installed in the instrument.\n2)Failed to read cell ID. |
1)Perform cell registration,exposure time adjustment, and wavelength calibration.\n2)Check the connection of the cell cable. |
56E2 |
PDA Detector: Temperature is increasing inside the instrument.[89D2] |
PDA Detector: Temperature is increasing inside the instrument. |
If the filter is clogged, replace the filter. |
56E3 |
PDA Detector: Lamp housing fan rotation error.[89D6] |
Lamp compartment fan malfunction |
Turn the power OFF and contact your Shimadzu representative. |
56E4 |
PDA Detector: Power heat exhaust fan rotation error.[89D7] |
Power heat exhaust fan malfunction |
Turn the power OFF and contact your Shimadzu representative. |
56E5 |
PDA Detector: Polychrometor is overheated.[89D8] |
Malfunction inside the instrument |
Turn the power OFF and contact your Shimadzu representative. |
56E6 |
PDA Detector: Heater sensor is open.[89D9] |
Malfunction inside the instrument |
Turn the power OFF and contact your Shimadzu representative. |
56E7 |
PDA Detector: Room sensor is open.[89DA] |
Malfunction inside the instrument |
Turn the power OFF and contact your Shimadzu representative. |
56E8 |
PDA Detector: Photo diode is saturated.[8945] |
The signal is saturated because the exposure time for the PDA elements is |
Perform exposure time optimization using SPD-M30A Utility. |
56E9 |
PDA Detector: IP resource contention.[8956] |
A device with the same IP address as the instrument exists on the same network. |
Change the IP address of the instrument or device to resolve the collision. |
56EA |
PDA Detector: Failed to detect Cell.[8964] |
Cell is not installed or the cell cable is not connected. |
Install the cell and connect the cable correctly. |
56EB |
PDA Detector: Abnormal high temperature is detected inside the instrument[89D3] |
1)The suction filter on the back of the instrument is clogged.\n2)Malfunction inside the instrument |
1)If the filter is clogged, replace the filter.\n2)Turn the power OFF and contact your Shimadzu representative. |
56EC |
PDA Detector: D2 lamp overvoltage.[89D4] |
1)D2 lamp malfunction\n2)Malfunction inside the instrument |
1)Replace the D2 lamp.\n2)Turn the power OFF and contact your Shimadzu representative. |
56ED |
PDA Detector: D2 lamp overvoltage[89D5] |
1)D2 lamp malfunction\n2)Malfunction inside the instrument |
1)Replace the D2 lamp.\n2)Turn the power OFF and contact your Shimadzu representative. |
56EE |
PDA Detector: Undefined warning.[89FB] |
A warning occurred in the PDA detector that is undefined in LabSolutions. |
Refer to the SPD-M30A Instruction Manual regarding error codes. If the LabSolutions version is not compatible with the PDA detector ROM version, update it to the most recent version. |
56EF |
PDA Detector: Undefined error(Level 1).[89FC] |
An error occurred in the PDA detector that is undefined in LabSolutions. |
Refer to the SPD-M30A Instruction Manual regarding error codes. If the LabSolutions version is not compatible with the PDA detector ROM version, update it to the most recent version. |
56F0 |
PDA Detector: Undefined error(Level 2).[89FD] |
An error occurred in the PDA detector that is undefined in LabSolutions. |
Refer to the SPD-M30A Instruction Manual regarding error codes. If the LabSolutions version is not compatible with the PDA detector ROM version, update it to the most recent version. |
56F1 |
PDA Detector: Undefined error(Level 3).[89FE] |
An error occurred in the PDA detector that is undefined in LabSolutions. |
Refer to the SPD-M30A Instruction Manual regarding error codes. If the LabSolutions version is not compatible with the PDA detector ROM version, update it to the most recent version. |
56F2 |
PDA Detector: Undefined fatal error.[89FF] |
An error occurred in the PDA detector that is undefined in LabSolutions. |
Refer to the SPD-M30A Instruction Manual regarding error codes. If the LabSolutions version is not compatible with the PDA detector ROM version, update it to the most recent version. |
56F3 |
PDA Detector: The CBM start signal is abnormal.[8957] |
A malfunction was detected while synchronizing time between the PDA detector and the CBM-20A. |
Confirm the network environment (cables, hubs, routers, etc.) and the ROM version of the CBM controller specified on the master server. If that does not resolve the problem, contact your Shimadzu representative. |
56F4 |
The exposure time of the installed cell is not adjusted. Set the exposure time. |
The exposure time of the installed cell is not adjusted. |
Set the exposure time. |
56F5 |
The wavelength of the installed cell is not calibrated. Perform wavelength calibration. |
The wavelength of the installed cell is not calibrated. |
Perform wavelength calibration. |
5808 |
Signal received from ALARM INPUT terminal. [2008] |
There is an error in the input of the ALARM INPUT terminal. |
Resolve the error according to instructions includeswith the external device connected to the terminal. |
580F |
Nonexistent vial is selected. Verify the rack type and vial number, then restart analysis. |
Nonexistent vial has been selected. |
Verify the rack type and vial number, then restart analysis. |
5810 |
Cannot detect needle home position in X (left-right) direction. Please press the RINSE button. |
Debris exists in XYZ direction for needle, or needle malfunction. |
Rinse theAutosampler._If this does not correct the problem, check whether debris exists or not.\nIf the error persists, turn the power OFF and contact your Shimadzu |
5811 |
Cannot detect needle home position in Y (front-back) direction. Please press the |
Debris exists in XYZ direction for needle, or needle malfunction. |
Rinse theAutosampler._If this does not correct the problem, check whether debris exists or not.\nIf the error persists, turn the power OFF and contact your Shimadzu |
5812 |
Cannot detect needle home position in Z (up-down) direction. Please press the RINSE button. |
Debris exists in XYZ direction for needle, or needle malfunction. |
Rinse theAutosampler._If this does not correct the problem, check whether debris exists or not.\nIf the error persists, turn the power OFF and contact your Shimadzu |
5813 |
Cannot detect low-pressure valve home position. Please press the RINSE button. |
Low-pressure valve or control board malfunction. |
Remedy this problem referring to the following procedure. 1. Check if the rotor seal has reached its service life. Replace the rotor seal if it has reached its service life. 2. Disassemble and clean. Sonicate the rotor seal ant remove the wear particles. (At this time, replace scratched seals with new seals.) If the above procedure does not remedy the problem, contact your Shimadzu representative. |
5819 |
An undefined rack was detected. Please set the rack correctly. [2119] |
Autosampler failed to detect the rack, because the rack has not been correctly installed to the back of rack plate. |
Check the rack No., and set the correct No. When using a user-defined rack, first perform teaching to set the rack information. |
581A |
Cannot detect high-pressure valve home position. Please press the RINSE button. |
Cannot detect high-pressure valve home position. |
Remedy this problem referring to the following procedure. 1. Check if the rotor seal has reached its service life. Replace the rotor seal if it has reached its service life. 2. Disassemble and clean. Sonicate the rotor seal ant remove the wear particles. (At this time, replace scratched seals with new seals.) If the above procedure does not remedy the problem, contact your Shimadzu representative. |
581B |
Cannot detect metering pump home position. Please press the RINSE button. [211B] |
Metering pump or control board malfunction. |
Rinse theAutosampler. If this does not correct the problem, turn the power OFF and contact your Shimadzu representative. |
581D |
Nonexistent vial is selected. Verify the rack type and vial number, then restart analysis. |
Nonexistent vial has been selected. |
Verify the rack type and vial number, then restart analysis. |
581F |
Cannot detect metering pump rotation sensor. Please press the RINSE button. [211F] |
Metering pump or control board malfunction. |
Remedy this problem referring to the following procedure. Check if the plunger seal has reached its service life. Replace the plunger seal if it has reached its service life. The torque required for turning the valve sometimes increases due to foreign objects in the rinse solution and impurities in the sample. Remove these foreign matters by disassembling and cleaning the valve. At this time, replace scratched seals with new seals. If the above procedure does not remedy the problem, contact your Shimadzu |
5821 |
No vial is set in the specified vial number. After rinsing, set the vial and restart analysis. [2191] |
1.No vial is set in the specified vial number.\n2. Cannot detect vial as its height is too short. |
Remedy this problem referring to the following procedure. 1. Set the vial to the specified rack position. 2. A probable cause is that a vial of height lower than a standard Shimadzu sample vial is in use. In this case, set the sample vial automatic detection function OFF referring to the Instruction Manual. If the above procedure does not remedy the problem, contact your Shimadzu |
5822 |
Autosampler unit leak sensor has detected a leak. [2192] |
There is a leak at a flow line connection of autosampler or the valve rotor seal. |
Check for liquid leakage from the following locations. 1. Injection port. If liquid is leaking from the injection port, replace the needle seal. 2. Valve piping joint. If liquid is leaking from the valve piping joint, re-tighten the piping, or, replace the piping and the joint parts with new parts. 3. Sample loop. If liquid is leaking from the sample loop, replace with new parts. 4. Leak sensor sensitivity adjustment. If the leak sensor is over-responding, calibrate the sensor, and re-adjust the threshold value. If the above procedure does not remedy the problem, contact your |
5824 |
Debris detected on needle. Please remove the debris from the inside of the Autosampler. [2194] |
Needle tip detects strong force. |
Remedy this problem referring to the following procedure. 1. Replace the needle.\n2. If the septum is not Shimadzu genuine parts, replace to Shimadzu genuine parts or replace softer one.\n3. Verify that the sample rack is installed correctly. \nIf the above procedure does not remedy the problem, contact your Shimadzu |
5825 |
The XY (front-back-left-right) drive unit encoder detected abnormal values. [2195] |
The XY (front-back-left-right) drive unit encoder or encoder slit may be defective. |
Contact your Shimadzu representative. |
5826 |
Autosampler door is open during analysis. Injections are not performed while the door is open. [2196] |
Autosampler door is open during analysis. |
Close theAutosampler door. |
5829 |
Needle malfunctions in X (leftright) direction. Please press the RINSE button. [2199] |
Debris exists in XYZ direction for needle, or needle malfunction. |
Rinse theAutosampler. If this does not correct the problem, check whether debris exists or not.\nIf the error persists, turn the power OFF and contact your Shimadzu |
582A |
Needle malfunctions in Y (frontback) direction. Please press the RINSE button. [219A] |
Debris exists in XYZ direction for needle, or needle malfunction. |
Rinse theAutosampler. If this does not correct the problem, check whether debris exists or not.\nIf the error persists, turn the power OFF and contact your Shimadzu |
582B |
Needle malfunctions in Z (updown) direction. Please press the RINSE button. [219B] |
Debris exists in XYZ direction for needle, or needle malfunction. |
Rinse theAutosampler. If this does not correct the problem, check whether debris exists or not.\nIf the error persists, turn the power OFF and contact your Shimadzu |
582C |
High-pressure valve malfunctions. Please press the RINSE button. [219C] |
High-pressure valve malfunction. |
Remedy this problem referring to the following procedure. 1. Check if the rotor seal has reached its service life. Replace the rotor seal if it has reached its service life. 2. Disassemble and clean. Sonicate the rotor seal remove the wear particles. (At this time, replace scratched seals with new seals.) If the above procedure does not remedy the problem, contact your Shimadzu representative. |
582D |
Low-pressure valve malfunctions. Please press the RINSE button. [219D] |
Low-pressure valve malfunction. |
Remedy this problem referring to the following procedure. 1. Check if the rotor seal has reached its service life. Replace the rotor seal if it has reached its service life. 2. Disassemble and clean. Sonicate the rotor seal remove the wear particles. (At this time, replace scratched seals with new seals.) If the above procedure does not remedy the problem, contact your Shimadzu representative. |
582E |
Metering pump malfunctions. Please press the RINSE button. |
Metering pump malfunction. |
Rinse theAutosampler. If this does not correct the problem, turn the power OFF and contact your Shimadzu |
582F |
Different types of racks are being used for the left and right racks. [219F] |
Different types of racks are being used for the left and right racks. |
Verify that the type of racks are same and slide the rack along the guides all the way to the back of the instrument. |
5830 |
Pump outlet pressure has exceeded specified P.Max. [2230] |
Pump outlet pressure has exceeded specified maximum pressure value P.Max. |
Check the flow line for clogging. If no clogging is found, the P.Max setting is inappropriate, and must be changed. |
5831 |
Pump outlet pressure is lower than specified minimum pressure value P.Min. [2231] |
Pump outlet pressure is lower than specified minimum pressure value P.Min. |
Check the flow line for leaks. If no leak is found, the P.Min setting is inappropriate, and must be changed. |
5832 |
Sensor cannot detect the home position for the pump drive motor. [2232] |
There is a possibility problem occurs in driving unit of solvent pump. |
Turn the power off and then on again, and press [Pump]. If the error message is displayed again, turn off the power and contact your Shimadzu representative.\nIf this error often occurs, turn the power OFF and contact your Shimadzu representative. |
5834 |
Pump unit leak sensor has detected a leak. [2234] |
Pump unit leak sensor has detected a leak. |
(1) See if there is a leak in the pump or degasser. Find and repair the leak, and dry the area around the leak sensor as necessary.\n(2) See if the sensitivity is set incorrectly. Adjust the leak sensor sensitivity as necessary. |
5835 |
Pump outlet pressure has exceeded 2 MPa during purge. |
Pump outlet pressure has exceeded 2 MPa during purge. |
Locate the clogged section of the tubing between the pump to theAutosampler's needle. |
5836 |
DIV Error. [2236] |
----- |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
5837 |
GEV A Error. [2237] |
----- |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
5838 |
GEV B Error. [2238] |
----- |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
5839 |
GEV C Error. [2239] |
----- |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
583A |
GEV D Error. [223A] |
----- |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
5840 |
Grating motor home position detector is not operating properly. [2340] |
Grating motor home position detector is not operating properly.Motor or control board may be defective. |
Contact your Shimadzu representative. |
5841 |
Filter motor home position detector is not operating properly. [2341] |
Filter motor home position detector is not operating properly.Motor or control board may be defective. |
Contact your Shimadzu representative. |
5843 |
D2 lamp is not detected. [2343] |
D2 lamp is not installed. |
(1) Verify that the D2 lamp is properly installed.\n(2) The D2 lamp connector may be disconnected. |
5844 |
D2 lamp is over current. [2344] |
Excessive current was applied to the D2 lamp.The consumable counters of D2 lamp may exceed the criteria. |
Replace the D2 lamp. Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu representative. |
5845 |
A lighting or circuit problem has occurred in the D2 lamp. [2345] |
A lighting or circuit problem has occurred in the D2 lamp.The consumable counters of D2 lamp may exceed the criteria, or control board may be defective. |
Replace the D2 lamp. Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu representative. |
584A |
The wavelength shift is greater than 1 nm. [234A] |
The wavelength shift is greater than 1 nm. |
1. Verify that cell is securely installed.\n2. Either the sample in the cell strongly absorbs visible light, or there are large bubbles in the cell. , So that the amount of light transmitted through the cell is extremely small at the D2 lamp's lambda maximum wavelength (656 nm).\n3. If there is an error in the wavelength calibration, perform wavelength calibration.\nIf the error persists, turn the power OFF and contact your Shimadzu representative. |
584B |
Internal fan (located at bottomcenter of instrument) is not operating properly. [234B] |
Internal fan (located at bottom-center of instrument) is not operating properly. The temperature of the lamp power supply has become abnormally high. |
Contact your Shimadzu representative. |
5855 |
Unable to detect cell. Verify that the cell is securely installed. |
The detector cell is not installed. |
Verify that the cell is securely installed. |
5856 |
Lamp housing cover is open. Verify that the lamp housing cover is closed. [23A6] |
Lamp housing cover is open. |
Verify that the lamp housing cover is closed. |
5858 |
Detector power system (board) is defective. [23A8] |
Detector power system (board) is defective.A/D converter may be defective. |
Contact your Shimadzu representative. |
5859 |
Internal fan (at rear of lamp housing) is not operating properly. Temperature becomes too high. [23A9] |
Internal fan (at rear of lamp housing) is not operating properly. The temperature inside the lamp housing has become abnormally high. |
Contact your Shimadzu representative. |
585A |
Internal mercury lamp is defective. An error occurred during wavelength check. [23AA] |
Internal mercury lamp is defective. A malfunction occurred during wavelength check. |
Perform calibration again and wavelength check. If this does not correct the problem, contact your Shimadzu representative. |
585B |
An error occurred during span checking. [23AB] |
A malfunction occurred during span check.Motor or control board may be defective. |
Contact your Shimadzu representative. |
5860 |
Oven temperature has exceeded T.Max. [2470] |
Oven temperature has exceeded T.Max value.\nTemperature sensor or control board may be defective, if the set value is correctly. |
Lower the set value for oven temperature, or set T.Max to a value 5C higher than the oven temperature value. (This error often occurs when the oven temperature initially overshoots the set temperature.) |
5861 |
The oven temperature detection value is too high. [2471] |
Abnormally high temperature detected during conversion by A/D converter.Temperature sensor is short circuit, or A/D converter may be defective. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu representative. |
5862 |
The oven temperature detection value is too low. [2472] |
Abnormally high temperature detected during conversion by A/D converter.Temperature sensor is open circuit, or A/D converter may be defective. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu representative. |
5863 |
Oven leak error [2473] |
----- |
Check for liquid leakage along the flow_path, etc., and wipe off any liquid around the leak sensor. If there is no liquid leakage, there is a problem in adjustment of the leak sensor sensitivity. Adjust the leak sensor sensitivity referring to the Instruction Manual. |
5867 |
Set temperature is less than 15 C below room temperature. [2477] |
Set temperature value is inappropriate, in one of following ways:\n(1) When use of cooler is enabled, set temperature is room temperature minus 10C or lower.\n(2) In other situations, set temperature is lower than the room temperature. |
Set temperature higher than room temperature minus 10C if using the cooler, or higher than room temperature if not using the cooler.\n(Temperature regulation is not stopped.) |
586A |
The room temperature detection value is too high. [247A] |
Abnormally high temperature detected during conversion by A/D converter._Temperature sensor is short circuit, or A/D converter may be defective. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu representative. |
586B |
The room temperature detection value is too low. [247B] |
Abnormally high temperature detected during conversion by A/D converter._Temperature sensor is open circuit, or A/D converter may be defective. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu representative. |
586C |
An unexpectedly high temperature was detected by the oven temperature sensor. [247C] |
Abnormally high temperature detected in temperature sensor._Temperature sensor is short circuit, or A/D converter may be defective. |
Turn the power OFF and contact your Shimadzu representative. |
586D |
The current value flowing to the Peltier unit in the oven unit is too low. [247D] |
Current flowing in column oven Peltier unit is abnormally low._Peltier fan or control board may be defective. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu representative. |
586E |
Column oven door is open. [247E] |
Column oven door was open for too long time while using suppressor. |
Close the door. |
5870 |
Sample rack temperature has exceeded T.Max. [25C0] |
Sample rack temperature has exceeded T.Max value.\nTemperature sensor or control board may be defective. |
Contact your Shimadzu representative. |
5871 |
Abnormally high temperature detected during conversion by A/D converter. [25C1] |
Abnormally high temperature detected during conversion by A/D converter._Temperature sensor is short circuit, or A/D converter may be defective. |
Contact your Shimadzu representative. |
5872 |
Abnormally low temperature detected during conversion by A/D converter. [25C2] |
Abnormally high temperature detected during conversion by A/D converter._Temperature sensor is open circuit, or A/D converter may be defective. |
Contact your Shimadzu representative. |
587A |
Abnormally high temperature detected in dehumidifier. [25CA] |
Abnormally high temperature detected during conversion by A/D converter._Temperature sensor is short circuit, or A/D converter may be defective. |
Contact your Shimadzu representative. |
587B |
Abnormally low temperature detected in dehumidifier. [25CB] |
Abnormally high temperature detected during conversion by A/D converter._Temperature sensor is open circuit, or A/D converter may be defective. |
Contact your Shimadzu representative. |
587C |
Abnormal high temperature of rack detected by temperature sensor. |
Abnormally high rack temperature detected in temperature sensor._Temperature sensor is short circuit, or A/D converter may be defective. |
Contact your Shimadzu representative. |
587D |
Current flowing in column sample rack Peltier unit is abnormally low. [25CD] |
Current flowing in dehumidifier Peltier unit is abnormally low._Peltier fan or control board may be defective. |
Contact your Shimadzu representative. |
587E |
Current flowing in column dehumidifier Peltier unit is abnormally low. [25CE] |
Current flowing in column dehumidifier Peltier unit is abnormally low. |
Contact your Shimadzu representative. |
588A |
A seal malfunction has occurred in the online degasser vacuum line. [26BA] |
Vacuum pressure of degasser connected to this instrument has been abnormal for certain time.\nDegasser module is defective. |
If [NG] is indicated for the vacuum pressure, turn the power OFF then ON again. If [OK] is indicated for the vacuum pressure, there is no problem since the online degasser is automatically restored. |
5891 |
Power supply cooling fan has stopped. [27B1] |
Power supply cooling fan has stopped._Fan cable or fan rotation sensor may be defective. |
Contact your Shimadzu representative. |
5892 |
Pump cooling fan has stopped. [27B2] |
Pump cooling fan has stopped._Fan cable or fan rotation sensor may be defective. |
Contact your Shimadzu representative. |
5893 |
Column oven cooling fan has stopped. [27B3] |
Column oven cooling fan has stopped._Fan cable or fan rotation sensor may be defective. |
Contact your Shimadzu representative. |
5894 |
Autosampler cooling fan has stopped. [27B4] |
Autosampler cooling fan has stopped._Fan cable or fan rotation sensor may be defective. |
Contact your Shimadzu representative. |
5895 |
Sample rack cooling fan has stopped. [27B5] |
Sample rack cooling fan has stopped._Fan cable or fan rotation sensor may be defective. |
Contact your Shimadzu representative. |
58A0 |
Power board is not detected. [28D0] |
Power board was not detected._Cable or control board may be defective. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
58A2 |
UV Sensor board is not detected. [28D2] |
UV sensor board was not detected._Cable or control board may be defective. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
58A3 |
Degasser is not detected. [28D3] |
Degasser was not detected._Cable or control board may be defective. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
58A4 |
Pump board 1 is not detected. [28D4] |
Pump board 1 was not detected._Cable or control board may be defective. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
58A5 |
Pump board 2 is not detected. [28D5] |
Pump board 2 was not detected._Cable or control board may be defective. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
58A6 |
Oven board is not detected. [28D6] |
Oven board was not detected._Cable or control board may be defective. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
58A7 |
Autosampler board 1 is not detected. [28D7] |
Autosampler board 1 was not detected._Cable or control board may be defective. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu representative. |
58A8 |
Autosampler board 2 is not detected. [28D8] |
Autosampler board 2 was not detected._Cable or control board may be defective. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu representative. |
58A9 |
Sample Cooler board is not detected. [28D9] |
Sample cooler board was not detected._Cable or control board may be defective. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu representative. |
58B1 |
Mobile phase A is under the alarm level. [2901] |
The level of mobile phase has gotten too low. |
Replace the mobile phase. |
58B2 |
Mobile phase B is under the alarm level. [2902] |
The level of mobile phase has gotten too low. |
Replace the mobile phase. |
58B3 |
Mobile phase C is under the alarm level. [2903] |
The level of mobile phase has gotten too low. |
Replace the mobile phase solution. |
58B4 |
Mobile phase D is under the alarm level. [2904] |
The level of mobile phase has gotten too low. |
Replace the mobile phase solution. |
58B5 |
Rinse liquid is under the alarm level. [2905] |
The level of rinse liquid has gotten too low. |
Replace the rinse liquid. |
58C0 |
Failed to write maintenance log. [3001] |
Cannot write to maintenance log._Control board may be defective. |
Turn the power OFF and contact your Shimadzu representative. |
58C1 |
Failed to read maintenance log. [3002] |
Cannot read maintenance log._Control board may be defective. |
Turn the power OFF and contact your Shimadzu representative. |
58C2 |
The maintenance log is full and log data cannot be entered. |
The maintenance log is full and log data cannot be entered. |
Turn the power OFF and then ON again. The full log is initialized. |
58C8 |
Failed to write error log. [3011] |
Cannot write to error log._Control board may be defective. |
Turn the power OFF and contact your Shimadzu representative. |
58C9 |
Failed to read error log. [3012] |
Cannot read error log._Control board may be defective. |
Turn the power OFF and contact your Shimadzu representative. |
58CA |
The error log is full and log data cannot be entered. [3013] |
The error log is full and log data cannot be entered. |
Turn the power OFF and then ON again. The full log is initialized. |
58D0 |
Failed to write operation log. [3021] |
Cannot write to operation log._Control board may be defective. |
Turn the power OFF and contact your Shimadzu representative. |
58D1 |
Failed to read operation log. [3022] |
Control board may be defective._Control board may be defective. |
Turn the power OFF and contact your Shimadzu representative. |
58D2 |
The operation log is full and log data cannot be entered. [3023] |
The operation log is full and log data cannot be entered. |
Turn the power OFF and then ON again. The full log is initialized. |
58D8 |
Failed to write validation log. [3031] |
Cannot write to validation log._Control board may be defective. |
If the log function is necessary, turn the power OFF and contact your Shimadzu representative. |
58D9 |
Failed to read validation log. [3032] |
Cannot read validation log._Control board may be defective. |
If the log function is necessary, turn the power OFF and contact your Shimadzu representative. |
58DA |
The validation log is full and log data cannot be entered. [3033] |
The validation log is full and log data cannot be entered. |
Turn the power OFF and then ON again. The full log is initialized, and log data can once again be entered. |
58E0 |
Writing error with ROM for saving detector parameters [3051] |
----- |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
58E1 |
Reading error with ROM for saving detector parameters |
----- |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
58E4 |
Invalid pump adjustment parameters have been initialized. [3061] |
Invalid pump adjustment parameters have been initialized.\nThis warning is rarely occurred, if the instrument has been turned off when writing the adjustment |
It is necessary to adjust again.\nTurn the power OFF and contact your Shimadzu representative. |
58E5 |
Invalid pump calibration data have been initialized. [3062] |
Invalid pump calibration data have been initialized.\nThis warning is rarely occurred, if the instrument has been turned off when writing the calibration data. |
Press the CE key to clear,andthen view the System Check screen. Correct the necessary pump items. |
58E6 |
The usage history of the pump consumables has been initialized to "0" due to an abnormality. |
The usage history of the solvent delivery pump consumables was initialized to "0" due to an abnormality. This warning is rarely occurred, if the instrument has been turned off when writing the consumable |
Turn the power OFF and contact your Shimadzu representative. |
58E8 |
Invalid autosampler adjustment parameters have been initialized. [3071] |
InvalidAutosampler adjustment parameters have been initialized.\nThis warning is rarely occurred, if the instrument has been turned off when |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu representatives. |
58E9 |
Invalid autosampler calibration data have been initialized. [3072] |
InvalidAutosampler calibration data have been initialized.\nThis warning is rarely occurred, if the instrument has been turned off when writing the calibration data. |
Press the CE key to clear,andthen view the System Check screen. Correct the necessaryAutosampler items. |
58EA |
The usage history of the Autosampler consumables has been initialized to "0" due to an abnormality. [3073] |
Parameter has been initialized to 0 (zero)._This warning is rarely occurred, if the instrument has been turned off when writing the consumable counters |
The next time the instrument is operated, wait for theAutosampler to come to a complete stop before turning OFF the instrument. |
58EC |
Invalid oven calibration data have been initialized. [3081] |
Invalid oven calibration data have been initialized.\nThis warning is rarely occurred, if the instrument has been turned off when writing the calibration data. |
Press the CE key to clear,andthen view the System Check screen. Correct the necessary oven items. |
58ED |
Invalid cooler calibration data have been initialized. [3091] |
Invalid cooler calibration data have been initialized.\nThis warning is rarely occurred, if the instrument has been turned off when writing the calibration data. |
Press the CE key to clear,andthen view the System Check screen. Correct the necessary cooler items. |
58F0 |
Communications to PC data buffer overflow. [4001] |
There is a PC data buffer overflow. |
Check the PC status and the cable connections. If the error persists, turn the power OFF and contact your Shimadzu representative. |
58F1 |
Communications to PC monitor buffer overflow. [4004] |
There is a PC monitor buffer overflow. |
Check the PC status and the cable connections. If the error persists, turn the power OFF and contact your Shimadzu representative. |
58F4 |
Timeout error occurred during communication (reception) between CPUs. [5001] |
Timeout error occurred during communication (reception) between CPUs. |
Turn the power OFF and contact your Shimadzu representative. |
58F5 |
Timeout error occurred during communication (transmission) between CPUs. [5002] |
Timeout error occurred during communication (transmission) between CPUs. |
Turn the power OFF and contact your Shimadzu representative. |
58F8 |
Retry limit is exceeded for reception at the RS-232C I/F. [6001] |
Retry limit has been exceeded for reception at the RS-232C interface. |
Check the PC status and the cable connections. If the error persists, turn the power OFF and contact your Shimadzu representative. |
58F9 |
Retry limit is exceeded for transmission at the RS-232C I/F. [6002] |
Retry limit has been exceeded for transmission at the RS-232C interface. |
Check the PC status and the cable connections. If the error persists, turn the power OFF and contact your Shimadzu representative. |
58FA |
Retry limit is exceeded for waiting for ACK at the RS-232C I/F after transmission. [6003] |
Retry limit has been exceeded for waiting for ACK at the RS-232C interface after transmission. |
Check the PC status and the cable connections. If the error persists, turn the power OFF and contact your Shimadzu representative. |
5901 |
Undefined error [0001] |
----- |
Turn the power OFF and contact your Shimadzu representative. |
5902 |
CTO: Cannot detect cooling unit. [0079] |
Cooler is not mounted or the cable is snapping. |
Turn the power OFF and contact your Shimadzu representative. |
5903 |
SIL: Operation mode is local. Set the operation mode to remote. |
Operation mode is local. |
Set the operation mode to remote. |
5904 |
SIL: The home position of the rack drive motor cannot be detected. [0095] |
----- |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu representative. |
5905 |
Failed to write maintenance log. [3001] |
Cannot write to maintenance log. |
Turn the power OFF and contact your Shimadzu representative. |
5906 |
Failed to read maintenance log. [3002] |
Cannot read maintenance log. |
Turn the power OFF and contact your Shimadzu representative. |
5907 |
The maintenance log is full and log data cannot be entered. |
The maintenance log is full and log data cannot be entered. |
Turn the power OFF and then ON again. The full log is initialized, and log data can once again be entered. |
5908 |
Failed to write error log. [3011] |
Cannot write to error log. |
Turn the power OFF and contact your Shimadzu representative. |
5909 |
Failed to read error log. [3012] |
Cannot read error log. |
Turn the power OFF and contact your Shimadzu representative. |
590A |
The error log is full and log data cannot be entered. [3013] |
The error log is full and log data cannot be entered. |
Turn the power OFF and then ON again. The full log is initialized. |
590B |
Failed to write operation log. [3021] |
Cannot write to operation log. |
Turn the power OFF and contact your Shimadzu representative. |
590C |
Failed to read operation log. [3022] |
Cannot read operation log. |
Turn the power OFF and contact your Shimadzu representative. |
590D |
The operation log is full and log data cannot be entered. [3023] |
The operation log is full and log data cannot be entered. |
Turn the power OFF and then ON again. The full log is initialized. |
590E |
Communications to PC data buffer overflow. [4001] |
There is a PC data buffer overflow. |
Check the PC status and the cable connections. If the error persists, turn the power OFF and contact your Shimadzu representative. |
590F |
Communications to PC monitor buffer overflow. [4004] |
There is a PC monitor buffer overflow. |
Check the PC status and the cable connections. If the error persists, turn the power OFF and contact your Shimadzu representative. |
5920 |
Transmission error. |
The error has occurred in communication to PC. |
Turn the power OFF and then ON again. |
5921 |
System checks timeout error occurred. [8001] |
System check timeout error has occurred. |
Turn the power of system controller and component OFF once and turn it ON again. |
5922 |
PDA Detector: Connection request received from more than one PDA detector. [80F0] |
Connection request received from more than one PDA detector. |
Check if the same master name is registered in other PDA detector. |
5923 |
PDA Detector: Failure to connect to PDA detector. [80F1] |
Failure to connect to PDA Detector. |
Check the network instruments such as network cables, hubs and routers. Turn the power OFF and then ON again. If the error persists, contact your Shimadzu representative. |
5924 |
PDA Detector: LAN timeout error is occurred (PDA detector). [80F2] |
LAN timeout error has occurred. |
Check the network instruments such as network cables, hubs and routers. Turn the power OFF and then ON again. If the error persists, contact your Shimadzu representative. |
5925 |
PDA Detector: Connection cancel because of changing PDA detector's IP address. [80F3] |
PDA's settings are changed. |
----- |
5926 |
PDA Detector: Connection cancel because of changing PDA detector's master system name. [80F4] |
PDA's settings are changed. |
----- |
5927 |
PDA Detector: Connection cancel because of changing PDA detector's system name. [80F5] |
PDA's settings are changed. |
----- |
5928 |
Mobile phase A is under the alarm level. [81D0] |
The remaining volume of mobile phase is less than the alarm level. |
Replace the mobile phase and reset the amount of mobile phase. |
5929 |
Mobile phase B is under the alarm level. [81D1] |
The remaining volume of mobile phase is less than the alarm level. |
Replace the mobile phase and reset the amount of mobile phase. |
592A |
Mobile phase C is under the alarm level. [81D2] |
The remaining volume of mobile phase is less than the alarm level. |
Replace the mobile phase and reset the amount of mobile phase. |
592B |
Mobile phase D is under the alarm level. [81D3] |
The remaining volume of mobile phase is less than the alarm level. |
Replace the mobile phase and reset the amount of mobile phase. |
592C |
Purge fail. Line pressure is too high. [81D4] |
Pump discharge pressure has exceeded [P-PMAX] during purging. |
Open the drain valve before pressing .\nIf an error message appears during auto-purging by system controller, inspect the clogging in the flow line. If no clogging is found, set the appropriate [P-FLOW] and [P- |
592D |
Communication error occurred between A and B pumps. [81F0] |
A transmission error has occurred between pumps A and B. |
Turn the power OFF and contact your Shimadzu representative. |
592E |
A seal malfunction has occurred in the online degasser vacuum line. [81F1] |
Vacuum pressure of degasser connected to this instrument has been abnormal for certain time. |
Check the vacuum pressure of degasser. If the vacuum pressure is [NG], turn the power OFF once and turn it ON again.\nIf the vacuum pressure is [OK], it is of no matter, for the degasser is returning to run automatically. |
592F |
SIL: Low-pressure valve's rotational position incorrect positions. [8513] |
Low-pressure valve's rotational position incorrect positions. |
Turn the power OFF and contact your Shimadzu representative. |
5930 |
SIL: System error. [8529] |
There is an error in theAutosampler's internal circuits. |
Turn the power OFF and contact your Shimadzu representative. |
5931 |
SIL: Cannot adjust metering pump position. [8562] |
Measuring pump does not operate correctly. |
Turn the power OFF and contact your Shimadzu representative. |
5932 |
SIL: Needle malfunctions in X (left-right) direction. [8563] |
Needle's X-axis_(sideways) movement is incorrect. |
Turn the power OFF and contact your Shimadzu representative. |
5933 |
SIL: Needle malfunctions in Y (front-back) direction. [8564] |
Needle's Y-axis_(forward/backward) movement is incorrect. |
Turn the power OFF and contact your Shimadzu representative. |
5934 |
SIL: Cooling unit malfunctions. [8565] |
There is an error in the sample cooler's cooling unit. |
Turn the power OFF and contact your Shimadzu representative. |
5935 |
SIL: Heater malfunctions. [8566] |
There is an error in the sample cooler's heating unit. |
Turn the power OFF and contact your Shimadzu representative. |
5936 |
SIL: Abnormal temperature detected by temperature sensor. |
There is an error in the sample cooler's temperature sensor. |
Turn the power OFF and contact your Shimadzu representative. |
5937 |
SIL: Rinse liquid is under the alarm level. [8568] |
The remaining volume of rinse liquid is less than the alarm level. |
Replace the rinse liquid and reset the amount of rinse liquid. |
5938 |
SIL: Needle protection error is occurred. [8569] |
A foreign substance was detected at the tip of the needle. |
Check for foreign substances inside theAutosampler. |
5939 |
SIL: Rack-changer malfunctions in X (left-right) direction. [85D0] |
Motor's X-axis movement_(i.e., taking out the rack) is notcorrected. |
Turn the power OFF and contact your Shimadzu representative. |
593A |
SIL: Rack-changer malfunctions in Z (up-down) direction. [85D1] |
Motor's Z-axis movement_(i.e. moving vertically) is notcorrected. |
Turn the power OFF and contact your Shimadzu representative. |
593B |
SIL: Rack-changer malfunctions in Y (front-back) direction. [85D2] |
Motor's Y-axis movement_(forward/backward) is notcorrected. |
Turn the power OFF and contact your Shimadzu representative. |
593C |
SIL: Rack-changer home position error in X (left-right) direction. [85D3] |
Motor's X-axis movement_(i.e., taking out the rack) is notcorrected. |
Turn the power OFF and contact your Shimadzu representative. |
593D |
SIL: Rack-changer home position error in R (rotation) direction. |
Motor's R-axis movement_(i.e., rotating the rack) is notcorrected. |
Turn the power OFF and contact your Shimadzu representative. |
593E |
SIL: Rack-changer home position error in Z (up-down) direction. [85D5] |
Motor's Z-axis movement_(i.e. moving vertically) is notcorrected. |
Turn the power OFF and contact your Shimadzu representative. |
593F |
SIL: Rack-changer home position error in Y (up-down) direction. [85D6] |
Motor's Y-axis movement_(forward/backward) is notcorrected. |
Turn the power OFF and contact your Shimadzu representative. |
5940 |
SIL: There is a problem in the lock sensor on the rack stack. |
----- |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
5941 |
SIL: Rack-changer error is occurred. [85D8]_(Reserved) |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5942 |
SIL: Rack-changer cannot release the rack. [85D9] |
Rack was not released correctly. |
Turn the power OFF and contact your Shimadzu representative. |
5943 |
SIL: Rack-changer cannot find the rack. [85DA] |
Error occurred while rack was being taken out. |
Turn the power OFF and contact your Shimadzu representative. |
5944 |
SIL: Rack-changer cannot set the rack. [85DB] |
Error occurred while rack was being set. |
Turn the power OFF and contact your Shimadzu representative. |
5945 |
SIL: Abnormal high temperature detected by rack-changer temperature sensor. [85DC] |
There is an error in the sample cooler's temperature sensor. |
Turn the power OFF and contact your Shimadzu representative. |
5946 |
SIL: Rack-changer cooling unit malfunctions. [85DD] |
Abnormality in cooler control unit. |
Turn the power OFF and contact your Shimadzu representative. |
5947 |
SIL: Changer-stack is opened. [85DE] |
Stack was been pulled out for more then 3 minutes. |
Set the rack and reset the power. Cooling by the cooler will stop. |
5948 |
SIL: Rack-changer error is occurred. |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5949 |
SIL: The specified rack is not set on the rack changer. [85E0] |
Therack specified for analysis is not set. |
Set the rack correctly and restart analysis. |
594A |
SIL: Rack-changer system error 1 has occurred. [85E1] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
594B |
SIL: Rack-changer system error 2 has occurred. [85E2] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
594C |
SIL: Rack-changer system error 3 has occurred. [85E3] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
594D |
SIL: Rack-changer system error 4 has occurred. [85E4] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
594E |
SIL: Rack-changer system error 5 has occurred. [85E5] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
594F |
SIL: Rack-changer system error 6 has occurred. [85E6] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5950 |
SIL: Rack-changer system error 7 has occurred. [85E7] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5951 |
SIL :Cannot set rack in current position on rackchanger. [85E8] |
A rack cannot be set in a position undergoing analysis. |
Either removesthe rack from the rack position for which the indicator is flashing, or replacesit with a condensationprevention rack. |
5952 |
SIL: Rack-changer system error 8 has occurred. [85E9] |
System error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5953 |
SIL: Rack-changer heater malfunctions. [85EA] |
There is an error in the sample cooler's heating unit. |
Turn the power OFF and contact your Shimadzu representative. |
5954 |
Detector: Failed to write to EEPROM. [870E] |
EEPROM write error is occurred. |
Turn the power OFF and contact your Shimadzu representative. |
5955 |
Detector:D2 lamp does not light. [8743] |
Fault in deuterium lamp or in its circuits. |
Turn the power OFF and contact your Shimadzu representative. |
5956 |
Detector: D2 lamp is over current. [8744] |
Abnormally high current flowing through lamp. |
Turn the power OFF and contact your Shimadzu representative. |
5957 |
Detector:Detector temperature has exceeded 63 C. [874B] |
Instrument interior temperature has risen to an abnormal level. |
Check to be sure that the rear fan can move, and that the exhaust vent and side air intake are not blocked. If this error is still displayed, turn the power OFF and contact your Shimadzu representative. |
5958 |
Detector:Cell Temperature setting is too low[87D0] |
This warning is displayed when the flow cell temperature fails to coincide with temperature settings after undergoing 5 minutes or more of temperature adjustment control. This is caused by setting the flow cell temperature at a valve too close to room temperature, or setting the column oven at a higher temperature |
Set the temperature setting to a value at least 5C above room temperature. If the warning is still displayed, set the flow cell temperature to a value near or above the column oven temperature. If the temperature setting satisfies the above requirements and the warning is still displayed, turn the power switch OFF and contact your Shimadzu representative. |
5959 |
CTO: Rotary valve home position error occurred. [8630] |
Rotation error in an automatic columnswitching valve. |
Turn the power OFF and contact your Shimadzu representative. |
595A |
CTO: Incorrect CMD address. [8631] |
----- |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
595B |
CTO: Failed to write to CMD. [8632] |
----- |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
595D |
The stack is open. Please close the stack in the rack-changer soon. [85EC] |
A stack was left open for more than 30 minutes and so the sample cooler has been turned OFF. |
Close the stack. |
595E |
The temperature control stopped because stack has been opened for more than 30 minutes. If the stack will be closed, the temperature control will start again. [85ED] |
A stack was left open for more than 30 minutes and so the sample cooler has been turned OFF. |
Close the stack to turn the sample cooler ON again. |
595F |
No rack is detected. Check the rack on the stack. If the rack is not set and analysis is performed, the sequence will be |
There is a stack for which a rack has not been set. When analysis is performed for this rack, [RC NO. RACK] error is displayed and the scheduled operation is |
Set a rack in the position that is waiting for analysis. When using the Rack Changer/C, in order to prevent condensation, set racks holding empty micro_titer plates or deep-well micro_titer plates in all the unused stacks. |
5960 |
Scheduled data acquisition ended since the rack was not set in place. [85F0] |
----- |
----- |
5961 |
Rack setting is doubled. Remove rack or Replace dummy rack.[85F1] |
The same rack has been set twice. After analysis is performed for this rack, [RC FULL RACK] error is displayed and the scheduled operation is stopped. |
Either removesthe rack from the rack position for which the indicator is flashing, or replacesit with a condensationprevention rack. |
5963 |
As the stack has been closed, the temperature control has been on again. [85EE] |
The stack was closed and so the sample cooler turned ON. |
----- |
5964 |
OptLink memory error. [2F01 - 2F0A] |
Internal error occurred while communicating with other units. |
Turn the power OFF once and turn it ON again. |
5965 |
Communication error. [2F10 - 2F3F] |
Communication error occurred while communicating with PC. |
Check the communication environment such as network/RS232C cable etc. |
5966 |
The IP address has been set to "192.168.200.99" as the CBM failed to obtain an IP address. |
CBM cannot obtain the IP address in DHCP mode. |
1. Check the network instruments such as network cables, hubs and routers.\n2. Confirm the DHCP server is operating.\n3. Check the DHCP server settings. |
5967 |
The CBM will be rebooted because of the failure to renew/rebind the IP address. [4102] |
CBM cannot rebind/renew the IP address in DHCP mode. |
1. Check the network instruments such as network cables, hubs and routers.\n2. Confirm the DHCP server is operating.\n3. Check the DHCP server settings.\n* We recommend that the lease time of the IP address from the DHCP server should be infinite. |
5968 |
The CBM has been rebooted because of the failure to renew/rebind the IP address. [4103] |
CBM cannot rebind/renew the IP address in DHCP mode. |
1. Check the network instruments such as network cables, hubs and routers.\n2. Confirm the DHCP server is operating.\n3. Check the DHCP server settings.\n* We recommend that the lease time of the IP address from the DHCP server should be infinite. |
5969 |
PDA detector cannot accept communication command. [8084] |
A PDA detector which is paired with CBM is controlled by PC as an other system. |
Reset the pair with CBM by PDA's web screen. |
596A |
PDA detector is too busy. [80F7] |
CBM is turned on while PDA detector which is stand-alone use is run.\nFailed the communication with PDA at analysis |
Stop PDA analysis or turn the power OFF once and turn it ON again.\nCheck communication environment. |
596B |
SIL: The reserved fatal error code 1 has been detected. [85F2] |
Additional instrument errors |
Refer to theAutosampler instruction manual. |
596C |
SIL: The reserved fatal error code 2 has been detected. [85F3] |
Additional instrument errors |
Refer to theAutosampler instruction manual. |
596E |
SIL: The reserved error code 2 has been detected. [85F5] |
Additional instrument errors |
Refer to theAutosampler instruction manual. |
5973 |
SIL: Rack Changer Disconnect Error.[85F4] |
The connection cable between the autosampler and the rack-change is not connected properly. Or the rack-changer is turned off. |
Check the rack-changer is turned on.\nIf the rack-changer is turned on, check the RS-232C cable between the rackchanger and the autosampler is connected properly. |
5984 |
Purge is performed, because pressure does not rise to "Minimum Pressure" (Pump A). |
Purge is executed, because pressure does not rise to "Minimum Pressure". |
Check the flow line for leaks. If no leak is found, the [P.Min] setting is inappropriate, and must be changed. |
5985 |
The system which can be registered on the group is 10 in the maximum. [80E0] |
11 or more systems are registered on one group. |
The system which can be registered on the group is 10 in the maximum. |
5986 |
Please confirm startup time. |
Startup time is already passed. |
Set correct startup time. |
5999 |
Too low the set temperature for flow cell. The flow cell cannot be kept in specified temperature. |
The cell temperature setting is too low.\nThe room temperature exceeds the cell temperature setting by 10 C. Can not control the cell temperature with the set |
Set the cell temperature higher than "room temperature minus 10 C". If the error persists, turn the power off and contact your Shimadzu representative. |
599A |
Too high the set temperature for flow cell. The flow cell cannot be kept in specified temperature. |
The cell temperature setting is too high.\nThe cell temperature setting exceeds the room temperature by 10 C. Can not control the cell temperature with |
Set the cell temperature lower than "room temperature plus 10 C". If the error persists, turn the power off and contact your Shimadzu representative. |
599B |
EEPROM parameters were lost. |
This error message means that the previous parameters or time programs were lost when the power was turned on. |
This error has no affect on the operation in the workstation. |
599C |
Xe lamp does not light. |
Fault in Xenon lamp or in its circuits. |
Turn the power off and turn it again, then light the Xenon lamp. Xenon lamp does not light without restarting the |
599D |
Invalid wavelength calibration |
Wavelength calibration data are abnormal. |
Execute the wavelength calibration. |
599E |
Invalid wavelength calibration |
Wavelength calibration data are abnormal. |
Execute the wavelength calibration. |
599F |
Invalid wavelength calibration |
Wavelength calibration data are abnormal. |
Execute the wavelength calibration. |
59A0 |
Invalid wavelength calibration |
Wavelength calibration data are abnormal. |
Execute the wavelength calibration. |
59A1 |
Hg lamp does not light. |
Fault in Hg lamp or in its circuits. |
Turn the power off and turn it again. If the error persists, contact your Shimadzu representative. |
59A2 |
The wavelength shift is greater than 2 nm. |
The wavelength shift is greater than 2 nm. |
(1) Verify that cell is securely installed.\n(2) Execute "WAVE CALIB" (in the "CALIBRATION" menu of the detector) and check the wavelength accuracy.\n(3)_Large bubbles may remain in the flow cell. Inject water, methanol or acetonitrile into the cell to push the bubbles out.\nIf the error persists after these actions, turn the power off and contact your Shimadzu representative. |
59A3 |
The wavelength shift is greater than 2 nm. |
The wavelength shift is greater than 2 nm. |
(1) Verify that cell is securely installed.\n(2) Execute "WAVE CALIB" (in the "CALIBRATION" menu of the detector) and check the wavelength accuracy.\n(3)_Large bubbles may remain in the flow cell. Inject water, methanol or acetonitrile into the cell to push the bubbles out.\nIf the error persists after these actions, turn the power off and contact your Shimadzu representative. |
59A4 |
The wavelength shift is greater than 2 nm. |
The wavelength shift is greater than 2 nm. |
(1) Verify that cell is securely installed.\n(2) Execute "WAVE CALIB" (in the "CALIBRATION" menu of the detector) and check the wavelength accuracy.\n(3)_Large bubbles may remain in the flow cell. Inject water, methanol or acetonitrile into the cell to push the bubbles out.\nIf the error persists after these actions, turn the power off and contact your Shimadzu representative. |
59A5 |
Flow cell is overheated. |
Flow cell temperature has risen to an abnormal level. |
Check that the exhaust vent and side air intake are not blocked, and the filters of the air intake (front and side) are not clogged. If this error persists, turn the power off and contact your Shimadzu representative. |
59A6 |
Lamp cover is open. |
Lamp cover is open. |
Attach the lamp cover securely. |
59A7 |
Fan for the Peltier unit has stalled. |
There is an error in the fan for the Peltier unit. |
Turn the power off and turn it again. If the error persists, contact your Shimadzu representative. |
59A8 |
Cooling fan for the Xe lamp power has stalled. |
There is an error in the cooling fan for the Xe lamp power supply. |
Check to be sure that the rear fan can move. Also check that the exhaust vent and side air intake are not blocked, and the filters of the air intake (front and side) are not clogged. If this error is still displayed, turn the power off and contact your Shimadzu representative. |
59A9 |
Lamp mirror home position error. |
Lamp mirror home position sensor does not operate properly. |
Turn the power off and turn it again. If the error persists, contact your Shimadzu representative. |
59AA |
ROM error occurred. |
ROM error. |
Turn the power off and turn it again. If the error persists, contact your Shimadzu representative. |
59AB |
RAM error occurred. |
RAM error. |
Turn the power off and turn it again. If the error persists, contact your Shimadzu representative. |
59AC |
Error in the flow cell's cooling unit. |
This error appears when the temperature of the flow cell does not reach the set temperature within 30 minutes after the power is turned on. |
Turn the power off and turn it again. If the error persists, contact your Shimadzu representative. |
59AD |
Error in the flow cell's heating unit. |
This error appears when the temperature of the flow cell does not reach the set temperature within 30 minutes after the power is turned on. |
Turn the power off and turn it again. If the error persists, contact your Shimadzu representative. |
59B1 |
Leak sensor wire is broken. |
Leak sensor malfunctions (probably due to broken / disconnected wire). |
Turn the power off and turn it again. If the error persists, contact your Shimadzu representative. |
59B2 |
Leak sensor is short-circuited. |
Leak sensor malfunctions. |
Turn the power off and turn it again. If the error persists, contact your Shimadzu representative. |
59B3 |
Constant pressure mode is not supported. |
A constant pressure pumping mode is set on the system controller with [1: 20A] set at [OP MODE]. |
Use the constant flow pumping mode. |
59B4 |
Abnormal high temperature is detected inside the instrument. |
Interior temperature has risen to an abnormal level. |
Check to be sure that the internal fan can move,and that the rear exhaust vent is not blocked. If this message still appears, turn the power OFF and contact your Shimadzu |
59B5 |
Motor cooling fan has stalled. |
The cooling fan is faulty. |
Turn the power OFF and contact your Shimadzu representative. |
59B6 |
Link is cut off during pumping and pumping is stopped automatically. |
The system controller is disconnected during the pumping under control of the system controller. |
Remove the cause of disconnection of the system controller. This error occurs only when the auxiliary function [LINK CHK] is enabled. |
59B7 |
Leak sensor is open or shortcircuit. |
The leak sensor detection value is abnormal. The leak sensor may be |
Turn the power OFF and contact your Shimadzu representative. |
59B8 |
Motor current error is detected. |
The motor current is abnormal. |
Turn the power OFF and contact your Shimadzu representative. |
59B9 |
Cannot write parameters in EEPROM. |
A nonvolatile memory read/write error has occurred. Parameters may not have been saved correctly. |
Turn the power OFF and contact your Shimadzu representative. |
59BA |
Gas sensor is open. |
The gas sensor detection value is abnormal. The gas sensor may be |
Turn the power OFF and contact your Shimadzu representative. |
59BB |
Heater failure. |
The heater has failed. |
Turn the power OFF and contact your Shimadzu representative. |
59BC |
Power heat exhaust fan rotation error. |
The power heat exhaust fan is not rotating correctly. |
If anything is interfering with the fan rotation, turnthe power OFF and remove the obstruction. Ifthe obstruction cannot be removed or if the erroroccurs from any other cause, contact yourShimadzu representative. |
59BD |
Post-column cooler rotation error. |
The post-column cooler is not rotating correctly. |
If anything is interfering with the fan rotation, turnthe power OFF and remove the obstruction. Ifthe obstruction cannot be removed or if the erroroccurs from any other cause, contact yourShimadzu representative. |
59BE |
Block cooling fan rotation error. |
The block cooling fan is not rotating correctly. |
If anything is interfering with the fan rotation, turnthe power OFF and remove the obstruction. Ifthe obstruction cannot be removed or if the erroroccurs from any other cause, contact yourShimadzu representative. |
59BF |
Heater L temperature has exceeded T.Max. |
Thermosensor readings are abnormally high. There may be a short circuit in the |
Turn the power OFF and contact your Shimadzu representative. |
59C0 |
Door sensor failure. |
The door sensor has failed. |
Turn the power OFF and contact your Shimadzu representative. |
59C1 |
Pretreatment file does not exist. |
No description of the specified pretreatment program is found. |
Check if the specified pretreatment program exists, and if it exists, specify the edit page number on the pretreatment screen for the autosampler method at the workstation LabSolutions, and check the contents of the pretreatment program. |
59C3 |
Sample cooling unit malfunction. |
There is an error in the sample cooler's cooling unit. |
Turn the power OFF and contact your Shimadzu representative. |
59C4 |
Dehumidification unit malfunction. |
Peltier dehumidifier temperature sensor error |
Turn the power OFF and contact your Shimadzu representative. |
59C5 |
Room temperature sensor malfunction. |
The room temperature sensor detects an abnormally high or low temperature. |
Turn the power OFF and contact your Shimadzu representative. |
59CC |
The vial number is not valid for the current plate on the tray. |
The invalid vial number is set. |
Check the plate and set the valid vial number. |
59CD |
Leak error is detected by a sensor near a High-pressure valve. |
abnormally high or low temperature. |
Check for liquid leakage from the following locations. \n1. Injection port. If liquid is leaking from the injection port, replace the needle seal. \n2. Valve piping joint. If liquid is leaking from the valve piping joint, re-tighten the piping, or, replace the piping and the joint parts with new parts.\n3. Sample loop. If liquid is leaking from the sample loop, replace with new parts. \n4. Leak sensor sensitivity adjustment. If the leak sensor is over-responding, calibrate the sensor, and re-adjust the threshold value. If the above procedure does not remedy the problem, contact your |
59CE |
Abnormal temperature detected by temperature sensor into rack plate L. |
There is an error in the sample cooler L's temperature sensor. |
Turn the power OFF and contact your Shimadzu representative. |
59CF |
Abnormal temperature detected by temperature sensor into rack plate M. |
There is an error in the sample cooler M's temperature sensor. |
Turn the power OFF and contact your Shimadzu representative. |
59D0 |
Abnormal temperature detected by temperature sensor into rack plate R. |
There is an error in the sample cooler R's temperature sensor. |
Turn the power OFF and contact your Shimadzu representative. |
59D1 |
The sequence has skipped because a rack is not adjusted. Adjust the rack. |
The analysis has been started by using the sample rack where the rack position is not adjusted. |
Adjust the rack, then restart analysis. |
59D2 |
Rack on rack plate L is not |
A not adjusted rack is set on rack plate L. |
Adjust the rack on rack plate L. |
59D3 |
Rack on rack plate M is not adjusted. |
A not adjusted rack is set on rack plate M. |
Adjust the rack on rack plate M. |
59D4 |
Rack on rack plate R is not adjusted. |
A not adjusted rack is set on rack plate R. |
Adjust the rack on rack plate R. |
59D5 |
Invalid rack ID. |
A rack with invalid rack ID is set. |
Confirm the rack ID, then set the correct rack ID. |
59D6 |
Racks with same rack ID are set. |
Racks with same rack ID are set. |
Set a rack with another rack ID. |
59D7 |
The temperature control of rack plate L has been turned off because no rack has been set for more than 30 minutes. If the rack is set, the temperature control will be on again. |
No rack has been set on rack plate L for more than 30 minutes. |
Set a rack on rack plate L. Alternatively stop the temperature control of rack plate L. |
59D8 |
The temperature control of rack plate M has been turned off because no rack has been set for more than 30 minutes. If the rack is set, the temperature control will be on again. |
No rack has been set on rack plate M for more than 30 minutes. |
Set a rack on rack plate M. Alternatively stop the temperature control of rack plate M. |
59D9 |
The temperature control of rack plate R has been turned off because no rack has been set for more than 30 minutes. If the rack is set, the temperature control will be on again. |
No rack has been set on rack plate R for more than 30 minutes. |
Set a rack on rack plate R. Alternatively stop the temperature control of rack plate R. |
59DA |
Invalid vial number. |
Invalid sample(vial) number. |
Confirm the sample number. Alternatively confirm whether a rack is set. |
59DB |
Leak error is detected by a sensor near a Low-pressure valve. |
Liquid leaks from a tube joint or a Lowpressure valve. |
Check for liquid leakage from the following locations.\n1. Valve piping joint. If liquid is leaking from the valve piping joint, re-tighten the piping, or, replace the piping and the joint parts with new parts.\n2. Column piping joint (when using CTO-30AS.) If liquid is leaking from the column piping joint, re-tighten the piping or, replace the piping and the joint parts with new parts.\n3. Leak sensor sensitivity adjustment. If the leak sensor is over-responding, calibrate the sensor, and re-adjust the threshold value. If the above procedure does not remedy the problem, contact your |
59E7 |
Pump:Leak is detected at the solenoid valve unit. Eliminate the leak in the interior or exterior. |
Leak sensor of solenoid valve unit has detected leakage.\nSensor is too sensitive. |
Inspect the leakage in the solenoid valve unit. Wipe away any liquid around the leak sensor.\nAdjust the leak sensor actuation level. |
5A00 |
SFC:Operation mode is local. Set the operation mode to remote. [8B80] |
Operation mode is local. |
Set the operation mode to remote. |
5A01 |
Could not detect the home position of the BPR motor.[8B90] |
Could not detect the home position of the BPR motor. |
Press the (power button) on the instrument to turn the power OFF, and then press the (power button) on the instrument again to turn the power ON. Next, press [pressure button] on the instrument. \nIf the error message is displayed again even when this operation is performed, press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
5A02 |
The BPR motor has slipped.[8B91] |
The BPR motor has slipped. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
5A03 |
BPR has Over-heated.[8B92] |
BPR has Over-heated. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
5A04 |
BPR heater controller error.[8B93] |
The BPR heater electrical controller is in error. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
5A05 |
BPR cover is open.[8B94] |
BPR cover is open. |
Attach the BPR cover securely. |
5A06 |
BPR temperature sensor is open or short-circuit.[8B95] |
Lamp housing temperature sensor is open or short-circuit. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
5A07 |
Pump head temperature sensor is open or short-circuit.[8BA0] |
Pump head temperature sensor is open or short-circuit. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
5A08 |
Ppump head cooling fan rotation error.[8BA1] |
The pump head cooling fan is not rotating correctly. |
If anything is interfering with the fan rotation, turn the power OFF and remove the obstruction. Ifthe obstruction cannot be removed or if the erroroccurs from any other cause, contact yourShimadzu representative. |
5A09 |
Coolant pump for cooling the pump head failure.[8AA2] |
The coolant pump for cooling the pump head has failed. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
5A0A |
SFC: Leak is detected. [8BB0] |
Eluent leak can be detected in the BPR. |
Stop the pump flow and then execute any action to stop the eluent leak. See detail Instrument manual. |
5A0B |
SFC: Leak sensor is open or short-circuit.[8BB1] |
The leak sensor detection value is abnormal. The leak sensor may be disconnected. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
5A0C |
SFC: Pump has turned off because the pump head cooling temperature was not stable or cooling was turned off.[8BA3] |
----- |
----- |
5A20 |
SFE:Operation mode is local. Set the operation mode to remote. |
Operation mode is local. |
Set the operation mode to remote. |
5A21 |
System Error[8B01] |
System Error |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
5A22 |
SFE: Incorrect rack |
SFE: Incorrect rack number.[8B05] |
Set values within the valid range. |
5A23 |
SIL: The rack is not set. [8B06] |
The rack is not set. |
Set the rack correctly and restart analysis. |
5A24 |
SFE: Rack Changer Disconnect Error.[8B07] |
The connection cable between the SFE and the rack-change is not connected properly. Or the rack-changer is turned off. |
Check the rack-changer is turned on.\nIf the rack-changer is turned on, check the RS-232C cable between the rackchanger and the SFE is connected properly. |
5A25 |
SFE:Needle home position error. [8B10] |
Move to direction was faulty. |
Check if nothing is stuck in the movable part.\nIf nothing is stuck, contact your Shimadzu representative. |
5A26 |
SFE: Motor has slipped. [8B11] |
SFE: Motor has slipped. |
Turn the power OFF and contact your Shimadzu representative. |
5A27 |
Debris detected on needle. Please remove the debris from the inside of the SFE. [8B12] |
Needle tip detects strong force. |
Remedy this problem referring to the following procedure. 1. Replace the needle.\n2. Verify that the sample rack is installed correctly. \nIf the above procedure does not remedy the problem, contact your Shimadzu |
5A28 |
Could not detect the home position of the SFE extraction valve.[8B20] |
Could not detect the home position of the SFE extraction valve. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
5A29 |
Could not detect the home position of the SFE vent valve.[8B21] |
Could not detect the home position of the SFE vent valve. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
5A2A |
SFE: Leak is detected. [8B30] |
Eluent leak can be detected in the SFE. |
Stop the pump flow and then execute any action to stop the eluent leak. See detail Instrument manual. |
5A2B |
SFE: Leak sensor is open or short-circuit.[8B31] |
The leak sensor detection value is abnormal. The leak sensor may be disconnected. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
5A2C |
SFE: CO2 gas leak is detected. [8B32] |
CO2 gas leak can be detected in the SFE. |
Stop the pump flow and then execute any action to stop the eluent leak. See detail Instrument manual. |
5A2D |
SFE: CO2 gas sensor is open or short-circuit.[8B33] |
The CO2 gas sensor detection value is abnormal. The CO2 gas sensor may be disconnected. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
5A2E |
SFE:No extraction vessel is set on the rack. [8B40] |
No extraction vessel was placed in the rack position. |
Set the extraction vessel properry.If the error message is displayed again, press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
5A2F |
Heater is turned off because the temperature of the extraction vessel exceeded the maximum temperature.[8B41] |
The temperature of the extraction vessel exceeded the [Maximum Temperature] of the instrument parameters. |
Maximum temperature must be specified 5 C higher than the extraction vessel temperature. |
5A30 |
SFE: Extraction vessel temperature sensor is open or short-circuit.[8B42] |
The extraction vessel temperature sensor detection value is abnormal. The CO2 gas sensor may be disconnected. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
5A31 |
Heater is turned off because the temperature of the pre-heaterl exceeded the maximum temperature.[8B43] |
The temperature of the pre-heater exceeded the [Maximum Temperature] of the instrument parameters. |
Maximum temperature must be specified 5 C higher than the extraction vessel temperature. |
5A32 |
SFE: Pre-heater temperature sensor is open or shortcircuit.[8B43] |
The pre-heater temperature sensor detection value is abnormal. The preheater temperature sensor may be |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
5A33 |
SFE: Low set temperature. [8B45] |
Set temperature value is inappropriate. Temperature is specified with lower than the room temperature plus 5C. |
Set temperature higher than room temperature plus 5C.\n(Temperature control is not stopped.) |
5A34 |
SFE: room temperature sensor is open or short-circuit.[8B50] |
The room temperature sensor detection value is abnormal. The room temperature sensor may be disconnected. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
5A35 |
SFE: The extraction vessel temperature has dropped. Please confirm the extraction vessel has been fastened.[8B46] |
----- |
Please confirm the extraction vessel has been fastened. |
5A36 |
SFE: The extraction vessel heater failure.[8B47] |
----- |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
5A37 |
SFE: Pre-heater failure.[8B48] |
----- |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
5A38 |
SFE: The extraction vessel temperature sensor wire is broken.[8B49] |
----- |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
5A39 |
SFE: The temperature compensation sensor failure.[8B4A] |
----- |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
5A3A |
Controller: Autosampler and SFE Unit are connected at the same time.[8B02] |
Controller: Autosampler and SFE Unit are connected at the same time.[8B02] |
----- |
7203 |
Download the Instrument Parameters. Data acquisition cannot be started. |
Data acquisition cannot be started since the instrument parameters have not been downloaded. |
Download the instrument parameters, and start data acquisition. |
7204 |
Data acquisition cannot be started. |
Data acquisition cannot be started for some reason. |
Close the sub-window, and perform the operation again. Try again after making sure that the network is connected correctly in the case of a network configuration. |
7211 |
The data have not been acquired from the instrument. The data acquisition is stopped. |
The buffer memory of the instrument is full since the data have not been acquired from the instrument. |
Check the communication environment and operating status of the instrument and the workstation.\nIf the error is not resolved, turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
7300 |
MS: Timeout occurred in the communication with the MS. |
There was no response from the MS within the specified period of time. |
Check the cable and power supply, turn OFF the power of all equipment including the PC, and restart them. |
7301 |
MS: Received analysis data is abnormal. |
----- |
Check the cable and power supply, turn OFF the power of all equipment including the PC, and restart them. |
7302 |
MS: Communication with MS was not achieved within the specified number of retry counts. |
Communication was attempted for the specific number of time and there was no response from the MS. |
Check the cable and power supply, turn OFF the power of all equipment including the PC, and restart them. |
7303 |
MS: A hardware error occurred during communication with the MS. |
A hardware error occurred during communication with the MS (mass spectrometer unit). |
Check the cable and power supply, turn OFF the power of all equipment including the PC, and restart them. |
7304 |
MS: MS unit is busy. |
The MS unit does not receive signals. |
Check the cable and power supply, turn OFF the power of all equipment including the PC, and restart them. |
7305 |
MS: A command that cannot be used for communication with the MS was used. |
Communication with the MS cannot be normally achieved for some reason. |
Check the cable and power supply, turn OFF the power of all equipment including the PC, and restart them. |
7306 |
Parameters to be set for the MS are not correct. |
Communication with the MS cannot be normally achieved for some reason. |
Check the cable and power supply, turn OFF the power of all equipment including the PC, and restart them. |
7307 |
MS: MS is not connected. |
Connection failed because communication with the MS cannot be established. |
Check the cable and power supply. |
7308 |
MS: An undefined error occurred during communication. |
Communication with the MS cannot be normally achieved for some reason. |
Turn OFF the power of all equipment including the PC, and restart them. |
7309 |
MS: Parameters cannot be set for the MS. |
Communication with the MS cannot be normally achieved for some reason. |
Turn OFF the power of all equipment including the PC, and restart them. |
730A |
MS: Connection to the instrument cannot be achieved. |
The process to connect to the instrument was suspended. |
Restart the PC. |
730B |
MS: The size of data received is different from the requested size. |
Communication with the MS cannot be normally achieved for some reason. |
Turn OFF the power of all equipment including the PC, and restart them. |
730C |
MS: Invalid parameters have been sent. |
This problem may occur when the system is unstable. |
Restart the PC and retry the operation. |
730D |
MS: Invalid parameter ID was used. |
This problem may occur when the system is unstable. |
Restart the PC and retry the operation. |
730E |
MS: The MS responded indicating an abnormal system |
The MS is not operating in a normal state. |
Restart the MS. |
730F |
MS: Insufficient vacuum level. |
This operation is disabled because the vacuum level has not reached the sufficient value. |
Perform the operation after the vacuum system becomes ready on the [System Control] sub-window of the vacuum system. |
7310 |
MS: Method or tuning condition is not set for the MS. |
Method or tuning information is not set for the MS correctly. |
Turn OFF the power of the PC and MS and restart them. |
7311 |
MS: Please provide maintenance service to the following pump. |
Operation time of the pump has exceeded the limit for maintenance specified by system configuration. |
----- |
7319 |
MS: An error occurred in the high-frequency power supply. |
The high-frequency power supply is out of tune. |
Perform auto-tuning. If the message is displayed even during auto-tuning, contact your Shimadzu representative. |
731A |
MS: An error occurred in the high-voltage power supply. |
The current limiter of the high-pressure power supply was activated. |
Perform auto-tuning. If the message is displayed even during auto-tuning, contact your Shimadzu representative. |
731C |
MS: Excessive ions entered the detector. |
The message indicates excessive ions entered the detector. In this state, the filament automatically goes OFF. The filament automatically goes ON after a specific period of time only when data is acquired. The message is often displayed when the detector voltage is too high. |
Check the solvent elution time and detector voltage. |
7320 |
MS: Turbo molecular pump 1 should be overhauled. |
The message indicates that the operation time of turbo molecular pump 1 has exceeded the limit for maintenance. Overhaul is recommended. |
Contact your Shimadzu representative. |
7328 |
MS: Loading of the 80C186 system program failed. |
Either PROGPATH has not been registered in the registry or memory is |
Close all unnecessary applications. Hard disk free space will be reserved as an additional memory. Then, retry |
7329 |
MS: Loading of the DSP system program failed. |
Memory is insufficient. |
First, close all unnecessary applications. Hard disk free space will be reserved as an additional memory. Then, |
732A |
MS: MS instrument type does not match. Please change the instrument control class to be generated. |
----- |
----- |
732B |
MS: Instrument initialization failed. |
The instrument is not operating in a normal state. |
Turn OFF the power of all equipment including the PC, and restart them. |
732C |
MS: Upper/lower limit value error. Set a value within the range min. to max. |
----- |
----- |
732D |
MS: MS connection failed. |
----- |
Check the cable and power supply and then restart the MS and the PC. |
7330 |
MS: The instrument control program version does not match. Restart the mass spectrometer by referring to Help. |
The version of the instrument control program downloaded on the mass spectrometer does not match the version of the program on the PC. |
Restart the mass spectrometer unit to clear the instrument control program. Then, download the instrument control program by connecting to the instrument via LCMSsolution. \n Operation procedures are described below. \n 1) Quit LCMSsolution. Save necessary data. \n 2) Turn OFF the power of the PC. \n 3) Turn OFF the power of the mass spectrometer unit and turn it ON several seconds later. \n 4) Turn ON the power of the PC. \n 5) Start the latest version LCMSsolution and connect it with the instrument correctly. If the former version LCMSsolution is started in this state, an instrument control program for a different version may be downloaded. \n Whether or not the connection is correct can be checked on [Status] in the |
7331 |
Auto-Tuning was terminated. |
----- |
----- |
7332 |
Save peak profiles and spectra of the standard sample in this |
The message is displayed when data is saved in manual tuning. |
To save peak profiles and spectra of the standard sample in this setting, automatically start auto-tuning with no |
7333 |
The current tuning condition has tuned this file. \nDo you invalidate the tuning results and change the tuning condition? |
The message is displayed when parameters are changed for a file where tuning has been already completed. |
When you do not want to invalidate the tuning results, do not change parameters. |
7334 |
Are you sure you want to delete the selected monitor group? |
The message is displayed when deleting the currently selected monitor group. |
Click [OK] to delete and [Cancel] not to delete. |
7335 |
The group name is already registered. |
Registration of a new monitor group was attempted under a name already |
Change the monitor group name. |
7336 |
The default setting group cannot be deleted. |
Deletion of a standard monitor group was attempted. |
The default monitor group cannot be deleted. |
7337 |
Groups are already registered to the maximum (16). |
Up to 16 monitor groups including 2 standard groups can be registered. |
Delete unnecessary monitor group and register new one. |
7338 |
The Calibration Table will be supplemented. \nDo you perform? |
The message is displayed for confirmation prior to the supplement of the m/z Calibration Table. |
Click [OK] to execute supplement and [Cancel] not to. |
7339 |
The Calibration Table will be initialized. \nAre you sure you want to perform? |
The message is displayed for confirmation prior to the initialization of the m/z Calibration Table. |
Click [OK] to initialize and [Cancel] not to. |
733A |
m/z values cannot be added. |
New m/z values could not be added because the m/z Calibration Table was full. |
Up to 14 m/z values can be added. |
733B |
Standard m/z value cannot be deleted. |
Deletion of the standard m/z value was attempted. |
Only added m/z values can be deleted. |
733C |
An error occurred during base level detection in Auto-Tuning. |
----- |
Contact your Shimadzu representative. |
733D |
An error occurred during initial adjustment in Auto-Tuning. |
There is no standard sample. |
Prepare sufficient amount of the standard sample. |
733E |
An error occurred during resolution adjustment in AutoTuning. \n(Detector sensitivity is insufficient.) |
----- |
Perform auto-tuning again. Clean the lens. |
733F |
An error occurred during resolution adjustment in AutoTuning. \n(Resolution cannot be narrowed down.) |
----- |
Perform auto-tuning again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
7340 |
An error occurred during sensitivity adjustment in AutoTuning. \n(Detector sensitivity is insufficient.) |
----- |
Perform auto-tuning again. Clean the lens. |
7341 |
An error occurred during m/z calibration in Auto-Tuning. \n(Detector sensitivity is insufficient.) |
----- |
Perform auto-tuning again. Clean the lens. |
7342 |
An error occurred during m/z calibration in Auto-Tuning. \n(No peak for calibration m/z value is found.) |
----- |
Perform auto-tuning again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
7343 |
An error occurred during resolution adjustment in AutoTuning. \n(Peak intensity is too |
----- |
Perform auto-tuning again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
7345 |
Received peak profile data is abnormal. |
There may be a problem in communication or MS system state. |
Restart the MS. |
7347 |
Tuning could not be completed. |
Inappropriate tuning condition may be set. |
Change tuning conditions. |
734A |
The low-mass value is set larger than the high-mass value. |
----- |
Set the low-mass value below the high-mass value. |
734D |
Auto-Tuning started. |
----- |
---- |
734E |
Auto-Tuning has been |
----- |
---- |
7351 |
MS: Fluid leakage |
Fluid leakage is detected. |
Stop the LC pump and take care of the fluid leakage. |
7353 |
MS: Overheat. (Interface) \n If the temperature does not go down in several minutes, turn OFF the power of the MS immediately. |
----- |
If the temperature does not go down in several minutes, immediately quit LCMSsolution and turn OFF the power of the MS.\nIt is no problem when changing APCI Interface. |
7354 |
MS: Overheat. (DL) \n If the temperature does not go down in several minutes, turn OFF the power of the MS immediately. |
----- |
If the temperature does not go down in several minutes, immediately quit LCMSsolution and turn OFF the power of the MS. |
7355 |
MS: Overheat. (Heat block) \n If the temperature does not go down in several minutes, turn OFF the power of the MS immediately. |
The heat block may be damaged or the peripheral equipment may have a problem. |
1. For safety, check for any abnormal conditions (smoke, strange odor) around the heat block. \n 2. If there is any abnormal condition or the temperature does not go down in several minutes, stop the instrument and turn OFF the power of the main unit. \n 3. Contact your Shimadzu |
7356 |
MS: Turbo pump does not become ready. |
----- |
Repair or part replacement may be necessary. Contact your Shimadzu representative. |
7358 |
MS: Vacuum level of the IG is insufficient. |
Vacuum level of the MS has lowered. |
Check for vacuum leakage on the MS. |
735C |
MS: Interface is not connected. |
----- |
Connect the interface. |
735D |
MS: An error occurred in the high-frequency power supply. (Main rod) |
The high-frequency power supply is out of tune. |
Perform auto-tuning. If the error persists, detailed inspection is necessary. Contact your Shimadzu representative. |
735E |
MS: An error occurred in the high-frequency power supply. (Q- |
The high-frequency power supply is out of tune. |
Perform auto-tuning. If the error persists, detailed inspection is necessary. Contact your Shimadzu |
735F |
MS: An error occurred in the high-frequency power supply. (Octopole) |
The high-frequency power supply is out of tune. |
Perform auto-tuning. If the error persists, detailed inspection is necessary. Contact your Shimadzu representative. |
7360 |
MS: An error occurred in the high-pressure power supply. |
The current limiter of the high-pressure power supply was activated. |
Perform auto-tuning. If the error persists, detailed inspection is necessary. Contact your Shimadzu |
7361 |
MS: An error occurred in the high-pressure power supply. (Conversion) |
The current limiter of the high-pressure power supply was activated. |
Perform auto-tuning. If the error persists, detailed inspection is necessary. Contact your Shimadzu representative. |
7362 |
MS: An error occurred in the high-pressure power supply. |
The current limiter of the high-pressure power supply was activated. |
Perform auto-tuning. If the error persists, detailed inspection is necessary. Contact your Shimadzu |
7363 |
MS: The PG is not connected or has been disconnected. |
The Pirani gauge to measure low vacuum levels has not been connected or the wiring has been disconnected. |
Contact your Shimadzu representative. |
7366 |
MS: The temperature sensor is disconnected. (Heat block) |
The heat block temperature sensor may be disconnected or its peripheral circuit may have a problem. |
Contact your Shimadzu representative. |
7367 |
MS: The temperature sensor is disconnected. (DL) |
The DL temperature sensor may be disconnected or its peripheral circuit may have a problem. |
Contact your Shimadzu representative. |
7368 |
MS: The temperature sensor is disconnected. (Interface) |
The interface temperature sensor may be disconnected or its peripheral circuit may have a problem.\nIt is no problem when changing APCI Interface. |
Contact your Shimadzu representative. |
736A |
MS: An abnormal condition is detected on the heater. (DL) |
The DL temperature control heater or its control circuit may have a problem. |
1. For safety, stop the instrument and turn OFF the power of the main unit. \n2. Contact your Shimadzu |
736B |
MS: An abnormal condition is detected on the heater. |
The interface temperature control heater or its control circuit may have a problem. |
1. For safety, stop the instrument and turn OFF the power of the main unit. \n2. Contact your Shimadzu |
736C |
MS: An abnormal condition is detected on the heater. (Heat |
The heat block temperature control heater or its control circuit may have a problem. |
1. For safety, stop the instrument and turn OFF the power of the main unit. \n2. Contact your Shimadzu |
736E |
MS: IG will be turned OFF because the IG vacuum level is insufficient. |
The vacuum gauge was turned OFF by the protection function since the vacuum level of the MS lowered. |
Check for vacuum leakage on the MS. |
736F |
MS: Auto Start of the vacuum system started. |
----- |
----- |
7370 |
MS: Auto Stop of the vacuum system started. |
----- |
----- |
7371 |
MS: Auto Start of the vacuum system was suspended. |
----- |
----- |
7372 |
MS: Auto Stop of the vacuum system was suspended. |
----- |
----- |
7373 |
MS: Rotary pump was started. |
----- |
----- |
7374 |
MS: Rotary pump was stopped. |
----- |
----- |
7375 |
MS: Turbo molecular pump 1 was started. |
----- |
----- |
7376 |
MS: Turbo molecular pump 1 was stopped. |
----- |
----- |
7377 |
MS: Turbo molecular pump 2 was started. |
----- |
----- |
7378 |
MS: Turbo molecular pump 2 was stopped. |
----- |
----- |
7379 |
MS: Leak valve has been |
----- |
----- |
737A |
MS: Leak valve has been closed. |
----- |
----- |
737B |
MS:DL heater was turned ON. |
----- |
----- |
737D |
MS: Heater of the APCI interface was turned ON. |
----- |
----- |
737F |
MS: Ion gauge has been lit up. |
----- |
----- |
7381 |
MS: I/F shutter has been opened. |
----- |
----- |
7383 |
MS: Nebulizer gas valve has been opened. |
----- |
----- |
7385 |
MS: Heater of the heat block was turned ON. |
----- |
----- |
7388 |
MS: The shutter is closed. Please open the shutter to perform data acquisition. |
----- |
Open the shutter prior to starting data acquisition. |
7389 |
MS: Detector was turned ON. |
----- |
----- |
738A |
MS: Detector was turned OFF. |
----- |
----- |
738B |
MS: The following heater will be turned OFF. |
----- |
----- |
738C |
MS: Operation is disabled because data acquisition or tuning is currently performed. |
----- |
----- |
738E |
MS: No tuning file. Perform AutoTuning or Manual Tuning. |
----- |
----- |
738F |
MS: Nebulizing gas flow rate may be insufficient or unstable. \n The LC pump was stopped and data acquisition was suspended. |
Data acquisition could not start due to insufficient nebulizing gas flow rate. The LC pump was stopped to prevent contamination of the interface. |
Check the nebulizing gas flow rate on the instrument monitor. If the gas flow rate does not increase, there may be a problem on gas piping or gas cylinder. Once sufficient nebulizing gas flow rate is achieved, data acquisition can |
7395 |
MS: The detector could not be turned ON during data acquisition. |
After the detector was turned ON, it was turned OFF by the hardware protection mechanism. |
The problem occurs mainly when vacuum level is not completely stable. Turn the ion gauge ON manually, make sure that vacuum level has stabilized, and then turn ON the detector manually. If the problem recurs, contact your Shimadzu representative. |
7396 |
MS: The high-frequency power supply could not be turned ON during data acquisition. |
After the detector was turned ON, it was turned OFF by the hardware protection mechanism. |
The problem occurs mainly when vacuum level is not completely stable. Turn the ion gauge ON manually, make sure that vacuum level has stabilized, and then turn ON the detector manually. If the problem recurs, contact your Shimadzu representative. |
7397 |
MS: The APPI lamp could not light up. |
There are several possible causes as follows: \n (1) The gas sealed in the lamp ran out. \n(2) The cable to be connected to the APPI interface is unconnected. \n(3) The cable connected to the APPI interface has been disconnected. |
Replace the APPI lamp with a new one. If the error persists even after the replacement, contact your Shimadzu representative. |
7398 |
MS: The lamp was turned OFF due to high temperature of APPI. |
The APPI interface is overheated. |
The lamp can light up when the APPI interface cools down. If the problem recurs, contact your Shimadzu |
7399 |
MS: The lamp has been lit again because the APPI temperature lowered. |
The APPI interface became hot and the lamp was turned OFF. Then, the lamp was automatically lit again because the temperature lowered. |
----- |
739A |
MS: An abnormal condition of the APPI temperature protection circuit was detected. \nThe detector was turned Off for |
The lamp is usually turned OFF when the APPI interface becomes hot. However, the lamp remained lit. |
Contact your Shimadzu representative. |
739B |
MS: No applicable command. |
A command not existing in the LC time program or MS time program was |
Enter a command according to the dropdown list. |
739C |
MS: There is a mistake in the program. |
Combination of rates and values is incorrectly set in the temperature or flow rate program. |
There is a mistake, for example, the temperature is set to become lower from a higher value to lower value even when a rate is a positive value. Correct the setting. |
739E |
MS: The starting time is set later than the ending time. |
The value for the starting time is set larger than the ending time. |
Set the value for the starting time smaller than the ending time. |
73A2 |
MS: Please set the m/z value to be measured. |
No m/z value to be measured is set on the table. |
Set an m/z value. |
73B5 |
MS: Please enter at least one data acquisition condition. |
No data acquisition condition is set for MS instrument parameters. |
Set data acquisition conditions. When you do not want to operate the MS, remove it from the system configuration. |
73B6 |
MS: Measurement time is too short. |
The same value is set both for starting and ending of measurement time. |
Set the value for the ending time larger than the starting time. |
73B8 |
MS: Change the oil of the rotary pump. |
Operation time of the rotary pump has exceeded the limit for maintenance specified by configuration. |
Change the oil of the rotary pump. |
73BA |
MS: The interface cover is open. |
----- |
Close the interface cover. |
73BB |
MS: The temperature sensor is short-circuited. (Interface) |
The interface temperature sensor may be short-circuited or its peripheral circuit may have a problem. |
Contact your Shimadzu representative. |
73BC |
MS: The temperature sensor is short-circuited. (DL) |
The DL temperature sensor may be shortcircuited or its peripheral circuit may have a problem. |
1. For safety, stop the instrument and turn OFF the power of the main unit. \n2. Contact your Shimadzu representative. |
73BD |
MS: The temperature sensor is short-circuited. (Heat block) |
Heat block temperature sensor may be short-circuited or its peripheral circuit may have a problem. |
1. For safety, stop the instrument and turn OFF the power of the main unit. \n2. Contact your Shimadzu representative. |
73BE |
MS: The temperature sensor is short-circuited. (RF capacitor) |
The RF capacitor temperature sensor may be short-circuited or its peripheral circuit may have a problem. |
1. For safety, stop the instrument and turn OFF the power of the main unit. \n2. Contact your Shimadzu representative. |
73BF |
MS: The temperature sensor is disconnected. (RF capacitor) |
The RF capacitor temperature sensor may be disconnected or its peripheral circuit may have a problem. |
Contact your Shimadzu representative. |
73C0 |
MS: Overheat. (RF capacitor) If the temperature does not go down in several minutes, turn OFF the power of the MS immediately. |
The RF capacitor temperature control heater or its control circuit may have a problem. |
1. For safety, check for any abnormal conditions (smoke, strange odor) around the heat block. \n 2. If there is any abnormal condition or the temperature does not go down in several minutes, stop the instrument and turn OFF the power of the main unit. \n 3. Contact your Shimadzu |
73C1 |
MS: CPLD watch dog time error |
The main unit’s CPU performed an unexpected behavior. |
Contact your Shimadzu representative. |
73C2 |
The detector will be turned OFF due to insufficient IG vacuum level. |
The detector was turned OFF by the protection function since the vacuum level of the MS was insufficient. |
Check for vacuum leakage on the MS. |
73C3 |
MS: Automatic start of the vacuum system, low vacuum error |
The low vacuum level did not reach the specified value in the specific period of time during automatic start. Automatic start is stopped. |
Check for vacuum leakage. Repair or part replacement may be necessary. Contact your Shimadzu representative. |
73C4 |
MS: Automatic start of the vacuum system, high vacuum error |
The vacuum level did not reach the specified value in the specific period of time during automatic start. Automatic |
Check for vacuum leakage. Repair or part replacement may be necessary. Contact your Shimadzu representative. |
73C5 |
MS: Automatic start of the vacuum system, IG error |
The IG could not light up in the specific period of time during automatic start. Automatic start is stopped. |
Check for vacuum leakage. Repair or part replacement may be necessary. Contact your Shimadzu representative. |
73C6 |
MS: Automatic start of the vacuum system, TMP error |
The rotation of the turbo pump did not stabilize in the specific period of time during automatic start. Automatic start is |
Check for vacuum leakage. Repair or part replacement may be necessary. Contact your Shimadzu representative. |
73C7 |
MS: Warning to signify that excessive time is being spent to wait for the temperature to lower when stopping the instrument. (Heat block) |
The temperature of the heat block did not lower enough in the specific period of time during automatic stop. |
1. Check if heat release is being prevented by an obstacle placed on the heater of the heat block. \n2. If the problem persists, contact your Shimadzu representative. |
73C8 |
MS: Warning to signify that excessive time is being spent to wait for the temperature to lower when stopping the instrument. |
The temperature of the DL did not lower enough in the specific period of time during automatic stop. |
1. Check if heat release is being prevented by an obstacle placed on the heater of the DL. \n2. If the problem persists, contact your Shimadzu representative. |
73C9 |
MS: Warning to signify that excessive time is being spent to wait for the temperature to lower when stopping the instrument. (Interface) |
The temperature of the interface heater did not lower enough in the specific period of time during automatic stop. |
1. Check if heat release is being prevented by an obstacle placed on the heater of the interface. \n2. If the problem persists, contact your Shimadzu representative. |
73CA |
MS: Detector protection error due to A/D saturation |
Excessively strong ion signals were inputted to the detector continuously. The switch has been set to OFF in order to protect the detector. |
The voltage on the detector is too high. Lower the voltage on the detector. When the solvent spray on the interface is uneven, the DL may suck in droplets. \n Adjust the nebulizing gas flow rate and LC pump flow rate. When an ESI interface is used, the tip of the quartz tube capillary in the interface may be worn. \n In this case, cut the tip of the capillary referring to the instruction manual. There has been a case where an electric discharge occurred between the capillary tip and DL (only when an ESI interface is used). \n Adjust the length of the capillary tip. Normally, approximately 0.5 mm should stick out from the tip of the |
73CB |
MS: Drying gas has run out. |
Drying gas may not be supplied to the instrument or the piping for drying gas may not be done properly. |
1. Check the gas supply source and piping. \n2. If the problem persists, contact your Shimadzu representative. |
73CC |
MS: The nebulizer gas nozzle is clogged. The nebulizer gas is not properly controlled. |
The nebulizer gas nozzle may be clogged. Perform maintenance on the nozzle. |
Check the nebulizer gas nozzle and clear any clogging.. If the error persists, contact your Shimadzu representative. |
73CD |
MS: The drying gas is not properly controlled. |
The drying gas flow path is clogged or its pressure sensor or peripheral circuit may have a problem. |
Contact your Shimadzu representative. |
73CE |
MS: The liquid temperature sensor is disconnected. |
The liquid leakage sensor may be disconnected or its peripheral circuit may have a problem. |
Contact your Shimadzu representative. |
73CF |
MS: The room temperature sensor is disconnected. |
The liquid leakage sensor may be disconnected or its peripheral circuit may have a problem. |
Contact your Shimadzu representative. |
73D0 |
MS: The liquid temperature sensor is short-circuited. |
The liquid leakage sensor may be disconnected or its peripheral circuit may have a problem. |
Contact your Shimadzu representative. |
73D1 |
MS: The temperature sensor is short-circuited. |
The liquid leakage sensor may be disconnected or its peripheral circuit may have a problem. |
Contact your Shimadzu representative. |
|
MS: FCV-20A error |
----- |
??? |
73D7 |
MS: Provide maintenance service to the rotary pump. |
Maintenance service is necessary for the operation of the rotary pump. |
Maintenance service is necessary for the rotary pump. \n Contact your Shimadzu representative. |
73D8 |
MS: Operation time of the detector has exceeded the standard replacement time. |
Operation time of the detector has exceeded the limit for maintenance specified by configuration. |
Replace the MS detector with a new one. |
73D9 |
MS: Operation time of the APPI lamp has exceeded the standard replacement time. |
Operation time of the APPI lamp has exceeded the limit for maintenance specified by configuration. |
Replace the APPI lamp with a new one. |
73DA |
MS: Turbo molecular pump does not become ready. |
----- |
Repair or part replacement may be necessary. Contact your Shimadzu representative. |
73DB |
MS: Even though data acquisition started, the MS detector did not start. |
MS data acquisition did not start. |
Check if the system controller relay is appropriate on system configuration and make sure that the cable is correctly connected. |
73DC |
MS: Firmware update failed. |
Firmware update failed. |
Contact your Shimadzu representative. |
73DD |
MS: To turn ON the drying gas, please close the source window. |
The drying gas ON command was executed when the interface cover was |
Close the interface cover and then execute the drying gas ON command. |
73DE |
MS: Measurement data could not be loaded in time. |
Measurement data could not be loaded on the PC in time and buffer in the instrument became full. A high-load application may be working on the PC. |
Quit the high-load application and perform the operation again. |
73DF |
Are you sure you want to delete the selected m/z? |
Deletion of m/z was attempted. |
Click [Yes] to delete and [No] not to delete. |
73E0 |
Use the file as the default tuning file? |
----- |
Click [Yes] to load the tuning file next time the [Tuning] window is opened. Click [No] otherwise. |
73E1 |
MS: An unknown interface is connected. |
An unknown interface is connected. |
Connect a supported interface. |
73E2 |
MS: To perform Auto-Tuning, use ESI or DUIS interface. |
Auto tuning cannot be performed when an APCI (APPI) interface is mounted. |
Change the interface to the one that auto-tuning can be performed with. |
73E3 |
An error occurred during sensitivity adjustment in Auto- Tuning. \n(Peak intensity is too |
----- |
Perform auto-tuning again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
73E4 |
An error occurred during m/z calibration in Auto-Tuning. \n(Peak intensity is too great.) |
----- |
Perform auto-tuning again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
73E5 |
An error occurred during speed calibration in Auto-Tuning. \n(Detector sensitivity is insufficient.) |
----- |
Perform auto-tuning again. Clean the lens. |
73E6 |
An error occurred during speed calibration in Auto-Tuning. \n(Peak intensity is too great.) |
----- |
Perform auto-tuning again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
73E7 |
An error occurred during speed calibration in Auto-Tuning. \n(No peak for calibration m/z value is found.) |
----- |
Perform auto-tuning again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
73E8 |
An error occurred during peak profile measurement in AutoTuning. \n(Detector sensitivity is insufficient.) |
----- |
Perform auto-tuning again. Clean the lens. |
73E9 |
An error occurred during peak profile measurement in AutoTuning. \n(Peak intensity is too |
----- |
Perform auto-tuning again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
73EA |
Registry values for the following items do not match between the server and client PCs. Contact a service person. \nSettling time \nPolarity switching time |
The following setting values are different between the server and client machines. \n Set ring time \n Polarity switching time |
Contact your Shimadzu representative. |
73EB |
MS: Data acquisition time could not be changed. |
The data acquisition extension command was executed when no data acquisition was performed. |
Execute the data acquisition extension command during a data acquisition. |
73EC |
MS: Cannot start during the process to stop the standard sample. |
The sample injection start command was executed during the process to stop the standard sample. |
Wait for a while. |
73ED |
MS: The operation is disabled during the injection of the standard sample. |
One of the following operations was attempted during the injection of the standard sample. \n1. Turn the drying gas On/Off\n2. Set the drying gas flow rate |
Wait for a while. |
73EE |
MS: The standard sample was stopped in order to change the drying gas flow rate. |
Drying gas flow rate was changed during the injection of the standard sample. |
----- |
73EF |
Because the error had occurred while running the Manual Tuning, the Manual Tuning was |
The error had occurred while executing the manual tuning. |
Refer to the content of the error help and address the error. Then performs the manual tuning again. |
73F0 |
MS: The source window is open. |
----- |
Close the source window. |
73F1 |
MS: The detector could not be turned ON. |
Although the detector was turned ON, it was turned OFF for the hardware protection. |
Decrease the detector voltage and wait a moment, then Operate again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
73F2 |
MS: The high-frequency power supply could not be turned ON. |
Although the power supplayr was turned ON, it was turned OFF for the hardware protection. |
Wait a moment and Operate again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
7701 |
MS: A command that cannot be used for communication with the MS was used. |
Communication with the MS cannot be normally achieved for some reason. |
Check the cable and power supply, turn OFF the power of all equipment including the PC, and restart them. |
7702 |
MS: Invalid parameters were sent. |
This problem may occur when the system is unstable. |
Restart the PC and retry the operation. |
7703 |
MS: Parameters to be set for the MS are incorrect. |
Communication with the MS cannot be normally achieved for some reason. |
Check the cable and power supply, turn OFF the power of all equipment including the PC, and restart them. |
7704 |
MS: The size of data received is different from the requested size. |
Communication with the MS cannot be normally achieved for some reason. |
Turn OFF the power of all equipment including the PC, and restart them. |
7705 |
MS: A hardware error occurred during communication with the MS. |
A hardware error occurred during communication with the MS (mass spectrometer unit). |
Check the cable and power supply, turn OFF the power of all equipment including the PC, and restart them. |
7706 |
MS: Acquisition can not be started with this acquisition |
The number of events are too much or measurement range of a scan is too wide. |
Please reduce the number of events or make measurement range of a scan narrow. |
7707 |
MS: The interface type is not |
----- |
----- |
7708 |
MS: The interface type is not |
----- |
----- |
7709 |
MS: The interface type is not |
----- |
----- |
770A |
MS: Open leakvalve. |
----- |
----- |
770B |
MS: The rotary pump is On. |
----- |
----- |
770C |
MS: The rotary pump is Off. |
----- |
----- |
770D |
MS: The turbo pump is On. |
----- |
----- |
770E |
MS: The turbo pump is in the state of the standby. |
----- |
----- |
770F |
MS: The turbo pump is Off. |
----- |
----- |
7710 |
MS: Vacuum level is outside the range. (PG) |
----- |
----- |
7711 |
MS: Vacuum level is outside the range. (IG) |
----- |
----- |
7712 |
MS: Overheat. |
----- |
----- |
7713 |
MS: The temperature sensor is disconnected or Interface is not connected. |
----- |
----- |
7714 |
MS: The temperature sensor is short-circuited. |
----- |
----- |
7715 |
MS: An abnormal condition is detected on the heater. |
----- |
----- |
7716 |
MS: The heater is Off. (DL) |
----- |
----- |
7717 |
MS: The temperature is outside the range. (DL) |
----- |
----- |
7718 |
MS: The temperature parameter self adjustment has already been |
----- |
----- |
7719 |
MS: Gas flow rate may be insufficient or unstable. |
----- |
----- |
771A |
MS: The gas pressure sensor has a problem. |
----- |
----- |
771B |
MS: Nebulizing Gas is Off. |
----- |
----- |
771C |
MS: An error occurred in the high-pressure power supply. |
----- |
----- |
771D |
MS: An error occurred in the high-frequency power supply. |
----- |
----- |
771E |
MS: The analysis is running. |
----- |
----- |
771F |
MS: The analysis is not running. |
----- |
----- |
7720 |
MS: The initial parameter has not been received. |
----- |
----- |
7721 |
MS: The Method parameter has not been received. |
----- |
----- |
7722 |
MS: There no event the specified No. |
----- |
----- |
7723 |
MS: The settling time is longer than the polarity switch waiting |
----- |
----- |
7724 |
MS: The scanspeed is outside of the range. |
The Settling time may have been changed. |
Check the change of the Settling time at Log Browser. |
7725 |
MS: The width of Maicroscan is outside the range. (SIM) |
----- |
----- |
7726 |
MS: The number of event is outside the range. |
The number of event exceeds the upper or lower limit. |
In [MS] tab page in the [Instrument Parameter View], add or remove events. |
7727 |
MS: The number of Channel is outside the range. (SIM) |
----- |
----- |
7728 |
MS: Auto MS/MS dependent event cannot be extended. |
----- |
----- |
7729 |
MS: A standard sample is being injected. |
----- |
----- |
772B |
MS: A standard sample does the stop procedure. |
----- |
----- |
772C |
MS: The loader for the program update is not started. |
----- |
----- |
772D |
MS: The ion gauge is On. |
----- |
----- |
772E |
MS: An abnormal condition is detected on the fan. (Turbo molecular pump) |
----- |
----- |
772F |
MS: An abnormal condition is detected on the fan. (Main source of electrical power) |
----- |
----- |
7730 |
MS: An abnormal condition is detected on the fan. (Ionization interface) |
----- |
----- |
7731 |
MS: The high-pressure power supply is On. |
----- |
----- |
7732 |
MS: The high-frequency power supply is On. |
----- |
----- |
7733 |
MS: The interface is not a type with a heater. |
----- |
----- |
7734 |
MS: The heated gas is not Off. |
Heating gas cannot be shut OFF, due to the high interface temperature. |
Wait for the interface to cool sufficiently before shutting the heating gas OFF. |
7735 |
MS: The heated gas is not Off. |
Heating gas cannot be shut OFF because the interface (heated ESI) temperature control is ON. |
Wait for the interface to cool sufficiently before shutting the heating gas OFF. |
7736 |
MS: The interface power is not On. |
The interface heater cannot be switched ON because heating gas is OFF. |
Switch the heating gas ON before switching the interface heater ON. |
7737 |
MS: Ms device is not active. Summed value of heating gas flow and drying gas flow is over the upper limit. |
Summed value of heating gas flow and drying gas flow set at interface tab or MS program is over the upper limit. |
Set the value so that the sum of heating gas flow and drying gas flow is under the upper limit, or disable the sum of gas flow limit using the service tool. |
7738 |
MS: The drying gas is not On. Summed value of heating gas flow and drying gas flow is over the upper limit. |
Summed value of heating gas flow and drying gas flow set at interface tab or MS program is over the upper limit. |
Set the value so that the sum of heating gas flow and drying gas flow is under the upper limit, or disable the sum of gas flow limit using the service tool. |
7739 |
MS: The heating gas is not On. Summed value of heating gas flow and drying gas flow is over the upper limit. |
Summed value of heating gas flow and drying gas flow set at interface tab or MS program is over the upper limit. |
Set the value so that the sum of heating gas flow and drying gas flow is under the upper limit, or disable the sum of gas flow limit using the service tool. |
773A |
MS: The number of survey scan events is outside the range. |
The number of survey scan events is outside the range. |
Please make the number of the survey scan events into 32 or less. |
773B |
MS: ""Same Precursor as above event"" is selected in the head dependent event. |
----- |
----- |
7750 |
MS: Interface is not connected. |
----- |
Connect the interface. |
7751 |
MS: The source window is open. |
----- |
Close the interface cover. |
7753 |
MS: An error occurred in the high-pressure power supply. |
The current limiter of the high-pressure power supply was activated. |
Perform auto-tuning. If the error persists, detailed inspection is necessary. Contact your Shimadzu |
7754 |
MS: The APPI lamp could not light up. |
There are several possible causes as follows: \n (1) The gas sealed in the lamp ran out. \n(2) The cable to be connected to the APPI interface is unconnected. \n(3) The cable connected to the APPI interface has been disconnected. |
Replace the APPI lamp with a new one. If the error persists even after the replacement, contact your Shimadzu representative. |
7755 |
MS: The lamp was turned Off due to high temperature of APPI. |
The APPI interface is overheated. |
The lamp can light up when the APPI interface cools down. If the problem recurs, contact your Shimadzu |
7756 |
MS: Overheat. (Interface) \n If the temperature does not go down in several minutes, turn Off the power of the MS immediately. |
----- |
If the temperature does not go down in several minutes, immediately quit LCMSsolution and turn OFF the power of the MS.\nIt is no problem when changing APCI Interface. |
7757 |
MS: The temperature sensor is disconnected or Interface is not connected.(Iinterface) |
The Interface temperature sensor may be disconnected or its peripheral circuit may have a problem. |
Contact your Shimadzu representative. |
7758 |
MS: The temperature sensor is short-circuited. (Interface) |
The interface temperature sensor may be short-circuited or its peripheral circuit may have a problem. |
Contact your Shimadzu representative. |
7759 |
MS: An abnormal condition is detected on the heater. |
The interface temperature control heater or its control circuit may have a problem. |
1. For safety, stop the instrument and turn OFF the power of the main unit. \n2. Contact your Shimadzu |
775A |
MS: Overheat. (DL) \n If the temperature does not go down in several minutes, turn Off the power of the MS immediately. |
----- |
If the temperature does not go down in several minutes, immediately quit LCMSsolution and turn OFF the power of the MS. |
775B |
MS: The temperature sensor is disconnected. (DL) |
The DL temperature sensor may be disconnected or its peripheral circuit may have a problem. |
Contact your Shimadzu representative. |
775C |
MS: The temperature sensor is short-circuited. (DL) |
The DL temperature sensor may be shortcircuited or its peripheral circuit may have a problem. |
1. For safety, stop the instrument and turn OFF the power of the main unit. \n2. Contact your Shimadzu representative. |
775D |
MS: An abnormal condition is detected on the heater. (DL) |
The DL temperature control heater or its control circuit may have a problem. |
1. For safety, stop the instrument and turn OFF the power of the main unit. \n2. Contact your Shimadzu |
775F |
MS: Overheat. (Heat block) \n If the temperature does not go down in several minutes, turn Off the power of the MS immediately. |
The heat block may be damaged or the peripheral equipment may have a problem. |
1. For safety, check for any abnormal conditions (smoke, strange odor) around the heat block. \n 2. If there is any abnormal condition or the temperature does not go down in several minutes, stop the instrument and turn OFF the power of the main unit. \n |
7760 |
MS: The temperature sensor is disconnected. (Heat block) |
The heat block temperature sensor may be disconnected or its peripheral circuit may have a problem. |
Contact your Shimadzu representative. |
7761 |
MS: The temperature sensor is short-circuited. (Heat block) |
Heat block temperature sensor may be short-circuited or its peripheral circuit may have a problem. |
1. For safety, stop the instrument and turn OFF the power of the main unit. \n2. Contact your Shimadzu representative. |
7762 |
MS: An abnormal condition is detected on the heater. (Heat |
The heat block temperature control heater or its control circuit may have a problem. |
1. For safety, stop the instrument and turn OFF the power of the main unit. \n2. Contact your Shimadzu |
7763 |
MS: Overheat. (Q1 RF capacitor) If the temperature does not go down in several minutes, turn Off the power of the MS immediately. |
The Q1 RF capacitor temperature control heater or its control circuit may have a problem. |
1. For safety, check for any abnormal conditions (smoke, strange odor) around the heat block. \n 2. If there is any abnormal condition or the temperature does not go down in several minutes, stop the instrument and turn OFF the power of the main unit. \n |
7764 |
MS: Overheat. (Q3 RF capacitor) If the temperature does not go down in several minutes, turn Off the power of the MS immediately. |
The Q3 RF capacitor temperature control heater or its control circuit may have a problem. |
1. For safety, check for any abnormal conditions (smoke, strange odor) around the heat block. \n 2. If there is any abnormal condition or the temperature does not go down in several minutes, stop the instrument and turn OFF the power of the main unit. \n |
7765 |
MS: Nebulizing gas flow rate may be insufficient or unstable. \n The LC pump was stopped and data acquisition was suspended. |
Data acquisition could not start due to insufficient nebulizing gas flow rate. The LC pump was stopped to prevent contamination of the interface. |
Check the nebulizing gas flow rate on the instrument monitor. If the gas flow rate does not increase, there may be a problem on gas piping or gas cylinder. Once sufficient nebulizing gas flow rate is achieved, data acquisition can |
7766 |
MS: The nebulizer gas nozzle is clogged. The nebulizer gas is not properly controlled. |
The nebulizer gas nozzle may be clogged. Perform maintenance on the nozzle. |
Check the nebulizer gas nozzle and clear any clogging.. If the error persists, contact your Shimadzu representative. |
7767 |
MS: Drying gas flow rate may be insufficient or unstable. \n The LC pump was stopped and data acquisition was suspended. |
Drying gas may not be supplied to the instrument or the piping for drying gas may not be done properly. |
1. Check the gas supply source and piping. \n2. If the problem persists, contact your Shimadzu representative. |
7768 |
MS: The drying gas is not properly controlled. |
The drying gas flow path is clogged or its pressure sensor or peripheral circuit may have a problem. |
Contact your Shimadzu representative. |
7769 |
MS: CID gas flow rate may be insufficient or unstable. |
CID gas may not be supplied to the instrument or the piping for drying gas may not be done properly. |
1. Check the gas supply source and piping. \n2. If the problem persists, contact your Shimadzu representative. |
776A |
MS: The CID gas is not properly controlled. |
The CID gas flow path is clogged or its pressure sensor or peripheral circuit may have a problem. |
Contact your Shimadzu representative. |
776B |
MS: The temperature sensor is short-circuited. |
The liquid leakage sensor may be disconnected or its peripheral circuit may have a problem. |
Contact your Shimadzu representative. |
776C |
MS: The room temperature sensor is disconnected. |
The liquid leakage sensor may be disconnected or its peripheral circuit may have a problem. |
Contact your Shimadzu representative. |
776D |
MS: The liquid temperature sensor is short-circuited. |
The liquid leakage sensor may be disconnected or its peripheral circuit may have a problem. |
Contact your Shimadzu representative. |
776E |
MS: The liquid temperature sensor is disconnected. |
The liquid leakage sensor may be disconnected or its peripheral circuit may have a problem. |
Contact your Shimadzu representative. |
776F |
MS: Fluid leakage. |
Fluid leakage is detected. |
Stop the LC pump and take care of the fluid leakage. |
7770 |
MS: Vacuum level of the IG is insufficient. |
Vacuum level of the MS has lowered. |
Check for vacuum leakage on the MS. |
7771 |
MS: IG will be turned Off because the IG vacuum level is insufficient. |
The vacuum gauge was turned OFF by the protection function since the vacuum level of the MS lowered. |
Check for vacuum leakage on the MS. |
7772 |
MS: The detector will be turned Off due to insufficient IG vacuum level. |
The detector was turned OFF by the protection function since the vacuum level of the MS was insufficient. |
Check for vacuum leakage on the MS. |
7774 |
MS: The PG is not connected or has been disconnected. |
The Pirani gauge to measure low vacuum levels has not been connected or the wiring has been disconnected. |
Contact your Shimadzu representative. |
7775 |
MS: Vacuum level of the PG is insufficient. Turbo pump does not become ready. |
Turbo pump does not become ready by the protection function since the vacuum level of the MS lowered. |
Check for vacuum leakage on the MS. When turbo pump still does not become a lady, please contact your Shimadzu representative. |
7776 |
MS: Turbo pump does not become ready. |
----- |
Repair or part replacement may be necessary. Contact your Shimadzu representative. |
7777 |
MS: Automatic start of the vacuum system, high vacuum error |
The vacuum level did not reach the specified value in the specific period of time during automatic start. Automatic |
Check for vacuum leakage. Repair or part replacement may be necessary. Contact your Shimadzu representative. |
7778 |
MS: Automatic start of the vacuum system, TMP error |
The rotation of the turbo pump did not stabilize in the specific period of time during automatic start. Automatic start is |
Check for vacuum leakage. Repair or part replacement may be necessary. Contact your Shimadzu representative. |
7779 |
MS: Automatic start of the vacuum system, IG error |
The IG could not light up in the specific period of time during automatic start. Automatic start is stopped. |
Check for vacuum leakage. Repair or part replacement may be necessary. Contact your Shimadzu representative. |
777A |
MS: Automatic start of the vacuum system, low vacuum error |
The low vacuum level did not reach the specified value in the specific period of time during automatic start. Automatic start is stopped. |
Check for vacuum leakage. Repair or part replacement may be necessary. Contact your Shimadzu representative. |
777B |
MS: Warning to signify that excessive time is being spent to wait for the temperature to lower when stopping the instrument. (Heat block) |
The temperature of the heat block did not lower enough in the specific period of time during automatic stop. |
1. Check if heat release is being prevented by an obstacle placed on the heater of the heat block. \n2. If the problem persists, contact your Shimadzu representative. |
777C |
MS: Warning to signify that excessive time is being spent to wait for the temperature to lower when stopping the instrument. |
The temperature of the DL did not lower enough in the specific period of time during automatic stop. |
1. Check if heat release is being prevented by an obstacle placed on the heater of the DL. \n2. If the problem persists, contact your Shimadzu representative. |
777D |
MS: Warning to signify that excessive time is being spent to wait for the temperature to lower when stopping the instrument. (Interface) |
The temperature of the interface heater did not lower enough in the specific period of time during automatic stop. |
1. Check if heat release is being prevented by an obstacle placed on the heater of the interface. \n2. If the problem persists, contact your Shimadzu representative. |
777E |
MS: An error occurred in the high-pressure power supply. (Conversion) |
The current limiter of the high-pressure power supply was activated. |
Perform auto-tuning. If the error persists, detailed inspection is necessary. Contact your Shimadzu representative. |
777F |
MS: An error occurred in the high-pressure power supply. |
The current limiter of the high-pressure power supply was activated. |
Perform auto-tuning. If the error persists, detailed inspection is necessary. Contact your Shimadzu |
7780 |
MS: Detector protection error due to A/D saturation. |
Excessively strong ion signals were inputted to the detector continuously. The switch has been set to OFF in order to protect the detector. |
The voltage on the detector is too high. Lower the voltage on the detector. When the solvent spray on the interface is uneven, the DL may suck in droplets. \n Adjust the nebulizing gas flow rate and LC pump flow rate. When an ESI interface is used, the ti |
7781 |
MS: An error occurred in the high-frequency power supply. |
The high-frequency power supply is out of tune. |
Perform auto-tuning. If the error persists, detailed inspection is necessary. Contact your Shimadzu |
7782 |
MS: An error occurred in the high-frequency power supply. (Multipole) |
The high-frequency power supply is out of tune. |
Perform auto-tuning. If the error persists, detailed inspection is necessary. Contact your Shimadzu representative. |
7783 |
MS: An error occurred in the high-frequency power supply. (Q1 Main rod) |
The high-frequency power supply is out of tune. |
Perform auto-tuning. If the error persists, detailed inspection is necessary. Contact your Shimadzu representative. |
7784 |
MS: An fan error occurred in the high-frequency power supply. (Q1 Main rod) |
The high-frequency power supply fan stopped. |
Contact your Shimadzu representative. |
7785 |
MS: An error occurred in the high-frequency power supply. (Collision Cell) |
The high-frequency power supply is out of tune. |
Perform auto-tuning. If the error persists, detailed inspection is necessary. Contact your Shimadzu representative. |
7786 |
MS: An fan error occurred in the high-frequency power supply. (Collision Cell) |
The high-frequency power supply fan stopped. |
Contact your Shimadzu representative. |
7787 |
MS: An error occurred in the high-frequency power supply. (Q3 Main rod) |
The high-frequency power supply is out of tune. |
Perform auto-tuning. If the error persists, detailed inspection is necessary. Contact your Shimadzu representative. |
7788 |
MS: An fan error occurred in the high-frequency power supply. (Q3 Main rod) |
The high-frequency power supply fan stopped. |
Contact your Shimadzu representative. |
7789 |
MS: FCV-20A error. |
----- |
----- |
778A |
MS: CPLD watch dog time error. |
The main unit’s CPU performed an unexpected behavior. |
Contact your Shimadzu representative. |
778B |
MS: Signal intensity is too high. Detector will be turned off for a few seconds.\n Suggest sample be diluted or detector voltage |
----- |
----- |
778C |
MS: An error occurred in the fan. (Interface) |
The interface fan stopped. |
Contact your Shimadzu representative. |
778D |
MS: An error occurred in the fan. (Turbo pump) The vacuum system was stopped in the emergency automatically. |
The vacuum system was stopped in the emergency automatically by the turbo pump fan stopped in the specific period of time. |
Repair or part replacement may be necessary. Contact your Shimadzu representative. |
778E |
MS: An error occurred in the fan. (Power) The vacuum system was stopped in the emergency automatically. |
The vacuum system was stopped in the emergency automatically by the power fan stopped in the specific period of time. |
Repair or part replacement may be necessary. Contact your Shimadzu representative. |
778F |
MS: Turbo pump will be turned Off due to insufficient PG vacuum level. |
The vacuum gauge was turned Off by the vacuum level of the MS lowered. |
Check for vacuum leakage on the MS. If turbo pump does not become ready even after retry, contact your Shimadzu representative. |
7790 |
MS: The fan has stopped. (Turbo pump) |
The fan has stopped. (Turbo pump) |
Contact your Shimadzu representative. |
7791 |
MS: The fan has stopped. |
The fan has stopped. (Power) |
Contact your Shimadzu representative. |
7792 |
MS: Overheat. (Interface) \n If the temperature does not go down in several minutes, turn OFF the power of the MS immediately. |
----- |
If the temperature does not decrease after waiting a few minutes, immediately exit LabSolutions LCMS and switch OFF the MS unit. \n If this message appears when the interface is being replaced, it is not an error. \n Connect the interface and switch the APCI heater ON. If an error does not occur at that point, then the system is functioning |
7793 |
MS: The temperature sensor is disconnected or Interface is not connected. (Iinterface) |
The interface temperature sensor may be disconnected or its peripheral circuit may have a problem. |
Contact your Shimadzu representative. |
7794 |
MS: The temperature sensor is short-circuited. (Interface) |
The interface temperature sensor may be short-circuited or its peripheral circuit may have a problem. |
Contact your Shimadzu representative. |
7795 |
MS: An abnormal condition is detected on the heater. |
The interface temperature control heater or its control circuit may have a problem. |
1. For safety, stop the instrument and turn OFF the power of the main unit. \n2. Contact your Shimadzu |
7796 |
MS: The heating gas flow rate may be insufficient or unstable. \n The LC pump was stopped and data acquisition was suspended. |
Heating gas may not be supplied to the instrument or the piping for heating gas may not be done properly. |
1. Check the gas supply source and piping. \n2. If the problem persists, contact your Shimadzu representative. |
7797 |
MS: The heating gas is not properly controlled. |
The heating gas flow path is clogged or its pressure sensor or peripheral circuit may have a problem. |
Contact your Shimadzu representative. |
77A0 |
MS: Provide maintenance service to the rotary pump. |
Maintenance service is necessary for the operation of the rotary pump. |
Maintenance service is necessary for the rotary pump. \n Contact your Shimadzu representative. |
77A1 |
MS: Change the oil of the rotary pump. |
Operation time of the rotary pump has exceeded the limit for maintenance specified by configuration. |
Change the oil of the rotary pump. |
77A2 |
MS: Turbo molecular pump should be overhauled. |
The message indicates that the operation time of turbo molecular pump has exceeded the limit for maintenance. Overhaul is recommended. |
Contact your Shimadzu representative. |
77A3 |
MS: Operation time of the detector has exceeded the standard replacement time. |
Operation time of the detector has exceeded the limit for maintenance specified by configuration. |
Replace the MS detector with a new one. |
77A4 |
MS: Operation time of the APPI lamp has exceeded the standard replacement time. |
Operation time of the APPI lamp has exceeded the limit for maintenance specified by configuration. |
Replace the APPI lamp with a new one. |
77A5 |
MS: Operation time of the APCI has exceeded the standard replacement time. |
Operation time of the APCI has exceeded the limit for maintenance specified by configuration. |
Replace the APCI with a new one. |
77A6 |
MS: It is time to replace the exhaust filter. Please maintain the rotary pump. |
Maintenance service is necessary for the operation of the rotary pump. |
Maintenance service is necessary for the rotary pump. \n Contact your Shimadzu representative. |
77B0 |
MS: Measurement data could not be loaded in time. |
Measurement data could not be loaded on the PC in time and buffer in the instrument became full. A high-load application may be working on the PC. |
Quit the high-load application and perform the operation again. |
77B1 |
MS: Because the error had occurred while executing the manual tuning, the manual tuning was discontinued. |
The error had occurred while executing the manual tuning. |
Refer to the content of the error help and address the error. Then performs the manual tuning again. |
77B2 |
MS: Timeout occurred in the communication with the MS. |
There was no response from the MS within the specified period of time. |
Check the cable and power supply, turn OFF the power of all equipment including the PC, and restart them. |
77B3 |
MS: An unknown interface is connected. |
An unknown interface is connected. |
Connect a supported interface. |
77B4 |
MS: The MS is not connected. |
Connection failed because communication with the MS cannot be established. |
Check the cable and power supply. |
77B5 |
MS: Operation is disabled because data acquisition or tuning is currently performed. |
----- |
----- |
77B6 |
MS: Cannot start during the process to stop the standard sample. |
The sample injection start command was executed during the process to stop the standard sample. |
Wait for a while. |
77B7 |
MS: Vacuum level is insufficient. |
This operation is disabled because the vacuum level has not reached the sufficient value. |
Perform the operation after the vacuum system becomes ready on the [System Control] sub-window of the vacuum system. |
77B8 |
MS: Method or tuning condition is not set for the MS. |
Method or tuning information is not set for the MS correctly. |
Turn OFF the power of the PC and MS and restart them. |
77B9 |
MS: Firmware update failed. |
Firmware update failed. |
Contact your Shimadzu representative. |
77BA |
MS: MS connection failed. |
----- |
Check the cable and power supply and then restart the MS and the PC. |
77BB |
MS: The MS unit is busy. |
The MS unit does not receive signals. |
Check the cable and power supply, turn OFF the power of all equipment including the PC, and restart them. |
77BC |
MS: Communication with MS was not achieved within the specified number of retry. |
Communication was attempted for the specific number of time and there was no response from the MS. |
Check the cable and power supply, turn OFF the power of all equipment including the PC, and restart them. |
77BD |
MS: There is no applicable command. |
A command not existing in the LC time program or MS time program was |
Enter a command according to the dropdown list. |
77BE |
MS: Received analysis data is abnormal. |
----- |
Check the cable and power supply, turn OFF the power of all equipment including the PC, and restart them. |
77BF |
MS: Connection to the instrument cannot be achieved. |
The process to connect to the instrument was suspended. |
Restart the PC. |
77C0 |
MS: The trouble occurs in the communication with the MS. Verify the USB cable connections and check that the power is ON. |
Connection failed because communication with the MS cannot be established. |
Verify the USB cable connections and check that the power is ON. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
77C1 |
MS: The heater is Off.(Q1 or Q3 RF capacitor) |
----- |
----- |
77C2 |
MS: The trouble occurs in the communication with the instrument. Please place a USB cable again after closing |
----- |
----- |
77C3 |
MS: Unable to connect to the MS, because communication settings are not correct. |
Communication settings are not correct. |
Switch ON the MS unit and connect the MS unit to the computer. Then open the USB device list in communication settings and select the MS unit to be |
77C4 |
MS: Unable to connect to the MS, because firmware version is not correct. |
The firmware version is different from the Mot file version. |
Start LabSolutions and connect to the instrument. The firmware is updated automatically when it connects to the instrument. |
77D0 |
MS: Even though data acquisition started, the MS detector did not start. |
MS data acquisition did not start. |
Check if the system controller relay is appropriate on system configuration and make sure that the cable is correctly connected. |
77D1 |
MS: The operation is disabled during the injection of the standard sample. |
One of the following operations was attempted during the injection of the standard sample. \n1. Turn the drying gas On/Off\n2. Set the drying gas flow rate |
Wait for a while. |
77D2 |
MS: To turn On the drying gas, please close the source window. |
The drying gas On command was executed when the interface cover was |
Close the interface cover and then execute the drying gas On command. |
77D3 |
MS: CID Gas valve is closed. |
The preset method file is set to use CID gas. However, currently the CID gas valve is closed. |
Open the CID gas valve, or change the method file to one that does not use CID gas. |
77D4 |
MS: CID Gas valve is closed. Do you wish to continue?this process? |
The preset method file is set to use CID gas. However, currently the CID gas valve is closed. |
Open the CID gas valve, or change the method file to one that does not use CID gas. |
77D5 |
MS is not ready, but do you want to start anyway? |
----- |
----- |
77D6 |
MS: Acquisition time is too long. |
Acquisition time exceeds the upper limit. |
Set a shorter Acquisition Time. |
77D7 |
MS: The size of data required for data acquisition is too large. |
The amount of memory required for data acquisition exceeds the limit. |
Set a shorter Acquisition Time. |
77D8 |
MS: Pressure value of PG is below the specified level. |
Possible causes are that the DL is clogged up or the DL plug is attached. |
Check the DL. \nIf it is clogged up, please clean it.\nIf the DL plug is attached, please take it off. \nIf the problem persists, please contact your local Shimadzu |
77D9 |
MS: Pressure value of PG is below the specified level. |
Possible causes are that the DL is clogged up or the DL plug is attached. |
Check the DL. \nIf it is clogged up, please clean it.\nIf the DL plug is attached, please take it off.\nIf the problem persists, please contact your local Shimadzu |
77DA |
MS: There is an event in which event time is longer than acquisition time. Please check the conditions of an event. |
There is an event in which event time is longer than acquisition time. |
Open the method file by the method editor, and adjust acquisition time. |
77DB |
MS: Summed value of heating gas flow and drying gas flow is over the upper limit. |
Summed value of heating gas flow and drying gas flow set at interface tab or MS program is over the upper limit. |
Set the value so that the sum of heating gas flow and drying gas flow is under the upper limit, or disable the sum of gas flow limit using the service tool. |
7800 |
MS: Auto start of the vacuum system started. |
----- |
----- |
7801 |
MS: Auto stop of the vacuum system started. |
----- |
----- |
7802 |
MS: Auto start of the vacuum system was suspended. |
----- |
----- |
7803 |
MS: Auto stop of the vacuum system was suspended. |
----- |
----- |
7804 |
MS: Turbo molecular pump was started. |
----- |
----- |
7805 |
MS: Turbo molecular pump was stopped. |
----- |
----- |
7806 |
MS: Rotary pump was started. |
----- |
----- |
7807 |
MS: Rotary pump was stopped. |
----- |
----- |
7808 |
MS: Leak valve was opened. |
----- |
----- |
7809 |
MS: Leak valve was closed. |
----- |
----- |
780A |
MS: CID Gas valve was opened. |
----- |
----- |
780B |
MS: CID Gas valve was closed. |
----- |
----- |
780C |
MS: The following heater will be turned Off. |
----- |
----- |
780D |
MS: High voltage/high frequency power supply was not able to be turned on because the vacuum of ion gauge is not good. |
The high-voltage/high-frequency power supply could not be turned ON since the MS vacuum level is insufficient. |
Check for vacuum leakage on the MS. |
7830 |
MS: Survey Event which does not have Dependent Events. Do you wish to continue? |
Data is acquired even though there are Survey Events without Dependent Events. |
To acquire data by Survey Events without Dependent Events, click [OK], and to set Dependent Events, click [Cancel]. |
7831 |
MS: CID gas is not used. Do you wish to continue? |
Data is acquired without using CID gas at MS/MS data acquisition at preset conditions. |
To acquire MS/MS data without using CID gas, click [OK], and to acquire MS/MS data using CID gas, click [Cancel] and change the setting. |
7832 |
MS: CID gas is used. Do you wish to continue? |
Data is acquired using CID gas at data acquisition at preset conditions Even though there is no MS/MS Event. |
To acquire MS data using CID gas, click [OK], and to acquire MS data without using CID gas, click [Cancel] and change the setting. |
7833 |
MS: Please set start m/z less than end m/z. |
By an event scanning, the start m/z value is set larger than the end m/z value. |
Set the start m/z value smaller than the end m/z value. |
7834 |
MS: Analog is existed. Survey event and Dependent event can not be copied. |
An attempt was made to paste a Survey Event or a Dependent Event. |
The Survey Event or Dependent Event cannot be executed since analog output is set to this method. |
7835 |
MS: Dependent Event can pasted to only Survey Event. |
An attempt was made to paste a Dependent Event to a different Survey |
Paste the copied Dependent Event to the same Survey Event. |
7836 |
MS: This event can not be pasted. Please check parameters of this event. |
The parameters of this method differ from those of the method for the pasted event. |
Change the method parameters. |
7837 |
MS: Acquisition Time is too short. |
Acquisition time is shorter than event time. |
Please set acquisition time longer than event time. |
7838 |
MS: If this checkbox is unchecked, dependent events are deleted. |
An attempt was made to delete the Survey Event settings. For this reason, the Dependent Event created based on this event also will be deleted. |
To delete the Dependent Event, click [OK]. To cancel the delete, click [Cancel]. |
7839 |
MS: If a survey event is deleted, dependent events are deleted, too. |
An attempt was made to delete the Survey Event. For this reason, the Dependent Event created based on this event also will be deleted. |
To delete the Dependent Event, click [OK]. To cancel the delete, click [Cancel]. |
783A |
MS: Event time is short. Pause time and dwell time of each channels are less than |
The recommended values for the pause time or dwell time could not be ensured since the event time is too short. |
Set a longer event time. |
783B |
MS: Scan Speed is less than 0u/sec. Please change end m/z more than start m/z or long sampling interval. |
The scan speed cannot be ensured since the event time is too short or the measurement end m/z value is the same as the measurement start m/z or less. |
Set a measurement end m/z value that is larger than the measurement start m/z value, or set a longer event time. |
783C |
MS: Scan speed is less than or equal to 15u/sec. Please change wide m/z range or short event |
The scan speed exceeds the lower limit value since the m/z range to measure is too narrow or the event time is too long. |
Set a wider m/z range to measure or set a shorter event time. |
783D |
MS: Scan speed is over 15000u/sec. Please change narrow m/z range or long event |
The scan speed exceeds the upper limit value since the m/z range to measure is too wide or the event time is too short. |
Set a narrower m/z range to measure or set a longer event time. |
783E |
MS: There are invalid commands. |
The time program is set with invalid commands. |
----- |
783F |
MS: The value of DL bias exceeds tolerance range.\nPlease adjust the result of the subtraction of Qarray bias and DL bias to become -240 to |
The DL bias setting range exceeds the tolerance range. |
Adjust so that the value obtained by subtracting the DL bias setting value from the Qarray bias setting value is in the range -240 V to 240 V. |
7840 |
MS: The sum total of pause time and dwell time is over maximum of event time. Please adjust the pause time and dwell time. |
MS: The total of the pause time and dwell time exceeds the maximum value of the event time. |
Adjust the pause time and dwell time values. |
7841 |
MS: Please set precursor m/z. |
The precursor m/z is not specified. |
----- |
7842 |
MS: Please set product m/z. |
The product m/z is not specified. |
----- |
7843 |
MS: Please set m/z of at least one channel. |
No m/z is specified. |
----- |
7844 |
MS: Please set start time before end time. |
The start time is set after the end time. |
----- |
7845 |
MS: Please set target ion and reference ion. |
The target ion or reference ion is not inputed. |
----- |
7846 |
MS: There is no event. MRM event is added automatically. |
This message is displayed when a method file with no events was opened. |
Clicking [OK] displays a method registered with one MRM. |
7847 |
MS: Interface voltage is adjusted. |
The value of the interface voltage set in each event or the time program is not suited to the operation mode of the |
This message notifies you that the voltage was automatically adjusted. |
7848 |
MS: DUIS corona needle voltage is adjusted. |
The value of the DUIS corona needle set in each event or the time program is not suited to the operation mode of the |
This message notifies you that the voltage was automatically adjusted. |
7849 |
MS: Please set the interval of 0.03 min. or more between ESI mode and APCI mode. |
The interval of the command for changing the DUIS operation mode is not 0.03 min or more.\nA 0.03 min or more interval must be provided when using the command for changing the DUIS operation mode in the time program. |
Set the interval of the command for changing the DUIS operation mode to 0.03 min or more. |
784A |
MS: Please set the time of ESI mode and APCI mode 0.03 min. or more. |
The interval of the command for changing the DUIS operation mode is set to less than 0.03 min.\nA 0.03 min or more interval must be provided when using the command for changing the DUIS operation mode in the time program. |
Set the interval of the command for changing the DUIS operation mode to 0.03 min or more. |
784B |
MS: If analog out is set, measurement time of all events will be set same value. Do you wish to continue? |
When analog output is set, the measurement times of all events are automatically changed to be the same. |
To set analog output, click [OK].\nTo cancel analog output, click [Cancel]. |
784C |
MS: measurement time of all events are set value of the selected event. Do you wish to |
The measurement time of all events is changed to the measurement time of the currently selected event. |
To change the measurement time of all events, click [OK].\nTo cancel the change, click [Cancel]. |
784D |
MS: Event time of all events are set value of the selected event. Do you wish to continue? |
The event time of all events is changed to the event time of the currently selected event. |
To change the event time of all events, click [OK].\nTo cancel the change, click [Cancel]. |
784E |
MS: Dwell time of all events are set value of the selected event. Do you wish to continue? |
The dwell time of all MRM Events and SIM Events will be changed. |
To change the dwell time of all MRM Events and SIM Events, click [OK].\nTo cancel the change, click [Cancel]. |
784F |
MS: Pause time of all events are set value of the selected event. Do you wish to continue? |
The pause time of all MRM Events and SIM Events will be changed. |
To change the pause time of all MRM Events and SIM Events, click [OK]. To cancel the change, click [Cancel]. |
7850 |
MS: Please reduce the measurement time or increase the event time of each event. |
The amount of memory required for data acquisition exceeds the limit. |
Shorten the data acquisition time of each event, or set a longer event time. |
7851 |
MS: Please change the start m/z. |
The value set to [Losses of] cannot be measured by the current measurement start m/z setting. |
Change the measurement start m/z setting. |
7852 |
MS: The setting of conversion dynode in the attenuation is changed from the default setting (using a tuning file). |
The setting of conversion dynode in the attenuation is changed from the default setting.\nThe height and area of a peak is influenced by the change of conversion dynode value. |
It is an intended change, click [OK], and to change the setting of conversion dynode, click [Cancel]. |
7853 |
MS: The event time of following events was not changed. |
Measurement is not possible because the total event time for the combined dwell time and pause time for the displayed events exceeds the allowable event time. |
Set event time settings so that the total combined event time for dwell and pause times does not exceed the allowable event time. |
7854 |
MS: Scan speed is over 30000u/sec. Please change narrow m/z range or long event |
The scan speed exceeds the upper limit value since the m/z range to measure is too wide or the event time is too short. |
Set a narrower m/z range to measure or set a longer event time. |
7855 |
MS: Calculated event time or dwell time is longer than the Maximum value. \nPlease shorten the target loop time. |
If Dwell time is calculated with the desired value of the set-up maximum loop time, it will exceed the upper limit of Event time or Dwell time. |
Please change the desired value of the maximum loop time into a small value. |
7856 |
MS: Calculated Dwell Time is shorter than the Minimum value. \nPlease lengthen the target loop time. |
If Dwell time is calculated with the desired value of the set-up maximum loop time, it will be less than the lower limit (0.8msec) of Dwell time. |
Please change the desired value of the maximum loop time into a large value. |
7857 |
MS: Scan speed is over 30000u/sec. Please change narrow m/z range or long event |
The scan speed exceeds the upper limit value since the m/z range to measure is too wide or the event time is too short. |
Set a narrower m/z range to measure or set a longer event time. |
7860 |
MS: Auto-Tuning was terminated. |
----- |
----- |
7861 |
MS: Auto-Tuning started. |
----- |
----- |
7862 |
MS: Auto-Tuning has been completed. |
----- |
----- |
7863 |
MS: An error occurred on the Instrument. |
----- |
----- |
7864 |
MS: An error occurred in AutoTuning. |
----- |
Perform auto-tuning again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
7865 |
MS: Please execute Auto-Tuning after opening the CID gas valve. |
----- |
----- |
7866 |
MS: The gas or the Temperature is not stable. |
----- |
----- |
7867 |
MS: The standard sample was not injected. |
----- |
Perform auto-tuning again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
7868 |
MS: The standard sample was not stopped. |
----- |
Contact your Shimadzu representative. |
7869 |
MS: An error occurred during initial adjustment in Auto-Tuning. |
----- |
Perform auto-tuning again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
786A |
MS: An error occurred during detector adjustment in Auto- Tuning. \n(No peak for Adjustment m/z value is found.) |
----- |
----- |
786B |
MS: An error occurred during sensitivity adjustment in Auto- Tuning. \n(No peak for Adjustment m/z value is found.) |
----- |
----- |
786C |
MS: An error occurred during sensitivity adjustment in AutoTuning. \n (Peak intensity is too great.) |
----- |
Perform auto-tuning again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
786D |
MS: An error occurred during resolution adjustment in AutoTuning. \n(Resolution cannot be narrowed down.) |
----- |
Perform auto-tuning again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
786E |
MS: An error occurred during resolution adjustment in AutoTuning. \n (Peak intensity is too great.) |
----- |
Perform auto-tuning again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
786F |
MS: An error occurred during m/z calibration in Auto-Tuning. \n(No peak for calibration m/z value is found.) |
----- |
Perform auto-tuning again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
7870 |
MS: An error occurred during m/z calibration in Auto-Tuning. \n (Peak intensity is too great.) |
----- |
Perform auto-tuning again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
7871 |
MS: An error occurred during speed calibration in Auto-Tuning. \n(No peak for calibration m/z value is found.) |
----- |
Perform auto-tuning again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
7872 |
MS: An error occurred during speed calibration in Auto-Tuning. \n (Peak intensity is too great.) |
----- |
Perform auto-tuning again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
7873 |
MS: An error occurred during peak profile measurement in Auto-Tuning. \n (Peak intensity is |
----- |
Perform auto-tuning again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
7874 |
MS: The MS does not become ready. |
There was no response from the MS within the specified period of time. |
Check the cable and power supply, turn OFF the power of all equipment including the PC, and restart them. |
7875 |
MS: Pressure value of PG is below the specified level. |
Possible causes are that the DL is clogged up or the DL plug is attached. |
Check the DL. \nIf it is clogged up, please clean it. \nIf the DL plug is attached, please take it off. \nIf the problem persists, please contact your local Shimadzu |
7c00 |
ELSD is not connected. |
ELSD is not connected. |
Check the instrument status and the cable connections. |
7c01 |
Make sure that the Mobile phase doesn't flow to ELSD before closing the valve of nebulizing gas. Do you close the gas valve? |
----- |
The instrument might be damaged when nebulizinggas is stopped while the mobile phase flow to ELSD. Make sure that the Mobile phase doesn't flow when shut the valve of nebulizinggas. |
7c02 |
The acquisition source of ELSD's output is not correct. Are you sure you want to continue? |
The acquisition source of ELSD's output is incorrect. |
Open the property window of ELSD, and set the collect source. It becomes impossible to use a part of function when continuing as it is. |
7D01 |
MS: A command that cannot be used for communication with the MS was used. |
Communication with the MS cannot be normally achieved for some reason. |
Check the cable and power supply, turn OFF the power of all equipment including the PC, and restart them. |
7D02 |
MS: Invalid parameters were sent. |
This problem may occur when the system is unstable. |
Restart the PC and retry the operation. |
7D04 |
MS: The size of data received is different from the requested size. |
Communication with the MS cannot be normally achieved for some reason. |
Turn OFF the power of all equipment including the PC, and restart them. |
7D05 |
MS: Acquisition can not be started with this acquisition |
----- |
----- |
7D26 |
MS: The number of event is outside the range. |
The number of event exceeds the upper or lower limit. |
In [MS] tab page in the [Instrument Parameter View], add or remove events. |
7D33 |
MS: The trouble occurs in the communication with the MS. |
The trouble occurred during communication with the MS (mass spectrometer). |
Verify cable connections and check that the power is ON. Shut down the entire system, including the PC, then restart. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu |
7D34 |
MS: Timeout occurred in the communication with the MS. |
There was no response from the MS within the specified period of time. |
Verify cable connections and check that the power is ON. Shut down the entire system, including the PC, then restart. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu |
7D35 |
MS: The trouble occurs in the communication with the MS. |
The trouble occurred during communication with the MS (mass |
Verify cable connections and check that the power is ON. Shut down the entire system, including the PC, then |
7D36 |
MS: The trouble occurs in the communication with the MS. |
The trouble occurred during communication with the MS (mass |
Verify cable connections and check that the power is ON. Shut down the entire system, including the PC, then |
7D37 |
MS: The trouble occurs in the communication with the MS. |
The trouble occurred during communication with the MS (mass |
Verify cable connections and check that the power is ON. Shut down the entire system, including the PC, then |
7D38 |
MS: Timeout occurred in the communication with the MS. |
There was no response from the MS within the specified period of time. |
Verify cable connections and check that the power is ON. Shut down the entire system, including the PC, then restart. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu |
7D39 |
MS: The trouble occurs in the communication with the MS. |
The trouble occurred during communication with the MS (mass |
Verify cable connections and check that the power is ON. Shut down the entire system, including the PC, then |
7D3A |
MS: The version of software and the device is different. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
7D3B |
MS: The version of software and the device is different. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
7D3C |
MS: The version of software and the device is different. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
7D3D |
MS: The version of software and the device is different. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
7D3E |
MS: The MS unit is not ready. |
The MS unit is not ready. |
Verify cable connections and check that the power is ON. Shut down the entire system, including the PC, then restart. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu |
7D3F |
MS: IP address does not match. |
The IP address set by the software is different from the one registered on the hardware. |
Set the IP address again to be consistent with the actual system configuration. In addition, confirm the connection between hardware. |
7D40 |
MS: The syringe pump is not connected. |
The syringe pump is not connected. |
Check the syringe pump status and the cable connections. |
7D41 |
MS: The syringe speed is outside the range. |
A value which is invalid for the syringe speed has been set. |
Set the value within the appropriate range. |
7D42 |
MS: Acquisition can not be started with this acquisition |
----- |
----- |
7D43 |
MS: The power of the board is Off. |
The power of the board is Off. |
Verify cable connections and check that the power is ON. Shut down the entire system, including the PC, then restart. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu |
7D44 |
MS: The trouble occurs in the communication with the MS. |
The trouble occurred during communication with the MS (mass spectrometer). |
Verify cable connections and check that the power is ON. Shut down the entire system, including the PC, then restart. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu |
7D69 |
MS: Collision gas flow rate may be insufficient or unstable. |
Collision gas may not be supplied to the instrument or the piping for drying gas may not be done properly. |
1. Check the gas supply source and piping. \n2. If the problem persists, contact your Shimadzu representative. |
7D92 |
MS: Probe 1 or Source cover is not connected. |
----- |
Connect the Probe 1 or close the Source cover. |
7D93 |
MS: Probe 2 is not connected. |
----- |
Connect the Probe 2. |
7D94 |
MS: Turbo pump does not become ready.Turbo pump will be turned OFF. |
----- |
Repair or part replacement may be necessary. Contact your Shimadzu representative. |
7D95 |
MS: Automatic start of the vacuum system, Vaccum error.The vaccum system was |
The vacuum level did not reach the specified value in the specific period of time during automatic start. Automatic |
Check for vacuum leakage. Repair or part replacement may be necessary. Contact your Shimadzu representative. |
7D96 |
MS: Turbo pump will be turned OFF due to insufficient vacuum |
The turbo pump was not turned ON since the vacuum level of the MS was |
Check for vacuum leakage on the MS. |
7D97 |
MS: Turbo pump does not become ready. |
----- |
Repair or part replacement may be necessary. Contact your Shimadzu representative. |
7D98 |
MS: An error occurred in forline pressure.Turn off the power of the MS. |
The MS was turned OFF since the vacuum level of the foreline was insufficient. |
Check for vacuum leakage on the MS. |
7D99 |
MS: Overheat. (Q3RF) \n If the temperature does not go down in several minutes, turn OFF the power of the MS immediately. |
The Q3 RF capacitor temperature control heater or its control circuit may have a problem. |
1. For safety, check for any abnormal conditions (smoke, strange odor) around the electric board. \n 2. If there is any abnormal condition or the temperature does not go down in several minutes, stop the instrument and turn OFF the power of the main unit. \n 3. Contact your Shimadzu |
7D9A |
MS: Overheat. (Q1RF) \n If the temperature does not go down in several minutes, turn OFF the power of the MS immediately. |
The Q1 RF capacitor temperature control heater or its control circuit may have a problem. |
1. For safety, check for any abnormal conditions (smoke, strange odor) around the electric board. \n 2. If there is any abnormal condition or the temperature does not go down in several minutes, stop the instrument and turn OFF the power of the main unit. \n 3. Contact your Shimadzu |
7D9B |
MS: Overheat. (M1RF) \n If the temperature does not go down in several minutes, turn OFF the power of the MS immediately. |
The M1 RF capacitor temperature control heater or its control circuit may have a problem. |
1. For safety, check for any abnormal conditions (smoke, strange odor) around the electric board. \n 2. If there is any abnormal condition or the temperature does not go down in several minutes, stop the instrument and turn OFF the power of the main unit. \n 3. Contact your Shimadzu |
7D9C |
MS: An abnormal condition is detected on the heater power supply. |
The heater power supply or its control circuit may have a problem. |
1. For safety, stop the instrument and turn OFF the power of the main unit. \n2. Contact your Shimadzu representative. |
7D9D |
MS: An error occurred in the high-frequency power supply. |
----- |
----- |
7D9E |
MS: An error occurred in the high-frequency power supply. |
----- |
----- |
7D9F |
MS: An error occurred in the high-frequency power supply. |
----- |
----- |
7DE0 |
MS: The card gage door is open. |
----- |
Close the card gage door. |
7DE1 |
MS: Curtain gas flow rate may be insufficient or unstable. \n The LC pump was stopped and data acquisition was suspended. |
Data acquisition could not start due to insufficient curtain gas flow rate. The LC pump was stopped to prevent contamination of the probe. |
Check the curtain gas flow rate on the instrument monitor. If the gas flow rate does not increase, there may be a problem on gas piping or gas cylinder. Once sufficient curtain gas flow rate is achieved, data acquisition can start. |
7DE2 |
MS: Curtain gas flow rate may be insufficient or unstable. \n The LC pump was stopped and data acquisition was suspended. |
Data acquisition could not start due to insufficient curtain gas flow rate. The LC pump was stopped to prevent contamination of the probe. |
Check the curtain gas flow rate on the instrument monitor. If the gas flow rate does not increase, there may be a problem on gas piping or gas cylinder. Once sufficient curtain gas flow rate is achieved, data acquisition can start. |
7DE3 |
MS: Exhaust Gas Pressure may be insufficient. |
----- |
----- |
7DE4 |
MS: Nebulizer gas and heating gas flow rate may be insufficient or unstable |
Data acquisition could not start due to insufficient flow rate of the nebulizer gas and heating gas. |
Check the flow rate of the nebulizer gas and heating gas on the instrument monitor. If the gas flow rate does not increase, there may be a problem on gas piping or gas cylinder. Once sufficient flow rate of the nebulizer gas and heating gas is achieved, data acquisition can start. |
7DE5 |
MS: The Vaccum gage has been disconnected or has been broken.The vaccum system was stopped. |
The vacuum gage is disconnected or its control circuit may have a problem. |
Turn the power OFF and contact your Shimadzu representative. |
7DE6 |
MS: Vacuum level is insufficient. |
This operation is disabled because the vacuum level has not reached the sufficient value. |
Perform the operation after the vacuum system becomes ready on the [System Control] sub-window of the vacuum system. |
7DA0 |
MS: Provide maintenance service to the rotary pump. |
Maintenance service is necessary for the operation of the rotary pump. |
Maintenance service is necessary for the rotary pump. \n Contact your Shimadzu representative. |
7DA1 |
MS: Change the oil of the rotary pump. |
Operation time of the rotary pump has exceeded the limit for maintenance specified by configuration. |
Change the oil of the rotary pump. |
7DA2 |
MS: Turbo molecular pump should be overhauled. |
The message indicates that the operation time of turbo molecular pump has exceeded the limit for maintenance. Overhaul is recommended. |
Contact your Shimadzu representative. |
7DA3 |
MS: Operation time of the detector has exceeded the standard replacement time. |
Operation time of the detector has exceeded the limit for maintenance specified by configuration. |
Replace the MS detector with a new one. |
7DB0 |
MS: Measurement data could not be loaded in time. |
Measurement data could not be loaded on the PC in time and buffer in the instrument became full. A high-load application may be working on the PC. |
Quit the high-load application and perform the operation again. |
7DB1 |
MS: Because the error had occurred while executing the manual tuning, the manual tuning was discontinued. |
The error had occurred while executing the manual tuning. |
Refer to the content of the error help and address the error. Then performs the manual tuning again. |
7DB2 |
MS: Timeout occurred in the communication with the MS. |
There was no response from the MS within the specified period of time. |
Check the cable and power supply, turn OFF the power of all equipment including the PC, and restart them. |
7DB3 |
MS: An unknown interface is connected. |
An unknown interface is connected. |
Connect a supported interface. |
7DB4 |
MS: The MS is not connected. |
Connection failed because communication with the MS cannot be established. |
Check the cable and power supply. |
7DB5 |
MS: Operation is disabled because data acquisition or tuning is currently performed. |
----- |
----- |
7DB7 |
MS: Vacuum level is insufficient. |
This operation is disabled because the vacuum level has not reached the sufficient value. |
Perform the operation after the vacuum system becomes ready on the [System Control] sub-window of the vacuum system. |
7DB8 |
MS: Method or tuning condition is not set for the MS. |
Method or tuning information is not set for the MS correctly. |
Turn OFF the power of the PC and MS and restart them. |
7DBA |
MS: MS connection failed. |
----- |
Check the cable and power supply and then restart the MS and the PC. |
7DBB |
MS: The MS unit is busy. |
The MS unit does not receive signals. |
Check the cable and power supply, turn OFF the power of all equipment including the PC, and restart them. |
7DBC |
MS: Communication with MS was not achieved within the specified number of retry. |
Communication was attempted for the specific number of time and there was no response from the MS. |
Check the cable and power supply, turn OFF the power of all equipment including the PC, and restart them. |
7DBD |
MS: There is no applicable command. |
A command not existing in the LC time program or MS time program was |
Enter a command according to the dropdown list. |
7DBE |
MS: Received analysis data is abnormal. |
----- |
Check the cable and power supply, turn OFF the power of all equipment including the PC, and restart them. |
7DBF |
MS: Connection to the instrument cannot be achieved. |
The process to connect to the instrument was suspended. |
Restart the PC. |
7DC2 |
MS: The trouble occurs in the communication with the instrument. Please place a USB cable again after closing |
The trouble occurred during communication with the MS (mass spectrometer). |
Verify cable connections and check that the power is ON. Shut down the entire system, including the PC, then restart. |
7DC3 |
MS: The scanspeed is outside of the range. |
The Settling time may have been changed. |
Check the change of the Settling time at Log Browser. |
7DC4 |
MS: Because the error had occurred while executing the analyzing, the analyzing was discontinued. |
An error occurred during execution of analysis. |
Refer to the Help content for the error and solve the problem. Then execute analysis again. |
7DC5 |
MS: The connection was released. |
An error occurred that prevented the connection. |
Refer to the Help content for the error and solve the problem. Then connect the cable again. |
7DC6 |
MS: Serial number is not set. |
The serial number is not set in the system configuration settings. |
Set the serial number by clicking [Auto Configuration] in the [System Configuration] sub-window with the MS and PC connected. |
7DC7 |
MS: Serial number does not match. |
The serial number in the system configuration settings does not match the serial number of the instrument. |
Update the serial number by clicking [Auto Configuration] in the [System Configuration] sub-window with the MS and PC connected. |
7DC8 |
MS: The syringe that cannot be used for the tuning is set. |
The flow rate cannot be enough to run he tuning in the current syringe. |
Replace the current syringe with new syringe, then run the tuning again. |
7DC9 |
MS: Timeout occurred in the communication with the MS. |
There was no response from the MS within the specified period of time. |
Verify cable connections and check that the power is ON. Shut down the entire system, including the PC, then restart. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu |
7DCA |
MS: A hardware error occurred. The MS does not become ready. |
The MS does not become ready. |
Restart the software. If the phenomenon reoccurs, turn OFF the instrument and then restart it. If the phenomenon still reoccurs, contact your Shimadzu representative. |
7DD0 |
MS: Even though data acquisition started, the MS detector did not start. |
MS data acquisition did not start. |
Check if the system controller relay is appropriate on system configuration and make sure that the cable is correctly connected. |
7DD1 |
The tuning file has not been tuned. Perform auto-tuning or manual tuning using the interface specified in the method file. |
The tuning file has not been tuned. |
Perform auto-tuning or manual tuning using the interface specified in the method file. Analysis cannot start if there are MS analyses, such as Q1SIM or Q1, and MSMS analyses, such as MRM or Product Ion Scan, in the same method and [Use Collision Gas] is turned off. Turn off the [Use Collision Gas] then start analysis. If MS analysis without collision gas is required, either remove the MS/MS acquisition parameters, perform Manual Tuning in MS/MS |
7DD3 |
MS: Collision gas valve is closed. |
The preset method file is set to use Collision gas. However, currently the Collision gas valve is closed. |
Open the Collision gas valve, or change the method file to one that does not use Collision gas. |
7DD5 |
MS is not ready, but do you want to start anyway? |
----- |
----- |
7E00 |
MS: Start of the vacuum system started. |
----- |
----- |
7E01 |
MS: Stop of the vacuum system started. |
----- |
----- |
7E02 |
MS: Exhaust Gas valve was opened. |
----- |
----- |
7E03 |
MS: Exhaust Gas valve was closed. |
----- |
----- |
7E0D |
MS: High voltage/high frequency power supply was not able to be turned on because the vacuum of ion gauge is not good. |
The high-voltage/high-frequency power supply could not be turned ON since the MS vacuum level is insufficient. |
Check for vacuum leakage on the MS. |
7E0E |
MS: This operation cannot be performed immediately after startup or shutdown operation. Wait for a while and perform the |
The MS unit cannot keep starting the startup or shutdown operations. |
Wait for a while, then start startup or shutdown operation. |
7E12 |
MS: Shutdown of the vacuum system has not finished. Turn OFF the power of the MS.(Rotary pump is also stopped.) |
The MS unit failed to complete shutdown operation. |
Turn the MS unit off. Accordingly, the rotary pump will be stopped. |
7E30 |
MS: Survey Event which does not have Dependent Events. Do you wish to continue? |
Data is acquired even though there are Survey Events without Dependent Events. |
To acquire data by Survey Events without Dependent Events, click [OK], and to set Dependent Events, click [Cancel]. |
7E31 |
MS: Collision gas is not used. Do you wish to continue? |
Data is acquired without using Collision gas at MS/MS data acquisition at preset conditions. |
To acquire MS/MS data without using Collision gas, click [OK], and to acquire MS/MS data using Collision gas, click [Cancel] and change the setting. |
7E32 |
MS: Collision gas is used. Do you wish to continue? |
Data is acquired using Collision gas at data acquisition at preset conditions Even though there is no MS/MS Event. |
To acquire MS data using Collision gas, click [OK], and to acquire MS data without using Collision gas, click [Cancel] and change the setting. |
7E33 |
MS: Please set start m/z less than end m/z. |
By an event scanning, the start m/z value is set larger than the end m/z value. |
Set the start m/z value smaller than the end m/z value. |
7E34 |
MS: Analog is existed. Survey event and Dependent event can not be copied. |
An attempt was made to paste a Survey Event or a Dependent Event. |
The Survey Event or Dependent Event cannot be executed since analog output is set to this method. |
7E36 |
MS: This event can not be pasted. Please check parameters of this event. |
The parameters of this method differ from those of the method for the pasted event. |
Change the method parameters. |
7E37 |
MS: Acquisition Time is too short. |
Acquisition time is shorter than event time. |
Please set acquisition time longer than event time. |
7E38 |
MS: If this checkbox is unchecked, dependent events are deleted. |
An attempt was made to delete the Survey Event settings. For this reason, the Dependent Event created based on this event also will be deleted. |
To delete the Dependent Event, click [OK]. To cancel the delete, click [Cancel]. |
7E39 |
MS: If a survey event is deleted, dependent events are deleted, too. |
An attempt was made to delete the Survey Event. For this reason, the Dependent Event created based on this event also will be deleted. |
To delete the Dependent Event, click [OK]. To cancel the delete, click [Cancel]. |
7E3A |
MS: Event time is short. Pause time and dwell time of each channels are less than |
The recommended values for the pause time or dwell time could not be ensured since the event time is too short. |
Set a longer event time. |
7E3B |
MS: Scan Speed is less than 0u/sec. Please change end m/z more than start m/z or long sampling interval. |
The scan speed cannot be ensured since the event time is too short or the measurement end m/z value is the same as the measurement start m/z or less. |
Set a measurement end m/z value that is larger than the measurement start m/z value, or set a longer event time. |
7E3C |
MS: Scan speed is less than or equal to 15u/sec. Please change wide m/z range or short event |
The scan speed exceeds the lower limit value since the m/z range to measure is too narrow or the event time is too long. |
Set a wider m/z range to measure or set a shorter event time. |
7E3D |
MS: Scan speed is over 15000u/sec. Please change narrow m/z range or long event |
The scan speed exceeds the upper limit value since the m/z range to measure is too wide or the event time is too short. |
Set a narrower m/z range to measure or set a longer event time. |
7E3F |
MS: The value of DL bias exceeds tolerance range.\nPlease adjust the addition of DL bias and Qarray |
The DL bias setting range exceeds the tolerance range. |
Adjust so that the value obtained by adding the Qarray bias setting value to the DL bias setting value is in the range 240 V to 240 V. |
7E40 |
MS: The sum total of pause time and dwell time is over maximum of event time. Please adjust the pause time and dwell time. |
MS: The total of the pause time and dwell time exceeds the maximum value of the event time. |
Adjust the pause time and dwell time values. |
7E41 |
MS: Please set precursor m/z. |
The precursor m/z is not specified. |
----- |
7E42 |
MS: Please set product m/z. |
The product m/z is not specified. |
----- |
7E43 |
MS: Please set m/z of at least one channel. |
No m/z is specified. |
----- |
7E44 |
MS: Please set start time before end time. |
The start time is set after the end time. |
----- |
7E45 |
MS: Please set target ion and reference ion. |
The target ion or reference ion is not inputed. |
----- |
7E49 |
MS: Please set the interval of 0.03 min. or more between ESI mode and APCI mode. |
The interval of the command for changing the DUIS operation mode is not 0.03 min or more.\nA 0.03 min or more interval must be provided when using the command for changing the DUIS operation mode in the time program. |
Set the interval of the command for changing the DUIS operation mode to 0.03 min or more. |
7E4A |
MS: Please set the time of ESI mode and APCI mode 0.03 min. or more. |
The interval of the command for changing the DUIS operation mode is set to less than 0.03 min.\nA 0.03 min or more interval must be provided when using the command for changing the DUIS operation mode in the time program. |
Set the interval of the command for changing the DUIS operation mode to 0.03 min or more. |
7E4B |
MS: If analog out is set, measurement time of all events will be set same value. Do you wish to continue? |
When analog output is set, the measurement times of all events are automatically changed to be the same. |
To set analog output, click [OK].\nTo cancel analog output, click [Cancel]. |
7E4C |
MS: Measurement time of all events are set value of the selected event. Do you wish to |
The measurement time of all events is changed to the measurement time of the currently selected event. |
To change the measurement time of all events, click [OK].\nTo cancel the change, click [Cancel]. |
7E4D |
MS: Event time of all events are set value of the selected event. Do you wish to continue? |
The event time of all events is changed to the event time of the currently selected event. |
To change the event time of all events, click [OK].\nTo cancel the change, click [Cancel]. |
7E4E |
MS: Dwell time of all events are set value of the selected event. Do you wish to continue? |
The dwell time of all MRM Events and SIM Events will be changed. |
To change the dwell time of all MRM Events and SIM Events, click [OK].\nTo cancel the change, click [Cancel]. |
7E4F |
MS: Pause time of all events are set value of the selected event. Do you wish to continue? |
The pause time of all MRM Events and SIM Events will be changed. |
To change the pause time of all MRM Events and SIM Events, click [OK]. To cancel the change, click [Cancel]. |
7E50 |
MS: Please reduce the measurement time or increase the event time of each event. |
The amount of memory required for data acquisition exceeds the limit. |
Shorten the data acquisition time of each event, or set a longer event time. |
7E51 |
MS: Please change the start m/z. |
The value set to [Losses of] cannot be measured by the current measurement start m/z setting. |
Change the measurement start m/z setting. |
7E52 |
MS: The setting of conversion dynode in the attenuation is changed from the default setting (using a tuning file). |
The setting of conversion dynode in the attenuation is changed from the default setting.\nThe height and area of a peak is influenced by the change of conversion dynode value. |
It is an intended change, click [OK], and to change the setting of conversion dynode, click [Cancel]. |
7E53 |
MS: Event cannot be added any further. |
The number of event exceeds the upper limit. |
----- |
7E54 |
MS: Channel cannot be added any further. |
The number of channel exceeds the upper limit. |
----- |
7E55 |
MS: The event time of following events was not changed.Event# %s |
The same event time could not be set since the upper limit of the scan speed has been exceeded. |
Set the event time individually or confirm the event time so that it does not exceed the scan speed before selecting [Set same event time]. |
7E60 |
MS: Auto-Tuning was terminated. |
----- |
----- |
7E61 |
MS: Auto-Tuning started. |
----- |
----- |
7E62 |
MS: Auto-Tuning has been completed. |
----- |
----- |
7E63 |
MS: An error occurred in AutoTuning. |
----- |
----- |
7E64 |
MS: Nebulizing Gas is Off. |
----- |
----- |
7E65 |
MS: The MS does not become ready. |
There was no response from the MS within the specified period of time. |
Check the cable and power supply, turn OFF the power of all equipment including the PC, and restart them. |
7E66 |
MS: The gas or the Temperature is not stable. |
----- |
----- |
7E67 |
MS: The standard sample was not injected. |
----- |
----- |
7E68 |
MS: An error occurred in AutoTuning. \n (Peak intensity is too great.) |
----- |
Perform auto-tuning again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
7E69 |
MS: An error occurred during detector adjustment in AutoTuning. \n(The optimum value is |
----- |
Perform auto-tuning again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
7E6A |
MS: An error occurred during resolution adjustment in AutoTuning. \n(Resolution cannot be narrowed down.) |
----- |
Perform auto-tuning again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
7E6B |
MS: An error occurred during resolution adjustment in AutoTuning. \n(Resolution cannot be narrowed down.) |
----- |
Perform auto-tuning again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
7E6C |
MS: An error occurred during m/z calibration in Auto-Tuning. \n(m/z calibration point is not found.) |
----- |
Perform auto-tuning again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
7E6D |
MS: An error occurred during speed calibration in Auto-Tuning. \n(m/z calibration point is not found.) |
----- |
Perform auto-tuning again. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu representative. |
7E6E |
MS: Scan Speed is less than 0u/sec. Please change short settling time or long event time. |
----- |
----- |
7E6F |
MS: Scan speed is less than or equal to 10u/sec. Please change long settling time or short event time. |
----- |
----- |
7E70 |
MS: Scan speed is over 5000u/sec. Please change short settling time or long event time. |
----- |
----- |
7E80 |
MS: The trouble occurs in the communication with the MS. |
The trouble occurred during communication with the MS (mass spectrometer). |
Verify cable connections and check that the power is ON. Shut down the entire system, including the PC, then restart. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu |
7E81 |
MS: The specified module could not be started. |
The execution file to be started is not found. A probable cause is that the file path has been changed or deleted. |
Contact your Shimadzu representative. |
7E82 |
MS: The trouble occurs in the communication with the MS. |
The trouble occurred during communication with the MS (mass spectrometer). |
Verify cable connections and check that the power is ON. Shut down the entire system, including the PC, then restart. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu |
7E83 |
MS: The trouble occurs in the communication with the MS. |
The trouble occurred during communication with the MS (mass spectrometer). |
Verify cable connections and check that the power is ON. Shut down the entire system, including the PC, then restart. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu |
7E84 |
MS: The trouble occurs in the communication with the MS. |
The trouble occurred during communication with the MS (mass spectrometer). |
Verify cable connections and check that the power is ON. Shut down the entire system, including the PC, then restart. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu |
7E85 |
MS: The trouble occurs in the communication with the MS. |
The trouble occurred during communication with the MS (mass spectrometer). |
Verify cable connections and check that the power is ON. Shut down the entire system, including the PC, then restart. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu |
7E86 |
MS: Timeout occurred in the communication with the MS. |
There was no response from the MS within the specified period of time. |
Verify cable connections and check that the power is ON. Shut down the entire system, including the PC, then restart. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu |
7E87 |
MS: The trouble occurs in the communication with the MS. |
The trouble occurred during communication with the MS (mass spectrometer). |
Verify cable connections and check that the power is ON. Shut down the entire system, including the PC, then restart. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu |
7E88 |
MS: The MS unit is not ready. |
The MS unit is not ready. |
Verify cable connections and check that the power is ON. Shut down the entire system, including the PC, then restart. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu |
7E89 |
MS: Invalid parameters were |
----- |
----- |
7E8A |
MS: Invalid parameters were |
----- |
----- |
7E8B |
MS: The MS unit is not ready. |
The MS unit is not ready. |
Verify cable connections and check that the power is ON. Shut down the entire system, including the PC, then restart. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu |
7E8C |
MS: Invalid parameters were |
----- |
----- |
7E8D |
MS: Invalid parameters were |
----- |
----- |
7E8E |
MS: Invalid parameters were |
----- |
----- |
7E8F |
MS: Invalid parameters were |
----- |
----- |
7E90 |
MS: The analysis is running or the MS unit is not ready. |
----- |
----- |
7E91 |
MS: Firmware backup failed. |
----- |
----- |
7E92 |
MS: Acquisition can not be started with this acquisition |
----- |
----- |
7E93 |
MS: Acquisition can not be started with this acquisition |
----- |
----- |
7E94 |
MS: Acquisition can not be started with this acquisition |
----- |
----- |
7E95 |
MS: Acquisition can not be started with this acquisition |
----- |
----- |
7E96 |
MS: Acquisition can not be started with this acquisition |
----- |
----- |
7E97 |
MS: Acquisition can not be started with this acquisition |
----- |
----- |
7E98 |
MS: Acquisition can not be started with this acquisition |
The number of events and channels may have exceeded their upper limits. |
Add or delete an event on the [MS] tab page in the [Instrument Parameter View]. |
7E99 |
MS: Acquisition can not be started with this acquisition |
----- |
----- |
7E9A |
MS: Acquisition can not be started with this acquisition |
----- |
----- |
7EB0 |
MS: An hardware error occurred in foreline |
The instrument error occurred. |
Verify cable connections and check that the power is ON. Shut down the entire system, including the PC, then restart. If the error persists even after retry, detailed inspection is necessary. Contact your Shimadzu |
7EB1 |
MS: A hardware error occurred during communication with the |
----- |
----- |
7EB2 |
MS: A hardware error occurred during communication with the |
----- |
----- |
7EB3 |
MS: A hardware error occurred during communication with the |
----- |
----- |
7EC0 |
MS: MS maintenance setting is made. |
This is a confirmation message. MS maintenance setting was successful. |
----- |
7EC1 |
MS: Failed to make MS maintenance setting. |
MS maintenance setting was failed. |
Solve the problem relating to the sub-message, and set again. If this does not remedy the problem, contact your Shimadzu representative. |
F100 |
The last batch run was not completed successfully. |
The system might have crashed during the last data acquisition. |
----- |
F101 |
The batch queue is corrupted. |
Failed to load the batch queue. |
Contact your Shimadzu representative. |
F200 |
Started data processing. |
Message for notifying that data processing was started |
----- |
F201 |
Some chromatograms have not been saved. |
An attempt was made to acquire chromatograms beyond the limit. |
Set the chromatograms to acquire to the limit or below. |
F202 |
Some chromatograms have not been saved. |
An attempt was made to acquire chromatograms beyond the limit. |
Set the chromatograms to acquire to the limit or below. |
F203 |
Some spectra have not been saved. |
An attempt was made to acquire spectrum (spectra) beyond the limit. |
Set the spectrum (spectra) to acquire to the limit or below. |
F204 |
Failed to acquisition data. |
An error occurred during data acquisition. |
If an error occurs on the instrument, make sure that the instrument is configured correctly. |
F205 |
Failed to output report. |
An error occurred during report output. |
Make sure that acquisition printer is set correctly. |
F206 |
Failed to process data. |
An error occurred during data processing. |
----- |
F207 |
Failed to output the summary report. |
An error occurred during summary report output. |
----- |
F208 |
Failed to output summary report. |
An error occured during summary report output. |
----- |
F209 |
Cannot find the background file. |
Background compensation was not performed because the specified background file was not found. |
----- |
F301 |
An error occurred on the instrument. |
An error occurred on the instrument. |
If there are additional messages, remedy according to the content. |
F302 |
This function is not supported by the current instrument driver. |
An invalid function has been invoked. |
Contact your Shimadzu representative. |
F303 |
A warning has been issued from the instrument. |
A warning has been issued from the instrument. |
If there are additional messages, remedy according to the content. |
F304 |
Because start the analysis run from the instrument during connecting, stopped analysis. |
Because start the analysis run from the instrument during connecting to the W/S, stopped analysis. |
----- |
F305 |
The method has been changed by the instrument. |
The method has been changed by the instrument. |
----- |
F306 |
The injection process timed out. |
Acquisition was started, but the instrument could not get readied for injection within the time-out setting. |
----- |
F307 |
Information has been issued from the instrument. |
Information has been issued from the instrument. |
If there are additional messages, remedy according the content. |
6500 |
The backup parameters of the controller are corrupted. The method file and sequence file were initialized. |
The backup parameters of the controller are corrupted. The method file and sequence file were initialized. |
Set the method and sequence again. |
6501 |
The backup parameters of the controller are corrupted. The method file and sequence file were initialized. |
The method and sequence are returned to the same condition as when the instrument was last started because the backup parameters of the controller are corrupted. |
Check the method and sequence. |
6502 |
Illegal backup parameter on the instrument. System information has been changed. |
Instrument information (system name, group name) was initialized because an error was found in the backup parameters on the instrument. |
Set the instrument information (system name, group name) again on the instrument. |
6503 |
Product type of the instrument is incorrect. |
Product type of the instrument is incorrect. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
6504 |
Connection between the external detector and the instrument was disconnected. |
Communication is no longer possible with the external detector connected to the optical cable connector on the instrument. |
Check whether the external detector power is ON. \nIf the external detector power is ON, check that the optical cable is connected correctly. |
6505 |
Communication between the instrument and the workstation was disconnected. |
Communication between the instrument and the workstation was disconnected. |
Check the network connection status. If the error is not resolved, contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
6506 |
The buffer memory of the controller has overflowed. |
The buffer memory used for transmitting data from the instrument to the workstation is full. |
Check the network connection status. If the error is not resolved, contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
6507 |
The buffer memory of the controller has overflowed. |
The buffer memory used for transmitting data from the instrument to the workstation is full. |
Check the network connection status. If the error is not resolved, contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
6508 |
The instrument has started with the IP address "192.168.200.96" because it failed to automatically obtain an IP address. |
The instrument has started with the IP address "192.168.200.96" because it failed to automatically obtain an IP address. |
Contact the network administrator. Set the IP address settings (automatic acquisition/manual setting) to the correct settings. |
6509 |
The instrument restarted because IP address renewal/refresh failed. |
IP address renewal/refresh failed. |
Contact the network administrator. Set the IP address settings (automatic acquisition/manual setting) to the |
650B |
The maximum number of systems (10) that can be registered in the same group was |
More than 10 instruments are connected to the same group. |
Check the number of instruments connected to the instrument group where the error has occurred. |
650C |
A timeout occurred in communication between the pump and controller. |
An error occurred in communication between the replacement pump and controller, and the communication was |
Clear the error. If the error occurs again, contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
650D |
An error occurred in communication between the pump and controller. |
An error occurred in communication between the replacement pump and controller, but the error recovered. |
Clear the error. |
650E |
A timeout occurred in communication between the autosampler or column oven and controller. |
An error occurred in communication between the autosampler or column oven and controller, and the communication was disconnected. |
Clear the error. If the error occurs again, contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
650F |
An error occurred in communication between the autosampler or column oven and |
An error occurred in communication between the autosampler or column oven and controller, but the error recovered. |
Clear the error. If the error occurs again, contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
6510 |
An error signal was input from an external device connected to the external input (Alarm Input) on the back of the instrument. |
An error signal was received from the external device connected to the external input terminal. |
Check in accordance with the handling instructions of the external device connected to the external input terminal. |
6511 |
The external detector set on the instrument is not connected. |
Acquisition was started when a detector that differs from the detector set in the simple parameters of the acquisition main screen on the instrument was connected. |
Either connect the detector that is set in the simple parameters of the acquisition main screen on the instrument, or change the settings in the simple parameters of the acquisition main screen on the |
6514 |
The system check timed out. |
The system check timed out. |
Execute system check again. If the error is not resolved, contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
6515 |
Purge was executed because the pump pressure reached the minimum pressure. |
Purge was executed because the replacement pump pressure reached the [minimum pressure (P.Min)] in the instrument parameters. |
Inspect for liquid leaks in the flow paths, and if there are no problems, adjust the [minimum pressure (P.Min)] in the instrument parameters to a suitable value. |
6516 |
The sequence injection volume setting is outside the setting range. |
The injection volume setting value exceeds the [Max Injection Volume] setting in the system configuration. |
Change the injection volume setting value to a value below the [Max Injection Volume] setting in the system configuration. |
6517 |
A vial number that is different to the vial number set in the acquisition conditions may have been notified to the instrument. |
A vial number that is different to the vial number set in the acquisition conditions may have been notified to the instrument. |
Clear the error and run the acquisition again. |
6518 |
A rack number that is different to the rack number set in the acquisition conditions may have been notified to the instrument. |
A rack number that is different to the rack number set in the acquisition conditions may have been notified to the instrument. |
Clear the error and run the acquisition again. |
6519 |
The temperature of the column oven was not stable. |
The column oven did not stabilize at the set temperature during validation and calibration for the instrument. |
Check whether there are any problems with the column oven. |
651A |
There is no sample vial set in the autosampler rack. |
There is no sample vial set at the vial number specified in the acquisition conditions. |
Set a sample vial at the vial number specified in the acquisition conditions. \n[Hint]\nIf [Vial Skip] in the system settings on the instrument is [Enabled], acquisitions for which there is no sample vial set are not executed, and the |
651B |
The temperature of the sample cooler was not stable. |
The sample cooler did not stabilize at the set temperature during validation and calibration for the instrument. |
Check whether there are any problems with the sample cooler. |
651C |
The instrument power was turned OFF during analysis. |
The instrument power was turned OFF during acquisition. |
Check whether the power was turned OFF via the instrument console or the work station. If the power was not turned off, check the condition of the power supply. |
651D |
An error occurred in communication between the detector and controller. |
An error occurred in communication between the detector and controller. |
Clear the error. If the error occurs again, contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
651E |
There is no sample vial set for the next acquisition in the autosampler rack. |
When overlapping injection is run, the sample vial is not set in the vial number specified for the next acquisition. |
Clear the error. If the error occurs again, contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
6530 |
An sealing ability error occurred in the vacuum flow path of the degassing unit. |
An sealing ability error occurred in the vacuum flow path of the degassing unit. |
Check the vacuum pressure of the degassing unit. If the vacuum pressure is "Fail", turn the power of the degassing unit [OFF] and then [ON] again in the instrument parameters. If the vacuum pressure is "OK", the degassing unit has automatically recovered and there is no problem. |
6531 |
The remaining amount of Mobile Phase A-1 liquid is low. |
The remaining amount of Mobile Phase A1 liquid is low. |
Replace the Mobile Phase, and set the Mobile Phase amount again. |
6532 |
The remaining amount of Mobile Phase B-1 liquid is low. |
The remaining amount of Mobile Phase B1 liquid is low. |
Replace the Mobile Phase, and set the Mobile Phase amount again. |
6533 |
The remaining amount of Mobile Phase C-1 liquid is low. |
The remaining amount of Mobile Phase C1 liquid is low. |
Replace the Mobile Phase, and set the Mobile Phase amount again. |
6534 |
The remaining amount of Mobile Phase D-1 liquid is low. |
The remaining amount of Mobile Phase D1 liquid is low. |
Replace the Mobile Phase, and set the Mobile Phase amount again. |
6535 |
The remaining amount of Mobile Phase A-2 liquid is low. |
The remaining amount of Mobile Phase A2 liquid is low. |
Replace the Mobile Phase, and set the Mobile Phase amount again. |
6536 |
The remaining amount of Mobile Phase B-2 liquid is low. |
The remaining amount of Mobile Phase B2 liquid is low. |
Replace the Mobile Phase, and set the Mobile Phase amount again. |
6537 |
The remaining amount of Mobile Phase C-2 liquid is low. |
The remaining amount of Mobile Phase C2 liquid is low. |
Replace the Mobile Phase, and set the Mobile Phase amount again. |
6538 |
The remaining amount of Mobile Phase D-2 liquid is low. |
The remaining amount of Mobile Phase D2 liquid is low. |
Replace the Mobile Phase, and set the Mobile Phase amount again. |
6539 |
The degassing unit is not connected. |
The degassing unit is not connected. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
653A |
The leak sensor of the pump is malfunctioning. |
The leak sensor of the pump is malfunctioning. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
653B |
The leak sensor of the pump has detected an abnormal. |
The leak sensor of the pump has detected an abnormal. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
653C |
The backup parameters of the pump are corrupted. |
The backup parameters of the pump are corrupted. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
653D |
The consumable count data of the pump is corrupted. |
The consumable count data of the pump is corrupted. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
653E |
The pump pressure exceeded the maximum pressure. Check the flow paths for blockages, and check the maximum pressure |
The replacement pump pressure exceeded the maximum pressure. |
Inspect for blockages in the flow paths, and if there are no problems, adjust the [maximum pressure] in the instrument parameters to a suitable value or reduce the replacement flow rate. |
653F |
The pump pressure did not reach the minimum pressure. Check the flow paths for leaks, and check the minimum pressure |
The replacement pump pressure did not reach the minimum pressure. |
Inspect for liquid leaks in the flow paths, and if there are no problems, adjust the [minimum pressure] in the instrument parameters to a suitable value. This error is not activated for one minute after the start of pumping. |
6540 |
Could not detect the home position of the pump motor. |
Could not detect the home position of the replacement pump motor. |
Press the (power button) on the instrument to turn the power OFF, and then press the (power button) on the instrument again to turn the power ON. Next, press [Pump] in the acquisition main screen on the instrument. \nIf the error message is displayed again even when this operation is performed, press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
6541 |
The leak sensor of the pump detected a liquid leak. |
The leak sensor of the pump detected a liquid leak. |
Inspect the flow paths, etc. for liquid leaks, and wipe away any liquids from around the leak sensor. |
6542 |
The remaining amount of Mobile Phase A liquid is low. |
The remaining amount of Mobile Phase A liquid is low. |
Replace the Mobile Phase, and set the Mobile Phase amount again. |
6543 |
The remaining amount of Mobile Phase B liquid is low. |
The remaining amount of Mobile Phase B liquid is low. |
Replace the Mobile Phase, and set the Mobile Phase amount again. |
6544 |
The remaining amount of Mobile Phase C liquid is low. |
The remaining amount of Mobile Phase C liquid is low. |
Replace the Mobile Phase, and set the Mobile Phase amount again. |
6545 |
The pressure increased during purge of the pump. |
The remaining amount of Mobile Phase D liquid is low. |
Replace the Mobile Phase, and set the Mobile Phase amount again. |
6546 |
The pressure increased during purge of the replacement pump. |
The pressure increased during purge of the pump. |
If an error message is displayed during the auto purge operation, inspect the flow paths for blockages. |
6547 |
Pump is stopped because a rotation problem was detected in the cooling fan motor for the |
Pump is stopped because a rotation problem was detected in the cooling fan motor for the pump. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
6550 |
There is a mistake in the analysis parameters or settings. |
There is a mistake in the analysis parameters or settings. |
Check whether there are any mistakes or discrepancies in the acquisition parameters. If no mistakes or discrepancies are found, press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
6551 |
The leak sensor of the autosampler low pressure valve has detected a liquid leak. There is a liquid leak from the valve or measuring pump, or from an |
There is a liquid leak from the low pressure valve or metering pump, or from a unit positioned above the low pressure value (high pressure valve, column oven or detector, etc.). |
Inspect the liquid leak from the low pressure valve or metering pump, or in a unit positioned above the low pressure value (high pressure valve, column oven or detector, etc.), and wipe away any liquid from around the leak sensor. |
6552 |
The leak sensor of the autosampler high pressure valve has detected a liquid leak. There is a liquid leak from the high pressure valve, or in an upper- |
There is a liquid leak from the high pressure valve, or from a unit positioned above the high pressure value (detector etc.). |
Inspect the liquid leak from the high pressure valve, or from a unit positioned above the high pressure value (detector etc.), and wipe away any liquid from around the leak sensor. |
6553 |
Could not detect the rotation direction home position of the autosampler measuring pump. |
Could not detect the rotation direction home position of the autosampler metering pump. |
Press the (power switch) to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
6554 |
Heater is turned off because voltage of the autosampler rack L cooling unit is too high. |
Heater is turned off because voltage of the autosampler rack L cooling unit is too high. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
6555 |
Heater is turned off because voltage of the autosampler rack R cooling unit is too high. |
Heater is turned off because voltage of the autosampler rack R cooling unit is too high. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
6556 |
Heater is turned off because voltage of the autosampler dehumidifier unit is too high. |
Heater is turned off because voltage of the autosampler dehumidifier unit is too high. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
6557 |
Heater is turned off because the autosampler rack L heater has overheated and the thermal fuse has blown. |
Heater is turned off because the autosampler rack L heater has overheated and the thermal fuse has blown. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
6558 |
Heater is turned off because the autosampler rack R heater has overheated and the thermal fuse has blown. |
Heater is turned off because the autosampler rack R heater has overheated and the thermal fuse has blown. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
6559 |
The leak sensor in the autosampler low pressure valve has detected an abnormal. |
The leak sensor in the autosampler low pressure valve has detected an abnormal. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
655A |
The leak sensor in the autosampler low pressure valve is malfunctioning. |
The leak sensor in the autosampler low pressure valve is malfunctioning. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
655B |
The leak sensor in the autosampler high pressure valve has detected an abnormal. |
The leak sensor in the autosampler high pressure valve has detected an abnormal. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
655C |
The leak sensor in the autosampler high pressure valve is malfunctioning. |
The leak sensor in the autosampler high pressure valve is malfunctioning. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
655D |
Heater is turned off because the temperature sensor in autosampler rack L is |
Heater is turned off because the temperature sensor in autosampler rack L is malfunctioning. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
655E |
Heater is turned off because the temperature sensor in autosampler rack L has detected |
Heater is turned off because the temperature sensor in autosampler rack L has detected an abnormal. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
655F |
Heater is turned off because the temperature sensor in autosampler rack R is |
Heater is turned off because the temperature sensor in autosampler rack R is malfunctioning. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
6560 |
Heater is turned off because the temperature sensor in autosampler rack R has detected |
Heater is turned off because the temperature sensor in autosampler rack R has detected an abnormal. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
6561 |
Heater is turned off because the temperature sensor in the autosampler dehumidifier unit is malfunctioning. |
Heater is turned off because the temperature sensor in the autosampler dehumidifier unit is malfunctioning. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
6562 |
Heater is turned off because the sensor in the autosampler dehumidifier unit has detected an abnormal. |
Heater is turned off because the sensor in the autosampler dehumidifier unit has detected an abnormal. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
6563 |
The backup parameters for the autosampler and column oven are corrupted. |
The backup parameters for the autosampler and column oven are corrupted. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
6564 |
The consumable count data for the autosampler and column oven is corrupted. |
The consumable count data for the autosampler and column oven is corrupted. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
6565 |
The front panel of the autosampler may not be installed correctly. Alternatively, the sensor is malfunctioning. |
The front panel of the autosampler may not be installed correctly. Alternatively, the sensor is malfunctioning. |
First, check whether the front panel is attached correctly. \nIf the error is not resolved, press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
6566 |
Heater is turned off because a rotation problem was detected in the cooling fan of the |
A rotation problem was detected in the cooling fan of the autosampler. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
6567 |
A rotation problem was detected in the exhaust heat fan at the top of the rear of the instrument. |
A rotation problem was detected in the exhaust heat fan at the top of the rear of the instrument. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
6568 |
Could not detect the home position of the autosampler high pressure valve. |
Could not detect the home position of the autosampler high pressure valve. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
6569 |
Teaching has not been performed for the autosampler |
Teaching has not been performed for the autosampler injection port. |
Adjust the position of the needle in relation to the injection port. |
656A |
Could not detect the X-direction home position of the autosampler needle. |
Could not detect the X-direction home position of the autosampler needle. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
656B |
Could not detect the Y-direction home position of the autosampler needle. |
Could not detect the Y-direction home position of the autosampler needle. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
656C |
Could not detect the Z-direction home position of the autosampler needle. |
Could not detect the Z-direction home position of the autosampler needle. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
656D |
Failed to be detected by the forward and backward home position sensor of the autosampler measuring pump. |
Failed to be detected by the forward and backward home position sensor of the autosampler metering pump. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
656E |
There is no sample vial set in the autosampler rack. |
There is no sample vial set at the vial number specified in the acquisition conditions. |
Set a sample vial at the vial number specified in the acquisition conditions. \n[Hint]\nIf [Vial Skip] in the system settings on the instrument is [Enabled], acquisitions for which there is no sample vial set are not executed, and the next acquisition is executed. |
656F |
Could not detect the home position of the autosampler low pressure valve. |
Could not detect the home position of the autosampler low pressure valve. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
6570 |
System Error |
System Error |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
6571 |
The temperature sensor in the autosampler dehumidifier unit has detected an abnormal value. |
The temperature sensor in the autosampler dehumidifier unit has detected an abnormal value. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
6573 |
Heater is turned off because the temperature sensor in autosampler rack plate L has detected an abnormal. |
Heater is turned off because the temperature sensor in autosampler rack plate L has detected an abnormal. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
6574 |
Heater is turned off because the temperature sensor in autosampler rack plate R has detected an abnormal. |
Heater is turned off because the temperature sensor in autosampler rack plate R has detected an abnormal. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
6575 |
Analysis stopped because the position of the needle in relation to the autosampler rack has not been adjusted. |
Acquisition stopped because the position of the needle in relation to the rack has not been adjusted. |
Adjust the position of the needle in relation to the rack. |
6576 |
The position of the needle in relation to the rack set in autosampler rack plate L has not |
The position of the needle in relation to the left side rack has not been adjusted. |
Adjust the position of the needle in relation to the left side rack. |
6577 |
The position of the needle in relation to the rack set in autosampler rack plate R has not |
The position of the needle in relation to the right side rack has not been adjusted. |
Adjust the position of the needle in relation to the right side rack. |
6578 |
The rack ID of the rack set in the autosampler is not correct. |
The rack ID is not correct. |
Set the correct rack ID for the rack. |
6579 |
Racks that have the same ID are set in the autosampler. |
Racks that have the same ID are set in the autosampler. |
Change the rack ID of either the left or right side rack. |
657A |
Rack plate L cooler was switched off because there was no rack set in autosampler rack plate L for more than 30 minutes. |
The left side sample cooler temperature adjustment was stopped because the left side rack was not set for more than 30 minutes. |
Sample cooler temperature adjustment restarts when a left side rack is set. |
657B |
Rack plate R cooler was switched off because there was no rack set in autosampler rack plate R for more than 30 minutes. |
The left side sample cooler temperature adjustment was stopped because the right side rack was not set for more than 30 minutes. |
Sample cooler temperature adjustment restarts when a right side rack is set. |
657C |
The setting for the autosampler sample vial number is outside the setting range. |
The vial number that is outside of the setting range for the set plate is specified in the sequence. |
Check the acquisition conditions. \nAlso, check whether the plate is set correctly. |
657D |
The autosampler needle Xdirection motor has slipped. |
The autosampler needle X-direction motor has slipped. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
657E |
The autosampler needle Ydirection motor has slipped. |
The autosampler needle Y-direction motor has slipped. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
657F |
The remaining autosampler rinse solution is low. |
The remaining autosampler rinse solution is low. |
Replenish the autosampler rinse solution. |
6580 |
When the autosampler needle was lowered, it hit a foreign object causing excessive force to be exerted on it. |
When the autosampler needle was lowered, it hit a foreign object causing excessive force to be exerted on it. |
Press the (power button) on the instrument to turn the power OFF, and check whether there is shielding material preventing the lowering operation of the needle inside the instrument. If the error is not resolved, press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
6590 |
The detector parameters are corrupted. |
There is a fault in the non-volatile memory of the detector or a non-volatile memory write error has occurred. |
Perform calibration (wavelength calibration, absorbance correction, linearity correction) of the detector. If the error is not resolved, press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
6591 |
The leak sensor of the detector detected a liquid leak. |
A liquid leak has occurred in the detector. |
Inspect the detector for liquid leaks, and wipe away any liquids from around the leak sensor. |
6592 |
The buffer memory of the detector has overflowed. |
The buffer memory used for transmitting data from the instrument to the workstation is full. |
Check the communication environment and operating status of the instrument and the workstation. \nIf the error is not resolved, press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
6593 |
Could not re-send the data because the buffer memory of the detector overflowed during analysis. |
Recovery failed when resending transmission data from the instrument to the workstation |
Check the communication environment and operating status of the instrument and the workstation. \nIf the error is not resolved, press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
6594 |
Light intensity was exceeded because the detector took a long accumulation time. |
Light intensity was exceeded because the detector took a long accumulation time. |
Calibrate the accumulation time. |
6595 |
The leak sensor of the detector is malfunctioning. |
The leak sensor of the detector is malfunctioning. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
6596 |
Communication between the instrument and the workstation has timed out. |
Communication between the instrument and the workstation was interrupted. |
Check the communication environment and operating status of the instrument and the workstation. \nIf the error is not resolved, press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
6597 |
The detector cell temperature does not reach the set temperature. |
The detector cell temperature does not reach the set temperature. |
Replace the temperature control cell. \nIf the error is not resolved, press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
6598 |
The IP address of the instrument is duplicated. |
The IP address of the instrument is duplicated on a different device. |
Change the IP addresses so there are no duplicates. \nIf the error is not resolved, press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
6599 |
Heater of the cell is turned off because the detector cell temperature has exceeded the set value and is increasing. |
There is a fault in the control circuit of the temperature control cell. |
Replace the temperature control cell. |
659A |
The detector cell temperature does not lower to the set temperature. |
The temperature control cell temperature does not lower to the set temperature even after 5 minutes have elapsed. |
Set the [Cell Temperature] settings in the instrument parameters to a higher setting. \nChange the cell temperature to a setting higher than room temperature + |
659B |
The setting of the detector cell temperature is outside the permissible range. |
The [Cell Temperature] settings in the instrument parameters are set to a value outside the setting range. |
Set the [Cell Temperature] of the instrument parameters to a value within the setting range. |
659C |
The leak sensor of the detector is malfunctioning. |
The leak sensor of the detector is malfunctioning. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
659D |
The lamp is turned off because the temperature sensor of the detector lamp housing is |
The lamp is turned off because the temperature sensor of the detector lamp housing is malfunctioning. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
659E |
The D2 lamp cannot light because the lamp current of the detector is too large. |
There is a fault in the D2 lamp or a fault in the lighting circuit. |
Replace the D2 lamp. \nIf the error is not resolved, press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
659F |
The D2 lamp cannot light because the lamp current of the detector is too small. |
A lighting defect has occurred due to a broken filament in the D2 lamp or a fault in the constant current circuit, a lighting malfunction has occurred due to the lamp's life span, or there is a lighting defect due to a fault in the lamp itself. |
Replace the D2 lamp. \nIf the error is not resolved, press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
65A0 |
The W lamp cannot light because the lamp current of the detector is too small. |
This is a lighting defect caused by a broken filament in the W lamp, or due to a fault in the lighting circuit. |
Replace the W lamp. \nIf the error is not resolved by replacing the W lamp, contact your Shimadzu Sales Office/distributor or a service company designated by |
65A1 |
The detector lamp housing cover is open. Turn the lamp off. |
The lamp house cover is open or there is a fault in the cover sensor. |
Close the lamp house cover. \nIf the error is not resolved, replace the cover sensor. |
65A2 |
The D2 lamp cannot light because the heater current of the detector is too large. |
There is a fault in the D2 lamp or a fault in the lighting circuit. |
Replace the D2 lamp. \nIf the error is not resolved, press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
65A3 |
The lamp is turned off because the temperature of the detector lamp housing is high. |
The lamp is turned off because the temperature of the detector lamp housing is high. |
Follow the procedures below in the following order. \n(1) Check whether the fan on the rear of the instrument is obstructed. \n(2) Clean or replace the filters on the rear, left side and lamp front side (rear of the lamp house cover) of the instrument. \n If the error is not resolved, press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
65A4 |
The W lamp cannot light because the lamp current of the detector is too large. |
There is a fault in the W lamp lighting circuit. |
Replace the W lamp. \nIf the error is not resolved, press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
65A5 |
The D2 lamp cannot light because the heater current of the detector is too small. |
There is a fault in the D2 lamp or a fault in the lighting circuit. |
Replace the D2 lamp. \nIf the error is not resolved, press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
65A6 |
The detector D2 lamp is not connected. |
The D2 lamp connector is not connected. \nAlternatively, the D2 lamp cable is broken or there is a fault in the board. |
Follow the procedures below in the following order.\n(1) Check the connection with the D2 lamp.\n(2) Replace the D2 lamp. \n(3) Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
65A7 |
The detector W lamp is not connected. |
The W lamp connector is not connected. \nAlternatively, the W lamp cable is broken or there is a fault in the board. |
Follow the procedures below in the following order.\n(1) Connect the W lamp connector. \n(2) Replace the W lamp. \n(3) If the error is not resolved, press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
65A8 |
The W lamp cannot light because the lamp voltage of the detector is too large. |
There is a fault in the lighting circuit. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
65A9 |
Could not detect the filter home position in the detector. |
Could not detect the filter home position in the detector. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
65AA |
The temperature of the detector board is high. |
The temperature of the detector board is high. |
Follow the procedures below in the following order. \n(1) Check whether the fan on the rear of the instrument is obstructed. \n(2) Clean or replace the filters on the rear, left side and lamp front side (rear of the lamp house cover) of the instrument. \n(3) If the error is not resolved, press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
65AB |
The D2 lamp cannot light because the lamp voltage of the detector is too large. |
There is a fault in the lighting circuit. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
65AC |
The D2 lamp cannot light because the heater voltage of the detector is too large. |
There is a fault in the lighting circuit. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
65AD |
The lamp is turned off because a rotation problem was detected in the detector fan. |
The lamp is turned off because a rotation problem was detected in the detector fan. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
65AE |
Heater is turned off because the temperature of the spectrometer is high. |
Heater is turned off because the temperature of the spectrometer is high. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
65AF |
Heater is turned off because the temperature sensor of the spectrometer is malfunctioning. |
Heater is turned off because the temperature sensor of the spectrometer is malfunctioning. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
65B0 |
Failed to receive the detector signal. |
Failed to receive the detector signal. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
65B1 |
Could not detect the filter home position in the detector. |
Could not detect the filter home position in the detector. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
65B2 |
Could not detect the grating home position in the detector. |
Could not detect the grating home position in the detector. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
65B3 |
The spectrometer temperature does not reach the set temperature. Heater is turned off. |
The spectrometer temperature does not reach the set temperature. Heater is turned off. |
Contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
65B4 |
Failed to receive the detector signal. |
Failed to receive the detector signal. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
65B5 |
Failed to read or write the detector calibration data. |
Failed to read or write the detector calibration data. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
65B6 |
Results of wavelength accuracy check at instrument startup differ by more than 1nm. There may be bubbles in the flow cell. |
Results of wavelength accuracy check at instrument startup differ by more than 1nm. There may be bubbles in the flow cell. |
Perform replacement to remove the bubbles, and then check the wavelength accuracy using the performance check of the instrument. |
65B7 |
The lamp is turned off because the temperature of the detector board is high. |
The fan and the inspiratory port are closed or a fault in the lighting circuit. |
Follow the procedures below in the following order. \n(1) Check whether the fan on the rear of the instrument is obstructed. \n(2) Clean or replace the filters on the rear, left side and lamp front side (rear of the lamp house cover) of the instrument. \n If the error is not resolved, press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
65B8 |
The instrument is turned off because the ADC module on the board is malfunctioning. |
The instrument is turned off because the ADC module on the board is malfunctioning. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
65B9 |
Heater is turned off because of malfunctioning. |
The heater of the spectrmeter is short or there is a fault in the circuit. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
65C0 |
The column oven door is open when the column oven is at high temperature. Close the column oven door as soon as possible in mind, such as burns. |
The column oven door is open when the column oven is above 50C. |
Clear the error. The alarm sound and error message disappear. Be careful of burns, etc., when performing this work. |
65C1 |
Heater is turned off because liquid leak was detected inside the column oven. |
Heater is turned off because liquid leak was detected inside the column oven. |
Follow the procedures below in the following order. \n(1) If organic solvents have leaked inside the column oven, take the necessary steps to prevent leakage from the pipes.\n(2) Check whether there are any organic solvent gases near the instrument. If organic solvent gases are found, thoroughly.\n(3) If there are no problems in the above two points, calibrate the sensitivity of the gas |
65C2 |
An error was detected with the flow channel selection valve in the column oven. |
An error was detected with the flow channel selection valve in the column oven. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
65C3 |
The gas sensor of the column oven has detected an abnormal. |
The gas sensor of the column oven has detected an abnormal. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
65C4 |
Heater is turned off because the thermal fuse of the column oven has blown. |
Heater is turned off because the thermal fuse of the column oven has blown. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
65C5 |
Heater is turned off because the Peltier unit of the column oven has detected an abnormal. |
Heater is turned off because the Peltier unit of the column oven has detected an abnormal. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
65C6 |
Heater is turned off because the Peltier unit of the column oven has detected an abnormal. |
Heater is turned off because the Peltier unit of the column oven has detected an abnormal. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
65C7 |
Heater is turned off because a rotation problem was detected in the Peltier unit fan(outside) of the oven. |
Heater is turned off because a rotation problem was detected in the Peltier unit fan(outside) of the oven. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
65C8 |
Heater is turned off because a rotation problem was detected in the Peltier unit fan(inside) of the oven. |
Heater is turned off because a rotation problem was detected in the Peltier unit fan(inside) of the oven. |
If obstructions such as pipes, etc. are preventing the fan from rotating, press the (power button) on the instrument to turn the power OFF, and remove the obstruction. If the obstruction cannot be removed or if the error is not resolved, contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
65C9 |
Heater is turned off because the Peltier unit of the column oven has detected an abnormal. |
Heater is turned off because the Peltier unit of the column oven has detected an abnormal. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
65CA |
The room temperature sensor of the instrument has detected an abnormal. |
The room temperature sensor of the instrument has detected an abnormal. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
65CB |
Heater is turned off because the temperature of the column oven exceeded the maximum temperature. |
The temperature of the column oven exceeded the [Maximum Temperature] of the instrument parameters. |
Set the setting values so that the difference between the column oven temperature setting value in the simple parameters and [Maximum Temperature] in the instrument parameters is approximately 5C. \nIf the difference is too small, the column oven temperature may exceed the [Maximum Temperature] set in the method parameters before reaching the column oven temperature setting |
65CC |
Heater is turned off because the temperature sensor of the column oven is malfunctioning. |
Heater is turned off because the temperature sensor of the column oven is malfunctioning. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
65CD |
Heater is turned off because the temperature sensor of the column oven is malfunctioning. |
Heater is turned off because the temperature sensor of the column oven is malfunctioning. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
65CE |
The column oven temperature setting is too low compared to the room temperature. |
The column oven temperature setting is too low compared to the room temperature. |
Check the column oven temperature setting value in the instrument parameters, and if the difference between the setting value and room temperature is greater than 10C, change the settings so that the difference is less that 10C. |
65CF |
Failed to write to the column management device (CMD). |
CMD is malfunctioning. |
If the error is not resolved, press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
65D0 |
Heater is turned off because the heater of the column oven is malfunctioning. |
Heater is turned off because the heater of the column oven is malfunctioning. |
Press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by |
65D1 |
The upper part of the column oven door may be open. Alternatively, the column oven door sensor is malfunctioning. |
The upper part of the column oven door may be open. Alternatively, the column oven door sensor is malfunctioning. |
First, check that the column oven door is closed correctly. \nIf the error is not resolved, press the (power button) on the instrument to turn the power OFF and contact your Shimadzu Sales Office/distributor or a service company designated by Shimadzu. |
7F01 |
UV instrument error occurred. |
An error was notified from the UV instrument. |
This error is accompanied by a message sent from the GC instrument. Check the content of this error message and remedy the error according to the Instruction Manual provided with the GC instrument. |
FFFF |
No Help is available. Refer to the displayed message. |
----- |
----- |