Recent activity by Rob Stimpson-
AUX3-AUX5 relay error [Error Code:1047]
Relay or heater control circuit has a problem. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.
-
INJ1-DET2 relay error [Error Code:1046]
Relay or heater control circuit has a problem. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.
-
COL relay error [Error Code:1045]
Relay or heater control circuit has a problem. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.
-
Power controller board com. failed [Error Code:4045]
POWER PCB may have failed. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.
-
DET#1 2 3 4 FPD cooling fan error [Error Code:4121 4122 4123 4124]
FPD cooling fan has a problem. Turn off the main power of the instrument and check the installation status of the cooling fan.If it is installed correctly, the hardware fails. Turn off the main...
-
SCD detector abs. press is high [Error Code:9079]
The SCD detector absolute pressure sensor has failed. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.
-
xxx sensor error [Error Code:1026 1027 1028 1029 1030 1031 1032 1033 1056]
Temperature sensor fails. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.
-
xxx sensor is down [Error Code:1018 1019 1020 1021 1022 1023 1024 1025 1055]
Temperature sensor fails. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.
-
xxx sensor is short circuit [Error Code:1010 1011 1012 1013 1014 1015 1016 1017 1054]
Temperature sensor fails. Turn off the main power of the instrument, stop gas supply, and contact your Shimadzu sales/service representative.
-
SCD furnace abs. press was exceeded upper limit [Error Code:6112]
The pressure of the furnace is high. The inner pyrotube may be clogged. Replace the inner pyrotube. If the error is displayed even after the tube is replaced, turn off the main power of the inst...