In the event of an error, LabSolutions may report an error code. Please quote this error code when reporting the issue to Shimadzu. A description of each error code is listed below:
Additional LabSolutions Error Codes (3006 - FFFF)
Code Message Reason Action
0100 |
Unexpected error occurred. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
0101 |
Failed. |
Failed to finish the processing. |
If any additional message is shown, take necessary actions according to it. |
0102 |
Invalid argument - Debug. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
0103 |
Insufficient memory. |
There is insufficient memory for the system to perform processing normally. Too many operations occurring in Postrun can require more memory than is |
Close all unnecessary applications. If this does not solve the problem, reboot the PC; then restart the software. |
0104 |
Failed to initialize OLE. OLE system may not be installed - |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
0105 |
Object does not support the interface. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
0106 |
Object does not support aggrigation. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
0107 |
The CLSID is invalid against the ProgID. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
0108 |
OLE system is already initialized. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
0109 |
CLSID is not registered correctly. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
010A |
Failed to write the information to the registry. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
010B |
Failed to read the information from the registry. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
010C |
The version for OLE library does not match. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
010D |
The file already exists. |
The file name specified already exists on the drive. |
Specify a different file; then execute the command again. |
010E |
File is not found. |
The specified file could not be found. |
Use Data Explorer or Windows Explorer to verify the existence and location of the file. Execute the command again, specifying the correct file and location. |
010F |
Failed to revert storages transaction. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
0110 |
Invalid function. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
0111 |
Insufficient disk space. |
There is insufficient space on the disk to save the file. |
Delete unnecessary files or back up files to free up space on the disk. |
0112 |
Failed to perform ActiveX function because of illegal call. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
0113 |
Failed to perform ActiveX function because of short of memory. |
The program failed to finish the processing properly because of insufficient memory available for it. |
If this error recurs, reboot the computer and try using the program again. |
0114 |
Failed to perform ActiveX function because of invalid property array index. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
0115 |
The specified module is not found. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
0300 |
Failed to open the method file. |
The Method File may be in use in another LabSolutions window. If this is not the case, the file may be corrupted. |
1. Try executing the command again after closing the other window.\n2. If the error occurs again, the method file is probably corrupted. Close all applications, and execute error checking at Windows disk properties. Then create a new Method file. |
0301 |
Failed to open the data file. |
The data file may be in use in another LabSolutions window. If this is not the case, the file may be corrupted. |
1. Try executing the command again after closing the other window.\n2. If the error occurs again, the data file is probably corrupted. Close all applications, and execute error checking at Windows disk properties. |
0302 |
Failed to open the batch file. |
The Batch file may be in use in another LabSolutions window. If this is not the case, the file may be corrupted. |
1. Try executing the command again after closing the other window.\n2. If the error occurs again, the method file is probably corrupted. Close all applications, and execute error checking at Windows disk properties. Then create a new Batch file. |
0303 |
Failed to open the tuning file. |
The Tuning File may be in use in another LabSolutions window. If this is not the case, the file may be corrupted. |
1. Try executing the command again after closing the other window.\n2. If the error occurs again, the tuning file is probably corrupted. Close all applications, and execute error checking at Windows disk properties. Then create a new Tuning file. |
0304 |
Failed to open the UV library file. |
The Library File may be in use in another LabSolutions window. If this is not the case, the file may be corrupted. |
1. Try executing the command again after closing the other window.\n2. If the error occurs again, the method file is probably corrupted. Close all applications, and execute error checking at Windows disk properties. Then create a new Library file. |
0305 |
Failed to open the report format file. |
The Report Format File may be in use in another LabSolutions window. If this is not the case, the file may be corrupted. |
1. Try executing the command again after closing the other window.\n2. If the error occurs again, the method file is probably corrupted. Close all applications, and execute error checking at Windows disk properties. Then create a new Report Format file. |
0306 |
Failed to open the configuration file. |
The Configuration File may be in use in another LabSolutions window. If this is not the case, the file may be corrupted. |
If the error occurs again, the configuration file is probably corrupted. Close all applications, and execute error checking at Windows disk properties. Then close Windows, and reboot. If the error still occurs, the application is not running properly. Try the solution described below.\nClose all system applications._Using Explorer, open the following directory, which is located on the drive where the software is installed. (Assume that this is the C: drive.) C:\LabSolutions\System. In this folder, rename the System Configuration File "LCSys1.cfg" to "LCSys1.bak" (if a file with this name already exists, use a different name).\nFinally, restart the Analysis Application. The System Configuration File will be created with default values.\nNote\nThis method deletes all of the System Configuration settings. Once normal operation is restored, all of the System Configuration settings must be entered again.\nNote\nFor more details on this procedure, refer to |
0307 |
Failed to save the method file. |
Either the Method file is corrupted, or there is insufficient memory. |
1. Try saving another Method file. If this is successful, then the first file is corrupted. Close all applications, and execute error checking at Windows disk properties. Then create a new Method file.\n2. If another Method file cannot be saved, there may be insufficient memory. Close all unnecessary applications. |
0308 |
Failed to save the data file. |
Either the data file is corrupted, or there is insufficient memory. |
1. Try saving another data file. If this is successful, then the first file is corrupted. Close all applications, and execute error checking at Windows disk properties.\n2. If another data file cannot be saved, there may be insufficient memory. Close all unnecessary applications. |
0309 |
Failed to save the batch file. |
Either the Batch file is corrupted, or there is insufficient memory. |
1. Try saving another Batch file. If this is successful, then the first file is corrupted. Close all applications, and execute error checking at Windows disk properties.\n2. If another Batch file cannot be saved, there may be insufficient memory. Close all unnecessary applications. |
030A |
Failed to save the tuning file. |
The tuning file may be damaged or memory is insufficient. |
1. If other tuning files can be saved, the file is damaged. Close all applications and check for a drive error. Then, create a new tuning file. \n2. If other tuning files cannot be saved either, memory may be insufficient. Close all unnecessary applications. Disk free space will be reserved as an additional memory. |
030B |
Failed to save the UV library file. |
Either the Library File is corrupted, or there is insufficient memory. |
1. Try saving another Library File. If this is successful, then the first file is corrupted. Close all applications, and execute error checking at Windows disk properties.\n2. If another Library File cannot be saved, there may be insufficient memory. Close all unnecessary applications. |
030C |
Failed to save the report format file. |
Either the Report Format File is corrupted, or there is insufficient memory. |
1. Try saving another Report Format File. If this is successful, then the first file is corrupted. Close all applications, and execute error checking at Windows disk properties. Then create a new Report Format File.\n2. If another Report Format File cannot be saved, there may be insufficient memory. Close all unnecessary applications. |
030D |
Failed to save the configuration file. |
Either the configuration file is corrupted, or there is insufficient memory. |
1. The configuration file may be corrupted. Close all applications, then execute error checking at Windows disk properties. Set up a new configuration, and save it to overwrite the existing file. 2. There may be insufficient memory. Close all unnecessary applications. |
030E |
Save current method file? |
Changes have been made to the Method file. Confirm whether to save the changes or close the file without saving the |
----- |
030F |
Save current data file? |
Changes have been made to the data file. Confirm whether to save the changes or close the file without saving the changes. |
----- |
0310 |
Save current batch file? |
Changes have been made to the Batch file. Confirm whether to save the changes or close the file without saving the |
----- |
0311 |
Save current tuning file? |
Changes have been made to the Tuning File. Confirm whether to save the changes or close the file without saving the |
----- |
0312 |
Save current library file? |
Changes have been made to the Library File. Confirm whether to save the changes or close the file without saving the |
----- |
0313 |
Method file is not found. |
The specified Method file could not be found. |
1. Verify the path (drive and folder) in the case of a file management configuration.\n2. Verify the Method file name.\nVerify the Method file name |
0314 |
Data file is not found. |
The specified data file could not be found. |
1. Verify the path (drive and folder) in the case of a file management configuration.\n2. Verify the data file name. |
0315 |
Batch file is not found. |
The specified Batch file could not be found. |
1. Verify the path (drive and folder)in the case of a file management configuration.\n2. Verify the Batch file name. |
0316 |
Tuning file is not found. |
The specified Tuning File could not be found. |
1. Verify the path (drive and folder).\n2. Verify the Tuning File name. |
0317 |
UV library file is not found. |
The specified Library File could not be found. |
1. Verify the path (drive and folder)in the case of a file management configuration.\n2. Verify the Library File |
0318 |
Report format file is not found. |
The specified report format file could not be found. |
1. Verify the path (drive and folder) in the case of a file management configuration.\n2. Verify the report format file name. |
0319 |
Configuration file is not found. |
The System Configuration File could not be found. |
Reinstall the software. Specify the Configuration Settings again. |
031A |
Failed to open the file. |
The specified file could not be found. |
1. Verify the path (drive and folder) in the case of a file management configuration.\n2. Verify the file name. |
031B |
Failed to save the file. |
Either the file is corrupted, or there is insufficient memory. |
1. Try saving another file. If this is successful, then the first file is corrupted. Close all applications, and execute error checking at Windows disk properties.\n2. If the other file cannot be saved, there may be insufficient memory. Close all unnecessary applications. |
031C |
Failed to get the file. |
The specified file could not be get. |
Use Data Explorer or Windows Explorer to verify the existence and location of the file in the case of a file management configuration. Execute the command again, specifying the correct file and location.Use Data Explorer to verify the existence and location of the file in the network configuration. The file used for the acquisition cannot be used by the analysis. |
031D |
Save current data file? |
Changes have been made to the data file. Confirm whether to save the changes or close the file without saving the changes. |
----- |
031E |
Successfully converted file from CLASS-5000 File. |
As your selected file was in the CLASS5000 format, it has been converted. |
----- |
031F |
Failed to convert file from CLASS-5000 file. |
The selected file was in CLASS-5000 format, but could not be converted to the new file format. |
Restart the PC and try again. If the attempt fails again, the file may be corrupted. You can verify whether the file can be opened in CLASS-5000. |
0320 |
This file has been converted from other file type. You must save it with a different file_name. |
The"Save"command cannot be used for a file that has just been converted from another format. |
Save the file with a new name using the"Save As"command. |
0321 |
This file is read-only. You must save it with a different file_name. |
This file has a read-only attribute. No changes can be saved. |
Either save the file using a different file name, or cancel the Save operation. |
0322 |
A file_name cannot contain any of the following characters: \n\\ / : * ? "" < > | |
The file name uses invalid characters. |
The characters \ / : * ? <>|"cannot be used in Windows file names. Do not use these characters in the file name. |
0323 |
Save current report format file? |
Changes have been made to the report format file. Confirm whether to save the changes or close the file without saving the changes. |
----- |
0326 |
Save report format in current data file? |
The report format in the data file has been changed. |
Specify whether or not to save the report format. |
0327 |
The file "PsiPort Instrument Method" is not found. |
The specified PsiPort Instrument Method file could not be found. |
1. Verify the path (drive and folder).\n2. Verify the PsiPort Instrument Method file name. |
0328 |
Please change file name length less or equal to 80, because selected file name is too long. |
Selected file name is too long. |
Please change file name length (with extension) less or equal to 80. |
0330 |
The specified drive does not |
The wrong drive was specified. |
Enter the correct drive name. |
0331 |
The specified folder does not |
The wrong directory (folder) was specified. |
Enter the correct folder name. |
0332 |
Specified directory does not exist. Create directory? |
The specified directory does not exist. |
To create the directory, click [OK]. To cancel the creation, click [Cancel]. Next, enter the correct folder, and execute |
0333 |
Failed to create the folder. |
The directory could not be created. |
1. Execute the command again.\n2. If the error re-occurs, create the folder using Explorer. |
0334 |
Insufficient disk space. |
There is insufficient space on the drive specified, so the file cannot be saved. |
1. If the specified drive is on the hard drive, empty the Recycle Bin.\n2. Back up unnecessary files, and delete them from the drive. |
0335 |
Failed to read file. |
Either the desired information does not exist in the file, or the file is corrupted. |
Execute the command again. If the error re-occurs, the file may be corrupted. Close all applications; then execute error checking at Windows disk properties. |
0336 |
Invalid file type. |
The wrong type of file was specified. |
Check the file type; then select the correct type of file. |
0337 |
Another application is using this file. |
The file is currently being used by another window. |
Close the window that is currently using the file; then open the file again. |
0338 |
Another application is using this file. Open as read only. |
This file is currently being used by another program, but it can be opened as a readonly file. |
To open the file as read-only, click on OK.\nNote\nWhen the file is opened with a read-only attribute, any changes to the file cannot be saved with the "Save" command. To save the changes, save the file with a different name, using the "Save_As" command. To open the file, change it, and save it with the same name, click on the Cancel button, close the application which is using the file or wait until the current processing is completed, and then open the file again. Alternatively, if the file attribute is read-only, the read-only attribute could be removed for the file; then execute the command to open the file. |
0339 |
This file is write-protected. |
Either the file is being used by another application, or the file has a read-only attribute to prevent overwriting the file. |
If the file is being used by another application, close that application; then re-open the file. If the file attribute is readonly, the read-only attribute could be removed for the file; then execute the command to open the file. |
033A |
This file is write-protected. Open as read only. |
Either the file is being used by another application, or the file has the read-only attribute. Confirm whether to open the file as read-only. |
To open the file as read-only, click OK.\nNote\nWhen the file is opened with the read-only attribute, any changes to the file cannot be saved by the "Save" command. To save the changes, save the file with a different name, using the "Save As" command. To open the file, change it, and save it with the same name, click on the Cancel button, close the application which is using the file or wait until the current processing is completed, and then open the file again. Alternatively, if the file attribute is Read-only, the read-only attribute could be removed for the file; then |
033B |
File with same name already exists. |
Another file with the same name as the specified file name exists on that drive. |
1. Specify a different file name.\n2. Specify a different drive. |
033C |
The same file exists. Are you sure you want to overwrite it? |
Another file having the same name as the specified filename exists on that drive. This message is to verify whether to save the new file by overwriting the existing file. |
Click Yes to save the file, overwriting the existing file. The information in the existing file is lost. To save the file with a different file name, click No, and then save the file with the "Save As" command. |
033D |
Cannot write to file |
Either there is insufficient disk space, or the file is corrupted. |
Increase the amount of available space on the disk, and execute the command again. To increase the amount of available space:_1. If the specified drive is on the hard drive, empty the Recycle Bin.\n2. Back up unnecessary files, and delete them from the drive. |
033E |
Failed to read the file. |
The file is corrupted. |
Execute the command again. If the same error occurs, close all applications; then execute error checking at Windows disk properties. |
033F |
Failed to open the file because too many files are open. Please close unused files. |
The maximum number of open files is determined by the system. |
There may be insufficient memory.Close unnecessary applications. |
0340 |
Value is less than minimum limit. |
A value below the range minimum was entered. |
Check the allowable setting range; then enter an appropriate value. To obtain the range, refer to |
0341 |
Value is greater than maximum limit. |
A value above the range maximum was entered. |
Check the allowable setting range; then enter an appropriate value. To obtain the range, refer to |
0342 |
Value is not correct. |
An unsuitable value was entered. |
Check the allowable setting range or maximum number of characters; then enter an appropriate value. For information, refer to LabSolutions Help. |
0343 |
Cannot change this information. Your changes are invalidated. |
An attempt was made to edit and save a parameter, but the user does not have the rights to edit that parameter. |
Check the user rights in the User Administration window. |
0344 |
Failed to create window. |
The disk is corrupted, or there is insufficient memory. |
1. The disk may be corrupted. Close all applications; then execute error checking at Windows disk properties. Then execute the command again. 2. There may be insufficient memory. Close all unnecessary applications. 3. Reinstall the software. |
0345 |
Failed to run_the application. |
The application could not be started due to a system problem. |
Try again. If the error re-occurs, close all applications, and restart Windows; then try to execute the command again. If the error still occurs, reinstall the software. |
0346 |
No chromatogram is loaded. |
Because a chromatogram is not open, the specified processing could not proceed. |
Open a data file, display the chromatogram to be processed, and select it. |
0347 |
No spectrum is loaded |
Because a mass spectrum is not open, the specified processing could not proceed. |
Open a data file and specify the mass spectrum. In Postrun analysis, the mass spectrum is displayed by double-clicking on the chromatogram. |
0348 |
Failed to access log file. |
The log file could not be accessed. |
If Windows NT is being used, verify whether the user can access the Log file directory. |
0349 |
Failed to write operation log. |
Either there is insufficient space on the specified drive to write to the file, or the user does not have the rights to access the Log file folder. |
If Windows NT is being used, verify whether the user has access to the Log file directory. To increase the amount of available space on the drive:_1. If the specified drive is on the hard drive, empty the Recycle Bin.\n2. Back up unnecessary files, and delete them from the drive. |
034A |
Backup for log files has been created. |
Every month, the first time the LabSolutions software is opened, the contents of the log files are compressed and backed up. The log file itself is |
The compressed log file is stored in a file named "LG(year)(month).lzh" in the Log folder of the software installation directory. |
034B |
Log file was compressed. |
A backup of the log file will be made and compressed when this software is started up at the beginning of the month. |
A backup of the log file will be made and compressed when this software is started up at the beginning of the month. The compressed log file is saved in the Log folder under the installation folder in the format |
034C |
Failed to create files on disk. |
Either there is insufficient memory, there is insufficient disk space, or the disk is writeprotected. |
1. If there is insufficient memory, close unnecessary applications.\n2. Verify whether there is insufficient space on the hard drive. To increase the amount of available space on the hard drive:\n1) If the specified drive is on the hard drive, empty the Recycle Bin.\n2) Back up unnecessary files, and delete them from the drive. Move |
034D |
Failed to load Dynamic Linked Library. |
The Dynamic Linked Library (DLL) does not exist, or it may be corrupted. |
1. The file may be corrupted. Close all applications, and then execute error checking at Windows disk properties.\n2. Reinstall the software. |
034F |
Terminated by User. |
Processing was stopped due to user |
----- |
0350 |
Failed to copy the file. |
Either there is insufficient space on the copy destination drive, or an application is using a file with the same name. |
If there is insufficient disk space, increase the available disk space.\nClose the application that is using the file with the destination file name; then copy the file. |
0351 |
Failed to delete the file. |
Either the file to delete does not exist, or it is being used by an application. |
Verify the existence of the file to be deleted. Close the application that is using the file; then delete the file. |
0353 |
Failed to create new project (folder). |
A project (folder) of the same name may already exist. |
Try creating the new folder using a different folder name. |
0354 |
This project(folder) already exists. |
A project (folder) of the same name already exists. |
Try creating the new folder using a different folder name. |
0355 |
Are you sure you want to delete this project (folder)? |
This is a prompt to confirm the deletion of the project (folder). |
Click the [OK] button to delete the project (folder) or the [Cancel] button to cancel. |
0356 |
Files exist in the project (folder). Must delete files before deleting project (folder). |
Existing files of a project must be deleted before the project can be deleted. |
Delete all the files in the project and then delete the folder. |
0357 |
Failed to remove the project(folder). |
The project has not been deleted. |
Reboot the computer, and then delete the project. If the project cannot be deleted, try using Windows Explorer to delete it. If deletion cannot be done using Windows Explorer, the current user may not have the authority to delete this project folder. Contact your network administrator to assign the necessary permissions. |
0358 |
File name is too long. |
Entered name is too long. |
Set a shorter file name. |
0359 |
Configuration settings are not correct. |
Instrument configuration settings do not match with the actual installation. |
Check the instrument configuration and optional settings. |
0360 |
Failed to display Help. |
The Help file could not be found. |
Verify that the LabSolutionsHelp.chm and LabSolutionsMessageHelp.chm files exist in the LabSolutions PROGRAM folder. If they cannot be found, reinstall the software. |
0361 |
The instrument is not connected, or is already in use. |
The analytical instrument is not connected, or is already in use. |
Make sure that the analytical instrument is turned ON and is running normally, and that it is not being used by others. Also, check the connection cables. |
0362 |
Once you apply audit trail to method file, you cannot cancel the function.\n Are you sure you want to apply? |
This warning states that the audit trail function cannot be canceled once it has been specified for a Method. |
Select "Yes" to specify the Audit Trail function for the Method, or "No" to cancel. |
0363 |
File saving was canceled by user. |
The file save operation was canceled in progress. |
----- |
0367 |
The application is running on Snap Shot mode right now. \n[File SaveAs] operation will terminate Snap Shot mode. \nDo you really want to save file even if you cancel the Snap Shot mode? |
You are trying to save a data file opened from Snap Shot as a new data file. If you continue to [Save As], Snap Shot mode will terminate and new [Save As] data file will be re-opened. |
Click [Cancel] to continue Snap Shot mode. Click [OK] to terminate Snap Shot mode and to save as new data file for analyzing new data file. It should be mentioned that a file name opened in Snap Shot mode could not be used for a [Save As] file name. |
0368 |
You cannot overwrite a file opened in Snapshot mode. |
A data file being viewed in Snap Shot mode cannot be saved with the same name. Click OK to cancel the save |
To save a Snap Shot data file, specify a different file name. |
036B |
Once you apply audit trail to configuration file, you cannot cancel the function. \nAre you sure you want to apply? |
This message means that once you apply audit trail to config file you cannot cancel the function. |
Click [Yes] to apply audit trail. Click [No] to cancel this function. |
036C |
Are you sure you want to delete the template file? |
You tried to delete a template file. |
Click [Yes] to delete template file. Click [No] to cancel. |
036D |
You cannot overwrite the existing file. |
Rewriting the existing file is prohibited by Policy setting. |
Save as other file name. |
036E |
Current method file will be saved. |
This message is displayed when the method file is modified at selecting "Plot" in Data Acquisition. |
Click [OK] to save the method file and to start plotting chromatogram. Click [Cancel] to cancel the save and plotting. |
036F |
System error occurred. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
0370 |
Please install Adobe Reader. |
Adobe Reader must be installed to read online manual. |
Install Adobe Reader along the procedure written in Operation Manual. |
0371 |
Succeeded to convert the specified files to Acrobat |
This message means specified file can be successfully converted to Acrobat |
----- |
0372 |
Failed to send e-mail. |
Fail to send mail by uncertain cause. |
Confirm send mail settings and network setting to send |
0373 |
E-mail settings are not correct. |
Send mail setting is invalid. |
Set right send mail setting. |
0374 |
Failed to load e-mail setting parameters. |
This error can not be assigned. |
Re-try to set send mail setting. If this problem cannot be solved by this procedure, re-start PC and try again to set |
0375 |
Disable to send e-mail function, on currently setting. |
Send mail setting is invalid. |
Set right send mail setting. |
0376 |
Failed to download to the instrument. |
Fail to download the instrument parameter to analytical instrument. |
Confirm system configuration setting followed by re-start the LabSolutions. In addition, confirm the instrument condition and connection between both instrument and PC. |
0377 |
Failed to load XML purser. |
XML conversion program is not installed correctly. |
Re-install the LabSolutions. |
0378 |
Failed to create the XML file. |
Fail to convert to XML file format. |
Restart the PC and try again. If the attempt fails again, the file may be corrupted. You can verify whether the file can be opened in LabSolutions. |
0379 |
There is no specified path. |
There is no specified path. |
Specify the existing path and try again. |
037A |
The data file name is not |
There is no specified data file. |
Specify the data file and try again. |
037B |
The destination file name is not specified. |
The file name for conversion of XML format is not specified. |
Specify the file name and try again. |
037C |
Failed to open the sequence file. |
This file has been used on other window. If the file is not in use, this may be corrupted. |
1. Close the file and try again.\n2. If the attempt fails again, this file may be corrupted. |
037D |
Failed to open the waveform file. |
This file has been used on other window. If the file is not in use, this may be corrupted. |
1. Close the file and try again.\n2. If the attempt fails again, this file may be corrupted. |
037E |
Invalid parameter for From field |
Parameter in "From" field is invalid. |
Input correct parameter. |
037F |
Invalid parameter for To field |
Parameter in "To" field is invalid. |
Input correct parameter. |
0380 |
Invalid server name. |
Parameter in "Server name" field is invalid. |
Input correct parameter. |
0381 |
Incorrect port number. |
Parameter in "Port No" field is invalid. |
Input correct parameter. |
0382 |
E-mail transmission test has been completed successfully. |
This message means Send Email Test can be successfully performed. |
----- |
0383 |
Data file has been converted to AIA ANDI file. |
This means that theLabSolutionsdata file was successfully converted. |
----- |
0384 |
Failed to convert file from data file. |
The selected file was in LabSolutions format, but could not be converted to the new file format. |
Reboot the PC and try again. If the attempt fails again, the file may be corrupted. You can verify whether the file can be opened inLabSolutions. |
0385 |
CHROMATOPAC file has been converted to LabSolutions file |
This message means specified file can be successfully converted to LabSolutions file format. |
----- |
0386 |
Failed to convert file from CHROMATOPAC file. |
Fail to convert from CHROMATOPAC file. |
Restart the PC and try again. If the attempt fails again, the file may be corrupted. You can verify whether the file can be opened inCHROMATOPAC. |
0387 |
CLASS-LC10 file has been converted to LabSolutions file. |
This message means specified file can be successfully converted toLabSolutionsfile format. |
----- |
0388 |
Failed to convert file from CLASS-LC10 file. |
The selected file was in CLASS-LC10 format, but could not be converted to the new file format. |
Restart the PC and try again. If the attempt fails again, the file may be corrupted. You can verify whether the file can be opened in CLASS-LC10. |
0389 |
CLASS-VP file has been converted to LabSolutions file. |
This message means specified file can be successfully converted toLabSolutionsfile format. |
----- |
038A |
Failed to convert file from CLASS-VP file. |
Fail to convert from CLASS-VP file. |
Restart the PC and try again. If the attempt fails again, the file may be corrupted. You can verify whether the file can be opened in CLASS-VP. |
038B |
AIA ANDI file has been converted to LabSolutions file. |
This means that the AIA ANDI file was successfully converted. |
----- |
038C |
Failed to convert file from AIA ANDI file. |
The selected file was in AIA ANDI format, but could not be converted to the new file |
Restart the PC and try again. If the attempt fails again, the file may be corrupted |
038D |
ASCII file has been converted to LabSolutions file. |
This means that theLabSolutionsdata file was successfully converted. |
----- |
038E |
Failed to convert file from ASCII file. |
Fail to convert from ASCII file. |
Restart the PC and try again. If the attempt fails again, the file may be corrupted. You can verify whether the file is created with predetermined format. |
038F |
Failed to open the UV spectrum file. |
This file has been used on other window. If the file is not in use, this may be corrupted. |
1. Close the file and try again.\n2. If the attempt fails again, this file may be corrupted. |
0390 |
Failed to register to the Agent database. |
Failed to register to the CLASS Agent because the some kind of errors occurred. |
Confirm the error contents in the Output Window, and then optimize the Agent setting. |
0391 |
Failed to connect to Shimadzu Authentication Server. |
Fail to connect to User Authentication Tool. |
The PC network may be disconnected. Confirm the PC network connection. |
0392 |
This application is already running. |
You try to execute the application which has been already executed. |
Close the application and then try again. |
0393 |
Once you apply this setting, you cannot cancel the function. Are you sure you want to apply? |
This message confirm you because you cannot unlock this setting after applying it. |
----- |
0394 |
Incorrect method section information |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
0395 |
Raw data information is incorrect. |
----- |
---- |
0396 |
Failed to write the file |
Either there is insufficient disk space, or the file is corrupted. |
Increase the amount of available space on the disk, and execute the command again. To increase the amount of available space:_1. If the specified drive is on the hard drive, empty the Recycle Bin.\n2. Back up unnecessary files, and delete them from the drive. |
0397 |
CLASS-LC10 file has been converted to JCAMP file. |
This message means specified file can be successfully converted to JCAMP file |
----- |
0398 |
Failed to convert file from CLASS-LC10 file. |
The selected file was in CLASS-LC10 format, but could not be converted to the new file format. |
Restart the PC and try again. If the attempt fails again, the file may be corrupted. You can verify whether the file can be opened in CLASS-LC10. |
0399 |
CLASS-VP file has been converted to JCAMP file. |
This message means specified file can be successfully converted to JCAMP file |
----- |
039A |
Failed to convert file from CLASS-VP file. |
Fail to convert from CLASS-VP file. |
Restart the PC and try again. If the attempt fails again, the file may be corrupted. You can verify whether the file can be opened in CLASS-VP. |
039B |
Batch File has been converted to ASCII file. |
Message for notifying that the specified file has been converted to an ASCII file. |
----- |
039C |
Failed to convert file from batch file. |
Batch File cannot be converted to ASCII file. |
Restart the PC and try again. If the attempt fails again,the batch file may be corrupted. Make sure that the file can be read normally by the application. |
039D |
Data File has been converted to ASCII file. |
Message for notifying that the specified file has been converted to an ASCII file. |
----- |
039E |
Failed to convert file from data file. |
Data File cannot be converted to an ASCII file. |
Restart the PC and try again. If the attempt fails again,the data file may be corrupted. Make sure that the file can be read normally by the application. |
039F |
An error occurred on the analytical instrument. Check the details of the error at [Output Window] or [Log Browser]. |
An error occurred on the analytical instrument. |
Check the details of the error at [Output Window] or [Log Browser]. |
03A0 |
Failed to open the MS U161library file. |
The Library File may be in use in another window. If this is not the case, the file may be corrupted. |
If the error occurs again, the Library File is probably corrupted. Close all applications, and execute error checking at Windows disk properties. Then create a new |
03A1 |
Failed to save the MS library file. |
Either the Library File is corrupted, or there is insufficient memory. |
1. Try saving another Library File. If this is successful, then the first file is corrupted. Close all applications, and execute error checking at Windows disk properties.\n2. If another Library File cannot be saved, there may be insufficient memory. Close all unnecessary applications. |
03A2 |
Save current MS library file? |
Changes have been made to the Library File. Confirm whether to save the changes or close the file without saving the |
----- |
03A3 |
MS library file is not found. |
The specified Library File could not be found. |
1. Verify the path (drive and folder).\n2. Verify the Library File name. |
03A4 |
CLASS-VP is not installed. |
Although CLASS-VP software should be installed to convert from CLASS-VP file to LCsolution file, this software has not been installed. |
You should install the CLASS-VP software to convert from CLASS-VP file to LCsolution file. |
03A5 |
No license to use the specified instrument exists. |
You can not use the specified instrument under the your software license. |
If you want to use this instrument, purchase the software license. |
03A6 |
Invalid CLASS-VP environment configuration file name. Click the [Tools] - [Option] menu, and set the"Instrument*.cfg" file in the window that opens. |
CLASS-VP environment configuration file name is invalid. |
Click the [Tools] - [Option] menu, and set the "Instrument*.cfg"file in the window that opens. |
03A7 |
No license to use the specified application exists. |
The application can not be executed because the software license has not been |
Please purchase the software license. |
03A8 |
The data file cannot be converted into a JCAMP format file. |
Data File cannot be converted into a JCAMP format file. |
Restart the PC and try again. If the attempt fails again, the data file may be corrupted. Make sure that the file can be read normally by the application. |
03A9 |
Data file has been converted to JCAMP file. |
Message for notifying that the specified file has been converted to JCAMP file. |
----- |
03AA |
The JCAMP format file cannot be converted. |
The selected file was in JCAMP format, but could not be converted to the new file format. |
Restart the PC and try again. If the attempt fails again, the file may be corrupted. Make sure that the file can be read normally by the application. |
03AB |
The JCAMP format file was converted into a data file. |
Message for notifying that the specified file has been converted to data file. |
----- |
03AC |
The CLASS-8000/LCMSsolution Ver.2.0x format data file cannot be converted. |
The CLASS-8000/LCMSsolution Ver.2.0x format data file cannot be converted. |
----- |
03AD |
The CLASS-8000/LCMSsolution Ver.2.0x format data file was converted. |
Message for notifying that the specified file has been converted to data file. |
----- |
03AE |
The CLASS-8000/LCMSsolution Ver.2.0x format method file cannot be converted. |
The CLASS-8000/LCMSsolution Ver.2.0x format method file cannot be converted. |
----- |
03AF |
The CLASS-8000/LCMSsolution Ver.2.0x format method file was converted. |
Message for notifying that the specified file has been converted to method file. |
----- |
03B0 |
The application is running on Snap Shot mode. [File Open] operation will terminate the Snap Shot mode. Do you wish to |
You attempted to open a file in the Snap Shot mode. |
Click [OK] to exit the Snap Shot mode and enter the usual Postrun mode. \nClick [Cancel] to stay in the Snap Shot mode. |
03B3 |
You cannot initialize the format while it is being edited. |
This message is displayed when you run [Initialize Format] after running [Edit Format] from the [File] menu - [Print Image xxx] (xxx differs depending on the application software). \n[Initialize Format] deletes the format file. But as the format file to be initialized is currently being used in [Edit Format], the deletion fails and thus the above message is resulted. |
To initialize the format, be sure to close the report creation screen, which is opened by [Edit Format], in advance. |
03B4 |
Default report format file is not found. |
The default report file specified on the [Report] page in the [Setting Options] subwindow does not exist. |
Select the [Option] command on the [Tools] menu, and set the default report file on the [Report] tab page in the [Setting Options] sub-window. |
03B5 |
The license information cannot be correctly verified. Make sure that the USB dongle is correctly attached. |
The license information cannot be correctly verified. |
Make sure that the USB dongle is correctly attached. |
03B6 |
ID setting is not correct. |
ID setting is incorrect. |
Set the ID again. |
03B7 |
Delimiter is not correct. |
Delimiter is incorrect. |
Enter the delimiter again. |
03B8 |
Please specify the file name. |
The file name is not set. |
Set the file name. |
03BB |
Save current method file and data file? |
The method file or the data file has been changed. Confirmation for saving changes and exiting, or exiting without saving the changes. |
To save the new settings and continue operation, click [Yes]. To continue operation without saving the new settings, click [No]. To cancel the operation, click [Cancel]. |
03BC |
Cannot perform since you do not have rights. |
You do not have the required rights. |
Check the rights. |
03BD |
Failed to upload instrument parameters. |
Could not upload instrument parameters. |
Check the details displayed with the message. |
03BE |
The content of the change was recorded. |
Message for notifying that the audit trail was recorded. |
----- |
03C0 |
This operation is not available, because another user occupies the instrument. |
The instrument was occupied by another user. |
This operation cannot be done because the instrument was occupied by another user. |
03C1 |
Instrument Type is not specified. Please set the type on Instrument Administration window. |
The type of instrument that you attempted to start up is not set. |
Open the [Instrument Administration] sub-window from the [Administration Tools] sub-window in the [LabSolutions Main] window, and set the instrument type for the corresponding instrument. |
03C2 |
A communication error occurred between modules. The application will be closed. |
A communications error has occurred between modules of the software. . |
Check whether the PC environment is an overload. |
03C3 |
Failed to save this file, because this file is not in the state which can be used. |
The file cannot be overwrite,because this file is not in the state which can be used. |
Save the file under a different name. |
03C4 |
Another user is editing the file. Please save as another file |
The file you attempted to save is being edited by another user. |
Save the file under a different name. |
03C5 |
Another user is using the file by Realtime Analysis. Please save as another file name. |
The file you attempted to save is being used for data acquisition by another user. |
Save the file under a different name. |
03C6 |
Realtime Analysis application is not running. |
The method file currently being used for data acquisition cannot be saved while Realtime Analysis has not started up. |
Start up Realtime Analysis. |
03C7 |
Another user is using the Realtime Analysis. |
The method file currently being used for data acquisition cannot be saved since another user is using Realtime Analysis. |
Wait for the other user to exit Realtime Analysis. |
03C8 |
Unable to save the file while Startup process of data acquisition is going. Try to save again after the process is done. |
The file cannot be used since the file is being used in start processing of data acquisition. |
Wait a while and save the file again after start processing of data acquisition ended. |
03C9 |
Unable to save the file while data processing is going. Please try to save again after the processing is done. |
The file cannot be saved since the file is being used in data processing after data acquisition. |
Wait a while and save the file again after data processing ended. |
03CA |
The file was modified by the data acquisition processing. Try to save again. |
The file at the save destination was updated by data acquisition processing while save processing was being performed. Save processing was canceled since the latest file was acquired to recreate the audit trail information. |
Save the file again. |
03D0 |
Failed to access to Analysis Server. |
Either your PC or the Analysis Server is disconnected from the network or the Analysis Server is currently down. |
They may have been disconnected from the network. Check to see if they are properly connected to the network. \nWhen the Analysis Server is down, contact your network |
03D1 |
Unable to connect to the instrument, because the licence for the instruments does not exist or the number of upper limit of instrument control license is |
Unable to connect to the Instrument because the licence for the instruments does not exist or the number of upper bounds of control licenses for the instruments which can be connected at the |
Start this Realtime Analysis after confirm the licence for the instruments and finishing the other Realtime Analysis currently connected to the instrument. |
03D2 |
Reset the used count for consumables |
----- |
----- |
03D3 |
Failed to convert CLASS-VP data file. Install CLASS-VP Ver.5.05 or later/Ver.6.13 or later. |
The data file cannot be converted since CLASS-VP is not installed or the currently installed version is old. |
Install CLASS-VP Ver.5.04 or later_(when using Ver.5 series), or Ver.6.12 or later_(when using Ver.6 series). When converting compressed PDA data, CLASS-VP Ver.5.05 or later/Ver.6.13 or later is necessary to open CLASS-VP data file. |
03E0 |
This file is created by the product of the previous version. Overwriting this file will not maintain compatibility with the older versions. Are you sure you |
The file you attempted to overwrite is in old format made on previous versions of this product. This warning can be hidden by the [Option] setting on the [Tools] menu. |
Save under a different name to leave behind files that can be used in previous versions. |
03E1 |
Batch processing will overwrite the files of old format. \nAre you sure you want to start? |
The files overwritten by batch processing contain files in old format. This warning can be hidden by the [Option] setting on the [Tools] menu. |
Back up the data file and method file that are currently used in batch processing. |
03E2 |
Failed to save the file, because the LCMS license is not installed. |
The LCMS license is not installed. |
Save under a different name, or install the LCMS license. |
03E3 |
LCMS-QP8000 data cannot be loaded. |
LCMS-QP8000 data is not supported. |
---- |
03E4 |
LCMS-IT-TOF data cannot be loaded. |
LCMS-IT-TOF data is not supported. |
---- |
03E5 |
Export is finished. |
----- |
----- |
03E6 |
LC/GCsolution data file was converted. |
----- |
----- |
03E7 |
LC/GCsolution data file cannot be converted. |
----- |
----- |
03E9 |
MS type was changed.\nPlease restart the Realtime Analysis window. |
The MS type was changed in the system configuration. So, the [Realtime Analysis] window must be restarted. |
----- |
03EA |
Cannot connect to the database, because of failure. Please log in again. |
----- |
----- |
03EB |
Please select the interface. |
----- |
----- |
03EC |
The Instrument Method ID is |
----- |
----- |
03ED |
The Data Processing Method ID is issued. |
----- |
----- |
03EE |
The Calibration Curve ID is |
----- |
----- |
03EF |
The Instrument Method ID is changed. |
----- |
----- |
03F0 |
The Data Processing Method ID is changed. |
----- |
----- |
03F1 |
The Calibration Curve ID is |
----- |
----- |
0501 |
Successfully registered to the Agent database. |
----- |
----- |
0502 |
Warning occurred at the time of the registration to the Agent database. |
It is possible to fail the registration to the CLASS Agent because the some kind of warnings occurred. |
Confirm the warning contents in the Log Browser, and then optimize the Agent setting. |
0503 |
Failed to register to the Agent database. |
Failed to register to the CLASS Agent because the some kind of errors occurred. |
Confirm the error contents in the Log Browser, and then optimize the Agent setting. |
0504 |
CLASS Agent user rights is not registered. Please run the Agent Manager and register a user account. |
The right to register the data to the CLASS Agent can not be specified. |
Specify the right in the CLASS Agent Manager. |
0505 |
Agent registration settings have been changed. |
----- |
----- |
0507 |
A forced exit occurred because freezing of LSSAgentRegister.exe was |
There was no response for the fixed period of time when registering data to the Agent database. |
Perform processing again after restarting the PC. If the error occurs even after this, contact your Shimadzu representative. |
0508 |
CLASS-Agent is not installed. |
Because CLASS-Agent is not installed, it is not possible to execute it. |
Please install CLASS-Agent. |
0701 |
CBM Error: Syntax error |
----- |
----- |
0702 |
CBM Error: Invalid value |
----- |
----- |
0704 |
CBM Error: File type is not of specified |
----- |
----- |
0707 |
CBM Error: Invalid file name |
----- |
----- |
0708 |
CBM Error: File already exists |
----- |
----- |
0709 |
CBM Error: Too many files |
----- |
----- |
070A |
CBM Error: File not found |
----- |
----- |
070B |
CBM Error: Directry full |
----- |
----- |
070D |
CBM Error: Invalid token |
----- |
----- |
070F |
CBM Error: Invalid operation |
----- |
----- |
0711 |
CBM Error: Operation break |
----- |
----- |
0712 |
CBM Error: Insufficient memory |
----- |
----- |
0713 |
CBM Error: Invalid path |
----- |
----- |
072A |
CBM Error: Port is not opend |
----- |
----- |
072B |
CBM Error: Not accepted |
Controller did not accept a command. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
072C |
CBM Error: Invalid function |
Invalid function has been invoked. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
072D |
CBM Error: Link error occurred |
Link error has occurred. |
(1) Check the instrument status and the cable connections.\n(2) Use faster PC, or decrease the number of systems or detectors to decrease transmission traffic.\nIf the error persists, turn the power OFF and contact your Shimadzu representative. |
073A |
CBM Error: No response |
There was no response from controller. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
0741 |
CBM Error: Write protected |
----- |
----- |
0743 |
CBM Error: Drive is not ready |
----- |
----- |
074C |
CBM Error: Disk full |
----- |
----- |
0750 |
CBM Error: Disk is not initialized |
----- |
----- |
0760 |
CBM Error: Data size does not match |
----- |
----- |
0780 |
CBM Error: Edit file # mismatches run file # |
----- |
----- |
0781 |
CBM Error: Upload buffer size is not enough |
----- |
----- |
07A0 |
CBM Error: Valid data file is not found |
----- |
----- |
07A1 |
CBM Error: Device is not installed |
----- |
----- |
07A2 |
CBM Error: Device is not |
----- |
----- |
07A3 |
CBM Error: Method is already |
----- |
----- |
07A4 |
CBM Error: Parameter is not |
----- |
----- |
07A5 |
CBM Error: Program size exceeds maximum memory |
----- |
----- |
07B0 |
CBM Error: Configuration settings do not match with the actual |
Configuration settings do not match the actual system. |
Reconfigure the system by auto-configuration. |
07B1 |
CBM Error:Invalid configuration parameter settings. |
Configuration parameter settings were not valid. |
Turn the power OFF and then ON again. If the error persists, turn the power OFF and contact your Shimadzu |
07B2 |
CBM Error: No device is assigned to the stream |
----- |
----- |
07B3 |
CBM Error: Device is not |
----- |
----- |
07B4 |
CBM Error: Controller type does not match with the actual HPLC |
Controller type does not match with the actual HPLC system. |
Select correct LC controller type in the instrument configuration. |
07C0 |
CBM Error: FOR-NEXT nest too deep |
----- |
----- |
07C1 |
CBM Error: FOR-NEXT nest does not match |
----- |
----- |
07C2 |
CBM Error: Max. # of files |
----- |
----- |
07D0 |
CBM Error: In pointer overruns out pointer of AD ring buffer |
----- |
----- |
07D1 |
CBM Error: Not enough data |
----- |
----- |
07D2 |
CBM Error: Level is out of range |
----- |
----- |
07D3 |
CBM Error: Data is stored in |
----- |
----- |
07E0 |
CBM Error: No data is stored in buffer |
----- |
----- |
07E1 |
CBM Error: Rate has changed |
----- |
----- |
07E2 |
CBM Error: Gain has changed |
----- |
----- |
07F0 |
CBM Error: Too many clients use CBM.OCX |
----- |
----- |
07FF |
Value is out of range. |
An out-of-range value was set to CBM_. |
Restart the CBM and PC. |
1000 |
Failed to set up the user. |
The new settings cannot be saved. A probable cause is a problem in the network connection in the case of a network configuration. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1001 |
Failed to get the user information. |
Failed to retrieve the information. A probable cause is a problem in the network connection in the case of a network configuration. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1002 |
Same user's name already exists. |
The same name as that of a currently registered user was entered. |
Enter another user name. |
1003 |
Same user's ID already exists. |
The same ID as that of a currently registered user was entered. |
Enter another user ID. |
1004 |
No rights of user administration. |
A user not allowed to perform user administration attempted to make a |
Log in as a user with rights. |
1005 |
Cannot perform, because other users are editing. |
There are several possible causes as follows.\n(1) When other user is editing.\n(2) When some problems occur |
(1) Try again after other user has finished editing.\n(2) Wait about ten minutes and try again. If this does not remedy the problem, contact your Shimadzu representative. |
1006 |
The specified user does not exist. |
The specified user does not exist. |
Close the sub-window, and select the user again. |
1007 |
Are you sure you want to delete the user account? |
Displayed for confirmation when a user is deleted. |
To delete, click [Yes], and to cancel deletion, click [No]. |
1008 |
Failed to delete the user account. |
Failed to delete. A probable cause is a problem in the network connection in the case of a network configuration. |
Delete again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1009 |
Failed to set up the user group. |
The new settings cannot be saved. A probable cause is a problem in the network connection in the case of a network configuration. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
100A |
Failed to getting the user group information. |
Failed to retrieve the information. A probable cause is a problem in the network connection in the case of a network configuration. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
100B |
Are you sure you want to delete the user group? |
Displayed for confirmation when a user group is deleted. |
To delete, click [Yes], and to cancel deletion, click [No]. |
100C |
The specified user group does not exist. |
The specified user group does not exist. |
Close the sub-window, and select the user group again. |
100D |
Failed to delete the user group. |
Failed to delete. A probable cause is a problem in the network connection in the case of a network configuration. |
Delete again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
100E |
Failed to set up the rights group. |
The new settings cannot be saved. A probable cause is a problem in the network connection in the case of a network configuration. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
100F |
Failed to get the rights group information. |
Failed to retrieve the information. A probable cause is a problem in the network connection in the case of a network configuration. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1010 |
The rights group has users._\nIf the group will be deleted, the rights will also be deleted from these users._\nAre you sure you want to continue? |
Displayed for confirmation when a rights group to which the user belongs is deleted. |
To delete, click [Yes], and to cancel deletion, click [No]. |
1011 |
The specified rights group does not exist. |
The specified rights group does not exist. |
Close the sub-window, and select the rights group again. |
1012 |
Failed to delete the rights group. |
Failed to delete. A probable cause is a problem in the network connection in the case of a network configuration. |
Delete again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1013 |
Failed to get Windows domain account. |
A probable cause is a problem in the connection with the domain controller. |
Try again after checking the network connection status with the domain controller. |
1014 |
Failed to update. |
The new settings cannot be saved. A probable cause is a problem in the network connection in the case of a network configuration. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1015 |
Failed to get login user information. |
Failed to retrieve the information. A probable cause is a problem in the network connection in the case of a network configuration. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1016 |
If the logout is forced, the files under edit will be discarded. Are you sure you want to continue? |
Displayed for confirmation when a forced logout is notified. |
Click [OK] to perform a forced logout, and click [Cancel] to cancel. |
1017 |
Failed to force logout. |
The forced logout cannot be notified. A probable cause is a problem in the network connection in the case of a network configuration. |
Try again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1019 |
The specified login user ticket is invalid. |
The currently operating application has not yet been logged into. |
Start the application again. |
101A |
The specified user already logs in. |
The ID of the currently logged in user was entered. |
Enter another user ID. |
101B |
It does not log in. |
There is no currently logged in user. |
Log in as a user. |
101C |
The specified user ID is invalid. |
User ID entry is not correct. |
Enter the correct user ID. |
101D |
No rights to perform the forced logout. |
A user without the [Perform Forced Logout] rights attempted to make a setting. |
Log in as a user with the [Perform Forced Logout] rights. |
101E |
The specified user is not |
The specified user is not registered. |
Enter the correct user ID. |
101F |
The password is invalid. |
Password entry is not correct. |
Enter the correct password. Be careful when entering since the password is case-sensitive. If you have forgotten the password, contact the system administrator. |
1020 |
This user is locked now. \nPlease call a system administrator. |
Login failed since the login retry limit set to the system policy was exceeded. Currently, login is not possible. |
Wait a while until the login lockout time has elapsed. |
1021 |
This PC is locked now. \nPlease call a system administrator. |
Login is not possible for a while since login is locked out when login fails for the preset number of times or more. |
You can log in again after the time set in the system policy has elapsed. |
1022 |
Your password has expired and must be changed. |
The expiry date for the password has been exceeded. |
Check the expiry date with the system administrator. Change the password. |
1023 |
The passwords you typed do not match. |
The new password does not match the password you typed for confirmation. |
Enter the correct password. |
1024 |
The password is too short. |
A password shorter than the currently set minimum length was entered. |
Check the minimum password length, and enter a password longer than this length. |
1025 |
The network error has been recovered._\nWhen you operate it, please log in again after closing all screens. |
The network error was recovered. |
When you operate it, please log in again after closing all screens. |
1026 |
The network error occurred. \nPlease log in again after closing all screens. |
There are screens left open by a user who is attempting to log in. |
When you operate it, please log in again after closing all screens. |
1027 |
Login. |
----- |
----- |
1028 |
Logout. |
----- |
----- |
1029 |
Number of login users exceeds maximum. |
The upper limit for the number of simultaneous logged in users was |
Log out currently logged in other users. |
102A |
The password does not meet the complexity requirement.\nSet a password containing at least 3 of the following types of characters.\n-Uppercase alphabet characters (A to Z)\nLowercase alphabet characters (a to z)\n-Decimal numbers (0 to 9)\n-Special characters other than alphanumerics (!, $, #, %, |
The password that you attempted to set does not satisfy the complexity requirements of the password set in the system policy. |
Set a password containing at least 3 of the following types of characters.\n-Uppercase alphabet characters (A to Z)\nLowercase alphabet characters (a to z)\n-Decimal numbers (0 to 9)\n-Special characters other than alphanumerics (!, $, #, %, etc.) |
102B |
It is during a password change prohibition period. |
The password cannot be changed since it is still in the password change prohibition period. (The password change prohibition period cannot be set on the current |
Change the password after [Minimum Password Age] currently set in the system policy has been exceeded. |
102C |
The same password exists in the history buffer. |
An attempt was made to set the same password as the password left in the history for the number of preset times when the duplicate password management count is set in the system policy. |
Set another password. |
102D |
Password change is required at the time of first time login after the password change by other |
Password change is required at the time of first time login after the password was changed by a user administrator. |
Change the password. |
102F |
Failed to configure the fingerprint authentication. |
The new settings cannot be saved. A probable cause is a problem in the network connection in the case of a network configuration. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1030 |
Failed to register the fingerprint. |
The fingerprint cannot be registered for some reason. |
Execute fingerprint registration again. |
1031 |
No fingerprint is registered. |
The fingerprint is not registered. |
Log in as a fingerprint-registered user. |
1032 |
No lockout PCs, now. |
----- |
----- |
1033 |
No lockout users, now. |
----- |
----- |
1034 |
Failed to get the PC lockout information. |
Failed to retrieve the information. A probable cause is a problem in the network connection in the case of a network configuration. |
Start again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1035 |
Failed to release of PC lockout. |
Cannot cancel. A probable cause is a problem in the network connection in the case of a network configuration. |
Release again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1036 |
Failed to get the user lockout information. |
Failed to retrieve the information. A probable cause is a problem in the network connection in the case of a network configuration. |
Start again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1037 |
Failed to release of the user lockout. |
Cannot cancel. A probable cause is a problem in the network connection in the case of a network configuration. |
Release again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1038 |
The instrument which can be used does not exist. |
The instrument cannot be used by the user. Cannot log in during a failure. |
----- |
1039 |
The password is too long. |
A password longer than the currently set maximum length was entered. |
Check the maximum password length, and enter a password shorter than this length. |
103A |
Old password is invalid. |
Old password is invalid. The password cannot be changed. |
Enter the correct old password. |
103B |
Are you sure you want to delete the rights group? |
Displayed for confirmation when a rights group is deleted. |
To delete, click [Yes], and to cancel deletion, click [No]. |
103C |
Open Application. |
----- |
----- |
103D |
Close Application. |
----- |
----- |
103E |
Lockout the user. |
Message for notifying that login was locked out since the number of login failure was more than the lockout threshold set in the system policy. |
Once a login is locked, the user cannot log in until the login lockout time has elapsed. Wait a while. |
103F |
Lockout the PC. |
Message for notifying that login was locked out since the number of login failure was more than the lockout threshold set in the system policy. |
Once a login is locked, the user cannot log in until the login lockout time has elapsed. Wait a while. |
1040 |
Failed to log_in the system. |
Could not log in for some reason. |
Confirm the setting of [Multi-User] in [Accouont]tab of the[Security Policy Settings] and log in again. |
1041 |
Please put the finger you want to register on the fingerprint reader. |
----- |
----- |
1042 |
It is necessary to read the fingerprint again for confirmation. Please put the finger again after separating from the reader. |
It is necessary to read the fingerprint again for confirmation. Please put the finger again after separating from the reader. Cannot register unless you once take your finger away from the reader. |
----- |
1043 |
It is already registered. Are you sure you want to register again? |
Displayed for confirmation when an attempt was made to register a fingerprint to a user whose fingerprint is already |
Click [Yes] to re-reregister fingerprints, and click [No] to cancel. |
1044 |
Add a user account |
----- |
----- |
1045 |
Edit a user account |
----- |
----- |
1046 |
Delete a user account |
----- |
----- |
1047 |
Add a user gourp |
----- |
----- |
1048 |
Edit a user gourp |
----- |
----- |
1049 |
Delete a user gourp |
----- |
----- |
104A |
Add a rights gourp |
----- |
----- |
104B |
Edit a rights gourp |
----- |
----- |
104C |
Delete a rights gourp |
----- |
----- |
104D |
Another user is editing it now. Please log_in again after a while. |
Cannot perform because another user is editing. |
Try again after another user has finished editing. |
104E |
Please register your fingerprint. |
Displayed when [OK] is clicked after registering the No.1 and No.2 fingers. |
Click [OK] after registering the No.1 and No.2 fingers. |
104F |
Failed to look up the Windows domain. |
A probable cause is a problem in the connection with the domain controller. |
Try again after checking the network connection status with the domain controller. |
1050 |
Scanning of fingerprint has been completed. |
----- |
----- |
1051 |
Failed to scan your fingerprint. |
Cannot read your fingerprint. |
Perform reading again. |
1052 |
If the user name is edited, the signature on data files acquired by the user in question may not be possible. Are you sure you want to continue? |
Displayed for confirmation when the user name was changed with system policy "Level 1 Accept only by the operator who acquired data" set to ON. |
To execute editing, click [Yes], and to cancel editing, click [No]. |
1053 |
Logged in during a failure. |
----- |
----- |
1054 |
Login has been canceled. |
----- |
----- |
1055 |
User ID or password is not correct. \nYou may pay attention to the password, which is case sensitive. |
User ID or password is incorrect. |
Enter the correct user ID and password. Be careful when entering since the password is case-sensitive. If you have forgotten the password, contact the system administrator. |
1056 |
User ID is not correct or the fingerprint is not registered. |
User ID is incorrect or the fingerprint is not registered. |
Log in as a fingerprint-registered user. |
1057 |
Forced logout. |
----- |
----- |
1058 |
The specified user group name already exists.\nPlease specify another name. |
Displayed for confirmation when there is a user group of the same name as the specified user group name. |
Change to another name. |
1059 |
The specified rights group name already exists.\nPlease specify another name. |
Displayed for confirmation when there is a rights group of the same name as the specified rights group name. |
Change to another name. |
1100 |
The specified project does not exist. |
The specified project does not exist. |
Close the sub-window, and select the project again. |
1101 |
No rights of system policy administration. |
A user without the [System Administration] rights attempted to make a setting. |
Log in as a user with the [System Administration] rights. |
1102 |
Cannot perform, because other users are editing. |
There are several possible causes as follows.\n(1) When other user is editing.\n(2) When some problems occur |
(1) Try again after other user has finished editing.\n(2) Wait about ten minutes and try again. If this does not remedy the problem, contact your Shimadzu representative. |
1103 |
Failed to get the system policy information. |
Failed to retrieve the information. A probable cause is a problem in the network connection in the case of a network configuration. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1104 |
Failed to set the system policy information. |
The new settings cannot be saved. A probable cause is a problem in the network connection in the case of a network configuration. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1105 |
Failed to get the project policy information. |
Failed to retrieve the information. A probable cause is a problem in the network connection in the case of a network configuration. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1106 |
Failed to set the project policy information. |
The new settings cannot be saved. A probable cause is a problem in the network connection in the case of a network configuration. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1107 |
Once you apply this setting, you cannot cancel the function. Are you sure you want to apply? |
Message for confirming whether or not to continue since this setting cannot be canceled once it is executed. |
To set, click [OK], and to cancel setting, click [Cancel]. |
1130 |
The specified instrument name already exists. \nPlease specify another name. |
The specified instrument name already exists. |
Change to another name. |
1131 |
Target instrument does not exist. |
The specified instrument does not exist. |
Close the sub-window, and select the instrument again. |
1132 |
The number of the registered instruments exceeds maximum. |
The number of instruments exceeds the maximum registrable number. |
Instruments beyond the upper limit cannot be registered. |
1133 |
Failed to get the instrument information. |
The information cannot be acquired for some reason. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1134 |
Failed to add the instrument. |
The instrument cannot be added for some reason. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1135 |
Failed to edit the instrument. |
The instrument cannot be edited for some reason. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1136 |
Failed to remove the instrument. |
The instrument cannot be deleted for some reason. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1137 |
Failed to move the instrument. |
The instrument cannot be moved for some reason. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1138 |
Failed to get the information. |
The information cannot be acquired for some reason. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1139 |
Failed to add. |
Cannot add for some reason. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
113A |
Failed to delete. |
Cannot delete for some reason. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
113B |
Failed to edit. |
Cannot edit for some reason. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
113C |
Failed to restore. |
Cannot restore for some reason. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
113D |
The instrument will be deleted. \nAre you sure you want to continue?\n(Method files, batch files, report format files, and the specific information to the instrument will be deleted.) |
Displayed for confirmation since Method files, batch files, report format files, and the specific information saved to each instrument is deleted when an instrument is deleted. |
To delete, click [OK], and to cancel deletion, click [Cancel]. |
113E |
No response. |
No response for some reason. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
113F |
Information will be added from the following contents. |
----- |
----- |
1140 |
Failed to get the information. \nDo you fill the lack of information as 'unknown'? |
Failed to retrieve the information from the server. To use the instrument, validated information must be acquired later. However, instrument information can be added if locations that cannot be acquired are set as "unknown". |
To add, click [OK], and to cancel addition, click [Cancel]. |
1141 |
Information will be edited by the following contents. |
----- |
----- |
1142 |
Failed to get the information. \nDo you edit after filling the lack of information as 'unknown'? |
Failed to retrieve the information from the server. To use the instrument, validated information must be acquired later. However, instrument information can be edited if locations that cannot be acquired are set as "unknown". |
To edit, click [OK], and to cancel edition, click [Cancel]. |
1143 |
The specified name already exists. \nPlease specify another |
The specified name already exists. |
Change to another name. |
1144 |
Are you sure you want to remove? |
Displayed for confirmation when an instrument is deleted. |
To delete, click [OK], and to cancel deletion, click [Cancel]. |
1145 |
The controller has some instruments. \nPlease remove the controller after removing the connected instruments. |
Some instruments are connected to the controller. |
Perform deletion of the controller after deleting the instruments. |
1146 |
The controller will be initialized with the backup data on the |
----- |
----- |
1147 |
The administration No. are different between the registration and the present. \nThe controller will be initialized with the backup data on the database. \nAre you sure you want to continue? (Definition information will be updated after Restore.) |
Data is initialized and a restore is performed using the backup data on the database when a restore is performed when the administration No. in the registered details and the administration No. of the actual instrument do not match. Displayed for confirming whether or not to perform a restore. |
To restore, click [OK], and to cancel restoration, click [Cancel]. |
1148 |
Failed to communicate with the controller to remove. \nThe specific information on the controller remains after removing. \nAre you sure you want to continue? |
Controller-specific information is left behind if a controller is deleted when communication with the controller to be deleted is not possible. To delete specific information, first enable communications with the controller to be deleted. |
To delete even with controller-specific information left behind, click [OK], and to cancel deletion, click [Cancel]. |
1149 |
Cannot perform, because other users are editing. |
There are several possible causes as follows.\n(1) When other user is editing.\n(2) When some problems occur |
(1) Try again after other user has finished editing.\n(2) Wait about ten minutes and try again. If this does not remedy the problem, contact your Shimadzu representative. |
114A |
The target PC does not exist. |
The specified computer does not exist. |
Close the sub-window, and select the computer again. |
114B |
The instrument is used for analysis. \n Please perform operation after stop. |
The instrument is used for analysis. This operation is currently not available. |
Try again after stopping execution. |
114C |
Failed to communicate with the controller to which the instruments are connected. \nThe specific information on the instruments remains after removing. \nAre you sure you want to continue? |
Instrument-specific information (files saved to each instrument or system configuration information) is left behind when communication is not possible with the controller connected to the instrument. To delete specific information, first enable communications with the controller to be deleted. |
To delete even with the controller-specific information left behind, click [OK], and to cancel deletion, click [Cancel]. |
114D |
Failed to get the column information. |
Failed to retrieve the information. A probable cause is a problem in the network connection in the case of a network configuration. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
114E |
Failed to set the column information. |
The new settings cannot be saved. A probable cause is a problem in the network connection in the case of a network configuration. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
114F |
No rights of instrument administration. |
A user without the [Change Instrument Settings] rights attempted to make a |
Log in as a user with the [Change Instrument Settings] rights. |
1150 |
The administration No. are different between the registration and the present._\nDo you move it anyway?_\n(Definition information will be updated after the move.) |
Displayed for confirming whether or not to move when the administration No. in the registered details of the acquisition controller and the administration No. of the actual instrument do not match. The registered details are updated when a move is performed in this state. |
To move, click [OK], and to cancel moving, click [Cancel]. |
1151 |
Add Acquisition controller |
----- |
----- |
1152 |
Edit Acquisition controller |
----- |
----- |
1153 |
Edit PC |
PC edit |
----- |
1154 |
Remove Acquisition controller |
----- |
----- |
1155 |
Restore Acquisition controller |
----- |
----- |
1156 |
Add Instrument |
----- |
----- |
1157 |
Edit Instrument |
----- |
----- |
1158 |
Remove Instrument |
----- |
----- |
1159 |
Move Instrument |
----- |
----- |
115A |
Please restart the acquisition controller. After the restart, click [OK] to start the Restore. |
Restart the acquisition controller. |
After the restart, click [OK] to start the restoration. |
115B |
Please restart the acquisition controller to use. |
Restart the acquisition controller. |
----- |
115C |
The number of the registered instrument type are over the limitation. |
The number of the registered instrument type are over the limitation. |
The instrument type which exceeds the upper limit cannot register. |
115D |
The specified instrument is |
The specified instrument is invalid. |
Please make the instrument valid. |
115E |
Files and the specific information on the deleted instrument cannot be recovered.\nAre you sure you want to delete the instrument? |
----- |
----- |
1161 |
Failed to get the project information. |
Failed to retrieve the project information. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1162 |
Failed to create the project information. |
Failed to create the project information. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1163 |
Failed to edit the project information. |
Failed to edit the project information. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1164 |
Failed to completely delete the project information. |
Failed to delete the project information. \nThe instrument, the data file, the meta file and etc included in project may be in use. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again.\nConfirm whether the instrument, the data file, the metafile and etc included in project are used, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1165 |
The specified project name already exists. \nPlease specify another name. |
Displayed for confirmation when there is a project of the same name as the specified project name. |
Change to another name. |
1166 |
The project will be deleted. \nThe meta file, which the project has, will also be deleted. Are you sure you want to continue? |
Displayed for confirmation when a meta file relating to the project is left behind when a project is deleted. |
To delete, click [Yes], and to cancel deletion, click [No]. This message is not displayed when executing again after deleting the meta file. |
1167 |
Cannot remove instrument from a project under analysis. |
Cannot delete the instrument from the project since the instrument is currently performing analysis. |
Delete the instrument after analysis is finished. |
1168 |
Cannot remove instrument from a project under analysis |
Cannot delete instrument from a project under analysis reservation. |
Delete the instrument after analysis is finished or after deleting analysis reservation. |
1169 |
Cannot remove instrument from a project under meta file editing. |
Cannot delete instrument from the project since there is an instrument which is editing the meta file. |
Delete the instrument after the meta file editing is finished. |
116A |
No rights of project administration |
A user without the [Change Project Settings] rights attempted to make a |
Log in as a user with the [Change Project Settings] rights. |
116B |
MDB file has been newly recreated. Because the Restore failed. |
Cannot restore the MDB file. A new file was made, and the settings were initialized. |
----- |
116C |
The specified contents will be canceled. \nAre you sure you want to continue? |
Displayed for confirmation when the set content is discarded. |
If it is OK to delete the settings, click [Yes], and if it is not OK, click [No]. |
116D |
The specified project does not exist. |
The specified project does not exist. |
Close the sub-window, and select the project again. |
116E |
Some policies will be set to the default in ER/ES mode, because the source project is not in ER/ES mode. |
Since the copy source project is not in the ER/ES mode, several policies of the project to be made are set as ER/ES mode defaults. This message is displayed for confirming whether or not to continue. |
To create, click [Yes], and to cancel creation, click [No]. |
116F |
Project ID is not specified. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1170 |
The specified project is already deleted. |
The specified project was deleted. |
Update to the latest information, or restart the application, and specify an existing project to execute. |
1171 |
The definition file of operation does not exist. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1172 |
Add project |
----- |
----- |
1173 |
Edit project |
----- |
----- |
1174 |
Delete project |
----- |
----- |
1175 |
Please restart the software to enable the deletion of the project. |
The list of porject of the mainwindow is not refresed after the deleteion of the project. |
Please restart the mainwindow. |
1176 |
The specified project name already exists. |
Displayed for confirmation when there is a project of the same name as the specified project name. |
|
1177 |
Cannot perform, because other users are editing. |
There are several possible causes as follows.\n(1) When other user is editing.\n(2) When some problems occur |
(1) Try again after other user has finished editing.\n(2) Wait about ten minutes and try again. If this does not remedy the problem, contact your Shimadzu representative. |
1190 |
No rights to set up the system. |
A user without the [Change System Settings] rights attempted to make a |
Log in as a user with the [Change System Settings] rights. |
1191 |
Cannot perform, because other users are editing. |
There are several possible causes as follows.\n(1) When other user is editing.\n(2) When some problems occur |
(1) Try again after other user has finished editing.\n(2) Wait about ten minutes and try again. If this does not remedy the problem, contact your Shimadzu representative. |
1192 |
Failed to save the setting. |
The new settings cannot be saved. A probable cause is a problem in the network connection in the case of a network configuration. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1193 |
Failed to get the information. |
Failed to retrieve the information. A probable cause is a problem in the network connection in the case of a network configuration. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1194 |
Too many items. |
Exceeds the number of items that can be entered. |
Enter again after checking the number of items that can be entered. The number of items that can be entered can be verified by Help in each window. |
1195 |
The specified project does not exist. |
The specified project does not exist. |
Close the sub-window, and select the project again. |
1196 |
The settings have been changed. |
Settings were changed in the data processing setting sub-window. |
----- |
1197 |
The Display Format of the selected instrument type is reset.\nAre you sure? |
Displayed for confirmation when the Display Format of the selected instrument type is reset. |
To initialize, click [OK], and to cancel the initialization, click [Cancel]. |
1198 |
The same print format already exists. Are you sure you want to overwrite it? |
|
|
1199 |
Are you sure you want to delete the selected print format? |
|
|
119A |
The same print format already exists. Are you sure you want to rename it? |
|
|
119B |
The print format has been saved. |
|
|
119C |
The print format has been |
|
|
119D |
The file name has been renamed. |
|
|
1200 |
Failed in transmission test. |
The event transmission settings test failed since settings were incorrect or a fault occurred in the network. |
Execute again after making sure that the event transmission settings are connect. \nExecute again after making sure that the network is connected correctly. \nExecute again after making sure that the transmission is not blocked by the security software installed on the PC. If the security software blocks the transmission, change the security software settings to avoid blocking. \nWith LabSolutions, event transmission is executed using "LSSProcessController.exe". \nIf this does not remedy the |
1201 |
Transmission test has been completed successfully. |
The transmission test was successful. |
----- |
1202 |
Failed to save the settings. |
The new settings cannot be saved. A probable cause is a problem in the network connection in the case of a network configuration. |
Retry the e-mail settings. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1203 |
Failed to get the information. |
Failed to retrieve the information. A probable cause is a problem in the network connection in the case of a network configuration. |
Retry the e-mail settings. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1204 |
Failed to send e-mail/message. |
E-mail messages cannot be sent for some reason. |
Check the e-mail message settings. Also, check the network environment for sending e-mails. |
1205 |
E-mail/message has been sent. |
E-mail message transmission was successful. |
----- |
1206 |
Failed to attach the file. |
Could not append the file when e-mail was sent. |
1. Check that the specified file is available. 2. When a data file is specified, send again after making sure that the network is connected correctly. |
1207 |
No operation rights. |
A user without rights attempted to make a setting. |
To set the entire system, log in as a user with the [System Administration] rights, and to set by individual users, log in as a user with the [Change User Mail Settings] rights. |
1208 |
This operation is not available since the server is not set up. Please set up the server. |
This operation is not available since the server has not been set. |
Set the server. |
1209 |
Failed to change the e-mail transmission flag for the sample information. |
Failed to change the mail transmission flag of the sample information. A probable cause is a problem in the network connection in the case of a network configuration. |
Change the mail transmission flag of the sample information again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1230 |
Printing was abnormally |
Printing ended abnormally for some |
Print again after checking the printer status. |
1231 |
Failed to get the information. |
Failed to retrieve the information. A probable cause is a problem in the network connection in the case of a network configuration. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1232 |
No rights to perform the validation. |
A user without the [Perform Validation Support] rights attempted to make a |
Log in as a user with the [Perform Validation Support] rights. |
1233 |
Cannot get PC list. |
Failed to retrieve the information. A probable cause is a problem in the network connection in the case of a network configuration. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1234 |
Failed to get PC information. |
Failed to retrieve the information. A probable cause is a problem in the network connection in the case of a network configuration. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1235 |
Are you sure you want to delete specified PC information? |
Displayed for confirmation when the information of a selected computer is |
To delete, click [Yes], and to cancel deletion, click [No]. |
1237 |
An error occurred when Integrity Check Table was opened. |
There is no check table, or the table is corrupted. |
Contact your Shimadzu representative. |
1238 |
Check was terminated. <<Integrity check itself failed>> |
----- |
----- |
1239 |
Check has been completed. <<Integrity check results passed>> \nPress [Browse...] to view the report. |
----- |
----- |
123A |
Check has been completed. <<Integrity check results failed>> \nPress [Browse...] to view the report. |
----- |
----- |
123B |
Failed to get the integrity check results. |
Cannot get the result of program files check for some reason. |
Try again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
123C |
An error occurred when the file to be checked was opened. |
There is no file to be checked, or the file is corrupted. |
Contact your Shimadzu representative. |
123D |
Failed to create the Integrity Check Table. |
A probable cause is that the file specified in the check criteria table or the file to be checked in the program file check is in |
Create the check table again after the use of corresponding file is finished. |
123E |
PC information has been deleted. |
Information on the computer is deleted. |
----- |
123F |
An error occurred when the file to create Integrity Check Table was opened. |
Cannot open the program file check table creation file. |
Please make sure that the file specified in the startup parameter exists to create the Integrity Check Table. And then try again. |
1260 |
Failed to copy the customization information. |
Failed to copy the customized information. A probable cause is a problem in the network connection in the case of a network configuration. |
Copy the customized information again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1261 |
The customization information for selected users will be initialized. \nAre you sure you want to continue? |
Displayed for confirmation when the customized information of the currently selected user is initialized. |
To initialize, click [OK], and to cancel the initialization, click [Cancel]. |
1262 |
Failed to initialize the customization information. |
Failed to initialize the customized information. A probable cause is a problem in the network connection in the case of a network configuration. |
Initialize the customized information again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1291 |
Cannot retrieve the screen information because of the communication disturbance. |
A cause is a problem in the network connection. |
Execute again after making sure that the network is connected correctly. |
1263 |
No rights of administration for customization information. |
A user without rights attempted to perform administration operations. |
Log in as a user with the [Manage Customizing Information] rights. |
12C0 |
initializing. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
12C1 |
Property classification does not match. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
12C2 |
The property does not exist. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
12D0 |
Failed to get the license information from the dongle. |
The dongle is not inserted. |
Try again after inserting the dongle. |
12D1 |
Failed to store the license information to the profile administration database. |
The license information cannot be stored to the profile administration database. |
Contact your Shimadzu representative. |
12D2 |
Failed to get the license information from the profile administration database. |
The license information cannot be acquired from the profile administration database. |
Contact your Shimadzu representative. |
12D3 |
The dongle is disconnected. |
The dongle became disconnected during startup. |
Insert the dongle. |
12D4 |
The license authentication was not accepted. |
The dongle is not inserted, or the license is not registered. |
Try again after inserting the dongle, or registering the license. |
12D5 |
The specified license does not exist. |
The specified license does not exist. |
Insert the licensed dongle, or register the license. |
12D6 |
A serial number is not assigned to the specified license. |
A serial No. is not assigned to the specified license. |
Contact your Shimadzu representative. |
12D7 |
The specified license is already authenticated. |
Authentication of an already authenticated license was requested. |
Contact your Shimadzu representative. |
12E0 |
A redundant HDD degeneracy occurred. |
Acquisition controller HDD failure. |
Replace the malfunctioning HDD. Note that the system can be used normally until the other HDD malfunctions. |
12E1 |
Failed to rebuild the redundant HDD. |
Rebuilding ended, however, there is an unreadable sector on the drive. |
Replace the malfunctioning HDD. Note that the system can be used normally until the other HDD malfunctions. |
12E2 |
A drop in HDD remaining space occurred. |
Too many files on the acquisition controller. |
Delete unwanted files. |
12E3 |
A drop in remaining memory occurred. |
Acquisition controller processing overload. |
Restart the acquisition controller and see what happens. If the problem is not remedied, reduce the system load, for example, by reducing the number of analytical instruments. |
12E4 |
An increase in the CPU load rate occurred. |
Acquisition controller processing overload. |
Restart the acquisition controller and see what happens. If the problem is not remedied, reduce the system load, for example, by reducing the number of analytical instruments. |
12E5 |
The drop in HDD remaining space has been restored. |
----- |
----- |
12E6 |
The drop in remaining memory has been restored. |
----- |
----- |
12E7 |
The increase in the CPU load rate has been restored. |
----- |
----- |
12E8 |
System self-diagnostic has been performed. |
System self analysis was executed. |
When [NG] is indicated for the HDD, delete unwanted files. When [NG] is indicated for MEM, restart the acquisition controller and see what happens. If the problem is not remedied, reduce the system load, for example, by reducing the number of analytical instruments. When [NG] is indicated for the CPU, restart the acquisition controller and see what happens. If the problem is not remedied, reduce the system load, for example, by reducing the number of analytical instruments. |
12E9 |
The RAS service has been |
----- |
----- |
1301 |
The folder specified in the Data Explorer property does not exist. \nPlease change. |
The folder specified with the Data explorer property does not exist. |
Try again after changing the folder. |
1302 |
The specified folder does not |
The specified folder does not exist. |
Change the folder. |
1303 |
Are you sure you want to delete this folder? |
Displayed for confirmation when a folder is deleted. |
To delete, click [Yes], and to cancel deletion, click [No]. |
1304 |
At least a file exists in this project (folder). \nYou have to delete all contained files to delete the |
A file exists in this project folder. |
Delete the project after deleting all contained files. |
1305 |
This project (folder) already exists. |
The project folder of the same name has already existed. |
Enter the name of the other project (folder). |
1306 |
File is not found. |
Cannot find the specified file. |
Update to the latest information, or restart the application, and specify an existing file to execute. |
1307 |
Failed to open the file. |
The file cannot be opened for some reason. |
1. Try again after updating the information. 2. If the attempt fails again, the file may be corrupted. Execute again after restarting the application. |
1308 |
Invalid file type. |
The wrong type of file was specified. |
Check the file type; then select the correct type of file. |
1309 |
Failed to move the file._\nA file system error occurred. |
Failed to move the file. A file system error has occurred. |
Try again after restarting the application. |
130B |
This folder already has a file of the same name. \nAre you sure you want to overwrite with the new file? |
Displayed for confirming whether or not to overwrite files when a file of the same name exists in the selected folder. |
To overwrite, click [Yes], and to cancel overwriting, click [No]. |
130C |
Failed to copy the file. \nA file system error occurred. |
Failed to copy the file. A file system error has occurred. |
Try again after restarting the application. |
130D |
If you change a file name extension, the file may become unusable. \nAre you sure you want to change it? |
Displayed for confirmation when an extension is changed. |
To change, click [OK], and to cancel the change, click [Cancel]. |
130E |
Cannot rename the file. \nA file with the name you specified already exists. \nPlease specify another name. |
The file with the same name as that you specified already exists. |
Specify another name. |
130F |
Failed to rename the file. \nA file system error occurred. |
Cannot rename file. A file system error has occurred. |
Try again after restarting the application. |
1310 |
Send the file to the Recycle Bin? |
Displayed for confirmation when a file is moved to the recycle bin. |
To move, click [Yes], and to cancel moving, click [No]. |
1311 |
Are you sure you want to delete the file? |
Displayed for confirmation when a file is deleted. |
To delete, click [OK], and to cancel deletion, click [Cancel]. |
1312 |
Failed to delete the file. \nA file system error occurred. |
Failed to delete the file. A file system error has occurred. |
Try again after restarting the application. |
1313 |
Failed to search. |
Failed to search. |
1. Perform the search again. 2. If the error occurs again, restart the application, and execute again. |
1314 |
The number of records hit in the search exceeds 1000. Are you sure you want to display all? |
Displayed for confirmation since it may take time to display all results when the number of records hit in the search |
Click [OK] to display all search results, and click [Cancel] to cancel. |
1315 |
The registration to database was successfully completed. |
----- |
----- |
1316 |
Failed to register to database. |
The file cannot be registered to database for some reason. |
Try again after making sure that the network is connected correctly. |
1317 |
The specified file does not exist. |
Cannot find the specified file. |
Update to the latest information, or restart the application, and specify an existing file to execute. |
1318 |
The specified file does not exist. |
Cannot find the specified file. |
Update to the latest information, or restart the application, and specify an existing file to execute. |
131A |
Failed to register to database. \nPlease specify one or more meta files. |
Failed to register to database. |
Specify one or more meta files. |
131B |
Failed to import the specified file. |
Failed to import the specified file. |
Specify a file matching the file type to register. |
131C |
Failed to import the file. \nThe file size is over 1G byte. |
Failed to import the file. Files over 1G byte cannot be imported. |
Select a file of 1G byte or less. |
131D |
The file size is over 50 M byte. \nAre you sure you want to import it? |
Displayed for confirmation since it may take time to import and disk space may be taken up when a file is over 50 Mbytes. |
To import, click [OK], and to cancel the import, click [Cancel]. |
131E |
The file has been registered. |
----- |
----- |
131F |
The file has been deleted. |
----- |
----- |
1320 |
The file has been renamed. |
----- |
----- |
1321 |
The rollback has been |
----- |
----- |
1322 |
The file has been copied. |
----- |
----- |
1324 |
Cannot handle the file, because analysis is running. |
Cannot operate file because it is being used in analysis. |
Perform the operation after data analysis is finished. |
1325 |
Another user is using this file. \nAre you sure you want to force the operation? |
Displayed for confirming whether or not to forcibly operate when the selected file is in use by another user. When a file is forcibly operated, file operations by the current user will be aborted. |
Click [Yes] to forcibly perform the operation, and click [No] to cancel. |
1326 |
The specified file does not exist. |
Cannot find the specified file. |
Update to the latest information, or restart the application, and specify an existing file to execute. |
1327 |
Failed to restore to the latest version. |
Cannot restore the selected file as a latest version file. A probable cause is a problem in the network connection in the case of a network configuration. |
Close the sub-window, and perform the operation again. Try again after making sure that the network is connected correctly in the case of a network configuration. |
1328 |
Failed to import the file since the extension of the specified file is not registered in the extension definitions on the selected tab |
Cannot import the selected file since it does not match the extension currently set to the system. |
Check the relationship of the file set to the system, and select a file with the appropriate extension. |
1329 |
The specified file is not related to any applications. |
The specified file is not made relationship to any application. The appropriate application cannot be started up. |
Check the relationship of the file set to the system, and make the relationship as required. |
132A |
The file on editing or analysis will be forced to release. The user editing the file will unable to save it. Are you sure you want to continue? |
Displayed for confirming whether or not to forcibly enable use of the selected file. When a file is forcibly enabled for use, the new edits made by the current user can no longer be saved. |
Click [OK] to forcibly enable use, and click [Cancel] to cancel. |
132B |
The file has been released. |
Displayed when a currently edited file or file used for data acquisition is forcibly enabled for use. |
----- |
132C |
The file is already released. |
An attempt was made to forcibly enable a currently edited file or file used for data acquisition for use, however, the file has already been released. |
----- |
132D |
Failed to release the file. |
An attempt was made to forcibly enable a currently edited file or file used for data acquisition, however, it could not be |
Try again after restarting the application. |
132e |
A hidden tab page has been specified. |
The tab specified as displayed is currently hidden. |
Display the specified tab page in the Data explorer property sub-window. |
132f |
The specified software is not found. |
The specified software is not registered to the system. |
Specify the software registered to the system. |
1330 |
Failed to restore to the latest version since the signature locks |
The signature-locked file cannot be restored to the latest version from a past |
To restore past versions of signature-locked files, withdraw the signature. |
1331 |
You cannot register the file opened as database. Try to register again after closing the |
An attempt was made to register the currently open file to database. |
Try again after closing the file. |
1332 |
You cannot register the file used in current acquisition as database. |
An attempt was made to register the file currently used for data acquisition to database. |
Try again after data acquisition is finished. |
1333 |
You cannot register in the opened database file. Try to register again after closing the |
An attempt was made to register in the opened database file. |
Try again after closing the database file. |
1334 |
Failed to rename the file since the extension of the specified file is not registered in the extension definitions for the currently specified file type. |
The extension of the specified file is not registered in the extension definitions for the currently specified file type. |
Register the extension of the specified file to the extension definitions for the currently specified file type. |
1335 |
This file is already exist. \nAre you sure you want to overwrite with the new file? |
Displayed for confirming whether or not to overwrite files when a file of the same name exists. |
To overwrite, click [Yes], and to cancel overwriting, click [No]. |
1336 |
Cannot rename the file on |
|
|
1337 |
Failed to save the specified file. |
The specified file cannot be saved for some reason. |
Try again after restarting the application. |
1401 |
Failed to read from the database. |
Failed to read the database. |
Check the status of the database and the network connection status, and restart the application. |
1402 |
Failed to read from the backup database. \nThis file is not a backup database. |
Failed to read the backup database. The specified file is not a backup database. |
Specify the backup database that was backed up on this system. |
1403 |
Please input a password. |
----- |
----- |
1404 |
Password is not correct. \nPlease make sure the selection of backup database and input the correct password. \nThe password is case-sensitive. |
Password is not correct. |
Check the selected backup database and input the correct password. The password is case-sensitive. |
1405 |
Printing was abnormally |
Printing finished abnormally. |
Print again after checking the status of the specified |
1406 |
Printing was abnormally |
Printing finished abnormally. |
Print again after checking the status of the specified |
1407 |
The specified file is not related to any applications. |
The specified file is not made relationship to any application. The appropriate application cannot be started up. |
Check the relationship of the file set to the system, and make the relationship as required. |
1408 |
The specified data will be deleted. \nAre you sure you want to continue? |
Displayed for confirmation when the specified data is deleted. |
To delete, click [OK], and to cancel deletion, click [Cancel]. |
1409 |
Failed to copy the specified file. |
Failed to copy the specified file. |
Try again after restarting the application. |
140A |
Failed to move the specified file. |
Failed to move the specified file. |
Try again after restarting the application. |
140B |
Failed to delete the specified file. |
Failed to delete the specified file. |
Try again after restarting the application. |
140C |
Failed to search. |
Failed to search. |
1. Perform the search again. 2. If the error occurs again, restart the application, and execute again. |
140D |
The number of records hit in the search exceeds 1000. Are you sure you want to display all? |
Displayed for confirmation since it may take time to display all results when the number of records hit in the search |
Click [OK] to display all search results, and click [Cancel] to cancel. |
140E |
Failed to sign on the specified sample information. |
Failed to sign the specified sample information for some reason. |
Try again after restarting the application. |
1410 |
Failed to change the reference flag for the sample information. |
Failed to change the inspection flag of the sample information for some reason. |
Try again after restarting the application. |
1411 |
Failed to change the transmission flag for the sample |
Failed to change the upload flag of the sample information for some reason. |
Try again after restarting the application. |
1412 |
Failed to export the specified file. |
Failed to export the specified file for some reason. |
Try again after restarting the application. |
1413 |
The specified file will be target for signature. \nAre you sure to continue? |
Displayed for confirmation when the specified file is targeted for signing. |
To use the specified file for signing, click [Yes], and to cancel, click [No]. |
1414 |
Failed to import the specified file. |
The specified file cannot be imported for some reason. |
Try again after restarting the application. |
1437 |
The table information has been exported. |
'----- |
'----- |
1438 |
Failed to export the Table Information. |
The export file specified on the [Export Table Information] page in the [Option] sub-window does not create. |
Select the [Option] command on the [Tools] menu, and set the export file on the [Export Table Information] tab page in the [Option] sub-window. |
1439 |
The result file already has a file of the same name. You cannot overwrite the file. |
Cannot overwrite the file which overwrite is prohibited. |
Change the file name. |
1415 |
The specified file does not exist. |
Cannot find the specified file. |
The specified file could be deleted. Search again. |
1416 |
Failed to restore the specified file. |
Failed to restore the specified file. A probable cause is a problem in the network connection in the case of a network |
Close the sub-window, and perform the operation again. Try again after making sure that the network is connected correctly in the case of a network configuration. |
1417 |
Failed to read from the backup database. \nThe specified file could be corrupt. |
Failed to read the backup database. The specified file could be corrupt. |
If the error occurs even after you have checked the specified file again before executing it, execute using another file. |
1418 |
Failed to import the file. \nThe file size is over 1G byte. |
Failed to import the file. Files over 1 Gbyte cannot be registered. |
Select a file of 1 Gbyte or less. |
1419 |
The file size is over 50M bytes. \n Are you sure you want to import it? |
Displayed for confirmation since it may take time to import and disk space may be taken up when a file is over 50 Mbytes. |
To import, click [OK], and to cancel the import, click [Cancel]. |
141A |
The file has been registered. |
----- |
----- |
141B |
The result file is imported. |
----- |
----- |
141C |
The file has been copied. |
----- |
----- |
141D |
The file has been moved. |
----- |
----- |
141E |
The file has been deleted. |
----- |
----- |
141F |
The result file has been deleted. |
----- |
----- |
1420 |
The data set has been copied. |
----- |
----- |
1421 |
The data set has been moved. |
----- |
----- |
1422 |
The data set has been deleted. |
----- |
----- |
1423 |
The signature has been |
----- |
----- |
1424 |
The signature has not been accepted. |
----- |
----- |
1425 |
The signature has been |
----- |
----- |
1426 |
The Web reference flag has been updated. |
----- |
----- |
1427 |
The setting flag of transmission to host is updated. |
----- |
----- |
1428 |
The rollback has been |
----- |
----- |
1429 |
The number of the filtering conditions reaches the maximum limit. \nDo you delete other conditions to save this condition? |
Displayed for confirming whether or not to delete other filtering conditions when the number of saves for the filtering conditions reaches the upper limit. |
To delete other filtering conditions to save these conditions, click [OK], and to cancel saving, click [Cancel]. |
142A |
Cannot handle the file, because analysis is running. |
Cannot operate file because it is being used in analysis. |
Perform the operation after analysis is finished. |
142B |
Another user is using this file. \nAre you sure you want to force the operation? |
Displayed for confirming whether or not to forcibly operate when the selected file is in use by another user. When a file is forcibly operated, file operations by the current user will be aborted. |
Click [Yes] to forcibly perform the operation, and click [No] to cancel. |
142C |
Failed to acquire the list of printers currently registered to Windows on the client computer. |
The list of printers currently registered to Windows on the client computer cannot be acquired. |
Contact your Shimadzu representative. |
142D |
The specified project already has a file of the same name. Are you sure you want to overwrite with the new file? |
Displayed for confirming whether or not to overwrite files when a file of the same name exists in the specified project. |
To overwrite, click [Yes], and to cancel the overwrite, click [No]. |
142E |
Failed to register the file since the extension of the specified file is not registered in the extension definitions for the currently specified file type. |
Cannot register the selected file since it does not match the extension currently set to the system. |
Check the relationship of the file set to the system, and select a file with the appropriate extension. |
142F |
Sample information has been changed. |
----- |
----- |
1430 |
Data information has been |
----- |
----- |
1431 |
The specified file is already |
The same file cannot be added. |
Specify another file. |
1432 |
The result file already has a file of the same name. Are you sure you want to overwrite with the new file? |
Displayed for confirmation when the same file is added. |
----- |
1433 |
Files with the same name as that of the signature target file cannot be imported. |
Signature files of the same name cannot be imported. |
Specify another file. |
1434 |
This may take time to display. Are you sure you want to |
Displayed for confirmation when there are too much selected sample information. |
To continue, click [OK], and to cancel the operation, click [Cancel]. |
1435 |
The selected sample information is changed. Are you sure you want to continue? |
Displayed for confirmation when the selected sample information is changed. |
To continue the change, click [OK], and to cancel the change, click [Cancel]. |
1436 |
Save to follwoing folder. Are you sure? |
Displayed for confirmation when the selected file is exported. |
To save to the specified folder, click [YES], and to save to the arbitrary folder, click [NO]. |
1440 |
Failed to restore the specified file. The file does not exist. |
Cannot find the specified file. |
Update to the latest information, or restart the application, and specify an existing file to execute. |
1441 |
Failed to restore the specified file. The file is being used by other applications. |
The specified file is used by other applications. |
Wait a while and try again. |
1442 |
The number of thumbnail lists exceeds 50. Are you sure you want to display all? |
Displayed for confirmation since it may take time to display all results when the number of thumbnail lists exceeded 50. |
Click [Yes] to display all search results, and click [No] to cancel. |
1443 |
The number of thumbnail lists exceeds 100. Failed to display the thumbnail lists. |
Displayed this message when the number of thumbnail lists exceeded 100. |
Adjust the number of thumbnail lists to 100 or less. |
1444 |
The number of search result exceeds 10000. Please change the search condition. |
Displayed this message when the number of search result exceeds 10000. |
Please change the search condition. |
1445 |
The number of search result exceeds upper limit. Please change the search condition. |
Displayed this message when the number of search result exceeds upper limit. |
Please change the search condition. |
1446 |
Please set [Max.# to List] to 10000 or less. |
Displayed this message when [Max.# to List] exceeded 10000. |
Please set [Max.# to List] to 10000 or less. |
1447 |
The specified data set will be deleted. \nAre you sure you want to continue? |
Displayed for confirmation when the specified data set is deleted. |
The relevant all data file is deleted. To delete, click [OK], and to cancel deletion, click [Cancel]. |
1448 |
Failed to change the specified sample information. |
A probable cause is a problem in the database access or in the network connection. |
Make sure that the network is connected correctly and try again after a while. |
1449 |
Cannot perform, because the specified file is under edit. |
Cannnot perform while other user is editing the file or the file is edited in other |
Try again after other user has finished editing. |
144a |
All PDF files have been printed |
----- |
----- |
144b |
Failed to print out PDF files of following data files. |
Windows defalut printer is unusable or the data file does not have PDF file. |
Check the windows default printer setting and if the data file has PDF file. |
144c |
Failed to load the settings. |
A probable cause is a problem in the database access or in the network connection. |
Make sure that the network is connected correctly and try again after a while. |
144d |
Failed to create a report set. |
|
|
144e |
Failed to register a report set. |
A probable cause is a problem in the database access or in the network connection. |
Make sure that the network is connected correctly and try again after a while. |
144f |
Failed to create work directories to create a report set. |
|
Try again. If this does not remedy the problem, restart the computer, and try the operation again. |
1450 |
Failed to read the report set file. |
|
Try again. If this does not remedy the problem, restart the computer, and try the operation again. |
1451 |
|
|
|
1452 |
|
|
|
1453 |
|
|
|
1454 |
|
|
|
1455 |
|
|
|
1456 |
|
|
|
1457 |
|
|
|
1458 |
|
|
|
1459 |
|
|
|
145a |
|
|
|
145b |
|
|
|
145c |
|
|
|
145d |
|
|
|
145e |
|
|
|
145f |
|
|
|
1500 |
Insufficient database connections. |
Temporarily insufficient connections due to accesses concentrating on the database. |
Wait a while and try again. |
1501 |
Failed to access database. |
Failed to access the database. |
Perform the operation again after making sure that the network is connected correctly. If this does not remedy the problem, restart the computer, and try again. |
1502 |
Failed to create the folder. |
Failed to create the folder. |
Perform the operation again after making sure that the network is connected correctly. If this does not remedy the problem, restart the computer, and try again. |
1503 |
Failed to create the folder. |
Failed to create the folder. |
Perform the operation again after making sure that the network is connected correctly. If this does not remedy the problem, restart the computer, and try again. |
1504 |
Failed to open the file. |
Failed to open the file. |
Perform the operation again after making sure that the network is connected correctly. If this does not remedy the problem, restart the computer, and try again. |
1505 |
Failed to read the file. |
Failed to read the file. |
Perform the operation again after making sure that the network is connected correctly. If this does not remedy the problem, restart the computer, and try again. |
1506 |
Failed to write the file. |
Failed to write to the file. |
Perform the operation again after making sure that the network is connected correctly. If this does not remedy the problem, restart the computer, and try again. |
1507 |
Failed to delete the file. |
Failed to delete the file. |
Perform the operation again after making sure that the network is connected correctly. If this does not remedy the problem, restart the computer, and try again. |
1508 |
The same file exists. |
The file with the same name as that you specified already exists. |
Specify another file name. |
1509 |
Deleting this file is prohibited. |
An attempt was made to delete a file that cannot be deleted. |
----- |
150A |
Getting file list is not started. |
----- |
----- |
150B |
Start/End settings of a getting range are not correct. |
Specification of an acquisition start/end range is invalid. |
Try again after making sure that the appropriate acquisition range is specified. |
150C |
File list information exists after specified end point. |
This indicates that there is also a matching file beyond the specified range. |
This need not be remedied since it is internally processed. |
150D |
The file does not exist. |
The specified file does not exist. |
Update to the latest information, or restart the application, and specify an existing file to execute. |
150E |
The file is already in use. |
The specified file is already in use. |
Specify again after another user has returned the file. |
150F |
The file is already in use for analysis. |
The specified file is already in use for analysis. The file cannot be in use since it is being used in analysis. |
Specify again after analysis is finished. |
1510 |
The using information is not in agreement. |
The using information of the specified file is not in agreement. The information cannot be saved to this file since it has been forcibly used by another user. |
Close the file without saving the information, or rename and save the file. To edit this file, open the file again in the Edit mode. |
1511 |
The using information has not been taken. |
The using information of the specified file is not in agreement. Cannot save the information. |
Close the file without saving the information, or rename and save the file. To edit this file, open the file again in the Edit mode. |
1512 |
It is going to use the old edition file to edit. |
Only the latest file of history-managed files can be edited. |
Update to the latest information, or restart the application, and specify the file again to execute. |
1513 |
Non-registered data exists. |
The project cannot be deleted if it contains unregistered data when the project is to be deleted. A probable cause is a problem in the network connection between the acquisition controller and the database |
Try again after checking the network connection. |
1514 |
Under file access. |
The project cannot be deleted if a file for that project is currently open when the project is to be deleted. |
Close the file currently being edited, and try again. |
1515 |
The file has already existed. |
The file name specified already exists on the drive. |
Specify a different file; then execute the command again. |
1517 |
The version administration methods do not match between projects. |
An attempt was made to operate across 2 projects. This operation is not available when each of the version administration settings does not match. |
Perform the operation after matching the version management settings of the project policy. |
1518 |
The specified profile data does not exist. |
----- |
----- |
1519 |
The latest version exists in the database. \nDo you use the latest version? |
Displayed for confirmation when the database contains a latest version file. |
To use the latest version file, click [Yes], and to cancel, click [No]. |
151A |
The file has been registered. |
----- |
----- |
151B |
The file has been deleted. |
----- |
----- |
151C |
The file has been renamed. |
----- |
----- |
151D |
Non-registered data has been registered into automatic registration queue. |
----- |
----- |
151E |
The signature is locked. |
The specified file is signature-locked. |
Operations cannot be performed on signature-locked files. |
151F |
Initializing. |
The application is being initialized. |
Wait a while and try again. |
1520 |
Cannot sign since the file is not the latest. |
The specified file cannot be signed since it is not the latest. |
Perform signature operations on the latest file. |
1521 |
Using the specified file is not required. |
Internal code |
Internal code |
1522 |
The project and instrument combination does not match. |
The combination of instruments does not match the project specified from the application. |
A probable cause is that the relationship between the project and the instrument is deleted. Check in the project management sub-window. |
1523 |
The file is already in use to the same user. |
The specified file is already in use by the same user. |
----- |
1524 |
|
|
|
1525 |
|
|
|
1526 |
|
|
|
1550 |
The MDB file is not found in the specified location. \nDo you copy the MDB file to the specified location? |
Displayed for confirmation when the MDB file cannot be found at the specified location. |
To copy the MDB file currently in use to the specified location, click [OK]. To cancel, click [Cancel]. |
1551 |
No rights to set_up MDB. |
A user without the [System Administration] rights attempted to make a setting. |
Log in as a user with the [System Administration] rights. |
1552 |
Failed to optimize/repair. |
Failed to optimize/repair. |
If this error occurs after executing again, the MDB file cannot be repaired. The MDB file can be made again by deleting the MDB file and restarting the system. |
1553 |
Stops changes to the MDB. |
----- |
----- |
1554 |
Use of the database has been started. |
----- |
----- |
1555 |
This operation is not available since data acquisition is in progress. To perform this operation, end all data |
This operation is not available since data acquisition is in progress. |
To perform this operation, end all data acquisitions. |
1556 |
Failed to start the service. \nPlease call a system |
Failed to start the service. |
If the error occurs even after specifying another MDB file and restarting, contact your Shimadzu representative. |
1557 |
Failed to start the MDB configuration. |
Failed to start the MDB configuration. |
Contact your Shimadzu representative. |
1558 |
Failed to copy the MDB file. |
Failed to copy the MDB file. |
Copy again. If this does not remedy the problem, restart the computer, and try setting again. |
1559 |
Failed to change the MDB. |
Failed to change the MDB. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. |
1560 |
The specified file is a certification DB of old form. \n It is necessary to extend it to a certification DB of new form. Extend it? OK? |
The specified file is a certification DB of old form. |
Extend a certification DB of old form to a certification DB of new form,click [OK].To cancel click [Cancel]. |
1561 |
Failed to extend a certification |
Failed to extend a certification DB. |
Contact your Shimadzu representative. |
1570 |
MSDE is not installed. |
MSDE is not installed. |
Try again after installing MSDE. |
1571 |
Please specify the physical path. |
The physical path is not specified. |
Specify the physical path. |
1572 |
The database already exists. \nAre you sure you want to delete the existing database and create |
Displayed for confirming whether or not to create a new database when a database is already created. |
To delete the existing database and create a new database, click [OK]. To cancel, click [Cancel]. |
1573 |
Database creation process will be started. \nAre you sure you want to continue? |
Displayed for confirmation when creating a database. |
To create the database, click [OK], and to cancel creation, click [Cancel]. |
1574 |
Failed to create the system administration database. |
The system administration information database cannot be created for some |
Contact your Shimadzu representative. |
1575 |
Failed to create the Log database. |
The log database cannot be created for some reason. |
Contact your Shimadzu representative. |
1576 |
Failed to create the Project database. |
The project database cannot be created for some reason. |
Contact your Shimadzu representative. |
1577 |
Failed to create the Profile database. |
The profile database cannot be created for some reason. |
Contact your Shimadzu representative. |
1578 |
Creation of a database has been completed successfully. |
----- |
----- |
1579 |
No rights to create MSDE database. |
A user without the [System Administration] rights attempted to make a setting. |
Log in as a user with the [System Administration] rights. |
157A |
Please specify the Windows domain name. |
The Windows domain name is not specified. |
Specify the Windows domain name. |
157B |
An error occurred. The [Create SQL Server Database] subwindow cannot be opened. |
Failed to open the [Create SQL Server Database] sub-window. |
Solve the problem relating to the sub-message, and set again. If this does not remedy the problem, restart the computer, and try setting again. |
157C |
Failed to connect to the existing authentication database. |
Failed to connect to the existing authentication database. |
Try again after checking the status of the existing authentication database and after making sure that the network is connected correctly. |
157D |
Both for the project and for the system administration/user authentication is already made. Cannot be made again individually since a data |
----- |
----- |
157E |
The selected database does not exist. |
The selected database does not exist. |
Select the appropriate service name. |
157F |
The specified folder does not exist. |
The folder specified to the physical path does not exist. |
Make the folder specified to the physical path. |
1580 |
Operation has been canceled. The database creation process is being terminated. |
This is displayed when [Cancel] was selected in the password entry subwindow. |
----- |
1581 |
The existing user authentication database is already expanded. It cannot be created again. |
The existing user authentication database has already been expanded. It cannot be created again. |
Try again after re-creating from the user authentication tool. |
1582 |
Failed to delete [managed folder of data files]. |
Failed to delete [managed folder of data files] for some reason. |
Contact your Shimadzu representative. |
1590 |
Oracle is not installed. |
Oracle is not installed. |
Try again after installing Oracle. |
1591 |
Please input maximum database size. |
The maximum database size is not entered. |
Enter the maximum database size. |
1592 |
Please select the drive, which creates the System-table space of Oracle. |
The drive which creates the system-table space of Oracle is not selected. |
Choose the drive which creates the system-table space of Oracle. |
1593 |
Please select an instance name (SID). |
The service name is not selected. |
Choose an instance name (SID). |
1594 |
Please select a Windows domain name. |
The Windows domain name is not selected. |
Choose a Windows domain name. |
1595 |
The database already exists. \nAre you sure you want to delete the existing database and create |
Displayed for confirming whether or not to create a new database when a database is already created. |
To delete the existing database and create a new database, click [OK]. To cancel, click [Cancel]. |
1596 |
Maximum database size is over the capacity of the logical HD |
Maximum database size is over the capacity of the logical HD drive. |
Change the maximum size of the database to a size that does not exceed free space on the logical HD drive. |
1597 |
Database creation process will be started. \nAre you sure you want to continue? |
Displayed for confirmation when creating a database. |
To create the database, click [OK], and to cancel, click [Cancel]. |
1598 |
Failed to generate the instance. |
Instances cannot be generated for some reason. |
Contact your Shimadzu representative. |
1599 |
Failed to create the system administration database. |
The system administration information database cannot be created for some |
Contact your Shimadzu representative. |
159A |
Failed to create the Log database. |
The log database cannot be created for some reason. |
Contact your Shimadzu representative. |
159B |
Failed to create the Project database. |
The project database cannot be created for some reason. |
Contact your Shimadzu representative. |
159C |
Failed to create the Profile database. |
The profile database cannot be created for some reason. |
Contact your Shimadzu representative. |
159D |
Creation of a database has been completed successfully. |
----- |
----- |
159E |
No rights to create Oracle database. |
A user without the [System Administration] rights attempted to make a setting. |
Log in as a user with the [System Administration] rights. |
159F |
An error occurred. The [Create Oracle Database] sub-window cannot be opened. |
The [Create Oracle Database] sub-window cannot be opened. |
Solve the problem relating to the sub-message, and set again. If this does not remedy the problem, restart the computer, and try setting again. |
15A0 |
Failed to connect to the existing authentication database. |
Failed to connect to the existing authentication database. |
Try again after checking the status of the existing authentication database and after making sure that the network is connected correctly. |
15A1 |
Please make sure that a version of Oracle compatible with the tool is installed. |
A version of Oracle compatible with this tool is not installed. This tool supports Oracle 10 or more. |
Make sure that a version of Oracle compatible with the tool is installed. |
15A2 |
Both for the project and for the system administration/user authentication is already made. Cannot be made again individually since a data |
----- |
----- |
15A3 |
The selected service name is not for an existing database. |
The selected service name is not an existing database. |
Select the appropriate service name. |
15A4 |
Failed to re-create the database since a user is currently logged in. Please make sure that the LSS service or acquisition server is not running. |
Cannot re-create the database since a user is currently logged in. |
Create the database again after making sure that the LSS service or acquisition server is not started up, and after stopping the service and server if they are started. |
15A5 |
The existing user authentication database is already expanded. It cannot be created again. |
The existing user authentication database has already been expanded. It cannot be created again. |
Try again after re-creating from the user authentication tool. |
15A6 |
Failed to delete [managed folder of data files]. |
Failed to delete [managed folder of data files] for some reason. |
Contact your Shimadzu representative. |
15B0 |
The connection test has succeeded. |
----- |
----- |
15B1 |
Failed to connect to the database. |
Cannot connect to the database server. |
Try again after making sure that the database has started up normally, and that the network is connected correctly. |
15B2 |
Connection with the database server has succeeded. \nConnection with existing authentication database server |
----- |
----- |
15B3 |
Connection with the database server has succeeded. \nHowever, connection with existing authentication database server has failed. |
Could connect to the database server. Cannot connect to the existing authentication database server. |
Try again after making sure that the existing authentication database has started up normally, and that the network is connected correctly. |
15B4 |
Connection with existing authentication database server has succeeded._\nHowever, connection with the database server has failed. |
Cannot connect to the database server. Could connect to the existing authentication database server. |
Try again after making sure that the database has started up normally, and that the network is connected correctly. |
15B5 |
Connection with the database_server has failed. \nConnection with existing authentication database server either. |
Cannot connect to the database server. Cannot connect to the existing authentication database server, either. |
Try again after making sure that the database server and the existing authentication database hasstarted up normally, and that the network is connected correctly. |
15B6 |
The connection test with the LabSolutions server has succeeded. |
----- |
----- |
15B7 |
The connection test with the LabSolutions server has failed. |
The connection test with the LabSolutions server has failed. |
Try again after making sure that the network is connected correctly. |
15B8 |
No rights to set up database connection. |
A user without the [System Administration] rights attempted to make a setting. |
Log in as a user with the [System Administration] rights. |
15B9 |
Successfully connected to the existing authentication database server. |
----- |
----- |
15BA |
Failed to connect to the existing authentication database server. |
Failed to connect to the existing authentication database server. |
Try again after making sure that the existing authentication database has started up normally, and that the network is connected correctly. |
15BB |
Setting of the LabSolutions server will be canceled. |
----- |
----- |
15BC |
Setting of the database server will be canceled. |
----- |
----- |
15BD |
Use of the LabSolutions server has been started. |
----- |
----- |
15BE |
Use of the existing authentication database server has been |
----- |
----- |
15BF |
The existing authentication database server has been changed. |
----- |
----- |
15C0 |
Use of the existing authentication database server has been canceled. |
----- |
----- |
15C1 |
Failed to start up [Database Connection Settings]. |
[Database Connection Settings] could not be started up for some reason. |
Contact your Shimadzu representative. |
15C2 |
[Database Connection Settings] is not available. |
The new settings cannot be saved. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. |
15D0 |
Please select backup folder. |
The backup folder is not selected. |
Select a backup folder. |
15D1 |
Please select the project to be backed up. |
The project to be backed up is not selected. |
Select the project to be backed up. |
15D2 |
Please input the reason to backup. \n(A password is required at the time of ER/ES |
The reason for the backup has not been entered. |
Input the reason for backup. (A reason is required at the time of ER/ES mode.) |
15D3 |
A database already exists. \nPlease change the folder to |
The selected storage destination folder already has a database. |
Change the storage destination to a folder not containing a database. |
15D4 |
The input capacity of a folder is not appropriate. \nPlease input appropriate capacity. |
The value set at [Disk Space per a Folder] is invalid. |
Enter again. |
15D5 |
Please input a backup start and end date correctly. |
The backup start date and an end date have not been entered correctly. |
Set a backup start date and an end date correctly. |
15D6 |
Please enter a password. \n(A password is required at the time of ER/ES mode) |
The password has not been entered. |
Enter a password. (A password is required at the time of ER/ES mode.) |
15D7 |
The password confirmation must match. |
Password is not correct. |
Enter the correct password for the confirmation password. |
15D8 |
Please specify a temporary folder for backup. |
The backup temporary folder is not selected. |
Select a backup temporary folder. |
15D9 |
An error occurred while making the temporary backup file. |
The backup temporary file cannot be created. |
Make sure that the temporary folder at the backup destination is not write protected, and that there is |
15DA |
Backup has been completed successfully. |
----- |
----- |
15DB |
Backup will be terminated. \nAre you sure you want to terminate? |
Displayed for confirmation when cancellation of the backup is selected. |
To cancel, click [Yes], and to continue, click [No]. |
15DC |
An error occurred during backup process. |
The backup file cannot be created. |
Make sure that the folder at the backup destination is not write protected, and that there is sufficient disk space. |
15DD |
Please insert the next media. |
Message for prompting replacement with the next storage media when the backup to the current media is completed. |
Replace with the next storage media, and click [OK]. |
15DF |
Backup has been completed successfully. \nPlease keep the printed result with the media. |
----- |
----- |
15E0 |
Backup was abnormally terminated. |
The backup ended abnormally for some reason. |
Back up again. If this does not remedy the problem, restart the computer, and try again. |
15E1 |
Please insert a media in a drive. |
A disk is not inserted in the drive. |
Insert a disk in the drive. |
15E2 |
No rights to perform the backup. |
A user without the [System Administration] rights attempted to make a setting. |
Log in as a user with the [System Administration] rights. |
15E3 |
Backup/Restore is already running from other PCs. |
Cannot be backed up or restored when backup/restoration are performed from another client computer. |
Try again after backup/restoration on another client PC is finished. |
15E4 |
Please select the source database to restore. |
The restoring source database is not selected. |
Select the restoring source database. |
15E5 |
Please enter a password. \n(A password is required at the time of ER/ES mode) |
The password has not been entered. |
Enter a password. (A password is required at the time of ER/ES mode.) |
15E6 |
The source database to restore cannot be opened. \nThe password is invalid. |
The restoring source database cannot be opened. Password is not correct. |
Enter the password that was set when the backup was made. |
15E7 |
Please input the reason to restore. \n(A password is required at the time of ER/ES |
The reason for the restoration has not been entered. |
Input the reason for restoration. (A reason is required at the time of ER/ES mode.) |
15E8 |
Restore may take from several minutes to several hours. Are you sure you want to start Restore? |
Restoration may take from several minutes to several hours. Displayed for confirming whether or not to start restoration. |
To restore, click [OK], and to cancel, click [Cancel]. |
15E9 |
Restore will be terminated. \nAre you sure you want to terminate? |
Displayed for confirmation when cancellation of the restoration is selected. |
To cancel, click [OK], and to continue, click [Cancel]. |
15EA |
Please insert the next media. |
Message for prompting replacement with the next storage media when the restoration from the current media is |
Replace with the next storage media, and click [OK]. |
15EB |
Restore has been completed successfully. |
----- |
----- |
15EC |
Restore was abnormally terminated. |
The restoration ended abnormally for some reason. |
Restore again. If this does not remedy the problem, restart the computer, and try again. |
15ED |
No rights to perform the Restore. |
A user without the [System Administration] rights attempted to make a setting. |
Log in as a user with the [System Administration] rights. |
15EE |
Checksum does not match. |
Invalid checksum. The data backup could be corrupt. |
Try again after checking the backup MDB. |
15EF |
Failed to backup the files since there is a file larger than the split size. |
There is a file larger than the split size of the backup folder. |
Try again after increasing the split size of the backup folder. |
15F0 |
Failed to open backup MDB. |
The backup MDB cannot be opened for some reason. |
Try again after checking the backup MDB. |
15F1 |
The corresponding table is not found in backup MDB. |
Failed to operate since the corresponding table could not be found in the backup |
Try again after checking the backup MDB. |
15F2 |
Backup may take from several minutes to several hours. Are you sure you want to start backup? |
Backup may take from several minutes to several hours. Displayed for confirming whether or not to start backup. |
To back up, click [OK], and to cancel, click [Cancel]. |
15F3 |
An error occurred during copying to the removable storage media. |
Cannot write the file to the removable storage media. |
Make sure that the removable storage media is not write protected, and that there is sufficient disk space. |
15F4 |
The data file has been deleted after it was backed up. |
----- |
----- |
15F5 |
A temporary folder will be created. Do you continue? |
Displayed for confirmation when creating a temporary folder. |
To create a temporary folder, click [OK], and to cancel, click [Cancel]. |
15F6 |
Failed to create the temporary folder. |
Failed to create the temporary folder. |
Try again after making sure that the folder is not writeprotected. |
15F7 |
The backup was terminated. |
----- |
----- |
15F8 |
The Restore was terminated. |
----- |
----- |
15F9 |
Too many target items. Please specify the period again. |
The number of logs to be backed up is too many. |
Try again after specifying the backup conditions again. |
15FA |
It is not possible to back up directly to the CD/DVD drive. |
----- |
----- |
1601 |
Data file name already exists in batch file registered into batch queue, or the batch file under running. |
Data file name already exists in batch file registered into batch queue, or the batch file under run. |
Specify another batch file name. |
1605 |
Failed to create the database file. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1606 |
Failed to opening the database file. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
160F |
Registration to batch queue has been completed successfully. |
----- |
----- |
1610 |
Failed to register to the batch queue because of the communication failure. |
Cannot register to the batch queue because of the communication disturbance. |
Try again after making sure that the network is connected correctly. |
1611 |
Cannot register batch queue, because of the instrument occupation. |
Cannot register batch queue, because of the instrument occupancy. |
Register after canceling instrument occupancy. Occupancy can be forcibly canceled by users with rights. |
1612 |
Failed to move the batch file because of the communication failure. |
Cannot move the batch file because of the communication disturbance. |
Try again after making sure that the network is connected correctly. |
1613 |
Specified batch file does not |
Specified batch file does not exist. |
Select the batch again after checking the latest |
1614 |
Cannot move batch file. |
Cannot move batch file. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1615 |
Failed to start the batch processing because of the communication failure. |
Cannot start the batch processing because of the communication obstacle. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1616 |
Specified batch file does not |
Specified batch file does not exist. |
Select the batch again after checking the latest |
1617 |
Cannot start batch. |
Batch cannot be started for some reason. |
Close the sub-window, and perform the operation again. Try again after making sure that the network is connected correctly in the case of a network configuration. |
1619 |
Failed to edit wait setup, because of the communication failure. |
Cannot edit wait setup, because of the communication obstacle. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
161A |
The specified batch file does not exist. |
Specified batch file does not exist. |
Select the batch again after checking the latest information. |
161B |
Cannot edit a wait setup. |
The waiting setup cannot be changed for some reason. |
Close the sub-window, and perform the operation again. Try again after making sure that the network is connected correctly in the case of a network configuration. |
161C |
Failed to edit batch file, because of the communication failure. |
Cannot edit batch file because of the communication failure. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
161D |
The specified batch file does not exist. |
Specified batch file does not exist. |
Select the batch again after checking the latest information. |
161E |
Cannot edit batch file. |
The batch file cannot be edited for some reason. |
Close the sub-window, and perform the operation again. Try again after making sure that the network is connected correctly in the case of a network configuration. |
161F |
Cannot edit batch file of the property of another project. |
Failed to edit the batch file of another project. |
Select a project registered with a batch file, and perform the operation. |
1620 |
Cannot edit batch file of the property of another user. |
----- |
----- |
1621 |
The batch file is already in use. |
The batch file is already in use. |
Try again after another user has finished editing the batch |
1622 |
The using information is not in agreement. |
The using information of the file for the specified batch is not in agreement. The information cannot be saved to this batch since it has been forcibly lent by another user. |
Close the batch without saving the information, or rename and save the batch. To edit this batch, try opening the batch again. |
1623 |
Failed to delete batch file because of the communication failure. |
Cannot delete batch file because of the communication disturbance. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1624 |
The specified batch file does not exist. |
Specified batch file does not exist. |
Select the batch again after checking the latest information. |
1625 |
Failed to delete the batch file. |
The batch file cannot be deleted for some reason. |
Close the sub-window, and perform the operation again. Try again after making sure that the network is connected correctly in the case of a network configuration. |
1626 |
Failed to take over analysis, because of the communication failure. |
Cannot take over analysis, because of the communication obstacle. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1627 |
The specified batch file does not exist. |
Specified batch file does not exist. |
Select the batch again after checking the latest information. |
1628 |
Cannot take over analysis. |
Cannot take over an analysis operator for some reason. |
Close the sub-window, and perform the operation again. Try again after making sure that the network is connected correctly in the case of a network configuration. |
1629 |
Cannot take over batch file of the property of another project. |
Cannot take over batch file of the property of another project. |
Select a project registered with a batch file, and perform the operation. |
162A |
Failed to get the newest information of the batch queue because of the communication failure. |
Cannot get the newest information of the batch queue because of the communication disturbance. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
162C |
Cannot remove instrument from a project under analysis. |
Cannot delete instrument from a project under analysis. |
Try again after analysis is finished. |
162D |
Cannot remove instrument from a project under analysis |
Cannot delete instrument from a project under analysis reservation. |
Try again after analysis is finished or after deleting analysis reservation. |
162E |
The definition file of operation does not exist. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
162F |
Cannot interrupt the batch under running, which has a summary report, QA/QC, or bracketing. |
----- |
----- |
1630 |
Priority run has been registered. |
----- |
----- |
1632 |
Failed to start batch processing. |
Failed to start batch processing. |
Execute the batch again after checking the instrument |
1633 |
Failed to acquire the Batch Table. |
Failed to acquire the Batch Table. |
Make sure editing is possible, and try again. If the message is displayed again, contact your Shimadzu |
1634 |
Failed to register the Batch Table. |
Failed to register to the Batch Table. |
Make sure editing is possible, and try again. If the message is displayed again, contact your Shimadzu |
1635 |
Failed to take over an analysis operator. |
Failed to take over an analysis operator. |
Make sure analysis can be taken over, and try again. |
1636 |
The batch table under the edit will be void. Are you sure you want to continue? |
An attempt was made to start a batch that is currently edited for a reserved batch. |
If it is OK to invalidate the Batch Table currently being edited, click [OK]. If it is not OK, click [Cancel]. |
1637 |
Cannot save the batch file because the batch processing is |
Cannot change the batch queue once batch processing has started. |
Save the edited batch under a different name, or discard the batch file. |
1638 |
Any priority run cannot be registered while any batch analysis with bracket setting is |
Cannot set priority run since bracket is set to the currently executing batch. |
----- |
1639 |
Any priority run cannot be registered while any priority run is running. |
Cannot interrupt a batch registered with priority run and the batch which became the waiting for resumption by interruption. |
----- |
163A |
Any priority run cannot be registered while any batch analysis with summary report setting is running. |
Cannot set priority run since summary report is set to the currently executing batch. |
----- |
163B |
Any priority run cannot be registered while any batch analysis with QA/QC setting is |
Cannot set priority run since QA/QC is set to the currently executing batch. |
----- |
163C |
Any priority run cannot be registered while any batch analysis with startup setting is |
Cannot set priority run since startup is set to the currently executing batch. |
----- |
163D |
Any priority run cannot be registered while any batch analysis with shutdown setting is |
Cannot set priority run since shutdown is set to the currently executing batch. |
----- |
163E |
Any priority run cannot be registered. |
An attempt was made to set priority run to a batch that does not accept priority run. If you don't have rights for the priority run, you cannot register the priority run against the analysis started by another user. |
----- |
163F |
Any priority run cannot be registered while any batch analysis with action setting is |
Cannot set priority run since action is set to the currently executing batch. |
----- |
1640 |
The batch cannot be deleted because the batch file is under the edit. |
The batch cannot be deleted when the batch file is being edited. |
----- |
1641 |
The batch queue status has become 'Waiting'. |
The state of the batch registered to the batch queue changed to a waiting state. |
----- |
1642 |
Batch has been removed from the batch queue. |
----- |
----- |
1643 |
An analysis has been taken over. |
----- |
----- |
1644 |
An analysis is taking over. Wait for a while. |
An analysis is taking over. |
Wait for a while. |
1645 |
Any priority run cannot be registered while any batch analysis with multi-data report is |
The priority run cannot be registered because the batch analysis with multi-data repor is running. |
----- |
1646 |
Any priority run cannot be registered while any batch analysis with acquisition cycle time tuning is running. |
The priority run cannot be registered because the batch analysis with acquisition cycle time tuning is running. |
----- |
1680 |
Failed to lock the screen. |
The screen cannot be locked for some reason. |
Try again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1682 |
Another user has released the screen lock. |
----- |
----- |
1683 |
Failed to release the screen lock. |
The screen cannot be unlocked for some reason. |
Try again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1684 |
Failed to get the automatic lock information. |
----- |
----- |
1685 |
Failed to register the window. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1686 |
Failed to cancel the registration of the windows. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1687 |
Cannot refer to rights of instrument control. |
The control rights of the instrument cannot be referenced for some reason. |
Try again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1688 |
Cannot get rights of instrument control. |
The control rights of the instrument cannot be acquired for some reason. |
Try again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1689 |
Cannot move rights of instrument control. |
The control rights of the instrument cannot be moved for some reason. |
Try again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
168A |
Cannot release rights of instrument control. |
The control rights of the instrument cannot be released for some reason. |
Try again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
168B |
Other users are controlling. |
The selected instrument cannot be controlled since it is controlled by another |
Try again after another user has released the control rights. |
168C |
Cannot refer to rights of instrument occupation. |
The instrument occupancy rights cannot be referenced for some reason. |
Try again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
168D |
Cannot get rights of instrument occupation. |
The instrument occupancy rights cannot be acquired for some reason. |
Try again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
168E |
Cannot release rights of instrument occupation. |
The instrument occupancy rights cannot be released for some reason. |
Try again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
168F |
The instruments are occupied by other users. |
The selected instrument cannot be controlled since it is occupied by another user. |
Try again after another user has released the occupancy. Occupancy can be forcibly released of you have [Release Occupied Instrument] rights. |
1692 |
Cannot get rights of instrument control from other projects. |
Cannot obtain the instrument control rights because the projects are different. |
To acquire control rights, perform the operation again after selecting a project that currently has control rights. |
1693 |
Cannot release, because of no rights of instrument control. |
This is displayed when an attempt was made to release instrument control rights even though you do not have these rights. |
The release operation is not required since control rights are already released. |
1694 |
The same user keeps the rights of instrument control on other |
The same user gets control rights on other PCs. |
To acquire the control rights, perform the operation again after releasing the control rights owned on the other |
1695 |
The rights of instruments control are acquired by the same user. |
----- |
----- |
1696 |
No rights to unlock the screen. |
A user without the [Release Windows Lock] rights attempted to make a setting. |
Log in as a user with the [Release Windows Lock] rights. |
1697 |
Notification of forced logout failed on an instrument(s). |
Notification of forced logout failed on an instrument(s). |
Try again after checking the instrument status. |
1698 |
Window lock was executed. |
----- |
----- |
1699 |
Window unlock was executed. |
----- |
----- |
1700 |
The specified instrument does not exist. |
The specified instrument does not exist. |
Contact your Shimadzu representative. |
1701 |
The process has been forced to terminate. |
----- |
----- |
1702 |
Failed to start the service. |
Failed to start the service. |
Contact your Shimadzu representative. |
1703 |
In current state, cannot start service. |
The present state cannot start service. |
Contact your Shimadzu representative. |
1704 |
Failed to stop the service. |
Failed to stop the service. |
Contact your Shimadzu representative. |
1705 |
In current state, cannot stop service. |
The present state cannot stop service. |
Contact your Shimadzu representative. |
1706 |
Failed to pause the distribution control process. |
Failed to pause the distribution control process. |
Contact your Shimadzu representative. |
1707 |
Failed to restart the distribution control process. |
Failed to restart the distribution control process. |
Contact your Shimadzu representative. |
1708 |
Time-out |
Time-out occurred in processing that starts/ends a process. |
----- |
1709 |
In current state, cannot perform distribution. |
Data acquisition is in process of ending. |
Wait a while before trying again. |
1710 |
The process cannot accept requests. |
The acq server is busy. |
Wait a while before trying again. |
1730 |
Failed to transfer the file. |
The file cannot be transferred for some reason. |
Try again after making sure that the network is connected correctly. |
1761 |
Failed to retrieve the list of the printers. |
Failed to retrieve the list of the printers. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1762 |
Failed to retrieve the list of the printers. |
Failed to retrieve the list of the printers. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1763 |
Failed to set_up printing. |
Failed to make print settings. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1764 |
Failed to display the propertysetting dialog. |
Cannot display the property setting dialog. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1765 |
Failed to access the PDF file. |
Cannot access PDF file. |
Contact your Shimadzu representative. |
1766 |
The specified printer is not found. |
The specified printer is not found. |
Specify the printer again. If this does not remedy the problem, restart the computer, and try setting again. |
1767 |
Failed to set_up printing. |
Failed to make print settings. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1768 |
Failed to transfer the PDF file. |
Failed to transfer the PDF file. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1769 |
Failed to create the LabSolutions printers list. |
Failed to create the LabSolutions printer list. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
176A |
Failed to retrieve the LabSolutions printer list. |
Failed to retrieve the LabSolutions printer list. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
176B |
Failed to refresh the LabSolutions printer list. |
Failed to refresh the LabSolutions printer list. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
176C |
No printer is registered. |
The printer is not registered. |
Register the printer to perform printing. |
176D |
Failed to get the printer. |
Failed to obtain the printer. |
Check the printer settings in the OS. |
176E |
No rights of printer administration. |
A user without the [System Administration] rights attempted to make a setting. |
Log in as a user with the [System Administration] rights. |
1801 |
Failed to create the Log database file. |
Log database file cannot be created. |
Try again after restarting the PC. If the attempt fails again, the file or disk may be corrupted. |
1802 |
Failed to open the Log database file. |
This file has been used on the other application. If the file is not in use, this may be corrupted. |
Try again after terminating the application in which the file is used. If the attempt fails again, the file or disk may be corrupted. |
1803 |
Log database password is invalid. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1804 |
The Log database file is already opened. |
This file has been used on the other application. |
Try again after terminating the application in which the file is used. |
1805 |
Failed to close the Log database file. |
Error on closing the log database file. |
Try again after restarting the PC. If the attempt fails again, the file or disk may be corrupted. |
1806 |
Failed to read/write the Log database file. |
Failed to process of log database file. |
Try again after restarting the PC. If the attempt fails again, the file or disk may be corrupted. |
1807 |
The Log database file is not opened. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1808 |
Failed to delete the log. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1809 |
Delete all event logs. Are you sure you want to continue? |
This message confirms that the log will be deleted by selecting Log Delete button. |
Click [Yes] to delete the log, otherwise click [No]. |
180A |
Cannot search because of the communication failure. |
Failed to search. |
1. Perform the search again. 2. If the error occurs again, restart the application, and execute again. Try again after making sure that the network is connected correctly in the case of a network configuration. |
180B |
Too many characters. |
Exceeds the number of characters which can be entered. |
Try again after checking the maximum number of characters. The maximum number of characters can be verified by Help in each window. |
180C |
Are you sure you want to delete? |
Displayed for confirmation when performing deletion. |
To delete, click [OK], and to cancel, click [Cancel]. |
180D |
Failed to delete. |
Failed to delete. A probable cause is a problem in the network connection in the case of a network configuration. |
Delete again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
180E |
Invalid type of Log database. |
A file that is not the log database of this system was selected. |
Select the appropriate log database. |
180F |
Audit Trail is not set up. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1810 |
Password is not correct. \nPlease make sure the selection of backup database and input the correct password. \nThe password is case-sensitive. |
Password is not correct. |
Check the selected backup database and input the correct password. The password is case-sensitive. |
1811 |
The number of records hit in the search exceeds limit. \nAre you sure you want to display all? |
Displayed for confirmation when the search results exceed the certain number. |
Click [Yes] to continue display, and click [No] to cancel. |
1812 |
The number of records hit in the search exceeds limit to display. Please specify appropriate condition for the filtering. |
Displayed for confirmation when the search results exceeded the display maximum number. |
Specify the filtering conditions again to reduce the number of search results, and try again. |
1813 |
Log has been deleted. |
----- |
----- |
1814 |
Log has been saved as a file. |
----- |
----- |
1902 |
Failed to run the application or open the file. |
The specified license does not exist, or a probable cause is that the software has not been installed correctly. |
Register the license or restart the computer. If this does not remedy the problem, contact your Shimadzu representative. |
1903 |
Failed to open the instrument screen. |
A fault has occurred on the database, or a probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1904 |
Failed to open the data screen. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1905 |
Failed to open the analysis tool screen. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1906 |
Failed to open the administration tool screen. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1907 |
Failed to open the manual screen. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1908 |
Failed to display the project selection dialog. |
Failed to retrieve the information. A probable cause is a problem in the network connection in the case of a network configuration. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
190A |
Failed to display the on-line manual. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
190B |
Failed to start up the LabSolutions. |
The LabSolutions connot be started up for some reason. |
Try again after a while. If this does not remedy the problem, restart the computer, and try the operation again. |
190C |
Failed to change the project. |
Failed to change the project. |
Try again. If this does not remedy the problem, restart the computer, and try setting again. |
190D |
The selected project is invalid. |
The selected project is invalid. |
Close the sub-window, and select the project again. |
190E |
Failed to create the work folder. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
190F |
Previous Project is not found. |
Computer Name may have been changed. |
If those projects displayed on the message box were created on this computer, please click [YES]. Those projects will be updated. If those projects were not created on this computer, please click [NO]. If you do not update, you will not be able to use those projects. And please note that it becomes impossible for you to update those projects if you create a new project without updating. |
1931 |
Failed to initialize the acquisition control component for status of instrument administration. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1932 |
Failed to update the operating status in the acquisition control component for status of instrument administration. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1933 |
Failed to notify the operating status to the acquisition control component for the status of instrument administration. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1934 |
Failed to initialize the Lss control component for status of instrument operation. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1935 |
Failed to update the operating status in the Lss control component for status of instrument administration. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1936 |
Failed to get the operating status of instruments. |
Failed to retrieve the information. A probable cause is a problem in the network connection in the case of a network |
Restart the computer, and try the operation again. Try again after making sure that the network is connected correctly in the case of a network configuration. |
1937 |
Failed to display the batch queue dialog. |
The [Batch Queue] sub-window cannot be displayed for some reason. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1938 |
Failed to display the Realtime Analysis screen. |
The [Realtime Analysis] program cannot be started up for some reason. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. |
1939 |
Failed to display the Offline Editor screen. |
The [Offline Editor] program cannot be started up for some reason. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. |
193A |
Failed to display the Postrun Analysis screen. |
The [Postrun Analysis] program cannot be started up for some reason. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. |
193B |
Failed to take over analysis. |
Cannot take over analysis for some reason. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
193C |
Failed to force the release of instrument occupation. |
The instrument occupancy cannot be forcibly released for some reason. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
193D |
Failed to update the latest information of the instrument_operating status. |
Failed to retrieve the information. A probable cause is a problem in the network connection in the case of a network |
Restart the computer, and try the operation again. Try again after making sure that the network is connected correctly in the case of a network configuration. |
193E |
Failed to filter the instrument operating status list. |
Failed to filter the instrument operating situation list for some reason. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. |
193F |
Failed to display the table style settings screen. |
The [Table Style] sub-window cannot be opened for some reason. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. |
1940 |
Failed to display the table. |
The table cannot be displayed for some reason. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. |
1941 |
Failed to display the icon. |
The icon cannot be displayed for some reason. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. |
1942 |
Failed to align the icons. |
The icons cannot be aligned for some reason. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. |
1943 |
Failed to display the layout. |
Cannot display according to the layout for some reason. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. |
1944 |
Failed to set the background. |
The background cannot be set for some reason. |
Make sure that the background file type is appropriate, and try the operation again. If this does not remedy the problem, restart the computer, and try again. |
1945 |
Failed to clear the background. |
The background cannot be cleared for some reason. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. |
1946 |
Failed to get the user property. |
Failed to obtain the property. A probable cause is a problem in the network connection in the case of a network configuration. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1947 |
Failed to register the user property. |
The property cannot be saved. A probable cause is a problem in the network connection in the case of a network configuration. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1948 |
The batch queue dialog is displayed. |
The [Batch Queue] sub-window is opened. This operation is currently not available. |
Close the open [Batch Queue] sub-window, and try the operation again. |
1949 |
Release of instrument occupation has been forced. |
----- |
----- |
1950 |
Failed to update automatically the instrument_operating status. |
The instrument operating situation cannot be automatically updated. |
Restart the computer, and try the operation again. Try again after making sure that the network is connected correctly in the case of a network configuration. |
1951 |
Failed to display the browser screen. |
The [Browser] program cannot be started up for some reason. |
Try again. If this does not remedy the problem, restart the computer, and try the operation again. |
1952 |
Please operate this instrument on the computer connecting it. |
This instrument can be operated on the computer connecting it only.\nThe operating sutuation can be diaplayed on the computer unconnecting the instrument. |
Please operate on the computer connecting the target instrument, or operate the instrument connecting with this computer. |
1953 |
The instruments are occupied by other users or projects. |
The instruments are occupied by other users or projects. |
|
1A00 |
No rights of system administration. |
A user without the [System Administration] rights attempted to make a setting. |
Log in as a user with the [System Administration] rights. |
1A01 |
Failed to get multi data report configuration information. |
Failed to retrieve the information. A probable cause is a problem in the network connection in the case of a network configuration. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1A02 |
Failed to register multi data report configuration information. |
Failed to register. A probable cause is a problem in the network connection in the case of a network configuration. |
Register again. If this does not remedy the problem, restart the computer, and try setting again. Try again after making sure that the network is connected correctly in the case of a network configuration. |
1A03 |
Multi data report configuration information has been registered. |
----- |
----- |
1A04 |
Failed to get the information about the data processing type. |
Failed to retrieve the information. A probable cause is a problem in the network connection in the case of a network configuration. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1A05 |
Failed to create the template file. |
Cannot create the template file. A probable cause is a problem in the network connection in the case of a network configuration. |
Create the file again. If this does not remedy the problem, restart the computer, and try setting again. Try again after making sure that the network is connected correctly in the case of a network configuration. |
1A06 |
No rights to create templates. |
A user without the [Create Template] rights attempted to make a setting. |
Log in as a user with the [Create Template] rights. |
1A07 |
Failed to read the template file. |
Failed to read the template file. A probable cause is a problem in the network connection. |
Try again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1A08 |
Failed to connect to the database. |
Failed to connect to the database. A probable cause is a problem in the database state or network connection. |
Try again after checking the database status and make sure that the network is connected correctly. If this does not remedy the problem, restart the computer, and try |
1A09 |
Failed to create the report file. |
Failed to create the report file. A probable cause is a problem in the network connection in the case of a network configuration. |
Create the file again. If this does not remedy the problem, restart the computer, and try setting again. Try again after making sure that the network is connected correctly in the case of a network configuration. |
1A0A |
Failed to register the report file. |
Failed to register the report file. A probable cause is a problem in the network connection in the case of a network configuration. |
Register again. If this does not remedy the problem, restart the computer, and try setting again. Try again after making sure that the network is connected correctly in the case of a network configuration. |
1A0B |
Failed to create the PDF file. |
The PDF file cannot be created for some reason. |
Try again after making sure that the software for making PDFs is installed correctly. |
1A0C |
Successfully registered the report file. |
----- |
----- |
1A0D |
The specified project is already deleted, or you do not have access rights. |
The specified project is deleted, or you do not have access rights. |
Close the sub-window, and select the project again. If you do not have access rights, contact the system administrator and add a user to the project. |
1A0E |
No data to output to the report file since the data type does not match. |
There is no data to output since there is no data type to output to the specified template file. |
Select a template file containing the data type to be output. |
1A0F |
Failed to create the PDF file since the tool to generate PDFs is not installed. |
The tool for making PDF files is not installed. |
Contact your Shimadzu representative. |
1A13 |
File access error. |
Cannot access the template file. |
Make sure that the corresponding template file is not used by another user. |
1A15 |
No data information in the selected data file. |
Displayed for confirmation when the selected data file does not have data |
Select a data file containing data information. |
1A16 |
Microsoft Excel is not installed. |
For this function, Microsoft Excel is necessary. |
Please purchase Microsoft Excel and install it into PC. |
1A17 |
The multi-data report can not be created as no peak is found. |
The multi-data report can not be created as no peak is found. |
----- |
1A18 |
The value more than 301 cannot be set for the "max ID". |
The value more than 301 cannot be set for the "max ID". |
Please register the template again by setting the "max ID" for the specified template. |
1A27 |
The sheet name cannot be set the same name as another sheet |
This sheet name is same as another sheet. |
Set other sheet name. |
1A28 |
Save current template file? |
Changes have been made to the template file. Confirm whether to save the changes or close the file without saving the |
----- |
1A29 |
Save the current multi-data report file? |
Changes have been made to the multidata report file. Confirm whether to save the changes or close the file without saving the changes. |
----- |
1A2A |
Failed to open the data file. The data file has unknown device |
The data file has unknown device type. |
Open other data file registered with device type. |
1A2B |
Failed to open the data file. |
The data file cannot be openen. |
Open other data file. |
1A2C |
Invalid Value is set. |
Invalid value is set. |
Set proper value. |
1A2D |
Failed to Register the multi-data report file since another report file that has the same name is registered simultaneously. |
Failed to Register the multi-data report file since another report file that has the same name is registered simultaneously. |
Retry to create multi data report. |
1A2E |
Specify font size from 1 to 409. |
Specify font size from 1 to 409. |
Specify font size from 1 to 409. |
1A2F |
This MultiDataReport template file or report file can't be opened in current configuration. |
This MultiDataReport template file or report file can't be opened in current configuration. |
Uncheck [Use Excel to generate reports.] checkbox in MultiDataReport Configuration to open this file. |
1A30 |
No rights of system administration. |
The authenticated user does not have the [System Administration] rights. |
Authenticate as a user with the [System Administration] rights. |
1A31 |
Agent Converter is already running. |
An attempt was made to start up Agent Converter multiple times. Agent Converter cannot be started up multiple times on a single computer. |
Do not start up Agent Converter multiple times. |
1A32 |
Are you sure you want to delete selected conversion settings? |
Displayed for confirmation when the selected conversion settings are deleted. |
To delete, click [OK], and to cancel, click [Cancel]. |
1A33 |
The conversion settings have been deleted. |
----- |
----- |
1A34 |
The specified Agent database is already registered. |
The Agent database to be converted was specified. The same Agent database cannot be set multiple times. |
Specify another Agent database, or change to the conversion settings currently set with the same Agent database. |
1A35 |
The electronic recording/electronic signatures (ER/ES) mode does not match. Please select a project in the |
The electronic recording/electronic signatures (ER/ES) mode does not match between the specified Agent database and the project. |
Specify the Agent database and project set with the same electronic recording/electronic signatures (ER/ES) mode. |
1A36 |
Input items are not specified. |
Clicked [OK] without setting items. |
Set the items and click [OK]. |
1A39 |
Failed to get the project information. |
Failed to retrieve the project information from the LabSolutions database. |
Contact your Shimadzu representative. |
1A3A |
Failed to connect to the Agent database. |
Failed to connect to the Agent database to be converted. Check the details in the output sub-message. |
Contact your Shimadzu representative. |
1A3B |
The database is not selected. |
Clicked [OK] without selecting the |
Select the database, and click [OK]. |
1A3C |
The connection test with the LabSolutions server has failed. |
The connection test with the LabSolutions server has failed. |
Contact your Shimadzu representative. |
1A3D |
Failed to connect to the database. |
Failed to connect to the database. Check the details in the output sub-message. |
Contact your Shimadzu representative. |
1A3E |
Please exit and restart conversion settings again with authentication. |
The connection between the LabSolutions server and the database server is not recognized since no connection has been established when LabSolutions is initially started up on an independent computer. |
Since the authentication sub-window opens after the system is restarted, conversion settings can be made after authentication. |
1A3F |
Failed to connect to the Agent database. |
Failed to connect to the Agent database to be converted. Check the details in the output sub-message. |
Contact your Shimadzu representative. |
1A40 |
Failed to delete the data. |
Data cannot be deleted on the Agent database to be converted. Check the details in the output sub-message. |
Contact your Shimadzu representative. |
1A41 |
The data has been deleted. |
Data was deleted on the Agent database to be converted. |
----- |
1A43 |
No compatible software for conversion. |
The software currently registered to the conversion source is not registered to LabSolutions. |
Contact your Shimadzu representative. |
1A44 |
No compatible Result Table type for conversion. |
The Result Table type required for conversion is not registered. |
Contact your Shimadzu representative. |
1A45 |
Failed to connect to the project. |
Failed to connect to the conversion destination project. Check the details in the output sub-message. |
Contact your Shimadzu representative. |
1A46 |
Failed to register the data. |
Failed to register data to the conversion destination project. Check the details in the output sub-message. |
Contact your Shimadzu representative. |
1A47 |
Failed to register since the registered data is signaturelocked. |
Data of same file identifier as the data you attempted to register is signature-locked. The signature-locked data will be deleted since it cannot be converted. |
----- |
1A49 |
Batch conversion has been completed. |
----- |
----- |
1A4A |
Conversion settings are not correct. |
The setting information for batch conversion is incorrect. |
Correct the setting information to the correct values. |
1A4B |
[Software Type] or [Software Name] is duplicated. |
[Software Type] or [Software Name] is duplicated. |
Correct the duplicated settings. |
1A4C |
In the manual entry mode, set a value smaller than 10000 to [Software Type]. |
Values 10000 or more are automatically checked when the information file of an analytical instrument is loaded. Cannot specify in manual registration mode. |
In the manual registration mode, set a value smaller than 10000. |
1A4D |
The information required for conversion is not specified. |
The definition information has been set from the information file of the analytical instrument, however, the information for conversion has not been entered. |
Enter information for conversion. |
1A4E |
Column Nos. are duplicated. |
Column Nos. are duplicated. |
Correct the duplicated settings. |
1A4F |
Column Nos. are not set in order from 1. |
Column Nos. are not set in order from 1. A No. is missing midway. |
Enter in order starting from 1. |
1A50 |
Are you sure you want to delete the selected row from Result Table format? |
Message for confirming whether or not to delete the selected row from the Result Table format. |
To delete, click [OK], and to cancel, click [Cancel]. |
1A51 |
The Result Table type is |
The Result Table type is duplicated. |
Correct the duplicated settings. |
1A52 |
In the manual entry mode, set a value smaller than 10000 to the Result Table type. |
Values 10000 or more are automatically checked when the information file of an analytical instrument is loaded. Cannot specify in manual registration mode. |
In the manual registration mode, set a value smaller than 10000. |
1A53 |
Are you sure you want to delete the currently displayed Result Table format? |
Message for confirming whether or not to delete the currently displayed Result Table format. |
To delete, click [OK], and to cancel, click [Cancel]. |
1A54 |
Failed to save to the software definition file. |
Cannot save the definition information to the software definition file. Probable causes are insufficient disk space or a |
Contact your Shimadzu representative. |
1A55 |
A service of the same name is already registered. |
A service of the same name is already registered. |
Specify another service name. |
1A56 |
The conversion settings have been added. |
----- |
----- |
1A57 |
The conversion settings have been edited. |
----- |
----- |
1A58 |
The conversion configuration has been edited. |
----- |
----- |
1A59 |
Failed to get the definition information. |
A probable cause is a problem in the network connection. |
Try again after making sure that the network is connected correctly. If this does not remedy the problem, contact your Shimadzu representative. |
1A5A |
Failed to set the definition information. |
A probable cause is a problem in the network connection. |
Try again after making sure that the network is connected correctly. If this does not remedy the problem, contact your Shimadzu representative. |
1A5B |
Failed to initialize. |
The sub-window cannot be displayed since the information required at startup could not be loaded correctly. |
Contact your Shimadzu representative. |
1A5C |
The project at the conversion destination is an MDB file. |
The project at the conversion destination is an MDB file. |
Set the project at the conversion destination to Oracle or MSDE. |
1A5D |
Failed to access the Oracle connection configuration file. |
Failed to access the Oracle connection configuration file. |
Contact your Shimadzu representative. |
1A5E |
Failed to run AgentConverter Service. |
Failed to change Service Information. |
Contact your Shimadzu representative. |
1A60 |
No rights to set the LIMS Connecting Option. |
Execution was performed by a user without rights for setting the LIMS setting. |
Log in as a user with rights. |
1A61 |
Failed to get the LIMS Connecting Option configuration information. |
A probable cause is a problem in the network connection. |
Try again after making sure that the network is connected correctly. If this does not remedy the problem, contact your Shimadzu representative. |
1A62 |
Failed to register the LIMS Connecting Option configuration information. |
A probable cause is a problem in the network connection. |
Try again after making sure that the network is connected correctly. If this does not remedy the problem, contact your Shimadzu representative. |
1A64 |
Successfully completed the file output from the LIMS Transfer. |
----- |
----- |
1A65 |
Failed to acquire the database. |
Failed to acquire the database. |
Try again after making sure that the network is connected correctly. If this does not remedy the problem, contact your Shimadzu representative. |
1A66 |
Failed to reference the folder. |
The folder set by the LIMS setting cannot be referenced. |
Try again after checking the settings. |
1A67 |
Failed to access the file. |
The file cannot be accessed, or the transmission file path is too long. |
Try again after checking the settings. |
1A68 |
Failed to perform batch processing. |
Failed to execute batch processing. |
Try again after checking the content set to the transmission batch file. |
1A69 |
The file has been output. |
The file was output to the folder where the transmission file was created. |
----- |
1A6A |
Successfully connected to the FTP server. |
----- |
----- |
1A6B |
Failed to connect to the FTP |
----- |
----- |
1A6C |
Failed to run LIMS Service. |
----- |
----- |
1A6D |
Successfully completed import file from LIMS Receiver. |
----- |
----- |
1A6E |
Failed to import file from LIMS Receiver. |
----- |
----- |
1A90 |
Mulit data report is not created. The template file is not fitted into the configuration. |
Mulit data report is not created. The template file is not fitted into the configuration. |
If [Use Excel to generate reports.] checkbox in MultiDataReport Configuration is checked, use Excel template file(.xls/.xlsx). If the checkbox is unchecked, use LabSolutions multi data report template file(.lmdt). |
1A91 |
Not find data that matches the search criteria. |
----- |
----- |
1A92 |
Successfully registered the template file. |
----- |
----- |
1A93 |
Failed to register the template file. |
Failed to register the template file for some reason. |
Try again after making sure that the network is connected correctly. If this does not remedy the problem, contact your Shimadzu representative. |
1A94 |
Some cells are removed; The number of rows or columns in the multi data report/template file exceed max number. |
Repeat row settings in the template file may be problem, or current configuration differs from configuration that was set when the template file was created. |
Create a new template file and try creating multi data report again. |
1A95 |
Failed to paste. |
Cannnot paste contents of clipboard to specified area. Size of data to paste may differ from that of specified area or a part of merged cell may be changed. |
Try the following operation and try to paste again;\nSpecify another area, specify a single cell, or split merged cells. |
1AB0 |
Failed to run Multi-Data Register Service. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1AB1 |
No license to use Multi-Data Register Option. |
No license to run. |
Please register the license. |
1AB2 |
The import folder specified for Public Text files does not exist. \nPlease change. |
The wrong directory (folder) was specified. |
Please set the correct folder name. |
1AB3 |
The import folder specified for data files does not exist. \nPlease change. |
The wrong directory (folder) was specified. |
Please set the correct folder name. |
1AB4 |
The import folder specified for log files does not exist. \nPlease change. |
The wrong directory (folder) was specified. |
Please set the correct folder name. |
1AB5 |
Failed to analyze. |
Invalid map information or import file |
Please check the map information and import settings. |
1AB6 |
Failed to print the PDF file. |
Printer settings are not correct. |
Please check the printer settings. |
1AB7 |
Failed to run external application before analysis.\nHowever, the registration continues. |
Settings to run external application before analysis are not correct. |
Please specify the correct application. |
1AB8 |
Failed to create the temporary folder for analysis. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1AB9 |
Failed to save the registration information. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1ABA |
Failed to move the files. |
The import folder does not exist or there is no right to access the import folder. |
Please check if there is a right to access the import folder. |
1ABB |
Files have been moved to the temporary evacuation folder. |
----- |
----- |
1ABC |
Some Mapping values exceeded the maximum number of character count.\nParts of the Mapping values have been |
----- |
----- |
1ABD |
Failed to output the PDF file. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1ABE |
Prohibited characters have been deleted. |
Following characters cannot be used: \ / : * ? < > " ' |
Do not use prohibited characters. |
1ABF |
The data file will not cooperate with logs. |
Multiple data files have been detected. |
Output only one data file at a time. |
1AC0 |
Prohibit characters are deleted. |
Following characters, \ / : * ? < > " ', are used. |
Do not use prohibit characters. |
1AC1 |
The data file will not cooperate with logs. |
Multiple data files are detected. |
Output only one data file in the same time. |
1AC2 |
Failed to run Multi-Data Register Registration Service. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1AC3 |
Failed to register data-related logs. These logs will be re- |
----- |
----- |
1AC4 |
Successfully registered. |
----- |
----- |
1AC5 |
Failed to run external application after registration.\nHowever, the registration continues. |
Settings to run external application after registration are not correct. |
Please specify the correct application. |
1AC6 |
Failed to register files. |
----- |
----- |
1AC7 |
Failed to register files.\nFiles have been temporarily |
A probable cause is that the files were broken. |
Please check if the files are correct and register again. |
1AC8 |
Failed to delete the registration information. |
A probable cause is that the registration information has already been deleted. |
If the registration information has not been deleted, please delete it again. |
1AC9 |
Successfully deleted the registration information. |
----- |
----- |
1ACA |
Canceled to modify the registration information. |
----- |
----- |
1ACB |
Failed to modify the registration information. |
A probable cause is that the registration information has already been modified. |
Please modify the registration information again. |
1ACC |
The registration information has been modified. |
----- |
----- |
1ACB |
The registered files are broken. |
----- |
----- |
1ACC |
Application log registration |
----- |
----- |
1ACD |
Message exceeds the maximum number of character count.\nA part of the message will be |
----- |
----- |
1ACE |
Sub message exceeds the maximum number of character count.\nA part of the sub message will be lopped off. |
----- |
----- |
1ACF |
The file name is renamed becauce of duplication. |
----- |
----- |
1AE0 |
Abnormalities occurred in the configuration settings. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1AE1 |
The terminal configuration file was imported. |
----- |
----- |
1AE2 |
The terminal configuration file was exported. |
----- |
----- |
1AE3 |
A target file or folder is not found. |
The wrong file or directory (folder) was specified. |
Please set the existing correct file or folder name. |
1AE4 |
No rights to access a file or |
There is no right to access the file or |
Please check the right to access the file or folder. |
1AE5 |
Acquire a terminal configuration file from another PC? |
----- |
----- |
1AE6 |
The terminal configuration file was acquired from another PC. |
----- |
----- |
1AE7 |
Failed to get the information from the database or file. |
Failed to access the database or file. |
In the case of network configuration, please check network connection. \nIf the network is normal, contact your Shimadzu representative. |
1AE8 |
Failed to save the information to the database or file. |
Failed to access the database or file. |
In the case of network configuration, please check network connection. \nIf the network is normal, contact your Shimadzu representative. |
1AE9 |
Failed to save to the database or file since it conflicted with other users. |
The operation to save to the database conflicted with other users. |
Please perform again after a short while. |
1AEA |
There is an item with an insufficient input. |
The item displayed in the message has not been inputted. |
Please fill in the item displayed in the message. |
1AEB |
There is an item with duplicated contents. |
The item displayed in the message has duplicated contents. |
Please check the contents of the item displayed in the message. |
1AEC |
The user is different from the selected one. |
Authentication was performed by the user different from the selected one. |
Please perform authentication by the selected user. |
1AED |
Data file that exceeds the loadable range has been loaded.\nPlease check the data. |
Data file that exceeds the loadable range in the data view has been loaded. |
Please review the display range of the configuration file. |
1AEE |
Failed to validate the data.\nPlease check the data |
Loaded data does not match the data validation string. |
Please check the data or contents in the data validation string. |
1AEF |
The specified path is invalid. |
The incorrect path has been set. |
Please check your entry of the item. |
1AF0 |
Abnormalities occurred in Map Editor. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1AF1 |
The Map information is already registered. |
The Map information has already been registered. |
Please change the map information name. |
1AF2 |
Failed to get the information from the database or file. |
Failed to access the database or file. |
In the case of network configuration, please check network connection. \nIf the network is normal, contact your Shimadzu representative. |
1AF3 |
Failed to save the information to the database or file. |
Failed to access the database or file. |
In the case of network configuration, please check network connection. \nIf the network is normal, contact your Shimadzu representative. |
1AF4 |
Failed to save to the database or file since it conflicted with other users. |
The operation to save to the database conflicted with other users. |
Please perform again after a short while. |
1AF5 |
There is an item with an insufficient input. |
The item displayed in the message has not been inputted. |
Please fill in the item displayed in the message. |
1AF6 |
There is an item with duplicated contents. |
The item displayed in the message has duplicated contents. |
Please check the contents of the item displayed in the message. |
1AF7 |
The specified sheet or page does not exist. |
The specified sheet or page does not exist. |
Please check the contents of the item. |
1AF8 |
Map information is not saved. Are you sure you want to save? |
----- |
----- |
1AF9 |
Must be saved before the operation. Are you sure you want to save? |
----- |
----- |
1AFA |
Are you sure you want to remove the Map information under the |
----- |
----- |
1AFB |
The Map Information file was exported. |
----- |
----- |
1AFC |
Are you sure you want to delete the selected rows? |
----- |
----- |
1AFD |
The map information has been registered. |
----- |
----- |
1AFE |
The map information has been edited. |
----- |
----- |
1AFF |
The map information has been deleted. |
----- |
----- |
1B00 |
Failed to display the automatic data registration setting dialog. |
Failed to display the automatic data registration setting dialog for some reason. |
try again. If this does not remedy the problem, restart the computer, and try setting again. |
1B01 |
No rights to set the automatic data registration setting. |
Setting was performed by a user without rights for setting the automatic data registration. |
Log in as a user with rights. |
1B02 |
The specified data is deleted. \nAre you sure? |
----- |
----- |
1B03 |
Though there are settings in 「Setting of each keyword」,there are no settings in「Setting of each |
----- |
----- |
1B04 |
Restart the PC to apply settings. |
----- |
----- |
1B05 |
Failed to register the project. |
The database does't start or there is a problem in the network. |
Confirm the start of the database and confirm whether there is problem in the network when the database start. |
1B06 |
Failed to load the demand text |
Cannot load the demand text file. |
Check the demand text file. |
1B07 |
Failed to export the sample information file. |
Cannot export the sample information file. |
Check the sample information file. |
1B08 |
Auto data registration cannot be run. |
Cannot run auto data registration |
Restart the computer.If this does not remedy the problem, contact your Shimadzu representative. |
1B09 |
Failed to save to reiterated file. |
Failed to save to reiterated file. |
Check the save folder. |
1B0A |
Failed to register since the registered data is signaturelocked. |
Data of same file identifier as the data you attempted to register is signature-locked. Cannot register the data which is signature-locked. |
----- |
1B0B |
Register file. |
----- |
----- |
1B0C |
Failed to start the application. |
Cannot start the application. |
Confirm whether the specified application can be started. |
1C00 |
Succeeded to register the file. |
----- |
----- |
1C01 |
Failed to map the analysis date and time. |
The data file has an invalid analysis date and time. |
Please check the analysis date and time. |
1C02 |
Failed to export the sample information file. |
Cannot export the sample information file. |
Check the sample information file. |
1C03 |
Failed to delete the file. |
there is no right to access the file or the file has been used. |
Pleas grant to access the file and close all applications possible to use the file. |
1C30 |
Failed to map the analysis date and time. |
There is a deficiency in the setting of the analysis date and time. |
Please check the settings of the analysis date and time. |
1F00 |
Invalid parameter. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1F01 |
It is not initialized. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1F02 |
It is already initialized. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1F03 |
Insufficient free space. |
Normal processing can no longer be executed due to insufficient memory required for program operation. |
Close other applications that are not required. There is sometimes not enough memory when the [Postrun Analysis] program is being run twice or more times. If this error cannot be remedied, shut down and restart the |
1F04 |
The network timeout occurred. |
A probable cause is that the network or server load has risen. |
Wait a while before trying the operation again. If this does not remedy the problem, contact your Shimadzu |
1F05 |
The networkerroroccurred. |
A probable cause is a problem in the network connection. |
Try again after making sure that the network is connected correctly. If this does not remedy the problem, contact your Shimadzu representative. |
1F06 |
Cannot operate, because of failure. |
Operations cannot be executed while a fault has occurred on the database. |
Try again after recovering from the database fault. |
1F07 |
The input value is not correct. |
An inappropriate value was entered. |
Check the sub-message, and correct to the appropriate |
1F08 |
Please input. |
Some items have not been entered. |
Set the appropriate value. |
1F09 |
This application is already running. |
You try to execute the application which has been already executed. |
Close the application and then try again. |
1F10 |
The printer driver is not installed. |
An appropriate driver for the specified printer has not been installed. |
Install the appropriate driver. |
1F1A |
Network configuration is not correct. |
A probable cause is a problem in the network configuration. |
Try again after checking the network configuration. |
1F1B |
The specified date in search condition is not correct. |
The search conditions start data is set after the end date. |
Set the correct date. |
1F1C |
Failed to register the file. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
1F1D |
The same file exists. Are you sure you want to overwrite it? |
A file of the same file name exists at the file output destination. |
To continue, click [Yes], and to cancel, click [No]. |
1F1E |
Failed to print. |
Failed to print. |
Try again after checking that the printer has started up and that the network is connected correctly. If this does not remedy the problem, contact your Shimadzu |
1F1F |
The specified file does not exist. |
The specified file does not exist. |
Try again after checking that the specified file is available. If this does not remedy the problem, contact your Shimadzu representative. |
1F20 |
A file name cannot contain any of the following characters:_\n\\ / : * ? "" < > | |
An attempt was made to execute after entering characters that cannot be used in file names (e.g. \ / : * ? < > | "). |
\ / : * ? The " < > | characters cannot be used as Windows (R) file names. Enter the file name again using characters other than these. |
1F21 |
No license. |
No license for performing execution. |
Register the license. |
1F22 |
Add/delete function may have failed. Please check this. |
A probable cause is a problem in the network connection. |
Try again after making sure that the network is connected correctly. If this does not remedy the problem, contact your Shimadzu representative. |
1F24 |
The specified file is not related to any applications. |
The specified file is not made relationship to any application. |
Try again after linking the file to the application. |
1F25 |
The content of the change has been recorded. |
Message for notifying that the audit trail was recorded. |
----- |
1F26 |
Failed to display the project selection dialog. |
Failed to retrieve the information. A probable cause is a problem in the network connection in the case of a network configuration. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
1F27 |
Please change file name length less or equal to 80, because selected file name is too long. |
Selected file name is too long. |
Please change file name length (with extension) less or equal to 80. |
1F28 |
No rights of windows administrator. |
You try to start up windows by the user who did not have rights of windows |
Start up windows by the user who have rights of windows administrator. |
2000 |
Assistant bar is not initialized. |
The operator cannot use the assistant bar without initializing it. |
This is a rare system error. Reboot the computer and try again. |
2040 |
This file (file name) has been used on (window name) Window by User [user name].\nClose this file or Open read only or Cancel? |
The selected file is in use in the window displayed in the message. |
To close the file in question, and then open it in the target window, select the [Close then Open] button, or, to open it as a read-only file in the target window, select the [Open as Read Only] button. Click Cancel to cancel the operation. |
2041 |
This file (file name) has been used on (window name) Window.\nClose this file or Open read only or Cancel ? |
The selected file is in use in the window displayed in the message. |
To close the file in question,andthen open it in the target window, select the [Close then Open] button, or, to open it as a read-only file in the target window, select the [Open as Read Only] button. Click Cancel to cancel the operation. |
2042 |
This file %s has been used on System#%ld %s %s Window by User [%s]._\nClose this file or Open read only or Cancel? |
The selected file is in use in the window displayed in the message. |
To close the file in question,andthen open it in the target window, select the [Close then Open] button, or, to open it as a read-only file in the target window, select the [Open as Read Only] button. Click Cancel to cancel the operation. |
2043 |
This file (file name) has been used on System# (system number) (window name) Window.\nClose this file or Open read only or Cancel ? |
The selected file is in use in the window displayed in the message. |
To close the file in question,andthen open it in the target window, select [Close then Open] button, or, to open it as a read-only file in the target window, select the [Open as Read Only] button. Click Cancel to cancel the operation. |
2044 |
%s application will be closed. |
The user has selected to close the application. |
Select "OK"_when to exit the application and "Cancel" to keep the application open. |
2045 |
To use (window name), please specify (window name) to the Available on the Application Windows tab. |
Though (window name) have been specified to the Available on the Assistant bar tab, they are not specified to the Available on the Application Windows tab. |
(Window name) have to specify to the Available not only on the Assistant bar tab, but also on the Application tab. Please specify those to the Available on the Application Windows tab. |
2080 |
Please set a value within the range 10 to 200. |
A value outside the range was entered. |
Enter a value within the range. |
2081 |
Failed to create object. |
The object could not be created. |
There may be insufficient memory; close any other applications or restart Windows. If the problem still persists, the registry may have become corrupted. |
2082 |
The position of the item is out of the paper. |
The item is positioned off of the page. |
Position the item onto the page. |
2083 |
More than 100 pages cannot be inserted. |
The operator tried to add more than 100 pages. |
The maximum number of editable pages is 100. You cannot exceed 100 pages. |
2084 |
The first page cannot be deleted. |
The operator tried to delete page 1. |
A report must contain at least one page. |
2085 |
Cannot launch LSSLaunchingManager.exe. |
The printing process failed. |
There may be insufficient memory; close any other applications. Restart Windows. |
2086 |
Please set 5mm or more for the item width. |
A width value less than 5 mm was entered. |
Set the width items to at least 5 mm. |
2087 |
Please set 5mm or more for the item height. |
A height value less than 5 mm was entered. |
Set the height of items to at least 5 mm. |
2088 |
Are you sure you want to delete current page? |
The operator has elected to delete the page. All the items on the page will also be |
Confirmation message. |
2089 |
The printing format will be changed with saving now. \nAre you sure you want to continue? |
The user tried to save a default format. |
Confirm message before to change. |
208A |
Please set positive number to report number. |
----- |
----- |
208B |
Distiller is not installed. |
For this function, Adobe Acrobat Distiller is necessary. |
Please purchase Adobe Acrobat and install it into PC. |
208C |
PDF Writer is not installed. |
For this function, Adobe Acrobat PDFWriter is necessary. |
Please purchase Adobe Acrobat and install it into PC.\n Acrobat PDFWriter will not be install with Standard Setup.\n Please install it with using Custom Setup. |
208D |
Post Script driver is not installed. |
PostScript Printer Driver, configured for this function, is not found. |
Please check "Administration Manual" and install/configure PostScript Printer Driver with the procedure on it. |
208E |
Fix the file_name in the report |
----- |
----- |
208F |
Print jobs are left._\nAre you sure you want to exit? |
----- |
----- |
2090 |
You must install printer driver prior to printing._\nTo install the printer driver, select Settings from the Start menu and click Printers. Double click on Add |
The print or print preview cannot be performed as no printer driver is installed. |
Select [Printers] from [Settings] via [Start] menu. Then add printer. |
2091 |
SkyPDF Pro is not installed. |
The SkyPDF printer could not be found at output to PDF. |
Install SkyPDF Pro. |
2095 |
LSSPrint.exe does not respond and has been shutdown.Canceled Printing. |
The total size of data files which is necessary for summary report exceeded the tolerance, or communication error between a PC and a printer or paper-out condition occurred. |
Reduce the number of data files to use for summary, confirm the connection between the PC and the printer, or feed the printer with a paper. |
209A |
Please check polarity. |
Polarity to output the report has not been selected. |
Please check polarity. |
209B |
Please select Acquisition Mode. |
Acquisition Mode to output the report has not been selected. |
Please select Acquisition Mode. |
209C |
The report format file can not be converted properly because there are unreadable report items. |
There are unreadable report items in the report format file. |
Delete the MS or GPC report items on the software which the report format file is created. |
2100 |
No ID# is selected in the Compound Table. |
For the information to be displayed, a compound must be selected from the Compound Table. |
Select the compound to display or process. |
2101 |
No data file is selected. |
Although the data file should be specified in order to operate the function, the data file has not been specified. |
Specify (or select) the data file. |
2102 |
No information in the Compound Table. |
Processing cannot continue because the Compound Table information has not been entered. |
Enter the target compound information in the Compound Table. |
2103 |
No data file is included in the calibration information. |
There is no data file record in the calibration curve information. |
The calibration curve was not created properly. Re-create the calibration curve either in batch processing or the calibration curve window. |
2104 |
This time range is shorter than data collection interval. |
The specified data collection interval is less than the range set in the manual peak integration settings; the interval essentially becomes 0. |
Select a wider range. |
2105 |
Are you sure you want to remove the file from the calibration curve? |
Confirm whether to remove the data file from the calibration point in the calibration curve. |
Upon confirmation, the data file is removed. All calibration points from that data file are removed; then the calibration curve is re-created. |
2106 |
Cannot add the selected files since the number of total files would exceed maximum for |
The maximum number of calibration averaging points per level is 10. |
The maximum number of averaging points per level is 10. |
2107 |
Calibration Curve Information (the Quantitative Parameters, Compound Table) of the data file does not match with the information of the method file._\nDo you copy the information of the method file to |
The Calibration Curve Information (Quantitative Parameters, Compound Table) does not match the information in the data file, and processing cannot be performed. |
Copy the information from the method file to the data file so that all the information corresponds correctly. When finished, save the method file along with the data file. |
2108 |
The [Method View] is now being edited. |
Because the Method View is in Edit Mode, files cannot be saved or closed. |
Switch the Method View to View Mode,andthen perform the operation again. |
2109 |
Drag the mouse and specify the peak range. |
---- |
Drag the mouse from the base line start time to the end time on the chromatogram. |
210A |
Cannot perform this process because file(s) are read only. |
The same data file may be open in another window, or the file attribute is read-only. Saving or processing cannot be |
Close the file that is open in the other window, or remove the read-only attribute. |
210B |
Manually detected peaks exist. These peaks will be deleted if peaks are integrating again. |
The chromatogram where peak integration is to be performed contains manually integrated peak results. |
Click on OK to delete the manual peak integration results and perform automatic peak integration. |
210C |
Please integrate all IDs before executing this command. |
All IDs must be integrated before attempting peak integration on an ID by ID |
Select "Peak Integration for all IDs" from the Quantitative menu. Peak integration can then be performed on an ID by |
210D |
Method file and data file(s) cannot be saved because read only file(s) exist. Are you sure you want to continue? |
The data file(s), to be added as calibration point, is read only, because it is opened in another program. So the operations in Calibration Curve window will not be saved to method or data file(s). |
To reverse the operations in calibration curve, click on the "Yes" button; to cancel, click on "No.". |
210E |
Number of channels in the method does not match in the data you want to add. You cannot add the data. |
Data Analysis window accepts the method which has the same configuration of data acquisition channels as current System Configuration. |
Reload proper method or change System Configuration. |
210F |
Selected baseline range is too short. |
This error occurs when the area selected by dragging after selecting [New Baseline] is smaller than the one selected by dragging after running [Manual Peak Integration] in the Quantitation View of the MS Data Analysis. \nAfter this message is displayed, your operation will be canceled. |
Drag to select an area larger than the area selected first. |
2110 |
This data file does not contain the data acquired by the currently active detector. Data cannot be added. |
The data file that you are attempting to add the calibration curve to does not have a currently active detector. |
Add a data file containing the detector to be operated. |
2111 |
This data file does not contain the data acquired by the currently active detector. Data cannot be added. |
The data file that you are attempting to add the calibration curve to does not have data acquired by the currently active detector. |
Add a data file containing data acquired by the detector to be operated. |
2112 |
Are you sure you want to remove all data files from the calibration curve? |
Confirm whether to remove the all data files from the calibration point in the calibration curve. |
Upon confirmation, the all data files are removed. All calibration points from that data files are removed; then the calibration curve is re-created. |
2113 |
The method file will also be updated at the same time. Are you sure you want to continue? |
The [Save Data and Method File] menu was selected. |
To continue processing, click [OK]. |
2114 |
The file must be selected to reference configuration information. |
A new method file was created by the started application without specifying the instrument. |
In the file selection dialog box that is next displayed, select the method file or data file which the environment settings to use are saved. |
2115 |
Because the data file of the same name exists in the folder of the method file, the specified data file cannot be added. |
It cannot be added the folder of the method file that is opening now because the data file of this name is in the same folder as the method file when it tried to add the data file of another folder. |
Please change the file name of the data file that wants to be added. Or, please change the file name to the data file of this name in the same folder as the method file or move the folder. |
2116 |
This file does not contain the data acquired by the currently active detector. Failed to open |
There is only an MS detector set for the specified file; there is no detector to perform operations on. |
Select a data file containing the detector to be operated. |
2117 |
Failed to decompress data. |
The compressed data file was not able to be progressed. |
----- |
2118 |
This data file was not able to record all the data at the acquisition. |
Some of the data have not been acquired from instrument. A probable cause is a problem in the network connection in the case of a network configuration. |
Confirm the connection between instruments. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
2120 |
The range of expansion/contraction cannot be set by overstepping the base |
The base point position cannot be overstepped for contraction and expansion along the time axis or the intensity axis. |
Drag the mouse so that the two necessary points are established for expansion and contraction in a single direction from the base point. |
2121 |
Change the chromatogram display to MIC? |
----- |
---- |
2122 |
Time range settings are not correct. |
The specified start time is larger than end time. |
Specify acquisition start or end time. |
2123 |
The time is out of range. |
The time range specified exceeds the allowable range. |
Use a lower time setting within a 0 to 2,147,483,647 range. |
2124 |
The number of data files for the operation must be 2 or less. |
----- |
----- |
2125 |
Please select 16 or less data to display. |
More than16 data are selected. |
Limit the number of display data to select up to 16. |
2126 |
Minimum and maximum intensity settings are not correct. |
In the intensity axis setting, the left value is greater than the right value. |
Set a left value less than the right value. |
2127 |
Set the noise calculation time range (start time to end time) same as or higher than the period |
The noise calculation time range (start time to end time) is set narrower than the period value. |
Set the noise calculation time range (start time to end time) same as or higher than the period value. |
2140 |
No compound object. System error occurred. |
The Compound Table could not be created. |
The system may not be operating properly. Restart the computer and try again. Should the problem persist, reinstall the software. |
2141 |
No qualitative/ quantitative frame. System error occurred. |
The program function required to open the Postrun window could not be found. |
Restart the computer and try again. Should the problem persist, re-install the software. |
2142 |
Sample information could not be retrieved. |
The sample information could not be obtained from the data file. |
Restart the computer and try again. Should the problem persist, try using a different data file. If the same error occurs for all of the data files, re-install the software. If the error is generated with for that data file only, the data file |
2143 |
Intended process was |
The user canceled the operation. |
----- |
2144 |
Method information is stored. |
The method file information in the data file was successfully saved to a separate method file. |
----- |
2145 |
Method information is retrieved. |
The method information from the specified file was written into the data file. |
----- |
2146 |
MCs are already registered to the maximum. Delete from fragment table before registration. |
The maximum number of MCs has been registered. |
Delete some MCs from the Fragment Table before registration. |
2147 |
Registered to the Spectrum |
----- |
----- |
2148 |
No CLASS-5000 related file to convert. |
----- |
----- |
2149 |
Batch information does not exist. |
Batch table information does not exist in the data file. |
The data file was created by of the following methods. Batch information cannot be extracted from these types of data files.\n1. After data was acquired by a single analysis, it was batch processed. File was created by converting a CLASS-LC10 data file. |
2150 |
Please close Report window before closing the window. |
The report window, which opened after data analysis, is still open. |
Close the report window first. |
2152 |
The registered standard spectrum will be deleted. Are you sure you want to continue? |
The standard spectrum used for the similarity peak identification method will be deleted. |
Click [Yes] to delete the standard spectrum, otherwise [No]. |
2153 |
Are you sure you want to return to the original data when data was acquired? |
The [File] - [Rollback to Original Data] command was selected at <Data Analysis>. |
Click [Yes] to return the content of the data file to the state immediately after data acquisition. |
2154 |
Original method does not exist. |
Performed [Rollback to Original Data] function, but it could not work, as no original method information. |
This is a rare system error. |
2155 |
Please perform peak integration again after setting delay time. |
----- |
----- |
2156 |
No spectrum exists. |
The operation can not be executed because the target spectrum has not been |
Specify the spectrum and then try again. |
2157 |
Remove reference |
----- |
----- |
2158 |
No sampler information. |
The currently open data file does not have sampler information. Sampler information cannot be saved. |
----- |
2159 |
Chromatogram selected as the reference data is 16 or less. |
----- |
----- |
215A |
A maximum of 80 chromatograms can be drawn in a single graph. Close reference data? |
This is displayed if the maximum number of chromatograms (80) that can be drawn in a single graph is exceeded when [Stack Chromatograms] is changed to [Overlay Chromatograms]. |
Click [Yes] to close the reference chromatogram. When clicking [No], the state does not change to [Overlay Chromatograms] and remains at [Stack Chromatograms]. |
215B |
%d files have been exported. |
Message for notifying that the spectrum of the row selected in the Peak Table was output as the nth JCAMP file to the directory of the currently open data file according to the [Save Spectrum at Selected Peak] popup menu on the [Peak Table] tab page in [Results View]. |
----- |
215C |
Are you sure you want to return to the original report format? |
----- |
----- |
2160 |
The compounds not under searching will be deleted from library. Deleted compounds cannot be recovered._\nAre you sure you want to continue? |
This is a prompt to confirm whether or not the compounds marked as being ignored in search results are erased from the library. |
To delete those compounds, click the [OK] button. This deletion cannot be undone. |
2161 |
The data files will not be saved, only the method file will be saved. |
You can not save the data file, because the data file is in read only or you may not allow to edit the file. Only the method file will be saved. |
----- |
2162 |
Method information could not be read from the selected method file. Please select another file. |
There are not consistent in detector configuration between current data file and the method file to be read. |
Select another file. |
2163 |
No UV library file is specified. |
The operation can not be executed because the library file has not been |
Specify the library file and then try again. |
2166 |
The MS library file cannot be loaded. The library file is used in other location. Try again after the use of the library file finishes. |
----- |
----- |
2167 |
Failed to copy the Compound Table. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
2168 |
The background compensation has been made on the data. Do you rewrite the background data in the data file? |
The background compensation was made on the currently open data file. |
To rewrite the background data in the data file, click [Yes]. To cancel, click [No]. |
2169 |
Delete background compensation? |
The background compensation was made on the currently open data file. |
To cancel background compensation of the data file, click [Yes]. To cancel, click [No]. |
216A |
Please input a folder name. |
----- |
----- |
216b |
You cannot create a new library file by selecting an existing one. |
A channel other than the extraction channel is not set to the Multi Chromatogram Table for the PDA detector when the Compound Table Wizard is |
Try starting [Compound Table Wizard] after setting a channel other than an extraction channel to the PDA Multi Chromatogram Table. |
216c |
Identify peak first. |
Any identified peak doesn't exist. |
Identify peak first. |
216d |
The retention index is not set in the compound table of the selected method file. |
Retention index does not exist in the method file, or the compound table is not specified. |
If the retention index does not exist, specify it. If the compound table is not specified, specify it. |
216e |
There is no identified peak in the selected data file. |
There is no identified peak in compound results table of selected data file. Or the compound information is not registered in compound table. |
Select that data file again after identifying analytes in compound table. To identify them. Process data on Data Analysis window of GCMS Postrun. Namely, load data file, create compound table and conduct 'Peak Integrate for all IDs' or other command of 'Quantitative' menu.\nOr, select the data file that has identified peaks. |
2182 |
Lambda Max/ lambda Min are not calculated. |
The lambda min/max of specified spectrum has not been calculated. |
Change the wavelength range for lambda min/max calculation in [spectrum] tab of data analysis parameter. |
2183 |
Chromatogram is out of stored wavelength range and peak cannot be detected. |
You try to extract the chromatogram at the wavelength where is out of wavelength range stored in data file. |
Confirm the wavelength range and try to extract the chromatogram within the wavelength range. |
2184 |
Peak integration has not been performed due to internal data error. |
A significant error occurred during peak detection.\nThere may be insufficient memory. |
Close all unnecessary applications. |
2185 |
Peak could not be detected due to peak integration error. |
A significant error occurred during peak detection.\nThere may be insufficient |
Close all unnecessary applications. |
2186 |
No PDA multi-chromatogram channel is set. |
When starting [Compound Table Wizard], channels other than extraction is not set to the PDA Multi-Chromatogram Table. |
Set channels other than extraction to the PDA MultiChromatogram Table, and start [Compound Table Wizard] again. |
2201 |
Failed to save acquired data. |
Some files may be corrupted. |
Re-install the application. |
2202 |
Invalid parameter ID was used. |
This error is rare. It may indicate that the system is unstable. |
Restart the PC and restart the software. |
2203 |
Designated component does not exist. |
The operation was executed toward the instrument which has not been exist. |
Confirm that system configuration on analytical instrument is the same as that on software. |
2204 |
Designated component is not connected. |
The operation was executed toward the instrument which has not been exist. |
Confirm that system configuration on analytical instrument is the same as that on software. |
2205 |
Failed to create data acquisition thread. |
Unexpected error occurred during data acquisition. |
Restart the PC and try again. |
2206 |
The specified system configuration does not match the hardware configuration. |
The information about the instrument configuration specified in this application is different from the hardware configuration. |
Specify the instrument configuration to be consistent with the hardware configuration. In addition, confirm the connection between instruments. |
2207 |
Data acquisition was stopped by instrument operation. |
----- |
----- |
2209 |
Some Instrument cannot start after the analysis runs. |
----- |
----- |
220C |
Please input the instrument |
The instrument name is not specified. |
Specify the instrument name. |
220D |
The instrument is not connected, or is already in use. |
The connection was broken when the PC tried to connect to the instrument. |
Check the operating environment including the instrument. |
220E |
Instruments configuration |
----- |
----- |
220F |
The instrument name cannot contain any of the following characters: \n\\ / : * ? "" < > | |
The instrument name uses invalid characters. |
|
2211 |
Combinations of the detectors are not correct.\nPlease change the detector mode or decrease the number of detectors. |
The combinations of the detectors are exceeding the limit of the software. |
When using the PDA detector as 2D mode, remove one detector from "Modules Used for Analysis". Or use the PDA detector as 3D mode. |
2280 |
This operation cannot be performed during data |
You try to execute the operation which prohibit during data acquisition. |
Try again after terminating data acquisition. |
2281 |
This operation can only be performed during data |
You try to execute the operation which prohibit during data acquisition. |
Try again after terminating data acquisition. |
2282 |
Function competing with this operation is working. |
You try to execute the operation which cannot be done because of conflict with the other one. |
Try again after terminating the other operation. |
2283 |
The instrument does not support this function. |
You try to execute the operation which has not been supported on the current hardware or software. |
The operation cannot be executed. |
2284 |
This function is running now. |
You try to execute the operation which is in processing. |
Try again after terminating the operation. |
2285 |
This operation cannot be performed in the state of Stand |
You try to acquire the data under not preparation of analysis. |
Prepare the analysis before analysis. |
2286 |
This operation cannot be performed in the state where Standby Acquisition is made. |
You try to execute the operation which cannot be done during the preparation of analysis. |
Execute the operation under the not preparation of analysis. |
2287 |
This operation was canceled by user. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
2288 |
Direct control command failed due to the communication error!_\nPlease check the instrument connection, and try again. |
The connection between the PC and analytical instrument or each instrument and controller may be wrong. |
Try again after reconstructing the connection accurately. |
228A |
The parameters are not specified. Please set the parameters. |
The parameter used for the operation to be executed has not been specified. |
Specify the appropriate parameters. |
228B |
Baseline Check cannot be performed because incorrect detector channel is specified. |
Baseline check can not be executed because the detector channel specified in baseline check parameter does not exist. |
Re-specify the detector channel in the baseline check parameter and then try again. |
228C |
The system check has been performed. |
----- |
----- |
228E |
The flow of nebulizing gas is insufficient. The LC pump will stop. |
The LC pump has been operating with insufficient flow of nebulizing gas. The LC pump was stopped to prevent the interface from being contaminated. |
Check the flow rate of the nebulizing gas on the Instrument Monitor. If the flow rate of the gas cannot be raised, there may be the cause in the piping or the cylinder for the gas. |
2290 |
The composition name is over the number of characters receivable with instrument. |
----- |
----- |
2293 |
The system check was |
----- |
----- |
2294 |
Failed to perform system check. |
----- |
----- |
2295 |
Do you save batch file and open Batch Editor Window? |
You're going to open Batch Editor Window on without saving a batch file. |
When [Yes] is clicked, Batch Editor Window is opened after saving the batch file. |
2296 |
[Method File] is empty.Do you add to Batch Table? |
[Add Batch Table] button is clicked without setting [Method File]. |
You choose [Yes] and set [Method File] in Batch Table, or choose [No] and set [Method File] in Batch Setting. |
2297 |
[Data File Name] is empty.Do you add to Batch Table? |
[Add Batch Table] button is clicked without setting [Data File Name]. |
You choose [Yes] and set [Data File Name] in Batch Table, or choose [No] and set [Data File Name] in Batch Setting. |
2301 |
There is little free disk space. Are you sure you want to acquire |
Very little disk space is available. |
Back up data, and delete unwanted files. |
2302 |
The data cannot be acquired due to insufficient disk space. |
Very little disk space is available. |
Back up data, and delete unwanted files. |
2303 |
The data acquisition was stopped due to insufficient disk space. |
Very little disk space is available. |
Back up data, and delete unwanted files. |
2034 |
The method dump was terminated because invalid |
Error on creation or save of the method file. |
Confirm that correct parameter is specified on the method file. If the attempt failed, re-create the method file again. |
2305 |
Level# is too large. Are you sure you want to continue? |
[Level#] set to the data acquisition settings exceeds [# of Calib. Levels] set in the quantitative parameters of the method. |
When [Yes] is selected, data acquisition is continued at the current settings. If necessary, correct the parameters after data acquisition. \n When [No] is selected, change [Level#] or [# of Calib. Levels]. \n To change the [# of Calib. Levels] settings, load the method file in the [Data Acquisition] or [Calibration Curve] windows, and change the settings in the sub-window for setting the quantitative parameters. |
2306 |
Please set the ISTD Amount. |
For quantitation of unknown samples by the internal standard method, an internal standard amount must be specified. |
Specify the internal standard amount in the sample login window. |
2307 |
Incorrect vial number. |
Invalid vial number is specified on the single run analysis or batch analysis. |
Specify the value suitable for Autosampler and sample rack used. |
2308 |
Please set the Sample Amount. |
For quantitation of unknown samples, the sample amount is required. |
Specify the sample amount in the sample login window. |
2309 |
Please set the Dilution Factor. |
For quantitation of unknown samples, the dilution factor is required. |
Specify the dilution factor in the sample login window. |
230A |
Failed to perform the single run. |
Failed to register the analysis to the batch queue because of invalid settings. |
Try again after correcting settings. |
230B |
Failed to change the end time of the data acquisition. |
You may have set a invalid value. |
Try again after correcting settings. |
230C |
Failed to plot. |
Either the system or the computer may have become unstable. |
1. Follow any instructions shown below the error message. If the problem re-occurs, proceed to remedy number 2.\n2. Restart the computer. Execute error checking at Windows disk properties. If the problem re-occurs, proceed to remedy number 3.\n OFF the instrument and reboot the computer. If the problem still persists, contact your Shimadzu Service Representative. |
230D |
Failed to stop the data acquisition. |
The system may have become unstable. |
1. Follow any instructions shown below the error message. If the problem re-occurs, proceed to remedy number 2.\n2. Reboot the computer. Execute error checking at Windows disk properties. If the problem re-occurs, proceed to remedy number 3.\n OFF the instrument and reboot the computer. If the problem still persists, contact your Shimadzu Service Representative. |
230E |
Stop the batch processing? |
When data acquisition is canceled during a batch run, select whether to cancel the batch analysis also. |
Select "Yes" to cancel the batch analysis as well, or "No" to proceed to the next row of the batch. |
2310 |
Please set a data file name. |
The data file name has not been specified on the single run or batch analysis. |
Try again after specifying the method file. |
2311 |
Please set a report format name. |
The report output has been checked, but the report file name has not been |
Specify the report file name or check off the report output. |
2312 |
You cannot close the window on data acquisition or realtime batch._\nPlease select [Stop] in the assistant bar, the toolbar or the menu. |
The data acquisition or the batch processing windows cannot be closed during data acquisition or batch processing |
First select "Stop" from the assistant bar, toolbar, or the menu; then close the window. |
2313 |
Incorrect tray number. |
The tray name specified on the single run or batch analysis has not been supported |
Specify the correct tray name suitable for Autosampler or sample tray used. |
2314 |
Time range settings are not correct. |
The specified start time is larger than end time. |
Specify acquisition start or end time. |
2315 |
The application is not selected. |
Application has not been specified in the |
Try again after specifying the type of application. |
2316 |
Instrument number is not selected. |
Instrument number has not been specified in the list. |
Try again after specifying the instrument number. |
2317 |
You cannot close the window during data plot._\nPlease select [Stop] in the data acquisition window or the menu. |
You try to exit the application under the plotting. |
Stop the plot by clicking the "Stop" button and then exit. |
2319 |
Current method file will be saved to run the data acquisition. |
You try to start the data acquisition under the method file has not been saved. |
If you apply the modification to the method file and then start the acquisition, click "OK". |
231A |
Memory for drawing the graph cannot be allocated.\nDrawing graph will be terminated. |
Memory for drawing graph can not be allocated. |
Restart the PC and try again. |
231B |
The specified background file does not exist._\nAre you sure you want to continue Single Run? |
The specified background file does not exist. |
If clicking "Yes", background compensation will not be executed. |
231C |
Lengthen the sampling period (shorten frequency) or shorten the analysis time. |
Data acquisition can not be executed because of too large memory consumption. |
Specify the larger sampling rate value or the smaller end time to reduce the memory consumption. |
231D |
The system configuration for this method is different from the current instrument configuration._\nAre you sure you want to adapt the configuration in the method to the current |
There are different in instrument configuration between the method file to be opened and the instrument setting. |
If clicking "Yes", the configuration setting in the method file will be replaced by the current instrument setting. |
231E |
Failed to acquire the data for graph display. |
Failed to acquire the data for drawing in the Graph View. |
Check the operating environment including the instrument. |
231F |
Set longer sampling period (lower frequency) or shorter maximum time for Baseline Check |
Baseline Check can not be executed because of too large memory consumption. |
Specify the larger sampling rate value or the smaller end time to or set shorter maximum time for Baseline Check Parameters reduce the memory consumption. |
2321 |
Chromatogram selected as the reference data is 16 or less. |
This is because 16 or more chromatograms have already been drawn as reference chromatograms at [Open Reference Data File] on the [File] menu. |
Try again after closing the reference chromatogram. |
2322 |
Set PsiPort Instrument Method file_name. |
The analysis cannot start as PsiPort Instrument Method file is not specified. |
Specify PsiPort Instrument Method file. |
2324 |
The [Data Acquisition] window is closed. Data acquisition cannot be started. |
Data acquisition cannot be started since the [Data Acquisition] window is not opened though the start button was pressed on the instrument side. |
Stop data acquisition on the instrument, open the [Data Acquisition] window, and click the start button. |
2325 |
The data acquisition sub-window is locked. Data acquisition cannot be started. |
Data acquisition cannot be started since the sub-window is locked though the start button was pressed on the instrument side. |
Stop data acquisition on the instrument, unlock the subwindow, and click the start button. |
2326 |
The popup sub-window is open. Data acquisition cannot be started. |
Data acquisition cannot be started since the popup sub-window is open though the start button was pressed on the instrument |
Stop data acquisition on the instrument, close the popup sub-window, and click the start button. |
2327 |
No method file is loaded. Data acquisition cannot be started. |
Data acquisition cannot be started since the method file is not opened in the [Data Acquisition] window though the start button was pressed on the instrument side. |
Stop data acquisition on the instrument, open the method file in the [Data Acquisition] window, and click the start button. |
2328 |
Failed to save the method file. Data acquisition cannot be started. |
Data acquisition cannot be started since the method file changed by the upload cannot be saved though the start button was pressed on the instrument side. |
Stop data acquisition on the instrument, enable writing to the method file, and click the start button. |
2329 |
No rights to perform Single Run. Data acquisition cannot be started. |
Data acquisition cannot be started since you do not have Single Run execution rights though the start button was pressed on the instrument side. |
Stop data acquisition on the instrument, and log in as a user with the Single Run execution rights. |
2330 |
The analysis program will be shut down. \n(Other user may start the analysis, or you open the analysis window with other PC)" |
Data acquisition in which another user was registered to the batch queue was started, or [Realtime Analysis] program currently logged in by the same user is in use on another computer. |
The program was ended. Start up the program after making sure that data acquisition is not being executed by another user. |
2331 |
You will take over data acquisition started by another user. When data acquisition is taken over, the currently logged in user name will be registered to the data file. Do you continue? |
Data acquisition started by another user is being executed on the instrument that you attempted to connect to. If the [Realtime Analysis] program is started up in this state, the currently logged in user will take that data acquisition. If data acquisition is taken over, the name of the user that took over data acquisition will be recorded to the data file that is obtained or the log that is recorded from here on. |
Select whether to start up the [Realtime Analysis] program by taking over data acquisition, or not to start up the [Realtime Analysis] program. |
2332 |
It is going to close the analysis window that is opening with other PC. Are you sure you want to continue? |
The [Realtime Analysis] program currently logged in by the same user is in use on another computer. Multiple [Realtime Analysis] programs cannot be started up simultaneously on a single instrument. Start up the [Realtime Analysis] program on this computer after exiting the [Realtime Analysis] program on the other computer. |
Select whether to close the [Realtime Analysis] program on another computer and start up the [Realtime Analysis] program on the current computer, or not to start up the [Realtime Analysis] program. |
2333 |
The system configuration for this template is different from the current instrument configuration._\nThe configuration in the method will be adapted to the current |
The system configuration set in the instrument settings differs from the system configuration saved in the template that you are attempting to open. |
Click [OK] to rewrite the system configuration in the method file with the current system configuration. |
2334 |
Download the Instrument Parameters?_\n(OK: Download and System On, Cancel: System On only) |
No instrument parameters have yet been downloaded, or the instrument parameters must be downloaded to perform data acquisition in accordance with instrument parameter settings that have been changed since downloading of the previous instrument parameters. |
Click [OK] to download the instrument parameters. |
2335 |
There is little free disk space. Are you sure you want to start the Realtime Batch? |
Very little disk space is available. |
Back up data, and delete unwanted files. |
2337 |
Are you sure you want to stop GC? |
Displayed for confirmation when a GC is stopped. |
To stop, click [Yes], and to cancel stop, click [No]. |
2338 |
Are you sure you want to stop GC Program simultaneously? |
Displayed for confirming whether or not to stop while executing the GC Program. |
To stop, click [Yes], and to cancel stop, click [No]. |
2339 |
The configuration file is still used. Restart the analysis. |
The analysis cannot start as the configuration file is still used in the last |
Restart the analysis. Restart the PC when it fails although you try multiple times. |
233A |
Can not shutdown Windows because LabSolutions is running. |
Windows cannot be shutdown because LabSolutions is connecting to the instrument. |
Shutdown Windows after stopping single run, batch analysis, or baseline check. The message is still displayed in spite of stopping all of them, select [Exit] menu displayed by right clicking the [LabSolutions Service] in the systray on the taskbar. |
2400 |
Start Batch Processing (Data Acquisition). |
This message confirms that batch processing has begun in the Real Time |
----- |
2401 |
Stop Batch Processing (Data Acquisition). |
This message confirms that batch processing was terminated in the Real Time Analysis Window. |
----- |
2402 |
End Batch Processing (Data Acquisition). |
This message informs the operator that batch processing in the Real Time Analysis Window has been completed. |
----- |
2403 |
Start Batch Processing (Postrun Analysis). |
This message confirms that batch processing has begun in the Postrun |
----- |
2404 |
Stop Batch Processing (Postrun Analysis). |
This message confirms that batch processing was terminated in the Postrun Analysis Window. |
----- |
2405 |
End Batch Processing (Postrun Analysis). |
This message informs the operator that batch processing in the Postrun Analysis Window has been completed. |
----- |
2406 |
Batch will start at Row# xxx (If you want to start at Row 1, please click No). |
This message confirms that the analysis will begin with the Batch Table row displayed. |
Click on "Yes" to begin with the row number displayed. Click on "No" to begin operation from line 1. Click on "Cancel" to terminate batch processing. Specify the starting Batch row number in the Start Row tab of the "Settings" window. (Select Settings from the table right- |
2407 |
Please specify method file. |
The method file name has not been specified in the Batch Table. |
Click on the arrow in the Method File column; then select the method file name. Alternately, type in the method file |
2408 |
Please specify data file. |
The data file name has not been specified in the Batch Table. |
Click on the arrow in the Data File column; then select the data file name. Alternately, type in the data file name. For data acquisition, a data file by that name will be created. For batch processing in the Postrun Analysis window, an existing data file will be re-processed. |
2409 |
Specify Report Format File. |
The report format file name has not been specified in the Batch Table. |
Click on the arrow in the Report File column; then select the report format file name. Alternately, type in the report format file name. |
240A |
Stop batch processing. Do you want to stop data acquisition immediately? \n[Yes] Stop data acquisition now.\n[No] Wait until data acquisition is completed.\n[Cancel] Cancel batch stop. |
Batch processing in progress in the Real Time Analysis window was terminated. Select how to terminate the data acquisition. |
To immediately terminate data acquisition in progress and stop batch processing, click on "Yes." To finish the current data acquisition run, then stop batch processing, click on "No." When the "Cancel" button is pressed, batch processing and data acquisition will continue. (To terminate the data acquisition for the current row only, and resume batch processing with the next row, click on the Real Time Analysis window to activate it, and select Stop from the |
240B |
Please enter batch table contents. |
Batch processing was initiated with an empty Batch Table. |
Complete the Batch Table information, or open a previously created batch file. |
240C |
Failed to create the QA/QC file. |
The operator attempted to create a QAQC file, but the output file could not be created. Another file with the same name may be in use in another window. |
When the file of the same name as the file to output by QA/QC processing is being used in another window, close that file. |
240D |
Failed in QA/QC Processing. |
The QAQC execution failed for an unknown reason. |
Check the QAQC settings. If correct, try the following:_1. Close the application, restart Windows, and execute error checking at Windows disk properties.\n2. Confirm that the data file used for the QAQC calculations can be opened properly. If not, the data file may be corrupt. |
240E |
Please do not use .htm as an extension. |
The .htm file extension has been specified as the output file name extension in the QAQC settings. |
The output file is a text file and cannot use the .htm file extension. Refer to QAQC settings for further information. |
240F |
Batch processing was stopped because of the system check result. |
A system check was carried out during batch processing and the results were evaluated. Batch processing was then terminated based on the Action settings. |
The evaluation parameters are specified in the Action column in the Batch Table. |
2410 |
Batch processing is paused because of the system check result. |
A system check was carried out during batch processing and the results were evaluated. Batch processing was then terminated based on the settings. |
The evaluation parameters are specified in the Action column in the Batch Table. |
2411 |
The user program has been performed because of the system check result. |
A system check was carried out during batch processing and the results were evaluated. The user program was then executed based on the Action settings. |
The evaluation parameters are specified in the Action column in the Batch Table. |
2412 |
Batch processing was stopped because of the test result. |
QAQC or system suitability was executed during batch processing and the results were evaluated. Batch processing was then terminated based on the Action |
The evaluation parameters are specified in the Action column in the Batch Table. |
2413 |
Batch processing is paused because of the test result. |
QAQC or system suitability was executed during batch processing and the results were evaluated. Batch processing was then paused based on the Action settings. |
The evaluation parameters are specified in the Action column in the Batch Table. |
2414 |
The user program has been performed because of test result. |
QAQC or system suitability was executed during batch processing and the results were evaluated. The user program was then launched based on the Action |
The evaluation parameters are specified in the Action column in the Batch Table. |
2415 |
The sample has been re-injected because of test result. |
This message means that sample was reinjected in accordance with the specification of Action triggered by QA/QC or system suitability result. |
----- |
2416 |
The sample of row# xxx was run because of test result. |
This message means that the analysis displayed row No. was executed in accordance with the specification of Action triggered by QA/QC or system suitability result. |
----- |
2417 |
The method file has been restored to the state before performing data acquisition because of test result. |
This message means that the method file was restored to the state before performing data acquisition in accordance with the specification of Action triggered by QA/QC or system suitability result. |
----- |
2418 |
Failed to run user program. |
The specified user program does not exist, or it may be unavailable. |
Confirm that the specified user program exists. Check whether it can be executed. The user program settings are specified in the Batch Table and in the Batch Action |
2419 |
[Initialize Calibration Curve] is not specified at Standard Type setting. Do you specify [Initialize Calibration Curve]? |
A "Standard"Sample Type creates a calibration curve. If the previous calibration curve is not cleared, new calibration points will be added to the previously created calibration curve. Confirm whether this is acceptable. |
Click on "Yes" to clear the previous calibration curve. This sets the first row of a series of "Standard" Sample Types to "Initialize Calibration Curve." If "No" is selected, new calibration points are added to the current method calibration curve. For further information, refer to the LabSolutions Help. |
241A |
The system configuration for this method is different from the current instrument configuration._\nThe configuration in the method will be adapted to the current instrument configuration. |
The configuration in the method is adapted to the current instrument configuration. |
The method file specified in the current Batch Table was created under a different system configuration. The method file system configuration will need to be changed so that it matches the current configuration. Parameters which differ from the current configuration are shown in brackets [ ].\nClick on the "OK" button to match the method file configuration to the current configuration, or click on "Cancel" to exit batch processing.\nBeginning of the batch processing is canceled in case of the batch analysis when the cancel button is clicked.The opening ofthe method file is stopped and it is closed the method file in case of the |
241B |
Level# is too large. |
[Level#] set to the data acquisition settings exceeds [# of Calib. Levels] set in the quantitative parameters of the method. |
If necessary, correct the parameters after data acquisition. |
241C |
The Injection Volume is invalid. |
An injection volume value which is invalid for the autosampler specified in the configuration has been set. |
Set the value within the appropriate range. |
241D |
The Vial# is invalid. |
An value which is invalid for the autosampler specified in the configuration has been set in the Batch Table vial |
Change the value to a valid value. |
241E |
The specified tuning file has not been tuned. Perform auto-tuning or manual tuning. |
The tuning file has not been tuned. |
Perform auto-tuning or manual tuning using the interface specified in the method file. |
241F |
The data file of the same name exists in the batch table. |
Two or more data files with same name have been set in the Batch Table. |
This is not allowed when acquiring data with a batch. However, files that are being reprocessed in a Batch Table can be specified more than once. |
2420 |
Failed to copy the report format file to data file. |
During batch processing, the report format file failed to be copied to the data file. |
The report format file may be corrupt. Close all applications and execute error checking at Windows disk properties. If the problem persists, re-create the report format file. |
2421 |
No file is identified. |
An attempt was made to edit the method of the clicked row or analyze its data, however, the file cannot be specified since, for example, the file name has not been set to that row. |
Make sure that the file on the clicked row is set correctly. |
2422 |
Daily Shutdown has been performed after batch processing ends. |
Daily Shutdown was executed after the completion of batch processing, as specified by selecting "Daily Shutdown after Last Batch" from the Batch menu. |
----- |
2423 |
You cannot close the windows while a batch is in progress. |
The operator attempted to close the batch processing window or the application while batch processing was in progress. |
----- |
2424 |
Batch processing will be stopped. |
Batch processing termination was executed while in progress in the Real Time Analysis window. This message is displayed when data acquisition has already been completed or before it starts. |
To terminate batch processing, click on "Yes." To allow the batch to continue, click on "No." |
2425 |
Data has not yet been acquired for selected data file. |
To activate Data Analysis windows, the data filename on batch table, which is paused, is selected. But it does not exist. |
This operation is not valid. |
2426 |
Data acquisition on the %dth row was stopped. |
This message means that the bath analysis was terminated by "Terminate" command or some kind of error. |
----- |
2427 |
Inappropriate Sample Type is set. Please confirm it in the QA/QC method parameter window. |
The sample type specified in the batch table is not included in the method. |
Check the sample type using in the QA/QC method parameter and then select a sample type that is included in the method. |
2428 |
Failed to back up the method file. |
Failed to back up method file. There is insufficient disk space or PC may be |
Try again this processing after confirming hard disk space and then re-starting PC. |
2429 |
Failed to restore the method file. |
Failed to restore method file. The method has been used in other program or PC may be unstable. |
Try again this processing after confirming the method has not been used in other program and then re-starting PC. |
242A |
The data file has been renamed because the same file_name already exists. |
When there exist same file name in starting data acquisition, the file name will be automatically renamed and the data acquisition will be continued. Such processing will be occurred in the case of re-injection in batch processing. |
----- |
242B |
Start Single Run. |
This message means that single run analysis was started. |
----- |
242C |
Stop Single Run. |
This message means that the single run analysis was terminated by "Terminate" command or some kind of error. |
----- |
242D |
End Single Run. |
This message means that single run analysis was terminated. |
----- |
242E |
Start Startup Processing. |
This message means that "Startup" processing was started. |
----- |
242F |
Stop Startup Processing. |
This message means that the "Startup" processing was terminated by "Terminate" command or some kind of error. |
----- |
2430 |
End Startup Processing. |
This message means that "Startup" processing was terminated. |
----- |
2431 |
Start Shutdown Processing. |
This message means that "Shutdown" processing was started. |
----- |
2432 |
Stop Shutdown Processing. |
This message means that the "Shutdown" processing was terminated by "Terminate" command or some kind of error. |
----- |
2433 |
End Shutdown Processing. |
This message means that "Shutdown" processing was terminated. |
----- |
2434 |
Failed to Startup. |
This message means that "Startup" processing was failed. |
Confirm the system configuration and instrument condition. |
2435 |
Failed to Shutdown. |
This message means that "Shutdown" processing was failed. |
Confirm the system configuration and instrument condition. |
2436 |
Invalid Vial# format. |
Vial # can be specified by "row "column"expression such as A10 or C15 when using sample rack allowing such a expression manner. Specified vial # is invalid or sample rack used may not be supported such a expression manner. |
Confirm the type of sample rack and specify the vial # correctly. |
2437 |
Are you sure you want to stop data acquisition? |
Software confirms you to stop data acquisition. |
Click [Yes] to stop data acquisition, otherwise click [No]. |
2438 |
Stop Data Acquisition. |
This message means that the single run analysis was terminated by "Terminate" command or some kind of error. |
----- |
2439 |
Number of channels in the data does not match in the method. |
There is difference in detector channel configuration between data and method file. |
You must use method file which was used in acquiring the data file or had same detector channel configuration as the data file. If you cannot find appropriately method file, you can create it by opening the data file with Data Analysis program and then selecting "Save Method As". |
243B |
Extended column settings are not correct. |
The setting of extended column used for batch add-in is invalid. |
Set the appropriate parameters. |
243C |
Failed to write the settings of the extended column to the method. |
Failed to write the settings of the extended column for the batch add-in to the method |
Enable writing to the method file. |
243D |
Current batch file will be saved to run batch processing. |
You try to start the batch process under the method file has not been saved. |
If you apply the modification to the method file and then start the batch process, click "OK". |
243E |
Start Data Acquisition. |
Message for notifying that data acquisition was started. |
----- |
243F |
Please set level number. |
Level # has not been specified. |
Specify the Level # in the batch table. |
2440 |
The instrument is performing other task. Data acquisition cannot be started. |
The data acquisition can not be executed because the instrument is not in "Ready". |
Operate under "Ready" state. |
2441 |
Failed to output ASCII file. |
----- |
----- |
2442 |
Failed to output AIA file. |
----- |
----- |
2443 |
Pause Batch Processing (Data Acquisition) |
Analysis batch processing has been paused by the user or the preset action triggered automatically. This message is for leaving a record in the log. |
----- |
2444 |
Resume Batch Processing (Data Acquisition) |
Analysis batch processing has been paused by the user or the preset action triggered automatically. This message is for leaving a record in the log. |
----- |
2445 |
The format specified in the "Selection Line(s)" is invalid. |
----- |
To specify consecutive lines, use a hyphen (-). Ex.) 2-5 \nTo specify separate lines, use commas (,) for delimiting line numbers. Ex) 3,5,7 \nYou can use a combination of both. Ex) 2-5,7,10,15-20 |
2446 |
Set Psiport Instrument Method filename. |
The analysis cannot start as PsiPort Instrument Method file is not specified. |
Specify PsiPort Instrument Method file. |
2447 |
The batch processing execution information will be saved to the current batch file. |
The current batch file must be saved since the execution information of batch processing has been changed. |
To save and execute the batch file, click [OK]. To cancel, click [Cancel]. |
2448 |
Please specify batch file. |
The batch file name has not been entered. |
Enter the batch file name. |
2449 |
A Repeat Batch Run is set. Are you sure you want to perform Realtime Batch? |
Message for notifying that [Repeat batch run] has been specified and started in realtime batch. |
To perform [Repeat batch run], click [OK]. To cancel, click [Cancel]. |
244A |
The data file name is not set to be automatically created. Are you sure you want to perform Realtime Batch? |
This message is displayed when the data file name has not been set to be automatically created by [Repeat batch run] in realtime batch. |
To performRealtime batch, click [OK]. To cancel, click [Cancel]. |
244B |
To perform the noise calculation by the EP settings, the name of the background file must be set. |
Noise calculation has been specified to be performed in the EP settings, however, the background file name is not set. |
To perform the noise calculation by the EP settings, set the name of the background file. |
244C |
Batch processing was stopped according to the action setting. |
Batch processing was canceled since a row set as [Test: Any] - [Action: Stop] to batch action was executed. |
----- |
244D |
Batch processing is paused according to the action setting. |
Batch processing was paused since a row set as [Test: Any] - [Action: Pause] to batch action was executed. |
----- |
244E |
The program has been performed according to the action setting. |
The user program was executed since a row set as [Test: Any] - [Action: Execute User Program] to batch action was |
----- |
244F |
Injection has been performed again according to the action setting. |
Message for notifying that re-injection of the sample was executed since a row set as [Test: Any] - [Action: Reinject] to batch action was executed. |
----- |
2450 |
The xxth row has been performed according to the action setting. |
Message for notifying that data acquisition of the currently displayed row No. was executed since a row set as [Test: Any] - [Action: Goto] to batch action was |
----- |
2451 |
The method file has been returned to the pre-execution state according to the action setting. |
Message for notifying that the method file was returned to the state before execution since a row set as [Test: Any] - [Action: Restore Method] to batch action was executed. |
----- |
2452 |
Invalid summary report settings. |
[Summary Start] or [Summary End] have not been specified on [Summary Type] column in the batch file. |
Please specify [Summary Start] or [Summary End]. |
2453 |
Method file specified for the Startup is not found. |
Method file specified for the Startup is not found. |
Confirm the method file name specified for the Startup. |
2454 |
Specified file type for the Startup is invalid. |
Method file specified for the Startup can not be opened. |
Confirm the method file specified for the Startup. |
2455 |
Method file specified for the Shutdown is not found. |
Method file specified for the Shutdown is not found. |
Confirm the method file name specified for the Shutdown. |
2456 |
Specified file type for the Shutdown is invalid. |
Method file specified for the Shutdown can not be opened. |
Confirm the method file specified for the Shutdown. |
2458 |
The total size of data files which is necessary for summary report exceeded the tolerance, so that failed to output summary report. |
The total size of data files which is necessary for summary report exceeded the tolerance, so that failed to output summary report. |
Please reduce the number of data files to use for summary. |
24A0 |
Failed to access to Analysis Server. |
Either your PC or the Analysis Server is disconnected from the network or the Analysis Server is currently down. |
They may have been disconnected from the network. Check to see if they are properly connected to the network. \nWhen the Analysis Server is down, contact your network |
24A1 |
The communication failure with the analytical server occurred. \nTo connect with the server properly, LabSolutions should be restarted._\nAre you sure you want to continue the processing? |
Either your PC or the Analysis Server is disconnected from the network or the Analysis Server is currently down. |
They may have been disconnected from the network. Check to see if they are properly connected to the network. \nWhen the Analysis Server is down, contact your network administrator. If you want to continue your operation under this condition, select [Yes]. |
24A2 |
Failed to get the system administration information. |
1. The network may have been disconnected. \n2. The system administration database file may be corrupted or being used by another application. |
1. Check that your PC or the one in which the system administration database file resides is connected to the network properly. \nRun the command again after the other application finishes it's processing. If the error recurs even after you re-run the command, the system administration database file may be corrupted. |
24A3 |
You are registering the same batch file as the already registered. Are you sure you want |
You attempted to register a batch file to the batch queue under the same name with one of the batch files already |
Do not give a duplicate name to a batch file to register with any of the names of batch files already registered without a compelling reason. |
24A7 |
The data file is not found on a certain row. Are you sure you want to continue Postrun Batch? |
This is displayed when there is a row where the data file set in postrun batch is not found. |
To continue processing, click [Yes]. To cancel, click [No]. |
24A9 |
Realtime Batch ended since the Goto row in the action settings were not selected as the execution row. |
An attempt was made to execute data acquisition on the specified row No. according to the specified action from the QA/QC check results, however, realtime batch ended since the Goto row was not included in the execution row. |
Set the row No. specified by Goto in batch action so that it is included in the executed rows. |
24AA |
Start row should be set to odd number._\nThis batch table will start from previous row. |
The specified start row is an evennumbered row in the case of batch processing on a dual line, or the start of the selected range in partial execution is |
This is no problem as it is since batch processing is started from the previous row. A warning message will no longer be output if the start row or start of the selected range in partial execution is set as an odd-numbered row. |
24AB |
In order to start the batch processing, you have to release the waiting state of the batch or to click the start button on the batch queue dialog. \nDo you open the batch queue dialog? |
Click "Yes"if you want to open batch queue dialog , click "No" if you don't want open it. |
----- |
24AC |
The batch processing registered in the queue waits to start until clicking start button on the batch queue dialog. |
The batch processing still remains in the cue. |
On the batch cue window, click [Start] button and select the batch processing of the state of the standby or delete it,if it is unnecessary. |
24AD |
Start from row# %d-1, 2 (If you start first row, please click ""No""). |
This message confirms that the analysis will begin with the Batch Table row displayed. |
Click on "Yes" to begin with the row number displayed. Click on "No" to begin operation from line 1. Click on "Cancel" to terminate batch processing. Specify the starting Batch row number in the Start Row tab of the "Settings" window. (Select Settings from the table right- |
24AE |
Please set the %s parameters of method files to unify. |
The difference of the parameter was found in some method files for the unit that has to be the same parameters. |
Please use only the method file as which the parameter of the unit is the same. |
24B0 |
Start row is over the maximum row number. Stop Batch Processing(If you start first row, |
The row number set in the "Start from #"(batch settings) exceeds the maximum row in the batch table. |
Check the number set in the "Start from #" in the batch settings, and set a number within the maximum row of the batch table. |
24B1 |
Batch processing was stopped because the start row is over the maximum row number. |
The row number set in the "Start from #"(batch settings) exceeds the maximum row in the batch table. |
Check the number set in the "Start from #" in the batch settings, and set a number within the maximum row of the batch table. |
24B2 |
A Single Run of dual line can't be processed under the current instrument configuration. |
It tried to execute a single run of dual line though a current system configuration was one line. |
Please execute a single run matched to the system configuration setting. Or, please set the system configuration matched to the setting of a single run. |
24B3 |
The dual line batch table can't be processed under the current instrument configuration. |
It tried to execute the batch processing with the batch table of two lines though a current system confgiuration setting was |
Please execute the batch processing matched to the system configuration setting. Or, please set the system configuration matched to the setting of the batch. |
24B4 |
Start/End Setting is invalid. Batch runs all rows. |
This message confirms that the analysis will run with the all Batch Table rows. |
Click on "OK" to run all rows.Click on "Cancel" to terminate batch processing. Specify the start/End Batch row number in the General tab of the "Settings" window. (Select Settings from the table right-mouse menu.) |
24B5 |
The file has been renamed because the file name is too long. |
When file name is too long in starting data acquisition, the file name will be automatically renamed and the data acquisition will be continued. Such processing will be occurred in the case of re-injection in batch processing. |
----- |
2501 |
There is no detected peak |
Peaks have not been detected for this data. |
Quantitative calculation or manual peak integration can be executed only when a peak is detected. Change the peak integration parameters in data analysis and adjust so that peaks are detected. |
2502 |
Peak Identification has not been performed. |
Peak identification has not been performed for this data. |
Until the peaks have been identified, neither quantitation nor manual peak processing can be performed. Perform peak identification from the Qualitative menu in the Data Analysis window. In this software, peak identification is performed automatically during the peak integration |
2503 |
Quantitative Processing has not been performed. |
Quantitation has not been performed for this data. |
Quantitate from the Quantitative menu in the Data Analysis window. |
2504 |
No Calibration Curve. |
The data cannot be quantitated without a calibration curve. |
Either create a calibration curve with Batch processing, or create one in the Calibration Curve window. For more information, see the LabSolutions Help. |
2505 |
Internal standard peak is not found. |
The peak associated with the internal standard has not been detected. |
If the Internal Standard Method is selected as the Quantitative Method, the internal standard peak must be identified in order to generate a calibration curve and calculate concentrations. In the Compound Table, confirm the ID number of the internal standard (ISTD or ISTD & Reference); then check that peak integration and peak identification have been conducted for that ID in the Data Analysis window.\nIf the peak has not been detected, change the peak integration parameters.\nIf the peak has been detected, but failed to be identified, check the |
2506 |
No peak is identified. |
The peak was not identified. |
1. Verify the retention time in the Compound Table2. Confirm that the peak area is not below the minimum peak |
2507 |
No item to print. |
Nothing has been selected for printing. |
Paste the report item correctly in the report format file; in the item properties, select "Print Item." |
250A |
Invalid peak type. |
The peak Type is incorrect. |
Check the Type in the Compound Table. In particular, when the Internal Standard Method is specified, an internal standard is required. |
250B |
Averaging repetitions for one point exceeds ten (10) |
More than 10 standards specify the same level. |
The maximum number of repetitions for averaging is 10. When using a batch, select Standard (Initialize Calibration Curve) as the Type to clear the previous calibration curve. Otherwise, new standards will continue to be averaged. Normally, the first standard in the Batch Table should initialize the calibration curve. |
250E |
An error occurred when Peak Integration was performed. |
Peak integration was not completed due to insufficient memory. |
1. Execute peak integration again. 2. If the error re-occurs, verify the peak integration parameters, and reduce the number of peaks to be detected. |
250F |
No data file is registered. |
Because there is no data specified for the calibration point, processing cannot proceed. |
The calibration point does not exist for the specified calibration curve. A calibration curve cannot be edited by removing a calibration point. In the Calibration Curve window, add another calibration point to the calibration curve. Or, generate the calibration curve using batch |
2510 |
Sample Amount is not specified. |
The Sample Amount has not been specified and quantitation cannot be performed. |
Enter a value other than 0 for Sample Amount. For Batch processing, set the Sample Amount in the Batch Table. In the Data Analysis window, open the Sample Information, and specify a Sample Amount. The quantitation requires a dilution factor. Verify that this value is not set to 0.\nThe final concentration is obtained with the following equation.\nFinal concentration = (Concentration calculated from calibration curve) ´ (Dilution factor) / (Sample Amount) |
2511 |
No peak is selected. |
The peak to be processed has not been selected. |
Select the target peak. |
2512 |
Calculate concentration for all compounds? |
Confirm whether to calculate the concentrations. "Yes" is recommended. Although this is not required if the data is associated with a standard, the processing takes very little time. |
Confirmation as to whether or not to perform quantitative calculation. For example, when the data is a standard sample, there is no need for quantitative calculation though processing takes up almost no time at all. Normally, select [Yes]. |
2513 |
No peak is detected. |
No peaks have been detected. |
Verify the peak integration parameters. Decrease the Width or Slope values to facilitate peak detection. |
2517 |
Internal Standard Compound(s) required when using Internal Standard Method. |
Although the Internal Standard Method was selected as the Quantitative Method, there is no internal standard listed in the Compound Table. |
Specify an internal standard in the Quantitation Table. The Type should be either ISTD or ISTD&Ref. |
2518 |
Number of detected peaks (or slices) exceeded maximum number. |
The number of peaks exceeded the maximum number that can be processed in one chromatogram |
The number of peaks detected has exceeded the maximum detectable number of peaks (1000 peaks). Reduce the number of peaks to be detected using the following examples as a reference. \n- Increase the value of Slope, a Peak Integration parameter. \n- Prevent unwanted peaks from being detected using the Integrate ON/OFF function in an integration time program. |
2519 |
The current parameters will be copied to all channels. \nDo you copy the time programs, too? |
Confirm whether to copy the peak integration parameters of the current channel to all the channels. |
Click "Yes" if you want to copy the parameters with the time program, or click "No" if you want to copy without the time program. The operation is cancelled if you click |
251A |
Only one internal standard can be registered in an ISTD group. |
Multiple internal standards were assigned to a single IS group. |
The ISTD Group number specified in the Compound Table groups compounds by ISTD group. Specify only one internal standard per group. Quantitation is performed with the internal standard for that group. |
251B |
No internal standard registered in the ISTD group. |
No internal standard was registered to the IS group. |
The ISTD Group number specified in the Compound Table groups compounds by ISTD group. Specify only one internal standard per group. Quantitation is performed with the internal standard for that group. |
251C |
Internal standard compounds are not required. |
If the Internal Standard Method was not specified as the Quantitative Method, there is no need to specify an internal standard. |
Do not use the internal standard as a Compound Table Type. |
251D |
Please set concentration. |
Without a specified concentration, the calibration curve cannot be generated. |
Each level in the Compound Table requires a concentration. Except for Level #1, a value of 0 cannot be |
251E |
Please set band width in the Quantitative Table |
Identification processing cannot be performed because a band time width was not specified. |
When Band is selected for the Identification Method, a Band Time is required. Specify the Band time in the Compound Table. Peaks with a retention time within the band will be identified. If the Compound Table Band column is hidden, select it for display from the Table Style |
251F |
Please set IS group number. |
No IS group number has been specified in the Compound Table. |
The ISTD Group number specified in the Compound Table groups compounds by ISTD group. Specify only one internal standard per group. Quantitation is performed with the internal standard for that group. |
2521 |
Retention time in quantitative table has been corrected. |
Because identification was successfully performed, the retention time for the standard was updated in the Compound |
----- |
2522 |
A level of calibration curve exceeds maximum set in method. |
When the calibration curve was applied, it was not consistent with the maximum number of calibration points set in the |
Create a new calibration curve with Batch processing. |
2523 |
Calibration curves have been updated in the data file. |
When the quantitative parameters and Compound Table are edited, the calibration curve may be initialized so that they all correspond. |
This is a confirmation message. |
2524 |
Some IDs have no compound name. These IDs will be deleted._\nAre you sure you want to continue? |
There is ID to which the component name is not set by editing the Compound Table. |
The component name is always necessary to make it recognize as a compound on the table. Please set it to the compound table. When Yes clicks on a button, the line without the compound name is deleted. It returns to former screen when No clicks on a button. |
2525 |
Quantitative Parameters and/or Compound Table in Data do not match with those in Method. |
The quantitative parameters in the Method file do not match those in either the data file to be processed or the Compound |
For all to correspond, the quantitative parameters in the Method, the Compound Table, and the calibration curve information are copied to the data file. |
2526 |
No information is specified in the Compound Table. |
Processing cannot continue because the Compound Table information has not been entered. |
Enter the target compound information in the Compound Table. |
2527 |
The [Method View] is in the Edit mode. Do you apply the new settings? |
The sub-window cannot be ended, or the file cannot be saved or loaded if the Compound Table remains in the Edit mode at [Data Analysis] - [Calibration]. You must select whether to apply or discard the content of the edited Compound Table. |
Select whether to apply or discard the new settings. Clicking the [Yes] button applies the new settings, and clicking the [No] button discards the new settings. |
2529 |
The [Method View] is now being edited. |
|
|
2580 |
Method file has been renamed. Settings will be cleared. |
----- |
----- |
2581 |
# of levels has been changed. Settings will be cleared. |
----- |
----- |
2591 |
Distillation Data Analysis |
Distillation data analysis error occurs. |
Please refer to chapter 7 in the SIMDIS GC Software |
2600 |
Invalid macro strings. |
Specified parameter or function is invalid. Following strings can be used as parameter and function.\n<Parameter>\nRetTime- >retention time\nArea->peak area\nHeight>peak height\nConc->sample |
Confirm that parameter and function can be inputted correctly. |
2601 |
Invalid calculation formula. |
There is invalid formula, such as divided by zero. |
Input adequate formula. |
2602 |
Failed to load the rack definition information. |
This operation is not available. A probable cause is a problem in the network connection in the case of a network configuration. |
Try again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
2603 |
Failed to change the rack definition information. |
The new settings cannot be saved. A probable cause is a problem in the network connection in the case of a network configuration. |
Set again. If this does not remedy the problem, restart the computer, and try setting again. Execute again after making sure that the network is connected correctly in the case of a network configuration. |
2604 |
Cannot open Rack Definition window while another user opens the window. |
Cannot open Rack Definition window while another user opens the window. |
Wait for the other user to close Rack Definition window. |
2642 |
Chromatogram of specified channel does not exist in buffer. |
A significant error occurred on displaying the chromatogram. There may be insufficient memory. |
Close all unnecessary applications. |
2643 |
Invalid data format. |
A significant error occurred during data acquisition. There may be insufficient memory. |
Close all unnecessary applications. |
2680 |
Save current layout file? |
The Layout File was changed. |
This confirms whether the file is to be saved with its changed contents and then closed, or closed without |
2681 |
Failed to open all of specified data files. |
The number of files specified exceeds maximum number. |
Decrease the number of data files. |
2682 |
Save current browsing file? |
The Browsing File was changed. |
This confirms whether the file is to be saved with its changed contents and then closed, or closed without |
2683 |
Delete Data File from Quantitative Table? Notes standard samples are removed. |
----- |
To delete the data file, click the [OK] button. To terminate, click the [Cancel] button. |
2684 |
This file cannot be registered as unknown sample since it is standard sample registered in another caliblation curve. |
It tried to register the file registered as a standard sample in either mode of MS data/LC data as an unknown sample. |
Please switch [MS Data] or [LC Data] sub-menu from [Display] menu, and confirm whether the file is registered as a standard sample in other modes. Please remove from a standard sample in other modes or make the |
2685 |
Cannot remove standard sample files. |
----- |
|
2686 |
Create a new method file? |
Message for confirming whether or not to create a new method file when making a new browsing file. |
To create a new method file, click [Yes], and to load an existing method file, click [No]. To cancel the operation, click [Cancel]. |
2688 |
The line of the specified data or method is not corresponding to the current line. |
The line of the data or method that tried to be opened is not corresponding to the line of method or data read by the Quant Browser now. In the Quant Browser, only the data or method of the same line can be |
Please open specified data or method after closing all data files on the Quant Browser. Or, please open another data or method which is corresponding to the current line. |
2689 |
The instrument type of the specified data is not corresponding to the current instrument type. |
The instrument type of the data that tried to be opened is not corresponding to the type of method or data read by the Quant Browser now. In the Quant Browser, only the data or method of the same type can |
Please open specified data after closing all data files on the Quant Browser. Or, please open another data which is corresponding to the current instrument type. |
268A |
Save current method file. Are you sure you want to continue? |
This message is displayed when the method file is modified in Quant Browser. |
Click [OK], method file is saved and browsing file is saved.\nClick [Cancel], method file is not saved and browsing file is not saved, either. |
2700 |
Baseline check result passed. |
----- |
----- |
2701 |
Baseline check result failed. |
----- |
----- |
2702 |
Startup was terminated with error. |
The startup was terminated by some kind of error. |
Confirm the setting of startup or the error from instrument if it exists. |
2703 |
Shutdown was terminated with error. |
The shutdown was terminated by some kind of error. |
Confirm the setting of shutdown or the error from instrument if it exists. |
2704 |
AutoPurge was terminated with error. |
The autopurge was terminated by some kind of error. |
Confirm the setting ofAutoPurge or the error from instrument if it exists. |
2705 |
An error occurred during download scan parameters to instrument. |
The setting of scan parameter is invalid or there is problem the connection between instruments. |
Try again after ensuring that the parameter settings and the connection between instruments. |
2706 |
An error occurred at start scanning. |
The setting of scan parameter is invalid or there is problem the connection between instruments. |
Try again after ensuring that the parameter settings and the connection between instruments. |
2707 |
An error occurred during scanning. |
The setting of scan parameter is invalid or there is problem the connection between instruments. |
Try again after ensuring that the parameter settings and the connection between instruments. |
2708 |
An error occurred at getting spectrum. |
The setting of scan parameter is invalid or there is problem the connection between instruments. |
Try again after ensuring that the parameter settings and the connection between instruments. |
2709 |
Startup was terminated. |
----- |
----- |
270A |
Baseline check result failed. |
----- |
----- |
270B |
Baseline check result failed. |
----- |
----- |
270C |
Baseline check result failed. |
----- |
----- |
270D |
Start the baseline check. |
This message is output to the log when the baseline check is started. |
----- |
270E |
Stop the baseline check. |
This message is output to the log when the baseline check is stopped by an error. |
----- |
270F |
Pass the result of baseline check. |
This message is output to the log when the baseline check ends and the check result is pass. |
----- |
2710 |
Fail the result of baseline check. |
This message is output to the log when the baseline check ends and the check result is fail. |
----- |
2711 |
Scan cannot start because Lamp is Off. |
Spectrum scan can not be executed because lamp is turned OFF. |
Try again spectrum scan after turning on the lamp. |
2713 |
AutoPurge will be performed by the method settings. Do you continue? |
This message is displayed when AutoPurge is started by the direct button on the instrument control bar. |
To save the method for the data acquisition child and execute an AutoPurge, click [OK]. To cancel, click [Cancel]. |
2715 |
On the SIL-10A Series, purge execution will not stop. Do you continue? |
----- |
----- |
2716 |
Start AutoPurge processing. |
----- |
----- |
2717 |
End AutoPurge processing. |
----- |
----- |
2718 |
Data file Restore has been completed successfully. |
----- |
----- |
2719 |
Failed to restore the data file. |
----- |
----- |
271A |
Buffering data has been deleted successfully. |
----- |
----- |
271B |
Failed to delete buffering data. |
----- |
----- |
271C |
Data cannot be acquired from the instrument._\nIf the power supply is on in the instrument, it continues to acquire data._\nRestore the data may be available by rebooting this application after the instrument finishes acquiring data._\nWait for the instrument to finish acquiring data without turning Off |
Data can no longer be acquired from the instrument due to a communications error in a configuration which allows data repairs. Note, however that if the instrument is turned ON, it continues to acquire data and stores the data. |
Restore the data by rebooting this application after the instrument finishes acquiring data. Wait for the instrument to finish acquiring data without turning OFF the instrument. |
2741 |
Specified wavelength is out of acquired wavelength range. |
----- |
----- |
2742 |
Status data is not correct. |
You try to display the status data which does not exist in data file. |
Specify the status trace which will be acquired from instruments. If the error has been occurred in spite of the appropriate setting, the data file may be corrupted. |
2743 |
PDA 3D data does not exist. |
Some kind of process for the PDA data was executed in spite of lack of PDA data in the data file. |
Close the PDA data analysis window. If the attempt failed, the file may be corrupted. |
2744 |
Overlapped storage time does not exist between two Data Files. |
The calculation between two PDA data (such as background correction) had been executed, but the calculation was failed because there was different in data analysis time between them. |
Specify another file for the calculation. |
2745 |
Sampling rates do not match between two data files. |
The calculation between two PDA data (such as background correction) had been executed, but the calculation was failed because there was different in sampling rate between them. |
Specify another file for the calculation. |
2746 |
Chromatogram of specified channel does not exist. |
The data processing had been executed to the channel which there has not exist in data file. |
Use method file in which channel information is consistent with that in data file. If you are not available the method file, you can create the method file by executing the "Save Method As" menu in data analysis window. |
2747 |
No PDA data. |
Some kind of process for the PDA data was executed in spite of lack of PDA data in the data file. |
Close the PDA data analysis window. If the attempt failed, the file may be corrupted. |
2781 |
Please set the 1st coefficient. |
----- |
----- |
2782 |
It needs Reference Compound(s) for Relative Retention Time Identification Method. |
At least one reference compound is necessary in [Type] column of compound table for "Relative Rt." as Peak Identification Method. |
Set "Reference" or "ISTD&Reference" to [Type] cell in compound table. |
2783 |
The [Method View] will be returned from the Edit mode to the View mode. The current calibration curve will be deleted since the parameters for making the calibration curve were changed. Are you sure you want |
The calibration curve cannot be created from the information contained in the current data if the basic settings of the calibration curve are changed since calibration curves are normally created from multiple data. |
Make a new calibration curve since the calibration curve will be deleted when processing is continued. |
2784 |
Group Calibration isn't supported for Internal Standard Method. |
Grouping before calibration cannot be used when [Quantitative Method] has been set to [Internal Standard]. |
This combination of setting is not supported. |
2785 |
Group Calibration isn't supported for Manual RF. |
Grouping before calibration cannot be used when [Curve Fit Type] has been set to [Mean RF]. |
This combination of setting is not supported. |
2786 |
Calibration Curves will be deleted._\nAre you sure you want to continue? |
Calibration curve stored in method file will be deleted because the parameter for the creation of calibration curve has been changed. |
Click [Yes] to delete the calibration curve in the method file. |
2787 |
[Force Through] cannot be selected in exponential |
This is Mathmatical restriction. |
Make another selection. |
2788 |
Please set # of Calib. Levels to 2 or more for exponential curve. |
This is Mathmatical restriction. |
Set 2 or more Calib Levels number. |
2789 |
No data file has been included in the calibration information. |
There is no data file record in the calibration curve information. |
The calibration curve was not created properly. Re-create the calibration curve either in batch processing or the calibration curve window. |
278A |
Current parameters are transferred to all detectors. |
----- |
----- |
278B |
Please put check_mark to at least one peak. |
----- |
----- |
278C |
When the method for calculating noise is rms, please select the whole range or specified range as the time range. |
----- |
----- |
278E |
The number of selected peaks exceeds the maximum number of lines of the compound table. |
In the Compound Table Wizard, the number of selected peaks exceeds the maximum number(=1000) of lines of the compound table. |
Please decrease the number of selected peaks to become 1000 or less. |
2800 |
The similarity calculation failed because the wavelength range was not overlapped. |
The wavelength range for similarity calculation has not been overlapped. |
Specify another file for similarity calculation or change the wavelength range in data analysis parameter. |
2842 |
Incorrect tray number. |
The tray name has not been supported. |
----- |
2843 |
Invalid sample rack set in method. Please confirm sample rack on 'Advanced' in Instrument Parameters View. |
Invalid sample rack set in method. |
Please confirm sample rack on 'Advanced' in Instrument Parameters View. |
2844 |
Acquisition is disabled for all detectors. Are you sure you want to continue data acquisition? |
You try to run the analysis under the Data Acquisition Off. |
If you click "Yes", Chromatogram data will not be saved in the acquired data. If you want to save them, click "No" and then check the detector check box in the Data Acquisition in the Instrument Parameter View. |
2845 |
Data acquisition has been skipped according to the judgment of the baseline check |
----- |
----- |
2846 |
Batch processing is paused because of the baseline check result. |
A baseline check was carried out during batch processing and the results were evaluated. Batch processing was then terminated based on the settings. |
The evaluation parameters are specified in the Action column in the Batch Table. |
2847 |
The baseline check parameter of a method needs to be set to perform a baseline check. |
Although you have specified the batch table to run the baseline check, the method has not been set to run the |
To run the baseline check, set up the noise or drift in the baseline check parameter setting in addition to the batch table setting. |
2848 |
Invalid sample rack set in method. Please confirm sample rack on the insturment method. |
Invalid sample rack set in method. |
Please confirm sample rack on the insturment method. |
2880 |
Are you sure you want to delete row(s) of selected spectra? |
The number of spectra to be displayed exceeds the upper limit. |
Check off the "display" check box in the spectrum table to reduce the number of spectra. |
2881 |
Are you sure you delete row(s) of selected spectra? |
You try to delete the selected row(s) in the spectrum table. |
The selected row(s) will be deleted when clicking [Yes]. |
2882 |
The specified UV library file is already registered. |
The specified library file name has been already specified. |
Delete another library file name. |
2900 |
The system suitability was executed. |
This message is output when the system suitability test was completed. |
----- |
2901 |
An error occurred during initialization of the system |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
2903 |
An error occurred when data was written to the work file of the system suitability. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
2904 |
An error occurred when data was read from the work file of the system suitability. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
2905 |
The system suitability test settings could not be acquired from the method file. |
The settings could not be acquired from the method file at execution of the system suitability test. |
Set execution of the system suitability test in the system suitability settings of the method file to use, and then set the parameters. |
2906 |
An error occurred when the summary calculation was performed. |
An error occurred when the summary calculation was performed. |
Check the used data file to see if it is corrupted, and also check the operating environment. Also, try re-installing this software. |
2907 |
An error occurred while the summary data was being |
An error occurred when the summary data was acquired. |
Check the used data file to see if it is corrupted, and also check the operating environment. Also, try re-installing this |
2908 |
An error occurred while the summary data was being output to file. |
Because the file to be output has already been opened, this error may be occurred. |
After closing the file, try again. |
2909 |
An error occurred while the system suitability test was being performed. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
290A |
An error occurred while the system suitability test results were being output to HTML file. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
290B |
An error occurred while the system suitability test results were being output to text file. |
Because the file to be output has already been opened, this error may be occurred. |
After closing the file, try again. |
290C |
An error occurred when the system suitability test results were being displayed. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
290D |
An error occurred while the system suitability test results were being printed. |
A probable cause is that the software has not been installed correctly. |
Contact your Shimadzu representative. |
2940 |
Invalid calibration data format. |
By import of the calibration curve table, the form of the text file is wrong. |
Correct the text file specified by import. |
2941 |
Retention time in The Calibration Table is duplicate. |
Calibration curve cannot be drawn since multiple data of same retention time are set in the Calibration Table. |
Calibration curves will not be drawn if calibration points of the same retention time are set. Modify the retention time. |
2942 |
It is sorted by time. |
Calibration points registered in the Calibration Table are not in ascending |
The system will sort the data automatically. |
2943 |
Calibration curve is not monotone decreasing. \nContinue to save? |
There is a part not monotone decreasing on calibration curve obtained by the selected approximation. |
Verify molecular weight range to be applied. It is no problem if the curve within the range is monotone decreasing. |
2944 |
Delete calibration point that made from deleted file? |
Confirmation to delete calibration point data made from the data file to be deleted. |
Proceed the operation after verification. |
2945 |
Could not draw calibration curve because not enough calibration points. |
Enough number of calibration points to draw calibration curve are not set in the Calibration Table. |
The number of calibration points to be set varies depending on the applied approximation. Linear" requires 2 points. "3rd order" requires 4 points or more." |
2946 |
Could not calculate molecular distribution because not set maximum slice number. |
Slice calculation required for GPC calculation cannot be executed because maximum slice number is set to 0 in [Advanced…] of Integration parameters. |
Set the maximum slice number (50000 is recommended). |
2949 |
Cannot perform GPC calculation because there is no peak(slice data). |
Molecular weight distribution calculation cannot be performed since no peak has been detected. |
Redo peak integration and detect the target component peak. |
294A |
Cannot compare M.W data with D.P data. |
The unit of molecular weight distribution is different between the data to be compared and the data already loaded. |
Set the unit of molecular weight distribution in both data to either molecular weight or degree of polymerization for comparison. |
294B |
No PDA multi-chromatogram channel is set. Set any one of multi-chromatogram channels (Ch#1 to 4). |
GPC Data Analysis will not be operated unless any one of Ch#1 to 4 is selected. |
Edit multi-chromatogram table in Real Time Analysis or Postrun Analysis, and select one from Ch#1 to 4. |
294C |
Cannot open file since no PDA multi-chromatogram channel is set. Set PDA multi-chromatogram channels(Ch#1 to 4) in the Method editor. |
GPC Data Analysis will not be operated unless any one of Ch#1 to 4 is selected. |
Edit multi-chromatogram table in Real Time Analysis or Postrun Analysis, and select one from Ch#1 to 4. |
294E |
Can not merge CLASS GPC method into LCsolution method file that is enabled audit trail. |
You cannot merge CLASS GPC method into LabSolutions methods in which audit trail function is enabled. |
Merge CLASS GPC method into another LabSolutions method in which the audit trail function is not enabled. |
294F |
Correction by GPC control sample is not set in the method although GPC control sample is set to the sample type in batch |
Even though the sample type in a Batch is set to GPC Control Sample, the registered method file is not set to perform Correction by GPC control sample. |
Specify the setting of Correction by GPC control sample to be consistent between in batch table and in method. |
2951 |
IS peak was not identified. Time correction is not applied to GPC calculation. |
Time correction using internal standard could not be executed because IS peak was not identified. |
Verify the retention time of IS peak and modify the set value so that the IS peak can be identified. |
2952 |
Control sample peak was not identified. The current correction coefficient will be applied. |
Time correction using control sample could not be executed because control sample peak was not identified. |
Verify that the retention time of control sample is appropriate and re-execute the batch. |
2953 |
Could not import data analysis parameters with another detector setting. |
If the detector setting in a method file is different from the one in current data, the method file will not be imported. |
Import method files with the same environment setting as the one in the current data. |
2954 |
First, open method file. |
The system cannot load selected data file because no method file is opened in GPC Calibration Curve. |
Open method file before opening a data file. |
2955 |
Unable to define marker position (Time) in Chromatogram View using the current calibration |
Marker line was moved to the position where molecular weight calculation cannot properly be performed on the current |
Verify that the calibration curve is monotone decreasing. |
2956 |
Confirm the settting of the T.LIMIT.\nOne or more calibration points exist within T.LIMIT.\nContinue to save? |
The calibration point exists within T.LIMIT. |
Check the T.LIMIT value or the calibration point, and if correct, click [Yes].\n[No] will be clicked if wrong. |
2957 |
Peak exists within T.LIMIT. Cannot calculate proper molecular weight for the slice |
A peak exists within T.LIMIT. |
About the peak before T.LIMIT, a molecular weight is not calculated correctly.\nPlease improve the calibration curve to calculate it correctly. |
2958 |
Data file(s) cannot be saved because read only data file(s) exist. Are you sure you want to continue? |
Data file(s) open by read only exist. |
A click of [Yes] will save a method file, without saving data file(s).\nA click of [No] will not save a method file. |
Comments
0 comments
Please sign in to leave a comment.